K181075 · Nobel Biocare AB · KCT · Sep 27, 2018 · General Hospital
Device Facts
Record ID
K181075
Device Name
PureSet Tray
Applicant
Nobel Biocare AB
Product Code
KCT · General Hospital
Decision Date
Sep 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Nobel Biocare PureSet Tray is used in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization pouch or sterilization wrap.
Device Story
PureSet Tray is a reusable surgical tray system for organizing and storing Nobel Biocare instruments during cleaning, sterilization, and surgical procedures. The device consists of perforated stainless steel bases, lids, PEEK grommets/sleeves, and silicone feet; some models include anodized aluminum plates. It is used in hospitals and healthcare facilities by clinical staff. The tray is not a sterile barrier; it must be enclosed in an FDA-cleared sterilization wrap or pouch before steam sterilization in standard autoclaves. The device protects instruments during transport and storage. By organizing tools, it facilitates efficient surgical workflows and ensures instruments are properly processed. The tray's design allows for sterilant penetration via an evenly distributed hole pattern. It is intended for repeated reprocessing cycles.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993), sterilization validation (AAMI/ANSI ST79, ISO 14937), handle durability (DIN 58952-3), repeated reprocessing, and shipping simulation (ASTM D4169-16).
Technological Characteristics
Materials: Stainless steel, PEEK (grommets, sleeves, locks), silicone elastomer (feet), anodized aluminum (plates). Design: Perforated base/lid, one or two-level configurations. Sterilization: Steam (pre-vacuum and gravity displacement). Connectivity: None. Software: None. Standards: AAMI/ANSI ST79, ISO 10993, ASTM D4169-16, DIN 58952-3.
Indications for Use
Indicated for use in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning, sterilization, and implant/prosthetic treatment. Intended for use with an FDA-cleared sterilization wrap or pouch. Validated for specific surgical instruments including torque wrenches, implant drivers, direction indicators, drills, screw taps, screw drivers, and irrigation needles, with maximum load capacities of 1635g, 1082g, or 945g depending on the tray model.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Nobel Biocare - Kit Boxes and Kit Plates (K163600)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of a human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
September 27, 2018
Nobel Biocare AB % Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA, LLC 22715 Savi Ranch Parkway Yorba Linda. California 92887
Re: K181075
Trade/Device Name: PureSet Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: Class II Product Code: KCT Dated: August 27, 2018 Received: August 29, 2018
Dear Charlemagne Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181075
Device Name PureSet Tray
#### Indications for Use (Describe)
The Nobel Biocare PureSet Tray is used in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw driver and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1082 grams (NobelActive/NobelParallel CC PureSet Tray) and 945 grams (NobelReplace CC PureSet Tray).
| Method | Steam Sterilization<br>(Moist Heat Sterilization) | |
|---------------------------------------------------|---------------------------------------------------|----------------------------|
| Cycle | Dynamic-Air-Removal<br>(fractionated vacuum) | Gravity-Displacement |
| Temperature | 132°C<br>(270°F) | 132°C<br>(270°F) |
| Exposure time for a single-<br>use pouched device | 4 minutes<br>(full-cycle) | 15 minutes<br>(full-cycle) |
| Minimum drying times | 20 minutes | 30 minutes |
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
മ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# I. SUBMITTER
Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Submitted by Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348
Date Prepared: September 25, 2018
II. DEVICE
Name of Device: PureSet Tray Common or Usual Name: Sterilization Tray Classification Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (21 CFR 880.6850) Regulatory Class: II Primary Product Code: KCT
# III. PREDICATE DEVICE
Primary predicate: Nobel Biocare - Kit Boxes and Kit Plates (K163600) This predicate has not been subject to a design-related recall.
Reference predicate: Medtronic Sofamor Danek - Medtronic Transportation/Sterilization Cassettes (K152241)
# IV. DEVICE DESCRIPTION
The PureSet Tray is a reusable surgical tray to be used in combination with Nobel Biocare surgical instruments and components. The PureSet Tray is used to store and organize the instruments and components during both the surgical and reprocessing procedures. The PureSet Trays are not intended to maintain sterility on their own. They are intended to be used in conjunction with an FDA cleared sterilization wrap/pouch.
Each PureSet Tray consist of multiple components designed to be integrated into a single unit, which protects the interior components during transportation, and storage. All the components of the PureSet Tray are perforated with an evenly distributed
{4}------------------------------------------------
hold pattern and are designed to be used for sterilization via steam sterilization. Since the PureSet Trays are perforated, an FDA cleared wrap or pouch must be used for sterilization purposes and to maintain the sterility of the contents. The PureSet Trays are designed to be used with standard autoclaves used in hospitals and healthcare facilities.
