K151075 · Taihao Medical, Inc. · LLZ · Jan 15, 2016 · Radiology
Device Facts
Record ID
K151075
Device Name
BR-ABVS Viewer 1.0
Applicant
Taihao Medical, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS. Caution: Federal law restricts this device to sale by or on the order of a physician.
Device Story
BR-ABVS Viewer 1.0 is a standalone software application for Windows-based computers; it processes 3-D breast ultrasound image volumes (DICOM format) acquired from the Siemens ACUSON S2000 ABVS. The device enables physicians to visualize images in any orientation (axial, sagittal, coronal) based on anatomical coordinate systems and DICOM tag spacing. It includes functionality to annotate markers and automatically generate reports containing sub-image and location information. Used in clinical settings by physicians, the device serves as an adjunct to mammography, aiding in the review and analysis of breast ultrasound data to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical trials performed. Validation consisted of a performance study comparing the subject device to the predicate (ABVS Workplace) using clinical 3-D breast ultrasound images acquired from a Siemens ACUSON S2000 ABVS. The study validated 3-D image loading, visualization across axial, sagittal, and coronal planes, and adherence to anatomical coordinate systems defined by DICOM tags.
Technological Characteristics
Standalone software for Windows-based PCs. Processes DICOM 3-D ultrasound volumes. Features include multi-planar reconstruction (axial, sagittal, coronal), marker annotation, and automated report generation. No specific hardware requirements beyond standard PC/review station. Software development follows documented design, verification, and validation processes with risk assessment controls.
Indications for Use
Indicated for physicians to assist in the review and analysis of 3-D breast ultrasound images generated by the Siemens ACUSON S2000 Automated Breast Volume Scanner (ABVS) as an adjunct to mammography screening.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Siemens ACUSON S2000 Automated Breast Volume Scanner (K081148)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the logo is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
TaiHao Medical, Inc. % Chiu S. Lin, Ph.D. President LIN & ASSOCIATES, LLC 9223 Cambridge Manor Court POTOMAC MD 20854
Re: K151075
Trade/Device Name: BR-ABVS Viewer 1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 15, 2015 Received: December 15, 2015
Dear Dr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151075
Device Name BR-ABVS Viewer 1.0
#### Indications for Use (Describe)
BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS. Caution: Federal law restricts this device to sale by or on the order of a physician.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary of Safety and Effectiveness Information
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92
## 5.1. Identification of Submitter:
Submitter: TaiHao Medical Inc.
Address: 7F., No.410, Sec. 5, Zhongxiao E. Rd., Xinyi Dist., Taipei City 110, Taiwan (R.O.C.)
Phone: 886-2-2858-2357
Contact: HSIN HUNG (Simon) LAI
Title: President
Phone: 886-2-2858-2357
Email: simonlai@taihaomed.com
Manufacturer: TaiHao Medical Inc.
US Agent and Contact: Chiu S. Lin, Ph.D.
LIN & ASSOCIATES, LLC
Address: 9223 Cambridge Manor Court
Potomac, MD 20854 U.S.A.
Phone: (0) 301-591-3895
Email: cslin@lin-associates.com
Date prepared: April 8, 2015
## 5.2. Identification of Product
Device Trade Name: BR-ABVS Viewer 1.0
Common and Usual Name: Ultrasound Image Display Device
Device Classification Name: Picture Archiving and Communication System
Regulation Number: 21 CFR 892.2050
{4}------------------------------------------------
### Classification Product Code: LLZ
Classification: Class II Radiology Devices Classification Panel:
Manufacturer: TaiHao Medical Inc.
### 5.3. Predicate Device
This subject software medical device is substantially equivalent to the devices listed below:
Model: ABVS Workplace
Manufacturer: Siemens Medical Solutions
510(k) Number: K092067, cleared on September 18, 2009
### 5.4. Device Description
BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software also automatically generates reports to provide the sub-image and location information of markers annotated during the image review.
#### 5.5. Indications for Use
BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS.
#### 5.6. Comparison with Predicate Devices
BR-ABVS Viewer 1.0 is substantially equivalent to ABVS Workplace with a general intended use for viewing and analyzing ultrasound image data to physicians. Minor technological characteristics differences do not raise any new questions of safety and effectiveness.
