K163600 · Nobel Biocare AB · KCT · Jan 22, 2018 · General Hospital
Device Facts
Record ID
K163600
Device Name
Kit Boxes and Kit Plates
Applicant
Nobel Biocare AB
Product Code
KCT · General Hospital
Decision Date
Jan 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Nobel Biocare Kit Boxes are used in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices between surgical uses. The Nobel Biocare Kit Boxes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. Sterilization validations for the worst-case Kit Box included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screw driver and irrigation needles. The Kit Boxes were validated for a maximum load of 469 grams including kit box and instruments.
Device Story
Kit Boxes are modular, reusable containers for organizing, sterilizing, and storing surgical instruments (e.g., drills, drivers, taps) used in Nobel Biocare implant procedures. System consists of a base (silicone over-molded on plastic) and a removable transparent plastic lid; includes an insertable Kit Plate with silicone grommets and tool holders. Used in healthcare facilities by clinical staff. Device is non-sterile upon receipt; requires steam sterilization (dynamic-air-removal or gravity-displacement) inside an FDA-cleared sterilization wrap. Provides organized storage for instruments between surgical uses. Benefits include instrument protection and streamlined sterilization workflows. No mechanical stress exposure; biocompatibility confirmed via cytotoxicity testing per ISO 10993-5.
Clinical Evidence
Bench testing only. Validation included steam sterilization cycle parameters (temperature, exposure time, drying time) and cytotoxicity testing per ISO 10993-5 to ensure no toxic leachable substances transfer to instruments during sterilization.
Technological Characteristics
Materials: Polymer, silicone, and aluminum (drill stop box only). Configuration: Perforated base, lid, and insertable plate with silicone grommets/tool holders. Sterilization: Steam (dynamic-air-removal and gravity-displacement). Reusable. Biocompatibility: ISO 10993-5 (cytotoxicity).
Indications for Use
Indicated for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices between surgical uses. Used in conjunction with FDA-cleared sterilization wrap. Validated for surgical instruments including torque wrenches, implant drivers, direction indicators, drills, screw taps, screw drivers, and irrigation needles.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Restore Modular Sterilization Tray System (K131455)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three intertwined snakes and a pair of wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2018
Charlemagne Chua Sr. Regulatory Affairs Manager Nobel Biocare USA, LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Re: K163600
Trade/Device Name: Kit Boxes Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: December 20, 2017 Received: December 21, 2017
Dear Charlemagne Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163600
Device Name Kit Boxes
#### Indications for Use (Describe)
The Nobel Biocare Kit Boxes are used in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices between surgical uses. The Nobel Biocare Kit Boxes are not intended on their own to maintain sterility, it is intended to be used in conjunction with a legally marketed, FDA-cleared sterilization wrap. Sterilization validations for the worst-case Kit Box included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screw driver and irrigation needles. The Kit Boxes were validated for a maximum load of 469 grams including kit box and instruments.
Sterilization Parameters
| Method | Steam Sterilization<br>(Moist Heat Sterilization) | |
|---------------------------------------------------|---------------------------------------------------|-----------------------------|
| Cycle | Dynamic-Air-Removal<br>(fractionated vacuum) | Gravity-Displacement |
| Temperature | 132°C (270°F) | 132°C (270°F) |
| Exposure time for a single-<br>use pouched device | 4 minutes<br>(full-cycle) | 15 minutes<br>(full-cycle) |
| | 134°C (273.2 °F) | |
| Exposure time for the<br>Overkill Approach | 2 minutes<br>(half-cycle) | 7.5 minutes<br>(half-cycle) |
| | 1.5 minutes<br>(half cycle) | |
| Minimum drying times<br>validated | 20 minutes | 30 minutes |
| | 10 minutes | |
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### K163600
#### I. Submitter
Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x7830 Fax: (714) 998-9348
Submitted for: Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Date Prepared : January 22, 2018
#### II. Device
| Device Proprietary Name: | Kit Boxes |
|--------------------------|---------------------|
| Common or Usual Name: | Sterilization Tray |
| Classification Name: | Sterilization Wraps |
| Regulation Number: | 21 CFR 880.6850 |
| Product Code: | KCT |
| Device Classification: | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary Predicate
- . Medtronic Transportation/Sterilization Cassettes, K152241, Medtronic Sofamor Danek USA, Inc.
