The Acessa ProVu System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance. The Acessa ProVu System includes optional electromagnetic guidance for enhancing the ultrasonic image of the Acessa ProVu Handpiece and for predicting its future path on a computer monitor screen which also shows the ultrasound B-scan image.
Device Story
Acessa ProVu System integrates RF ablation, ultrasound visualization, and electromagnetic guidance into a single console. Used in laparoscopic procedures by physicians to treat symptomatic uterine fibroids. Input: RF power (up to 200W, 460kHz) and ultrasound signals. Operation: Handpiece electrodes deployed in tissue; RF energy delivered in Temperature Control, Power Control, or Coagulation modes. Electromagnetic tracking monitors the spatial relationship between the transducer and handpiece, projecting a virtual 3D representation and predicted path onto a hospital-provided monitor alongside the ultrasound B-scan. Output: Real-time visualization of ablation path and tissue features. Benefits: Enhanced guidance for precise tissue ablation. Safety: System monitors parameters; automatically stops energy if limits exceeded; user can manually override via foot pedal.
Clinical Evidence
Bench testing only. Testing included biocompatibility, transducer use-life, electrical safety (ANSI/AAMI/ES 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-2-37), software verification/validation, console/guidance system testing, benchtop ablation performance, and mechanical/functional qualification of handpieces, pads, and transducers.
Technological Characteristics
Monopolar RF ablation system (200W, 460kHz, sinusoidal). Electromagnetic tracking (NDI V3) for spatial visualization. Ultrasound beamformers/drivers. Connectivity: USB port, hospital-provided monitor. Software: Major level of concern. Standards: ANSI/AAMI/ES 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-2-37.
Indications for Use
Indicated for percutaneous, laparoscopic coagulation and ablation of soft tissue, specifically symptomatic uterine fibroids, in patients requiring such treatment under laparoscopic ultrasound guidance.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 28, 2018
Acessa Health, Inc. Kim Rodriguez President & CEO 7004 Bee Cave Road, Bldg. 3, Suite 200 Austin, TX 78746
Re: K181124
> Trade/Device Name: Acessa ProVu System Regulation Number: 21 CFR§ 884.4160 Regulation Name: Unipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: HFG, OEW, IYO, ITX Dated: August 31, 2018 Received: August 31, 2018
Dear Kim Rodriguez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181124
Device Name Acessa ProVu System
Indications for Use (Describe)
The Acessa Pro Vu System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.
The Acessa ProVu System includes optional electromagnetic guidance for enhancing the ultrasonic image of the Acessa ProVu Handpiece and for predicting its future path on a computer monitor screen which also shows the ultrasound B-scan image.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="text-decoration: none;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> </span> </span> </span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Acessa ProVu System is provided below.
| | Primary | Secondary |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Submitter's Name: | Acessa Health Inc. | Acessa Health Inc. |
| Submitter's Address: | 7004 Bee Cave Rd,<br>Bldg. 3, Suite 200,<br>Austin, TX 78746 | 7004 Bee Cave Rd,<br>Bldg. 3, Suite 200,<br>Austin, TX 78746 |
| Submitter's Telephone: | (877) 412-3828 | (877) 412-3828 |
| Submitter's Fax: | (925) 605-0327 | (925) 605-0327 |
| Contact Name: | Kim Rodriguez | Brian J. Bergeron |
| Date Summary was Prepared: | September 27, 2018 | |
| Trade or Proprietary Name: | Acessa ProVu System | |
| Common or Usual Name: | Radiofrequency Ablation System | |
| Primary Product Code: | HFG (coagulator, laparoscopic, unipolar (and<br>accessories)) | |
| Regulation Name and Number: | Class II per 21 CFR 884.4160: Unipolar Endoscopic<br>Coagulator-Cutter and Accessories | |
| Secondary Product Codes: | OEW, IYO, ITX | |
| Classification Panel: | Obstetrics and Gynecology | |
| Legally Marketed<br>(Unmodified) Devices to<br>Which Substantial Equivalence<br>is Claimed: | Primary Predicate:<br>K132744 Acessa Guidance System<br>Secondary Predicates:<br>K121858 The Acessa System<br>K163121 Telemed SmartUS Ultrasound System<br>The predicate devices have not been subject to a design<br>related recall. | |
{4}------------------------------------------------
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Acessa ProVu System provides radiofrequency (RF) ablation, ultrasound visualization and guidance within a single console and includes additional accessories. The Acessa ProVu System consists of a console with push buttons to control menu functionality. The console contains the following hardware and electronic components:
- RF Ablation System 1.
- Ultrasound Visualization System 2.
