BioZorb LP Marker

K152070 · Focal Therapeutics · NEU · Aug 26, 2015 · General, Plastic Surgery

Device Facts

Record IDK152070
Device NameBioZorb LP Marker
ApplicantFocal Therapeutics
Product CodeNEU · General, Plastic Surgery
Decision DateAug 26, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4300
Device ClassClass 2

Indications for Use

The BioZorb LP Marker is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures.

Device Story

BioZorb LP Marker is an implantable radiopaque device used to mark soft tissue sites for future visualization. It consists of a permanent titanium marker component and a bioabsorbable polylactic acid (PLA) spacer. The device is implanted manually during open surgical procedures and sutured in place. Once implanted, it provides a radiographically visible reference point for imaging modalities including mammography, ultrasound, X-ray, CT, and MR. By marking the site, the device facilitates accurate localization for subsequent medical procedures. The bioabsorbable component resorbs over one year or more, leaving the permanent marker in place. The device is provided sterile for single use.

Clinical Evidence

Bench testing only. Performance data included simulated use, mechanical integrity, imaging assessment, and MR compatibility testing to demonstrate substantial equivalence to the predicate device.

Technological Characteristics

Implantable radiopaque marker. Materials: Titanium (marker clip) and bioabsorbable polylactic acid (PLA) spacer. Form factor: Low profile (circular/oval). Principle: Radiographic visualization via permanent marker element in bioabsorbable spacer. Sterilization: Radiation. Single-use, sterile.

Indications for Use

Indicated for radiographic marking of soft tissue sites, including those requiring marking for future medical procedures, in patients undergoing soft tissue interventions.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name encircling a symbol of a stylized human figure in profile. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the circular seal. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 26, 2015 Focal Therapeutics Mr. George Hermann President 4370 Alpine Road #101 Portola Valley, California 94028 Re: K152070 Trade/Device Name: BioZorb LP Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: July 24, 2015 Received: July 27, 2015 Dear Mr. Hermann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {1}------------------------------------------------ adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152070 Device Name BioZorb LP Marker | Indications for Use (Describe) | The BioZorb LP Marker is indicated for radiographic marking of sites in soft tissue. | |--------------------------------|--------------------------------------------------------------------------------------| |--------------------------------|--------------------------------------------------------------------------------------| In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | <div> <span> <span style="font-size: 12pt;"> <span style="font-family: Symbol;"> <span style="color: black;">☑</span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | | <div> <span> <span style="font-size: 12pt;"> <span style="font-family: Symbol;"> <span style="color: black;">☐</span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 1. 510(k) Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT: | Focal Therapeutics | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | DATE PREPARED: | 08/14/15 | | CONTACT PERSON: | George Hermann<br>Focal Therapeutics<br>4370 Alpine Rd. #101<br>Portola Valley, CA 94028<br>Phone: 650.530.2394<br>Fax: 650.530.2397 | | TRADE NAME: | BioZorb™ LP Marker | | COMMON NAME: | Implantable Radiographic Marker | | CLASSIFICATION<br>NAME: | Implantable Clip, 21 CFR, 878.4300 | | DEVICE<br>CLASSIFICATION: | Class II | | PRODUCT CODE | NEU | PREDICATE DEVICES: (primary) BioZorb Marker (K143484) #### Substantially Equivalent To: The modified BioZorb Marker (BioZorb LP Marker) is substantially equivalent in intended use, principal of operation and technological characteristics to the BioZorb Marker cleared under premarket notification K143484. #### Description of the Device Subject to Premarket Notification: The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 year or more and a permanent component (titanium). The BioZorb LP Marker is provided sterile for single use and is implantable. ## Indication for Use: The BioZorb LP Marker is indicated for radiographic marking of sites in soft tissue. In addition, the Marker is indicated in situations where the soft tissue site needs to be marked for future medical procedures. {4}------------------------------------------------ ### Technical Characteristics: The modified BioZorb Marker has the same or similar physical and technical characteristics as the predicate device. Low profile markers have been added to the BioZorb product family. | | Subject device | Predicate device | |------------------------------------------|------------------------------|-----------------------------------------------------------------------------------------| | | BioZorb Marker,<br>Modified | BioZorb Marker<br>(K143484) | | Overall Technological<br>Characteristics | SAME | Radiographically visible permanent marker<br>element(s) in bioabsorbable polymer spacer | | Principle of Operation | SAME | Marker is positioned into tissue site for<br>radiographic visualization of tissue site | | Visualization<br>Compatibility | SAME | Mammography<br>Ultrasound<br>X-Ray<br>CT<br>MR | | Device Shape | Low Profile (circular, oval) | Spiral | | Materials of<br>Construction | SAME | Titanium (marker clip),<br>bioabsorbable polymer (spacer) | | Overall Device Length | SAME | 2-5 cm | | Typical Anatomical<br>Treatment Site | SAME | Soft tissue including breast | | Method of Marker<br>Deployment | SAME | Manual, open surgical | | Marker Stability | SAME | Sutured in place | | Provided sterile | SAME | Yes | | Sterilization method | SAME | Radiation | #### Performance Data: All necessary testing has been performed for the modified BioZorb Marker to assure substantial equivalence to the predicate device and demonstrate the device performs as intended. The device design was qualified through the following tests: - Simulated Use - Mechanical Integrity - . Imaging Assessment - MR Compatibility {5}------------------------------------------------ The modified BioZorb Marker met all specified criteria and did not raise new safety or performance questions. ### Basis for Determination of Substantial Equivalence: The Indication/Intended Use and the fundamental scientific technology of the modified device have not been changed and are the same as those described in the unmodified predicate device. The modified BioZorb Marker device is found to have a safety and effectiveness profile that is the same as the predicate device and is determined by Focal Therapeutics to be substantially equivalent to the BioZorb Marker (K143484).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...