K131500 · Aesculap, Inc. · HCH · Jan 24, 2014 · Neurology
Device Facts
Record ID
K131500
Device Name
YASARGIL ANEURYSM CLIPS AND CLIP APPLIERS
Applicant
Aesculap, Inc.
Product Code
HCH · Neurology
Decision Date
Jan 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Yasargil Aneurysm Clips are intended for occlusion of cerebral aneurysms in either a temporary or permanent manner. They are applied with Aesculap clip appliers, which contain titanium alloy or phynox jaws.
Device Story
Yasargil Aneurysm Clips and Clip Appliers are surgical instruments used by neurosurgeons for temporary or permanent occlusion of cerebral aneurysms. Clips are manufactured from titanium alloy or phynox (cobalt alloy); appliers consist of stainless steel bodies with titanium or phynox jaws. The device is used in a clinical/OR setting. The surgeon manually selects and applies the appropriate clip size (3-40 mm) and style (straight, curved, angled, bayonet, T-bar, fenestrated) using the applier to occlude the aneurysm. The device is MR Conditional in 3-Tesla MRI environments. Benefits include secure vessel occlusion during neurosurgery.
Clinical Evidence
Bench testing only. Biomechanical testing performed per ISO 9713, including closing force measurements and tensile testing of the blade-to-spring bond. MRI compatibility testing conducted per ASTM F2119, F2182, F2213, and F2052. Results demonstrate the device is MR Conditional in 3-Tesla environments and meets performance requirements.
Technological Characteristics
Materials: Titanium alloy (ISO 5832/3) or Phynox cobalt alloy (ISO 5832/7). Clip sizes: 3-40 mm. Applier lengths: 50 mm, 90 mm, 110 mm. Sensing/Actuation: Mechanical spring-loaded clip closure. Connectivity: None. Sterilization: Not specified. MR Conditional (3-Tesla) per ASTM F2503.
Indications for Use
Indicated for patients requiring temporary or permanent occlusion of cerebral aneurysms during neurosurgical procedures.
Regulatory Classification
Identification
An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.
{0}------------------------------------------------
## 510(k) SUMMARY (as required by 21 CFR 807.92) Yasargil Aneruysm Clips and Clip Appliers
# K131500
# JAN 2 4 2014
## COMPANY:
Aesculapp, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 2916714
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
|--------------------|----------------------------------------------------------------------------------------------|
| TRADE NAME: | Aesculap Yasargil Aneurysm Clips and Clip Appliers |
| COMMON NAME: | Aneurysm Clip and Clip Applier |
| REGULATION NUMBER: | 882.5200- Clip, Aneurysm<br>882.4175- Applier, Aneurysm Clip |
| PRODUCT CODE: | HCH, HCI |
| REVIEW PANEL: | Neurology |
### SUBSTANTIAL EQUIVALENCE
Aesculap, Inc. believes that the Aesculap Yasargil Aneurysm Clips and Clip Appliers are substantially to Aesculap's current Yasargil Aneurvsm Clips and Clip Appliers (K043041, K003519. K002871. K984109. K983758. K970050. K922272. K913765. K833652. K833651 and K833650).
### DEVICE DESCRIPTION
The Yasargil Aneurysm Clips are designed for temporary or permanent occlusion of vessels during neurosurgical procedures. Thev are manufactured from either titanium alloy according to ISO 5832/3 or phynox (cobalt alloy) per ISO 5832/7. The clips range in size from 3 mm to 40 mm and are available in straight, curved, angled, bayonet. T-bar, offset T-bar, concave T-bar, fenestrated, and non-fenestrated styles.
The Yasargil Clip Appliers are manufactured from stainless steel (body) with either a titanium allov or phynox jaw. The appliers are available in short (50mm), standard (90 mm) and long (110 mm) lengths, as well as straight, angled and bayonet shapes.
### INDICATIONS FOR USE
The Yasargil Aneurysm Clips are intended for occlusion of cerebral aneurysms in either a temporary or permanent manner. They are applied with Aesculap clip appliers, which contain titanium alloy or phynox jaws.
