K132744 · Halt Medical, Inc. · HFG · May 27, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K132744
Device Name
ACESSA GUIDANCE SYSTEM
Applicant
Halt Medical, Inc.
Product Code
HFG · Obstetrics/Gynecology
Decision Date
May 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4160
Device Class
Class 2
Indications for Use
The Acessa Guidance System is indicated for enhancing the ultrasonic image of the Acessa Handpiece and for predicting its future path on a computer monitor screen which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended as an optional accessory for use during the Acessa System procedure.
Device Story
Acessa Guidance System uses electromagnetic tracking to visualize surgical instruments during laparoscopic ultrasound-guided procedures. System components include a Guidance Controller, a disposable Guidance Ultrasound Transducer Sleeve, and a Table Top Field Generator (TTFG). The TTFG generates a magnetic field detected by sensors within the transducer sleeve and the Acessa Handpiece. The controller processes these signals to draw virtual representations of the handpiece and transducer on a computer monitor, overlaying the handpiece's predicted path onto the B-scan ultrasound image. This allows the physician to navigate the handpiece accurately relative to anatomical features during radiofrequency ablation. The system is an optional accessory used in clinical settings by physicians. It benefits patients by providing real-time spatial guidance, potentially increasing procedural precision and safety.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety (ANSI/AAMI ES 60601-1), EMC (EN/IEC 60601-1-2), and programmable electrical medical systems (IEC 60601-1-4) testing. Biocompatibility was verified per ISO 10993-1, -5, -10, and -11. System performance was validated through Guidance Sleeve end cap tensile testing and worst-case accuracy comparison testing in a simulated clinical environment. Software validation was performed by the predicate device manufacturer.
Technological Characteristics
Electromagnetic tracking system; components include Guidance Controller, disposable Ultrasound Transducer Sleeve, and Table Top Field Generator. Connectivity: Wired (4.5m Field Generator cable, Guidance Handpiece Cable). Power: AC medical grade. Software: Programmable electrical medical system. Biocompatibility: Compliant with ISO 10993 standards. Safety: Compliant with ANSI/AAMI ES 60601-1, EN/IEC 60601-1-2, and IEC 60601-1-4.
Indications for Use
Indicated for enhancing ultrasonic visualization and predicting the path of the Acessa Handpiece during the Acessa System radiofrequency ablation procedure. Intended for use by clinicians in a surgical setting.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
{0}------------------------------------------------
K132744
MAY 27 2014
Acessa
guidance
## Section 5 - 510(k) Summary
.
As Required By 21 CFR 807.92
| Date of Summary Preparation | May 22, 2014 |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter and Owner's Name and Address | Halt Medical, Inc.<br>131 Sand Creek Road, Suite B<br>Brentwood, CA 94513<br>Main: (925) 634-7943<br>Fax: (925) 634-7841 |
| Contact Person | Clarisa A. Tate<br>VP of RA & QA, Halt Medical, Inc.<br>Office: (925) 271-0626<br>e-mail: ctate@haltmedical.com |
| Trade Name | Acessa Guidance System |
| Common Name | Acessa Guidance System or electromagnetic tracking system |
| Classification Name | Unipolar endoscopic coagulator cutter and accessories |
| Classification | Class II |
| Product Code | HFG, OEW, IYO |
| Classification Panel | Obstetrics and Gynecology |
| Classification Regulation | 21 CFR §884.4160 |
| Legally Marketed Device to which<br>substantial equivalence is claimed | AIM System (K121479) |
| Intended Use | The Acessa Guidance System is indicated for enhancing the ultrasonic image of<br>the Acessa Handpiece and for predicting its future path on a computer monitor<br>screen which also shows the image of a B-scan (or similar display) of a medical<br>ultrasound imaging system. The device is intended as an optional accessory for<br>use during the Acessa System procedure. |
| Device<br>Description | The Acessa Guidance System consists of the following components:<br>Guidance Controller (Model Number 5100): Contains the tracking system and software used to run the system, with attached 4.5m Field Generator cable. Guidance Ultrasound Transducer Sleeve (Model Number 5500): A disposable sleeve that houses the ultrasound transducer, the magnetic guidance sensor and guidance display control buttons. Guidance Field Generator (Model Number 5200): The TTFG (Table Top Field Generator) generates a magnetic field that is picked up by the magnetic guidance sensors in the Handpiece and the Ultrasound Transducer Sleeve. Guidance Handpiece Cable (Model Number 5400): Connects a Guidance Handpiece to the Acessa Generator and the Guidance Controller. Power Cord (Model Number 4110): A medical grade power cord that provides AC power to the Controller. NOTES: |
| | • The Acessa Guidance System must be used with laparoscopic ultrasound.<br>Laparoscopic ultrasound equipment is not included with the Acessa<br>Guidance System.<br>• The Acessa Guidance System may only be used with the Acessa System, a<br>radiofrequency ablation device. |