The comparison table between our device and the predicate devices is provided below:
| | | <b>BR-ABVS Viewer 1.0</b> | <b>ABVS Workplace</b> |
|--|--|---------------------------|-----------------------|
|--|--|---------------------------|-----------------------|
{5}------------------------------------------------
| Manufacturer | TaiHao Medical Inc. | Siemens Medical Solutions |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K151075 | K092067 |
| Device Common Name | Picture archiving and<br>communications system | Picture archiving and<br>communications system |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Regulation Name | Picture archiving and<br>communications system | Picture archiving and<br>communications system |
| Product Code | LLZ | LLZ |
| Indications for Use | BR-ABVS Viewer 1.0 is<br>intended as a standalone<br>software device installed on a<br>standalone windows-based<br>computer to assist the<br>physician to visualize any<br>orientation of three-<br>dimensional (3-D) breast<br>ultrasound images generated<br>by Siemens ACUSON S2000<br>Automated Breast Volume<br>Scanner, ABVS (cleared in<br>K081148). The device is<br>indicated for use to assist the<br>physicians in their review and<br>analysis of the 3-D breast<br>ultrasound images generated<br>by ABVS. | ABVS Workplace is intended<br>to display ultrasound images<br>of the breast acquired from B-<br>mode imaging using an<br>automatic or handheld<br>scanning linear transducer.<br>The images may be reviewed<br>and analyzed by the<br>physician. The ABVS<br>Workplace is indicated for use<br>as an adjunct to<br>mammography. The ABVS<br>Workplace is not intended to<br>be used as a replacement for<br>screening. |
| Functional Capability of<br>Image Processing | BR-ABVS Viewer 1.0<br>provides visualization of any<br>orientation of 3-D image. | The ABVS Workplace applies<br>post-processing algorithms<br>based on the nipple location.<br>A reverberation removal<br>algorithm determines tissue<br>contact areas. This suppresses<br>reverberation artifacts from<br>the non-contact area. A<br>proprietary adaptive nipple<br>shadow reduction tool<br>analyzes the data volume and<br>enhances structures in the<br>retroareolar area to improve<br>visualization of this typically<br>challenging anatomical<br>region. |
| Software Design | Image processing techniques | Same |
| Platform | Window-based | Same |
| Operating System | Standard PC or review station | Own workstation, which<br>includes IT hardware and pre-<br>installed software |
| Clinical Application | As an adjunct to<br>mammography screening | Same |
| Image Type to Be Processed<br>by The Device | Automated breast ultrasound<br>images generated by Siemens<br>ACUSON S2000 Automated<br>Breast Volume Scanner,<br>ABVS (cleared in K081148). | Same |
| Image Format | DICOM images acquired on<br>Siemens ACUSON S2000<br>Automated Breast Volume | Same |
| | Scanner, ABVS | |
| Automatically Generating<br>Report | Yes | Yes |
| Performance Testing to<br>Support SE Determination | Results from Software<br>Validation Report of BR-<br>ABVS Viewer 1.0 | From the 510(k) Summary<br>that is available on the FDA<br>database, it appears that no<br>data from performance testing<br>were submitted. |
{6}------------------------------------------------
{7}------------------------------------------------
#### 5.7. Performance Standards
No applicable FDA performance standards have been issued.
#### 5.8. Software
Software development for BR-ABVS Viewer 1.0 follows documented processes for software design, verification and validation testing. A risk assessment has been completed to identify potential design hazards. Appropriate steps have been taken to control all identified risks for this type of image viewing and quantification device.
#### 5.9. Summary of Performance Data to Support Substantial Equivalence
BR-ABVS Viewer 1.0 displays the 3-D image volume by axial, sagittal, and coronal plane according to the anatomical coordinate system to provide an overall observation. The actual image size is obtained by considering the spacings of three axes specified in the standard DICOM tags. TaiHao Medical Inc. has conducted a performance study to validate and assess the performance of BR-ABVS Viewer 1.0 for its-intended use. An actual clinical image generated by a Siemens ACUSON S2000 Automated Breast Volume Scanner in 2014 was used to do the comparison testing between BR-ABVS Viewer 1.0 and the predicate (ABVS Workplace, K092067) in terms of substantial equivalence in 3-D image loading.
#### 5.10. Conclusions
The intended use, technological characteristics, and major functionality of BR-ABVS Viewer 1.0 are similar to the predicate device and no new issues of safety or effectiveness are introduced by using this device. The performance data generated, as described, demonstrates that our software device is as
{8}------------------------------------------------
safe and effective, as compared to the predicate. Therefore we believe BR-ABVS Viewer 1.0 is Substantially Equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.