{1}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The Yasargil Aneruysm Clip Appliers are substantially equivalent to the predicate Aesculap Yasargil Aneruysm Clips and Clip Appliers. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate devices through comparison in design, principles of operation, intended use, and materials. The Yasargil Aneruysm Clips and Clip Appliers device characteristics are summarized below.
| | New device | Predicate device Yasargil |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Yasargil Aneruysm Clips and<br>Clip Appliers | Aneruysm Clips and Clip<br>Appliers<br>K043041nc003519/K002871<br>K984109/K983758nc970050<br>K940970/K922272nc913765<br>K833652/K833651nc833650 |
| Indications | The Yasargil Aneurysm Clips<br>are intended for occlusion of<br>cerebral aneurysms in either a<br>temporary or permanent<br>manner. They are applied<br>with Aesculap clip appliers,<br>which contain titanium alloy<br>or phynox jaws. | The Yasargil Aneurysm Clips<br>are intended for occlusion of<br>cerebral aneurysms in either a<br>temporary or permanent<br>manner. They are applied with<br>Aesculap clip appliers, which<br>contain titanium alloy or<br>phynox jaws. |
| Aneurysm Clip: | | |
| Material | Titanium Alloy (Ti6Al6V) or<br>Phynox (CoCr) | Titanium Alloy (Ti6Al6V) or<br>Phynox (CoCr) |
| Sizes | 3-40 mm (blade length) | 2.8-40 mm (blade length) |
| Design | Fenestrated and non-<br>fenestrated | Fenestrated and non-<br>fenestrated |
| Blade Style | Straight, curved, offset T-bar,<br>convex T-bar, concave T-bar,<br>or angled | Straight, curved, bayonet, T-<br>bar or angled |
| Clip Applier: | | |
| Jaw Material | Titanium Alloy (Ti6Al6V) or<br>Phynox (CoCr) | Titanium Alloy (Ti6Al6V) or<br>Phynox (CoCr) |
| Body Material | Stainless Steel | Stainless Steel |
| Style | Straight, angled, bayonet | Straight, angled |
| Length | 50 mm, 90mm, and 110 mm | 50 mm, 90mm, and 110 mm |
{2}------------------------------------------------
Page 3 of 3
### PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s)" was completed for Aesculap Yasargil Aneruysm Clips and Clip Appliers. Biomechanical testing results demonstrate the Aesculap Aneruysm Clips and Clip Appliers are substantially equivalent when compared to our aneurvsm clips and clip appliers currently on the market. Preclinical testing was performed to demonstrate that the Yasargil Aneruvsm Clips and Clip Appliers performs as intended and is safe, as effective, and performs as well as the predicate devices. Testing was conducted in accordance with ISO 9713. The titanium clip closing force tolerance was within 7.5% of the nominal closing force. The Phynox clip closing force tolerance deviated from the standard, as a result, the Phynox clip closing force tolerance was outside the 7.5% nominal closing force. Tensile testing was conducted to validate the T-bar clip strength of the bond between the blade and the spring wire. The Phynox clips withstand 400N and the titanium clips withstand 210N. The tensile test failure is at the closing loop which is the thinnest point of the clip. Tensile testing was performed to simulate the applied force to the weld. The weld withstands 30N of force. In addition testing was performed according to the following MRI standards:
- . ASTM F2119 Evaluation ofMR Image Artifacts
- . ASTM F2182 Measurement of Radio Frequency Induced Heating During Magnetic Resonance Imaging
- . ASTM F2213 Qualitative Measurement of Magnetically Induced Torque in the Magnetic Resonance Environment
- ASTM F2052 Measurement of Magnetically Induced Displacement Force on the in . the Magnetic Resonance Environment
The results and evaluation conclude that the device is MR Conditional in 3-Tesla Magnetic . Resonance Imaging (MRI) systems according to ASTM F 2503 and is substantially equivalent to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 24, 2014
"Aesculap", Inc. c/o Ms. Kathy A. Racosky 3773 Corporate Parkway Center Valley, PA 18034
Re: K131500
Aesculap Yasargil Aneurysm Clips and Clip Appliers Regulation Number: 21 CFR 882.5200 Regulation Name: Aneurysm Clip Regulatory Class: II Product Code: HCH, HCI Dated: December 12, 2013 Received: December 13, 2013
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
## Page 2 - Ms. Kathy A. Racosky
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/Medica}Devices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
# Carlos L. Pena -S
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
## 510(k) Number (if known) K131500
### Device Name
Yasargil Aneurysm Clips and Clip Appliers
#### Indications for Use (Describe)
The Yasargil Aneurysm Clips are intended for occlusion of cercbral aneurysms in either a temporary or permanent manner. They are applied with Acsculap clip appliers, which contain titanium alloy or phynox jaws.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Carlos L. Pena -
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.