K180641 · GE Healthcare · LLZ · May 4, 2018 · Radiology
Device Facts
Record ID
K180641
Device Name
Invenia ABUS Viewer
Applicant
GE Healthcare
Product Code
LLZ · Radiology
Decision Date
May 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Image display of Automated Breast Ultrasound (ABUS), as well as other multi-modality DICOM medical images for diagnostic and non-diagnostic review purposes. The device will be used by trained healthcare professionals.
Device Story
Invenia ABUS Viewer is a software-based medical imaging display system; receives, transmits, and displays DICOM medical images from Invenia ABUS Scan Station and other ultrasound/imaging acquisition devices. Used by trained healthcare professionals in clinical settings for diagnostic and non-diagnostic review. Enables image review, reporting, and measurement of breast ultrasound studies. Facilitates clinical decision-making by providing visualization of multi-modality images; supports workflow efficiency in radiology environments.
Clinical Evidence
No clinical data required; substantial equivalence supported by non-clinical bench testing, including risk analysis, unit-level module verification, system integration testing, and final acceptance/performance validation.
Technological Characteristics
Software-based image display system; utilizes DICOM (NEMA PS 3.1-3.20) for connectivity and image handling. Complies with ISO 14971 (risk management) and IEC 62304 (software life cycle). Operates on a standard platform for multi-modality image review, measurement, and reporting.
Indications for Use
Indicated for trained healthcare professionals to display Automated Breast Ultrasound (ABUS) and other multi-modality DICOM medical images for diagnostic and non-diagnostic review.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive WAUWATOSA WI 53226
May 4, 2018
Re: K180641
Trade/Device Name: Invenia ABUS Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 8, 2018 Received: March 12, 2018
Dear Bryan Behn:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters, giving it a classic and recognizable appearance.
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K180641
Device Name Invenia ABUS Viewer
#### Indications for Use (Describe)
Image Display of Automated Breast Ultrasound (ABUS), as well as other multi-modality DICOM medical images for diagnostic and non-diagnostic review purposes. The device will be used by trained healthcare professionals.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular frame. The background of the circle is filled with a solid color, while the letters and the frame are in a contrasting color, creating a visually distinct and recognizable emblem.
# GE Healthcare 510(k) Premarket Notification Submission
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: March 8, 2018 GE Healthcare Submitter: 9900 Innovation Drive Wauwatosa, WI 53226
| Primary Contact Person: | Bryan Behn<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)247-5502<br>F:(414)918-8275 |
|---------------------------|--------------------------------------------------------------------------------------------------|
| Alternate Contact Person: | Elizabeth Gemmel<br>T: (262)788-3816 |
| Device: | Trade Name: Invenia ABUS Viewer |
| Common/Usual Name: | Invenia ABUS Viewer |
| Classification Names: | Class II |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code: | Picture archiving and communication system. 21 CFR 892.2050 LLZ |
| Primary Predicate Device: | K151075 BR-ABVS Viewer 1.0 |
| Reference Predicate Device(s): | K150420 Centricity Universal Viewer |
| Device Description: | The Invenia ABUS Viewer is intended for the image display of breast ultrasound images. The Invenia ABUS Viewer can received, transmit, and display medical images and associated patient medical information from the Invenia ABUS Scan Station as well as from other ultrasound or image acquisition devices. |
| Indications for Use: | Image display of Automated Breast Ultrasound (ABUS), as well as other multi-modality DICOM medical images for diagnostic and non-diagnostic review purposes. The device will be used by trained healthcare professionals. |
Invenia ABUS Viewer employs the same fundamental scientificInvenia ABUS Viewer employs the same fundamental scientific technology as its predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and there are decorative flourishes or swirls around the letters within the circle.
### Technology:
Determination of Comparison to Predicates Substantial Equivalence:
The proposed Invenia ABUS Viewer is a new platform substantially equivalent to the predicate devices. The following is an overview of the similarities between the proposed Invenia ABUS Viewer and the predicate BR-ABVS Viewer 1.0 (K151075). The systems are all intended for picture archiving and communication systems.
- The Invenia ABUS Viewer and predicate BR-ABVS . Viewer 1.0 software have the same clinical applications.
- Both software systems operate/devices on the same ● operating platform.
- Invenia ABUS Viewer and predicate have the same image ● processing techniques.
- The software systems/devices process the same type of ● images.
- The Invenia ABUS Viewer and predicate BR-ABVS ● software systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The Invenia and predicate operate with the same main ● interfaces of connectivity and archiving
The reference predicate, Centricity Universal Viewer, and Invenia ABUS Viewer are also similar in that they display medical images from various sources and utilize DICOM.
Summary of Non-Clinical Tests:
The Invenia ABUS Viewer and its applications comply with voluntary standards:
- ISO14971, Application of risk management to medical devices: Second edition 2007
- NEMA PS 3.1 3.20 (2011), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)
- IEC 62304, Medical device software Software life cycle ● processes
The following quality assurance measures are applied to the development of the system:
- Risk Analysis ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script. The letters are enclosed within a circular shape, and there are decorative swirls or flourishes around the outer edge of the circle.
# GE Healthcare 510(k) Premarket Notification Submission
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification)
- Final Acceptance Testing (Validation) ●
- Performance testing (Verification) ●
- Safety testing (Verification)
### Summary of Clinical Tests:
The subject of this premarket submission, Invenia ABUS Viewer, did not require clinical studies to support substantial equivalence.
- GE Healthcare considers the Invenia ABUS Viewer to be as safe, Conclusion: as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.