| | The Acessa Guidance System uses electromagnetic tracking technology to track<br>the positions of the Guidance Ultrasound Transducer Sleeve and the Guidance<br>Handpiece shaft and draws virtual representations of them in their spatial<br>relationship, so that a physician can predict the Guidance Handpiece shaft's<br>future path in relation to the features in the ultrasound slice. The Acessa<br>Guidance System is considered an optional accessory to procedures where<br>ultrasound is currently used for visualization, such as the Acessa System<br>procedure. |
| Technological<br>Characteristics<br>Compared to<br>Predicate<br>Devices | The design features and principal modes of operation of the Acessa Guidance<br>System is the equivalent to the commercially available AIM System. Both<br>products are configured the same and the software is made by the predicate<br>device's company which they based on their cleared device. |
| | Application of electromagnetic tracking technology with both devices within a<br>surgical procedure is completed in the same manner. Substantial equivalence is<br>established with respect to the same indication for use, principal design,<br>software used, performance, and safety requirements. |
| Performance<br>Testing | The Acessa Guidance System was subjected to electrical and safety testing<br>according to risks assessments performed based on the differences with the<br>predicate device. EMC and safety testing were completed on the device as<br>required by ANSI/AAMI ES 60601-1:2005 with A2:2010 3rd edition Medical Device<br>Equipment Part 1 General Requirements for Safety, EN/IEC 60601-1-2:2007 3rd<br>edition Medical Electrical Equipment General Requirements for Basic Safety and<br>Essential Performance Collateral Standard Electromagnetic Compatibility<br>Requirements and Tests, IEC 60601-1-4:2000 1st edition Medical Electrical<br>Equipment Par 1-4: General Requirements for Safety - Collateral Standard:<br>Programmable Electrical Medical Systems. The device passed all testing. |
| | A series of biocompatibility testing also demonstrated that the device with its<br>additional materials are safe, suitable, and appropriate for their intended use<br>and in compliance with ISO 10993-1:2009 4th edition Biological Evaluation of<br>Medical Devices Part 1 Evaluation and Testing within a Risk Management<br>Process, ISO 10993-5:2009 3rd edition Biological Evaluation of Medical Devices<br>Tests for In Vitro Cytotoxicity, ISO 10993-10:2010 2nd edition Biological<br>Evaluation of Medical Devices Tests for Irritation and Delayed-Type<br>Hypersensitivity, and ISO 10993-11:2006 2nd edition Biological Evaluation of<br>Medical Devices Tests for Systemic Toxicity. |
| | The Acessa Guidance System was also subjected to Guidance Sleeve end cap<br>tensile testing as well as system worse-case accuracy comparison testing in a |
| | simulated clinical environment. The device met all criteria. |
| | Software validations were also conducted, which was conducted by the same<br>manufacturer of the predicate device. |
| Conclusion | The Acessa Guidance System is substantially equivalent to the legally marketed<br>medical device as demonstrated by the technological characteristics comparison<br>and performance testing completed for this device. |
{1}------------------------------------------------
.
Image /page/1/Picture/1 description: The image shows the logo for Acessa guidance. The word "Acessa" is written in a cursive font, with the "A" being larger than the other letters. Below "Acessa" is the word "guidance" in a smaller, sans-serif font. The logo is simple and clean, with a focus on the company name.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Acessa Guidance. The word "Acessa" is written in a cursive font, with the "A" being the largest letter and the other letters gradually decreasing in size. Below "Acessa", the word "guidance" is written in a smaller, sans-serif font. The overall design is simple and professional.
:
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an eagle or bird in flight, with three curved lines representing wings or feathers. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
## DEPARTMENT OF HEALTII & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2014
Halt Medical, Inc. Clarisa A. Tate Vice President of Regulatory Affairs & Quality Assurance 131 Sand Creek Road, Suite B Brentwood, CA 94513-2040
Re: K132744
> Trade/Device Name: Acessa Guidance System Regulation Number: 21 CFR§ 884.4160 Regulation Name: Unipolar endoscopic coagulator cutter and accessories Regulatory Class: II Product Code: HFG Dated: April 25, 2014 Received: April 28, 2014
Dear Clarisa A. Tate,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Clarisa A. Tate
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a bold, sans-serif font. The letters "R" and "E" are partially obscured by a decorative, textured design. The name is written in black ink on a white background.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):
Device Name: Acessa Guidance System
Indications for Use: The Acessa Guidance System is indicated for enhancing the ultrasonic image of the Acessa Handpiece and for predicting its future path on a computer monitor screen which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended as an optional accessory for use during the Acessa System procedure.
Prescription Use ____X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2014.05.27 16:03:04 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.