Reference Devices:
- Restore Modular Sterilization Tray System, K131455, Restore Medical Solutions, Inc. ●
{4}------------------------------------------------
#### IV. Device Description
The Kit Boxes are a module-based kit system that can be organized in different ways according to the user's needs. A Module is composed by a Kit Box and an inserted Kit Plate. The Module can be used as a cleaning and sterilization container. Additionally, it is used for presentation and storage of products. These products are sold as standalone articles (Drills, Screw Taps, Depth Gauge, Implant Drivers, Torque Wrench etc.) and are available in a variety of combinations for different implant systems.
The Modules are intended for storing, organizing and sterilizing the instruments used during implant/prosthetic/retrieval treatment with Nobel Biocare implants.
The Module have 2 different sizes: The Main Module, and the Sub-Module.
#### Main Module
The Main Module measures 170 mm x 145 mm x 57 mm in size and consists of a Kit Box and a Kit Plate to be inserted. The Kit Box consists of a base with support (stand) and a removable transparent plastic lid. The base is made of silicone over molded on a grey or red plastic base. The inserted Kit Plate snaps into the base of the Kit Box and consists of a pattern of silicone Grommets and tool holders for instruments.
To accommodate the different systems, the Kit Plate is available with unique patterns, different color printing, different size of Grommets and different size tool-holders to meet the end user's needs.
#### Sub-Module
The Sub-Module (smaller Module) measures 170 mm x 57 mm and consists of Kit Box and a Kit Plate to be inserted. The Kit Box consists of a base with support (stand) and a removable transparent plastic lid. The base is made of silicone over molded on a grey or red plastic base. The inserted Kit Plate snaps into the base of the Kit Box and consists of a pattern of silicone Grommets and tool holders for instruments.
To accommodate the different systems, the Kit Plate is available with unique patterns, different color printing, different size of Grommets and different size tool-holders to meet the end user's needs.
{5}------------------------------------------------
# V. Indications for Use
The Nobel Biocare Kit Boxes are used in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices between surgical uses. The Nobel Biocare Kit Boxes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. Sterilizations for the worst-case Kit Box included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screw driver and irrigation needles. The Kit Boxes were validated for a maximum load of 469 grams including kit box and instruments.
| Method | Steam Sterilization<br>(Moist Heat Sterilization) | |
|---------------------------------------------------|---------------------------------------------------|-----------------------------|
| Cycle | Dynamic-Air-Removal<br>(fractionated vacuum) | Gravity-Displacement |
| Temperature | 132°C (270°F) | 132°C (270°F) |
| Exposure time for a single-<br>use pouched device | 4 minutes<br>(full-cycle) | 15 minutes<br>(full-cycle) |
| Exposure time for the<br>Overkill Approach | 2 minutes<br>(half-cycle) | 7.5 minutes<br>(half-cycle) |
| Minimum drying times<br>validated | 20 minutes | 30 minutes |
#### Sterilization Parameters
# VI. Comparison of Technological Characteristics
The subject and primary predicate devices share the following characteristics:
- . Similar Indications for Use
- Similar materials of construction
- Configuration
- Sterilization Methods
- Sterilization Parameters
- Reusable
- Microbial Barrier Properties ●
{6}------------------------------------------------
# Discussion
The Indications for Use statement between the subject and predicate devices is similar and does not differ. Both the subject and predicate devices are intended to organize, enclose, sterilize, transport and store medical devices and other instrumentation between surgical and other medical uses.
The subject devices are technologically different from the predicate devices as follows:
- Dimension .
- Design .
A comparison of the subject and predicate devices is provided in the table below.