- 3. Guidance System
{5}------------------------------------------------
The following accessories connect to the console:
- Dual Foot pedal (one for RF ablation and one for coagulation) O
- Video Cable O
- Power Cord O
- Acessa Pads and Acessa Pad Cable O
- O Acessa ProVu Handpiece and Acessa ProVu Handpiece Cable
- O Acessa Transducer, either:
- Acessa ProVu Transducer used with Acessa ProVu Transducer Sleeve ● (embedded with sensor), or
- Acessa ProVu Transducer with (embedded) sensor
- Acessa Table Top Field Generator or Acessa ProVu Planar Field Generator O
- Monitor (hospital-owned accessory monitor, not provided by Acessa) O
The Acessa ProVu System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control, Power Control and Coagulation Mode. The graphical user interface (GUI) is displayed on a hospital-owned monitor. The system enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the Acessa ProVu Handpiece placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the monitor. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal.
The Acessa ProVu System uses electromagnetic tracking technology to track the positions of the Acessa ProVu Transducer used with Acessa ProVu Transducer Sleeve (or the Acessa ProVu Transducer with embedded sensor) and the Acessa ProVu Handpiece shaft and draws virtual representations of them in their spatial relationship, so that a physician can predict the Acessa ProVu Handpiece shaft's future path in relation to the features in the ultrasound slice. The use of the guidance functionality is considered an optional accessory to procedures where ultrasound is currently used for visualization, such as the Acessa System procedure.
{6}------------------------------------------------
## INTENDED USE/INDICATIONS FOR USE
The Acessa ProVu System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.
The Acessa ProVu System includes optional electromagnetic guidance for enhancing the ultrasonic image of the Acessa ProVu Handpiece and for predicting its future path on a computer monitor screen which also shows the ultrasound B-scan image.
The subject and predicate device have the same intended use.
| | Acessa ProVu<br>System<br>(K181124) | Acessa Guidance<br>System<br>(K132744) | The Acessa<br>System<br>(K121858) | SmartUS EXT-<br>1M/3M<br>(K163121) |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| General Information | | | | |
| Classification<br>Regulation | 21 CRF §884.4160 | 21 CRF §884.4160 | 21 CRF §884.4160 | 21 CRF §892.1550 |
| Classification Name | Unipolar endoscopic<br>coagulator cutter and<br>accessories | Unipolar endoscopic<br>coagulator cutter and<br>accessories | Unipolar endoscopic<br>coagulator cutter and<br>accessories | Ultrasonic pulsed<br>doppler imaging<br>system |
| Product Code | HFG - coagulator,<br>laparoscopic, unipolar<br>(and accessories) | HFG - coagulator,<br>laparoscopic, unipolar<br>(and accessories) | HFG - coagulator,<br>laparoscopic, unipolar<br>(and accessories) | IYN |
| Secondary Product<br>Codes | OEW, IYO, ITX | OEW, IYO | N/A | ITX, IYO |
| Product Class | II | II | II | II |
| Intended Users | Health Care<br>Professionals | Health Care<br>Professionals | Health Care<br>Professionals | Health Care<br>Professionals |
| Classification Panel | Obstetrics and<br>Gynecology | Obstetrics and<br>Gynecology | Obstetrics and<br>Gynecology | Radiology |
| System Accessories | Field Generator<br>Ultrasound Transducer<br>RF Handpiece<br>Dual foot pedal<br>Single use pads<br>Handpiece Cable<br>Pad Cable<br>Power Cord | Field Generator<br>Ultrasound Transducer<br>RF Handpiece<br>Dual foot pedal<br>Single use pads<br>Handpiece Cable<br>Pad Cable<br>Power Cord | Field Generator<br>Ultrasound Transducer<br>RF Handpiece<br>Dual foot pedal<br>Single use pads<br>Handpiece Cable<br>Pad Cable<br>Power Cord | |
| RF Functionality | | | | |
| Operating Principle | Monopolar | Monopolar | Monopolar | |
| Max Power<br>Delivery | 200 W into 50-80 Ω | 200 W into 50-80 Ω | 200 W into 50-80 Ω | |
| Frequency | 460 kHz | 460 kHz | 460 kHz | |
| Wave Form | Sinusoidal | Sinusoidal | Sinusoidal | |
| Temperature<br>Display Range | 15 - 125°C | 15 - 125°C | 15 - 125°C | |
| Impedance Range | 0-511 Ohms | 0-511 Ohms | 0-511 Ohms | |
| Duration Range | 0.1 to 12 minutes | 0.1 to 12 minutes | 0.1 to 12 minutes | |
| Temperature<br>Display | Handpiece and Pad<br>thermocouples | Handpiece and Pad<br>thermocouples | Handpiece and Pad<br>thermocouples | |
| | Acessa ProVu<br>System<br>(K181124) | Acessa Guidance<br>System<br>(K132744) | The Acessa<br>System<br>(K121858) | SmartUS EXT-<br>1M/3M<br>(K163121) |
