K180640 · Elucent Medical, Inc. · NEU · Jun 4, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180640
Device Name
SmartClip Soft Tissue Marker
Applicant
Elucent Medical, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Jun 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The implanted SmartClip™ is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
Device Story
SmartClip is a sterile, single-use, permanently implantable soft tissue marker; preloaded in a 15ga stainless steel introducer needle delivery system. Marker consists of a Parylene C coated glass vial; retained in needle by biocompatible bone wax plug. Physician deploys marker into soft tissue (e.g., lung, GI, breast) via stylet-lock mechanism. Marker provides permanent radiographic, mammographic, ultrasound, and MRI visibility. Device facilitates localization of tissue sites for subsequent medical procedures. No software or algorithm involved.
Clinical Evidence
Bench testing only. Testing included simulated use, mechanical integrity, imaging assessment (X-ray, mammography, ultrasound, MRI), MR compatibility, biocompatibility, packaging, and sterilization validation. No clinical data presented.
Technological Characteristics
Permanent soft tissue marker; Parylene C coated glass vial. Delivery system: 15ga stainless steel needle (10.8 cm) with depth reference marks. Sterilization: Ethylene oxide (SAL 10^-6). Imaging compatibility: X-ray, mammography, ultrasound, MRI. Biocompatible materials. No software or electronic components.
Indications for Use
Indicated for radiographic marking of soft tissue sites for future medical procedures in patients requiring tissue localization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2018
Elucent Medical, Inc. Ms. Debra Kridner VP Regulatory and Quality 7480 Flying Cloud Drive, Suite 101 Eden Prairie, Minnesota 55344
Re: K180640
Trade/Device Name: SmartClip Soft Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: May 21, 2018 Received: May 22, 2018
Dear Ms. Kridner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180640
Device Name
SmartClip™ Soft Tissue Marker
Indications for Use (Describe)
The implanted SmartClip™ is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92
| Date Prepared: | May 18, 2018 |
|-----------------------------|---------------------------------------------------------------------------------------------|
| Applicant: | Elucent Medical, Inc.<br>7480 Flying Cloud Drive, Suite 110<br>Eden Prairie, MN 55344 USA |
| Official Correspondent: | Debra Kridner<br>Tel: (844) 417 1700 / Fax: (952) 314-7105<br>debra.kridner@elucent.com |
| Trade Name: | SmartClip™ Soft Tissue Marker |
| Common Name: | Implanted, Radiographic, Tissue Marker |
| Classification Number: | 21 CFR 878.4300 |
| Classification Name: | Implantable Clip |
| Device Class: | Class II |
| Product Code: | NEU |
| Review Panel: | General and Plastic Surgery |
| Predicate Device (primary): | Focal Therapeutics BioZorb Marker - K143484/K152070 |
| Reference Device: | Cianna Medical Permanent Tissue Marker - K132463 |
## Indications for Use for the Device Subject to this 510(k) Notification:
The SmartClip™ is indicated for radiographic marking of sites in addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.
## Device Description for the Device Subject to this 510(k) Notification:
The SmartClip™ is an ethylene oxide sterile, single use device composed of a soft tissue marker preloaded in a Delivery System. The marker is intended to be placed within soft tissue. The marker is visible using radiography (including mammographic imaging), ultrasound and MRI.
Image /page/3/Figure/8 description: This image shows a medical device with labels pointing to its different parts. The introducer needle length is labeled as 10.8 cm. Other labeled parts include the unlock button, deployment plunger, stylet, bone wax (3-4 mm), and SmartClip (8 mm x 1.4 mm).
Image /page/3/Figure/9 description: The image shows the word "SmartClip" with the trademark symbol next to it. The text is in a serif font and is black. The background is white.
#### Marker:
{4}------------------------------------------------
#### K180640 pg. 2 of 3
A sterile, single use device permanently implanted marker is approximately 8 mm wide. The marker is placed into the barrel of the introducer and maintained in place prior to insertion by a biocompatible bone wax plug. The marker can be implanted into various types of soft tissue (e.g., lung, gastrointestinal system, and breast) and subsequently be detected by means of radiography (including mammographic imaging), ultrasound and MRI.