# V. INDICATIONS FOR USE
The Nobel Biocare PureSet Tray is used in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization pouch or sterilization wrap. Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screw driver and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 gram (Trefoil PureSet Tray), 1082 grams (NobelActive/NobelParallel CC PureSet Tray) and 945 grams (NobelReplace CC PureSet Tray).
| Method | Steam Sterilization<br>(Moist Heat Sterilization) | |
|-----------------------------------------------------|---------------------------------------------------|----------------------------|
| Cycle | Dynamic-Air-Removal<br>(fractionated vacuum) | Gravity-Displacement |
| Temperature | 132°C<br>(270°F) | 132°C<br>(270°F) |
| Exposure time for a<br>single-use pouched<br>device | 4 minutes<br>(full-cycle) | 15 minutes<br>(full-cycle) |
| Minimum drying<br>times | 20 minutes | 30 minutes |
{5}------------------------------------------------
#### VI. Comparison of Technological Characteristics
# Comparison of intended use and indication for use statement
| Technological<br>characteristics | Subject Device | Primary Predicate | Reference Predicate | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | PureSet Tray | Kit Boxes and Kit Plates (K163600) | Medtronic Transportation/ Sterilization<br>Cassettes (K152241) | |
| | The Nobel Biocare PureSet Tray is<br>intended for use in healthcare facilities to<br>store and organize Nobel Biocare<br>surgical instruments and components<br>during cleaning/sterilization and during<br>implant/prosthetic treatment. The Nobel<br>Biocare PureSet Trays are not intended<br>on their own to maintain sterility; it is<br>intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>sterilization pouch or sterilization wrap. | The Nobel Biocare Kit Boxes are intended<br>for use in healthcare facilities to organize,<br>enclose, sterilize, transport, and store<br>medical devices between surgical uses.<br>The Nobel Biocare Kit Boxes are not<br>intended on their own to maintain sterility;<br>it is intended to be used in conjunction with<br>a legally marketed, validated, FDA-cleared<br>sterilization wrap. Sterilization validations<br>for the worst-case Nobel Biocare Kit Box<br>(170 mm x 145 mm x 57) included surgical<br>instruments such as torque wrenches,<br>implant drivers, direction indicators, drills,<br>etc. | The Medtronic Transportation/Sterilization<br>Cassettes are intended for use in healthcare<br>facilities to organize, enclose, sterilize,<br>transport, and store medical devices and<br>other instrumentation between surgical and<br>other medical uses. The Medtronic<br>Transportation/Sterilization Cassettes are not<br>intended on their own to maintain sterility; it<br>is intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>sterilization wrap. | |
| Technological<br>characteristics | Subject Device | Primary Predicate | Reference Predicate | |
| | PureSet Tray | Kit Boxes and Kit Plates (K163600) | Medtronic Transportation/ Sterilization<br>Cassettes (K152241) | |
| Indication for<br>use statement | The Nobel Biocare PureSet Tray is used<br>in healthcare facilities to store and<br>organize Nobel Biocare surgical<br>instruments and components during<br>cleaning/sterilization and during<br>implant/prosthetic treatment. The Nobel<br>Biocare PureSet Trays are not intended<br>on their own to maintain sterility; it is<br>intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>sterilization pouch or sterilization wrap.<br>Sterilization validations for the worst-<br>case PureSet Tray included surgical<br>instruments such as torque wrenches,<br>implant drivers, direction indicators,<br>drills, screw taps, screw driver and<br>irrigation needles. The PureSet Trays<br>were validated for a maximum load of<br>1635 grams (Trefoil PureSet Tray), 1082<br>grams (NobelActive/NobelParallel CC<br>PureSet Tray) and 945 grams<br>(NobelReplace CC PureSet Tray). | The Nobel Biocare Kit Boxes are used in<br>healthcare facilities to organize, enclose,<br>sterilize, transport, and store medical<br>devices between surgical uses. The Nobel<br>Biocare Kit Boxes are not intended on their<br>own to maintain sterility; it is intended to be<br>used in conjunction with a legally<br>marketed, validated, FDA-cleared<br>sterilization wrap. Sterilization validations<br>for the worst-case Nobel Biocare Kit Box<br>(170 mm x 145 mm x 57mm) included<br>surgical instruments such as torque<br>wrenches, implant drivers, direction<br>indicators, drills, etc. The Kit Boxes were<br>validated for a maximum load of 469<br>grams including kit box and instruments. | Cycle Tempe-<br>rature Exposure<br>time Minimum<br>dry time | |