| Set Power Display | Yes, coagulation only | Yes, coagulation only | Yes, coagulation only | |
| Duration of Energy<br>Delivery Displayed | Yes | Yes | Yes | |
| Control Modes | Temperature, Manual,<br>Coagulation | Temperature, Manual,<br>Coagulation | Temperature, Manual,<br>Coagulation | |
| Power Level<br>Controls | Yes, coagulation only | Yes, coagulation only | Yes, coagulation only | |
| Ablation Duration<br>Controls | Yes | Yes | Yes | |
| Maximum Ablation<br>Duration | 12 minutes at target<br>time | 12 minutes at target<br>time | 12 minutes at target<br>time | |
| Power Supply (PS) | 100-240 V, 50/60 Hz<br>switchable PS | 100-240 V, 50/60 Hz<br>switchable PS | 100-240 V, 50/60 Hz<br>switchable PS | |
| Peak Voltage | 230 V | 230 V | 230 V | |
| Manual Mode<br>Power Limit | 15 W | 10-15 W | 10-15 W | |
| Number of Serial<br>Communication<br>Ports | Single USB port | Single USB Port | Single USB Port | |
| Pump used with<br>Generator | None | None | None | |
| Display | Hospital-provided<br>Monitor | LCD/Touch Screen | LCD/Touch Screen | |
| User Control<br>Interface | System settings<br>controlled at the<br>Console. | System settings<br>controlled with the<br>handpiece buttons and<br>at the RF Generator<br>LCD/Touch Screen. | System settings<br>controlled with the<br>handpiece buttons and<br>at the RF Generator<br>LCD/Touch Screen. | |
| Guidance Functionality | | | | |
| Tracking<br>Technology | Electromagnetic | Electromagnetic | | |
| Accuracy | ±10 mm | ±10 mm | | |
| Views | Provide a<br>geometrically accurate<br>"3D" representation of<br>the Transducer and<br>Handpiece.<br>Display a 2D inset<br>view of the ultrasound<br>video, with the added<br>feature of the target<br>zone overlaid on it and<br>the laparoscopic<br>ultrasound transducer | Provide a<br>geometrically accurate<br>"3D" representation of<br>the Transducer and<br>Handpiece.<br>Display a 2D inset<br>view of the ultrasound<br>video and the<br>laparoscopic<br>ultrasound transducer | | |
| Tracking System | NDI V3 | NDI V2 | | |
| Ultrasound Functionality | | | | |
| Ultrasound System | Ultrasound beam<br>formers and drivers | Ultrasound beam<br>formers and drivers<br>(Off the shelf Aloka 6<br>Ultrasound system<br>K093488) | Ultrasound beam<br>formers and drivers<br>(Off the shelf Aloka 6<br>Ultrasound system<br>K093488) | Ultrasound beam<br>formers and drivers |
| Ultrasound<br>Transducer | Acessa ProVu<br>Transducer (Model<br>Number: 7500 or<br>7700) | Aloka UST 5526L-75<br>Transducer<br>(K040719) | Aloka UST 5526L-75<br>Transducer<br>(K040719) | Aloka UST 5526L-75<br>Transducer<br>(K040719) |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
{7}------------------------------------------------
{8}------------------------------------------------
The subject device is a combination system of the previously cleared devices that includes an RF ablation system (K121858), guidance control (K132744) and an OEM ultrasound component (K163121). As evidenced by the table above, there are differences in technological characteristics between the subject and predicate device. These differences do not raise different questions of safety and effectiveness.
## SUMMARY OF PERFORMANCE DATA
To further support a determination of substantial equivalence, non-clinical bench testing was conducted to support the subject device. The specific types of non-clinical testing conducted are listed below.
- Biocompatibility Testing O
- Use Life Testing of Transducers o
- Electrical Safety and Electromagnetic Compatibility testing per: O
- ANSI/AAMI/ES 60601-1:2005 + A2:2010 + A1:2012 .
- IEC 60601-2-2: 2017 ●
- IEC 60601-1-2: 2014
- . IEC 60601-2-37: 2015
- Software Validation and Verification Testing per FDA's Guidance for Industry and O FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
- Acessa ProVu Console and Guidance System Level Testing O
- Benchtop Ablation Testing to validate that the ablations created by the subject device O are substantially equivalent to the predicate.
- Acessa ProVu Handpiece Performance Testing to qualify and quantify the O mechanical and functional properties of the Handpiece.
- Acessa Pads Performance Testing to qualify and quantify the mechanical and O functional properties of the Pads.
- Acessa ProVu Transducer Performance Testing, including Acoustical and Thermal O Measurement, in conformance with IEC 60601-2-37
- Acessa ProVu Transducer Sleeve Testing to qualify and quantify the mechanical and O functional properties of the Transducer Sleeve.
# CONCLUSION
The performance data demonstrate that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.