Venera 508 Deep Vein Thrombosis (DVT) Prevention System
K180389 · Suzhou Minhua Medical Apparatus Supplies Co., Ltd. · JOW · Jun 29, 2018 · Cardiovascular
Device Facts
Record ID
K180389
Device Name
Venera 508 Deep Vein Thrombosis (DVT) Prevention System
Applicant
Suzhou Minhua Medical Apparatus Supplies Co., Ltd.
Product Code
JOW · Cardiovascular
Decision Date
Jun 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This system can be applied in either home or clinical settings to: - . Aid in the prevention of DVT - . Enhance blood circulation - . Diminish post-operative pain and swelling - . Reduce wound healing time - . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs - . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Device Story
Portable pneumatic compression system; consists of pump and cuff assemblies. Operates by cyclically inflating/deflating cuffs to stimulate blood flow via simulated muscle contractions. Provides 50mmHg pressure followed by 50-second deflation. Used in home or clinical settings; operated by patients or clinicians. Output is mechanical compression of extremities; aids in DVT prevention, circulation enhancement, pain/swelling reduction, and wound healing. Benefits patient by reducing venous stasis and associated complications during periods of immobility.
Clinical Evidence
Bench testing only. No clinical or animal studies conducted. Performance verified via pressure accuracy, pressure switch, cycle time, alarm function, and battery life testing. Biocompatibility (cytotoxicity, sensitization, irritation) and electrical safety/EMC (IEC 60601 series) testing performed.
Indicated for patients requiring DVT prophylaxis or treatment of venous/arterial stasis, ulcers, chronic venous insufficiency, and lower limb edema. Suitable for home or clinical use by patients expecting long periods of stationarity.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
Cirona 6300 Deep Vein Thrombosis(DVT) Prevention System (K151189)
Submission Summary (Full Text)
{0}------------------------------------------------
June 29, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Suzhou Minhua Medical Apparatus Supplies Co. Ltd. Alex Wang Regulatory Affairs Specialist No.789, Wu Fang Road, Friendship Industrial Zone Song Ling Town, Wu Jiang Suzhou, 215222 China
Re: K180389
Trade/Device Name: Venera 508 Deep Vein Thrombosis (DVT) Prevention System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 15, 2018 Received: June 4, 2018
Dear Alex Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Nicole M. Gillette -S 2018.06.29 08:28:57 -04'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180389
#### Device Name
Venera 508 Deep Vein Thrombosis Prevention Therapy System
#### Indications for Use (Describe)
The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).
This system can be applied in either home or clinical settings to:
- . Aid in the prevention of DVT
- . Enhance blood circulation
- . Diminish post-operative pain and swelling
- . Reduce wound healing time
- . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
- . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510K Summary
## Submitter:
Suzhou Minhua Medical Apparatus Supplies Co. Ltd.
No.789, Wu Fang Road, Friendship Industrial Zone, Song Ling Town,
Wu Jiang District, Su Zhou City, Jiangsu Province, China
Postal Code: 215222
Tel: +86-0512-63091066
Contact Person: Alex Wang
Date Prepared: Jan 15, 2018
# Device:
| Common Names: | Intermittent Pneumatic Compression Device |
|-------------------|----------------------------------------------------|
| Proprietary Name: | Venera™ 508 Deep Vein Thrombosis Prevention System |
| Regulatory Class: | II |
| Product Code: | JOW |
## Predicate Devices:
The Venera 508 Deep Vein Thrombosis Prevention System is equivalent to the following:
| Predicate Device | Manufacturer | 510(k)# |
|------------------------------------------------------------|------------------------|---------|
| Cirona 6300 Deep Vein<br>Thrombosis(DVT) Prevention System | Devon Medical Products | K151189 |
# Device Description
The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by
{4}------------------------------------------------
simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.
# Indications for Use:
The Venera™ 508 Deep Vein Thrombosis Prevention System is a portable treatment system intended to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).
This system can be applied in either home or clinical settings to:
- Aid in the prevention of DVT .
- . Enhance blood circulation
- . Diminish post-operative pain and swelling
- . Reduce wound healing time
- . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
- As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
## Technological Characteristics:
Below is a table of comparison for the technological characteristics against the predicate device:
| Predicate | Cirona 6300 DVT System |
|------------------------|------------------------|
| Indication for Use | S |
| Components | S |
| Material | SE |
| Biocompatibility | SE |
| Principle of Operation | SE |
| Treatment parameters | S |
| User Interface | S |
*SE – Substantial Equivalent *S - Same
{5}------------------------------------------------
The manufacturer believes that the technological characteristics of the Venera™ 508 Deep Vein Thrombosis Prevention System are substantially similar to those of the predicate device. Venera 508 DVT Prevention System has similar components to its predicate device and identical principles of operation.
# Performance Tests
To verify that the device design met its function and performance requirements, samples of the device underwent function and mechanical testing.
The following tests were conducted:
| Function Performance Tests |
|-----------------------------------|
| Pressure Accuracy Test Report |
| Pressure Switch Test Report |
| Venera 508 Cycle Time Test Report |
| Alarm Function Test Report |
| Battery Life Test Report |
The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective.
## Biocompatibility
Biocompatibility test evaluation for Venera 508 DVT Prevention System is done in accordance to the FDA Good Laboratory Practice quidelines. The following tests were done:
- Cytotoxicity い
- Sensitization ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ
- Irritation い
## Sterilization and Shelf Life
Sterilization and shelf life are not applicable to Venera 508 DVT Prevention System.
#### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC tests were conducted according to the following standards:
- IEC 60601-1: 2005 + A1:2012 Medical Electrical Equipment Part 1: General requirements , for basic safety and essential performance
{6}------------------------------------------------
- IEC 60601-1-2: 2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General requirements ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ for safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
- IEC 60601-1-11: 2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
## Software Verification and Validation
Software verification and validation were conducted and documentation is provided. The software was considered as a Moderate level of concern, since a failure or latent flaw in the software could directly or indirectly result in minor injury to the patient or operator.
## Animal Study and Clinical Study
No animal study or clinical study was conducted.
# Statement of Substantial Equivalence
The Venera 508 Deep Vein Thrombosis Prevention System is substantially equivalent in its intended use, technology, function, operating parameters to predicate devices that are currently commercially available and in distribution, and does not raise any new concerns regarding safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.