| | | Method<br>Steam Sterilization<br>(Moist Heat Sterilization) | Gravity<br>Displace-<br>ment 250° F<br>(121°<br>C) 30<br>Minutes 30<br>Minutes | |
| | | Cycle Dynamic-<br>Air-<br>Removal<br>(fractionat<br>ed<br>vacuum) Dynamic-<br>Air-<br>Removal<br>(fractionat<br>ed<br>vacuum) Gravity-<br>Displacem<br>ent | Gravity<br>Displace-<br>ment 270° F<br>(132°<br>C) 15<br>Minutes 30<br>Minutes | |
| | | | Gravity<br>Displace-<br>ment 275° F<br>(135°<br>C) 10<br>Minutes 30<br>Minutes | |
| | | | Dynamic-<br>Air-<br>Removal 270º F<br>(132°<br>C) 4<br>Minutes 30<br>Minutes | |
| | | Method<br>Steam Sterilization<br>(Moist Heat Sterilization) | Dynamic-<br>Air-<br>Removal 275° F<br>(135° C 3<br>Minutes 30<br>Minutes | |
| | | Temperat<br>ure 132°C<br>(270°F) 134°C<br>(270°F) 132°C<br>(270°F) | The Medtronic Transportation/ Sterilization<br>Cassettes are intended for use in healthcare<br>facilities to organize, enclose, sterilize,<br>transport, and store medical devices and<br>other instrumentation between surgical and<br>other medical uses. The Medtronic<br>Transportation/ Sterilization Cassettes are<br>not intended on their own to maintain sterility;<br>it is intended to be used in conjunction with a<br>legally marketed, validated, FDA-cleared<br>sterilization wrap. | |
| | | Cycle Dynamic-<br>Air-<br>Removal<br>(fractionat | Sterilization validations for the worst-case<br>Medtronic Transportation/Sterilization | |
| | | Gravity-<br>Displacement | | |
| | | Exposure<br>time for a<br>single-<br>use 4 minutes<br>(full-<br>cycle) 3 minutes<br>(full-cycle) 15<br>minutes<br>(full-cycle) | | |
| Technological<br>characteristics | Subject Device | | Primary Predicate | Reference Predicate |
| | PureSet Tray | | Kit Boxes and Kit Plates (K163600) | Medtronic Transportation/ Sterilization<br>Cassettes (K152241) |
| | ed<br>vacuum) | | pouched<br>device | Cassette (22.75 x 11.26 x 5.5 inches)<br>included implants and common surgical<br>instruments such as rasps, drivers, trials,<br>handles, inserters, probes, drills, etc. The<br>validated total weight was 28.4 lbs. The<br>validated worst-case loading configurations<br>of the Medtronic Transportation/ Sterilization<br>Cassette included the following worst-case<br>lumen dimensions. |
| | Temperature | 132°C<br>(270°F) | Exposre<br>time for<br>the<br>Overkill<br>Approach | |
| | | 132°C<br>(270°F) | 2 minutes<br>(half<br>cycle) | -363 x 1.575 mm<br>-247.5 x 4.1 mm<br>Device List:<br>1850060 Case- Triple Generic Outer Base<br>(22.74 x 11.260 x 5.040 inches) |
| | Exposure<br>time for a<br>single-use<br>pouched<br>device | 4 minutes<br>(full-cycle) | 1.5<br>minutes<br>(half<br>cycle) | 1850064 Lid- Generic Outer Lid (22.75 x<br>11.260 x 0.470 inches)<br>7022101L Tray Lid (21 x 10.13 x 0.075<br>inches)<br>P1850061 Tray 1 (20.75 x 9.79 x 1.32<br>inches) |
| | | 15 minutes<br>(full-cycle) | 7.5<br>minutes<br>(half<br>cycle) | P1850062 Tray 2 (21 x 10.13x 1.69 inches)<br>P1850063 Tray 3 (21 x 10.13 x 1.38 inches)<br>7059532 Large Caddy (9.47 x 6.37 x 1.3<br>inches) |
| | Minimum<br>drying times | 20<br>minutes | Minimum<br>drying<br>times | 7059532L Large Lid (5.85 x 4.725 x 0.095<br>inches) |
| | | 30 minutes | 20<br>minutes | P213018 Small Caddy (2 x 1.5 x 1.025<br>inches) |
| | | | 10<br>minutes | P9213018 Small Lid (2 x 1.29 x 0.095<br>inches) |
| | | | 30<br>minutes | |
{6}------------------------------------------------
Nobel Biocare Traditional 510(k) Notification (K181075) PureSet Tray September 2018
{7}------------------------------------------------
{8}------------------------------------------------
| Technological | Subject Device | Primary Predicate | Reference Predicate | | | |
|-----------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------|-----------------------------------|--------------------------------------|
| characteristics | PureSet Tray | Kit Boxes and Kit Plates (K163600) | Medtronic Transportation/ Sterilization<br>Cassettes (K152241) | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.