## Delivery System:
The Delivery System consists of a stylet, a stylet-lock to prevent and 15ga introducer needle. The stainless steel needle is approximately 10.8 cm long with 1cm depth reference marks. The marker is preloaded inside the needle and retained by a bone wax plug. When the stylet is completely depressed the marker and bone wax plug, are deployed from the end of the needle.
| Comparison of<br>Subject Device<br>to Predicate<br>Device | Subject Device<br>(under review) | Primary Predicate Device<br>Focal Therapeutics BioZorb Marker<br>(K143484/K152070) | Reference Device:<br>Cianna Medical Permanent<br>Tissue Marker<br>(K132463) |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The SmartClip™ is indicated for<br>radiographic marking of sites in<br>soft tissue. In addition, the<br>Marker is indicated in situations<br>where the soft tissue site needs to<br>be marked for future medical<br>procedures. | The BioZorb LP Marker is indicated<br>for radiographic marking of sites in<br>soft tissue. In addition, the Marker is<br>indicated in situations where the soft<br>tissue site needs to be marked for<br>future medical procedures | The Cianna Medical<br>Permanent Tissue Marker is<br>intended to mark tissue<br>during a percutaneous<br>breast biopsy procedure and<br>be permanently visible by<br>radiography and visible for<br>up to thirty days by<br>ultrasound. |
| Intended Use | Implanted Soft Tissue Marker | Implanted Soft Tissue Marker | Implanted Soft Tissue -<br>"Breast" Marker |
| Anatomical Site | Soft Tissue | Soft Tissue | Soft Tissue - Breast |
#### Indications for Use Comparison:
# Technological and Performance Characteristics Comparison:
| Comparison of<br>Subject Device to<br>Predicate Device | Subject Device<br>(under review) | Primary Predicate Device<br>Focal Therapeutics BioZorb<br>Marker<br>(K143484/K152070) | Predicate Device:<br>Cianna Medical Permanent<br>Tissue Marker<br>(K132463) |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Overall Technological<br>Characteristics | Radiographically visible<br>permanent marker | Radiographically visible<br>permanent marker | Radiographically visible<br>permanent marker |
| Principle of Operation | Marker is positioned into<br>tissue site for radiographic<br>visualization of tissue site | Marker is positioned into tissue<br>site for radiographic<br>visualization of tissue site | Marker is positioned into<br>tissue site for radiographic<br>visualization of tissue site |
| Visualization<br>Compatibility | X-Ray, Mammography<br>Ultrasound, MR | X-Ray, Mammography<br>Ultrasound, CT, MR | Radiographic and<br>Ultrasound (30 days) |
| Deployment Method | Introducer | Manual, open surgical | Introducer |
| Implant Duration | Permanent | Permanent | Permanent |
| Marker Materials | Parylene C coated glass vial | Titanium (marker clip),<br>bioabsorbable polymer (spacer) | Titanium beads and Nitinol |
| Sterility | Terminal sterilization by<br>ethylene oxide, sterility<br>assurance level 10-6 | Terminal sterilization by<br>radiation, sterility assurance<br>level 10-6 | Terminal sterilization by<br>ethylene oxide, sterility<br>assurance level 10-6 |
| Biocompatibility | Passed | Passed | Passed |
# Performance Testing:
{5}------------------------------------------------
All necessary testing has been performed for the SmartClip™ to assure substantial equivalence to the predicate devices and demonstrate the device performs as intended.
#### The device design was qualified through the following tests:
- Simulated Use .
- . Mechanical Integrity
- . Imaging Assessment
- MR Compatibility
- Biocompatibility
- Packaging
- Sterilization ●
#### Conclusion:
The SmartClip™ met all specified criteria and did not raise new safety or performance questions.
#### Basis for Determination of Substantial Equivalence:
The Indications for Use, Intended Use, Fundamental Scientific Technology and Principles of Operation for the SmartClip™ are the same as those described for the predicate device. The SmartClip™ is found to have a safety and effectiveness profile that is the same as the predicate device and is determined by Elucent Medical to be substantially equivalent.
In summary, the SmartClip™ has the following similarties to the predicate device which have previously received 510(k) clearance:
- Has the same indications for use
- Has the same intended use
- Has the same implant duration
- Used in the same anatomical site
- Uses similar technological characteristics
- Uses the same principles of operation
- Uses biocompatible materials
- Has the same sterility assurance level
Therefore, the conclusions drawn from the and clinicaltests (biocompatibility) demonstrate that the devices as safe, as effective, and performs as well as the beatly marketed device predicate. The SmartClip™ is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.