{7}------------------------------------------------
| | Kit Boxes | PRIMARY PREDICATE<br>Medtronic Transportation / Sterilization<br>Cassettes<br>(K152241) | REFERENCE DEVICE<br>Restore Modular Sterilization Tray System<br>(K131455) |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Design | A base, inserted plate, support(stand) and a<br>lid | A base, a lid with a locking latch and<br>individual inserts | A base, a lid and individual inserts |
| Dimensions | 170 mm x 145 mm x 57 mm (main)<br>170 mm x 70 mm x 57 mm (sub)<br>310 mm x 180 mm x 70 mm (kit organizer) | 22.75 in. x 11.26 in. x 5.51 in. | 10 in. x 20 in. x 4 in.<br>10 in. x 10 in. x 4 in. |
| Configuration | Perforated bases, lids and silicone grommets | Perforated bases, lids, and inserts | Perforated bases, lids, and inserts |
| Device Material | Aluminum (Drill Stop Box only), Polymer<br>and Silicon | Thermoplastic polymer, aluminum and<br>stainless steel | Thermoplastic polymer, aluminum and<br>stainless steel |
| Perfect<br>Perforation | Evenly distributed hole pattern | Evenly distributed hole pattern | Evenly distributed hold pattern |
| Sterilization<br>Method | Pre-Vacuum<br>Gravity Displacement | Pre-Vacuum<br>Gravity Displacement | High Vacuum Steam |
| Sterilant<br>Penetration | Yes | Yes | Yes |
| Microbial Barrier<br>Properties | To be used with an FDA cleared<br>sterilization wrap/pouch | To be used with an FDA cleared<br>sterilization wrap | To be used with an FDA cleared<br>sterilization wrap |
| Reusable | Yes | Yes | Yes |
| Material<br>Compatibility<br>with Sterilization<br>Method | Yes | Yes | Yes |
{8}------------------------------------------------
| | Kit Boxes | | | PRIMARY PREDICATE<br>Medtronic Transportation / Sterilization<br>Cassettes<br>(K152241) | | REFERENCE DEVICE<br>Restore Modular Sterilization Tray System<br>(K131455) |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------------------|-----------------------------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Parameters | Pre-Vacuum:<br>Temp 270° F<br>Exposure Time 4 minutes<br>Pre-vacuum: 3 times < 60 mbar<br>Drying Time: 30 minutes<br>Cooling Time: 10 minutes at room<br>temperature | Cycle | Tempe-<br>rature | Exposure<br>time | Minimum<br>dry time | High Vacuum (pre-vacuum, three pulse,<br>standard): Temp: 270° F<br>Exposure Time: 4 minutes<br>Cycle Dry Time (wrapped): 20 minutes<br>(minimum)<br>Cool Time: Varies according to load contents |
| | Gravity:<br>Temp 270° F<br>Exposure Time: 15 minutes<br>Pre-vacuum: N/A Drying<br>Time: 30 minutes<br>Cooling Time: 10 minutes at room<br>temperature | Gravity<br>Displace<br>-ment | 250° F<br>(121° C) | 30<br>Minutes | 30<br>Minutes | |
| | | Gravity<br>Displace<br>-ment | 270° F<br>(132° C) | 15<br>Minutes | 30<br>Minutes | |
| | | Gravity<br>Displace<br>-ment | 275° F<br>(135° C) | 10<br>Minutes | 30<br>Minutes | |
| | | Dynamic<br>-Air-<br>Removal | 270° F<br>(132° C) | 4<br>Minutes | 30<br>Minutes | |
| | | Dynamic<br>-Air-<br>Removal | 275° F<br>(135° C) | 3<br>Minutes | 30<br>Minutes | |
{9}------------------------------------------------
| | Kit Boxes | PRIMARY PREDICATE<br>Medtronic Transportation / Sterilization<br>Cassettes<br>(K152241) | REFERENCE DEVICE<br>Restore Modular Sterilization Tray System<br>(K131455) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Indications for<br>Use Statement | The Nobel Biocare Kit Boxes are used in<br>healthcare facilities to organize, enclose,<br>sterilize, transport, and store medical devices<br>between surgical uses. The Nobel Biocare Kit<br>Boxes are not intended on their own to<br>maintain sterility; it is intended to be used in<br>conjunction with a legally marketed,<br>validated, FDA-cleared sterilization wrap.<br>Sterilization validations for the worst-case Kit<br>Box included surgical instruments such as<br>torque wrenches, implant drivers, direction<br>indicators, drills, screw taps, screw driver and<br>irrigation needles. The Kit Boxes were<br>validated for a maximum load of 469 grams<br>including kit box and instruments. | The Medtronic Transportation/Sterilization<br>Cassettes are intended for use in healthcare<br>facilities to organize, enclose, sterilize,<br>transport, and store medical devices and other<br>instrumentation between surgical and other<br>medical uses. The Medtronic Transportation/<br>Sterilization Cassettes are not intended on<br>their own to maintain sterility; it is intended<br>to be used in conjunction with a legally<br>marketed, validated, FDA-cleared<br>sterilization wrap.<br><br>Sterilization validations for the worst case<br>Medtronic Transportation/Sterilization<br>Cassette (22.75 x 11.26 x 5.5 inches)<br>included implants and common surgical<br>instruments such as rasps, dirvers, trials,<br>handles, inserters, porbes, drills, etc. The<br>validated total weight was 28.4 lbs. The<br>validated worst case loading configurations<br>of the Medtronic Tranportation/ Sterilziation<br>Cassette included the following worst case<br>lumen dimensions.<br>-363 x 1.575 mm<br>-247.5 x 4.1 mm | The Restore Modular Sterilization Tray<br>System is intended for use in healthcare<br>facilities to organize, enclose, sterilize,<br>transport, and store medical devices and other<br>instrumentation between surgical and other<br>medical uses when used in conjunction with a<br>FDA cleared sterilization wrap.<br><br>The Modular Sterilization Tray System may<br>be used with common surgical instruments<br>such as hammers, rasps, drivers, chisels,<br>ringed-handled instruments, rongeurs,<br>pickups, including but not limited to short-<br>lumened devices etc. The validated total<br>weight for tray and contents was 20 pounds. | |
| Reusable Medical Device Challenge<br>Conditions | | | | |
| | | | | |
| | | Minimum<br>Inner<br>Diameter | Maximum<br>Length | Number of<br>lumens |
| | | 1.8 mm | 4 inches | 2 |
| | | High Vacuum (pre-vacuum, three pulse, | | |
{10}------------------------------------------------
| Kit Boxes | PRIMARY PREDICATE<br>Medtronic Transportation / Sterilization<br>Cassettes<br>(K152241) | REFERENCE DEVICE<br>Restore Modular Sterilization Tray System<br>(K131455) |
|-----------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| | Device List: | standard): |
| | 1850060 Case- Triple Generic Outer Base | Temp: 270° F |
| | (22.74 x 11.260 x 5.040 inches) | Exposure Time: 4 Minutes |
| | 1850064 Lid- Generic Outer Lid (22.75 x | Cycle Cry Time (wrapped): 20 minutes |
| | 11.260 x 0.470 inches) | (minimum) |
| | 7022101L Tray Lid (21 x 10.075 x 0.075 | Cool Time: Varies according to load contents |
| | inches) | |
| | P1850061 Tray 1 (20.75 x 9.79 x 1.32 inches) | Note: Cool drafts from air ducts or other air |
| | P1850062 Tray 2 (21 x 10.13 x 1.69 inches) | currents should be avoided during the cooling |
| | P1850063 Tray 3 (21 x 10.13 x 1.38 inches) | phase to avoid post-sterilization moisture |
| | 7059532 Large Caddy (9.47 x 6.37 x 1.3 | caused by rapid cooling syndrome. |
| | inches) | |
| | 7059532L Large Lid (5.85 x 4.725 x 0.095 | |
| | inches) | |
| | P213018 Small Caddy (2 x 1.5 x 1.025 | |
| | inches) | |
| | P9213018 Small Lid (2 x 1.29 x 0.095 | |
| | inches) | |
{11}------------------------------------------------
#### Discussion
As seen above, the differences between the subject and predicate devices are the design and dimensions of the kit box/plates. Both the subject and predicate devices are intended to organize, enclose, sterilize, transport and store medical devices and other instrumentation between surgical and other medical uses.
#### VII. Performance Data
Kit Boxes are intended for storing and organizing instruments used during implant/ prosthetic/retrieval treatment with NobelBiocare implants. The following performance data is being provided or relied upon in support of the substantial equivalence determination. Additionally, due to the intended use of the subject device, the Kit Boxes and Kit Plates are not exposed to mechanical stress. No additional performance testing is required.
### Sterilization
The Kit Boxes are supplied non-sterilized at the health care facility by the enduser. The testing validation confirmed the parameters for temperature, exposure time, dry time and cool time.
### Shelf-Life
The Kit Boxes are reusable, non-sterile devices, there is no expiration date identified.
#### Biocompatibility Testing
Kit Boxes a are considered accessories for medical devices, which do not come in to direct contact with the human body; however, they are intended for storing and subsequently sterilizing the respective medical devices and should be evaluated according to AAMI/ANSI/ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
Surgical and restorative kit boxes pose a potential risk of contaminating the respective biocompatible medical devices with leachable substances released during the sterilization process. Therefore, Cytotoxicity testing in accordance with AAMI/ANSI/ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity was performed and confirmed that no toxic leachable substances in a water-based environment were transferred to the medical device.
# VIII. Conclusion
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, subject Kit Boxes and Kit Plates are substantially equivalent to, and are as safe and as effective as, the legally marketed predicate device, cleared under K152241 under regulation 21 CFR 880.6850, product code KCT.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.