Cirona 6300 Disposable Deep Vein Thrombosis (DVT) Prevention System
Applicant
Devon Medical Products
Product Code
JOW · Cardiovascular
Decision Date
Sep 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cirona 6300 disposable deep vein thrombosis prevention system is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This device can be used in the home or clinical settings to: - Aid in the prevention of DVT - Enhance blood circulation - Diminish post-operative pain and swelling - Reduce wound healing time - Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs - As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Device Story
Portable pneumatic compression system; consists of pump and sleeve assemblies. Operates by noninvasively inflating/deflating garment to simulate muscle contractions and stimulate blood flow in extremities. Provides 50mmHg pressure followed by 50-second deflation cycle. Used in home or clinical settings; operated by patient or clinician. Output is mechanical compression; assists in DVT prevention, circulation, and edema reduction. Benefits patient by reducing DVT risk and promoting healing.
Clinical Evidence
No clinical data or animal studies were conducted. Evidence consists of bench testing, including pressure accuracy, pressure switch, cycle time, alarm function, battery life, and garment burst testing. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per GLP.
Indicated for patients requiring DVT prophylaxis, enhancement of blood circulation, reduction of post-operative pain/swelling, wound healing, or treatment of venous/arterial/diabetic ulcers, stasis dermatitis, chronic venous insufficiency, and lower limb edema. Suitable for patients expecting long periods of immobility. Prescription required.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Devon Medical Products Steve Xu Regulatory Affairs Specialist 1100 First Avenue, Suite 202 King Of Prussia, Pennsylvania 19406
Re: K151189
> Trade/Device Name: Cirona 6300 Disposable Deep Vein Thrombosis (DVT) Prevention System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: August 7, 2015 Received: August 10, 2015
Dear Steve Xu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151189
Device Name
Cirona 6300 Disposable Deep Vein Thrombosis Prevention System
Indications for Use (Describe)
The Cirona 6300 disposable deep vein thrombosis prevention system is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).
This device can be used in the home or clinical settings to:
- Aid in the prevention of DVT ●
- . Enhance blood circulation
- Diminish post-operative pain and swelling .
- Reduce wound healing time .
- Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
- As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5
#### 510K Summary
#### Submitter:
Devon Medical Products
1100 First Avenue, Suite 202
King of Prussia, PA 19406
Phone: 610.757.4127
610.930.4035 Fax:
Contact Person: John Siegel, COO
Date Prepared: September 18, 2015
## Device:
Common Names: Intermittent Pneumatic Compression Device
Cirona™ 6300 Disposable Deep Vein Thrombosis Prevention System Proprietary Name:
Regulatory Class II
Product Code: JOW
### Predicate Devices:
The Cirona 6300 Deep Vein Thrombosis Prevention System is equivalent to the following:
| Predicate Device | Manufacturer | 510(k)# |
|------------------|----------------------|---------|
| Vena Pro | Innovamed Health LLC | K133274 |
### Device Description
Premarket notification device:
Cirona 6300 Disposable Deep Vein Thrombosis Prevention System
The Cirona™ 6300 disposable deep vein thrombosis prevention system (refer as Cirona 6300 below) is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by simulating blood flow in the extremities (simulating muscle contractions). The Cirona 6300 system consists of a pair of pump and sleeve assemblies.
{4}------------------------------------------------
# Section 5
The device will alternatively inflate and deflate the garment (sleeve) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.
# Intended Use:
The Cirona 6300 disposable deep vein thrombosis prevention system is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).
This device can be used in the home or clinical settings to:
- . Aid in the prevention of DVT
- Enhance blood circulation
- . Diminish post-operative pain and swelling
- Reduce wound healing time
Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg . ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
- As a prophylaxis for DVT by persons expecting to be stationary for long periods of time .
# Technological Characteristics:
Below is a table of comparison for the technological characteristics against the predicate device:
| Predicate | Innovamed Vena Pro |
|------------------------|--------------------|
| Indication for Use | S |
| Components | S |
| Material | SE |
| Biocompatibility | SE |
| Principle of Operation | SE |
| Pressure | S |
| User Interface | S |
*SE - Substantial Equivalent *S - Same
{5}------------------------------------------------
# Section 5
The manufacturer believes that the technological characteristics of the Cirona 6300 Disposable Deep Vein Thrombosis Prevention System are substantially similar to those of the predicate device. Cirona 6300 has very similar components to its predicate devices and very similar principles of operation.
# Performance Tests
To verify that the device design met its function and performance requirements, samples of the device underwent function and mechanical testing.
The following tests were conducted:
| Function Performance Tests | |
|----------------------------|------------------------------------------|
| TR36.0001 | Pressure Accuracy Test Report |
| TR36.0002 | Pressure Switch Test Report |
| TR36.0003 | Cirona 6300 Cycle Time Test Report |
| TR36.0004 | Alarm Function Test Report |
| TR36.0005 | Battery Life Test Report |
| TR36.0010 | Cirona 6300 Garment Burst Testing Report |
The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is substantially equivalent to the predicate.
### Biocompatibility
Biocompatibility test evaluation for Cirona 6300 is done in according to the FDA Good Laboratory Practice. The following tests were done:
- . Cytotoxicity
- . Sensitization
- Irritation
# Sterilization and Shelf Life
Sterilization and shelf life is not applicable to Cirona 6300.
{6}------------------------------------------------
# Electrical Safety and Electromagnetic Compatibility (EMC)
EMC tests were conducted according to the following standards:
- IEC 60601-1: 2005+C1:2006+C2:2007 Medical electrical equipment Part 1: General . requirements for basic safety and essential performance
- IEC 60601-1-2: 2007 Medical electrical equipment Part 1-2: General requirements for basic ● safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-1-6:2010 Medical electrical equipment Part 1-6: General requirements for safety -Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
- IEC 60601-1-11: 2010 Medical electrical equipment Part 1-11: General requirements for . basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
## Software Verification and Validation
Software verification and validation was conducted and documentation is provided. The software was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in serious injury to the patient or operator.
### Animal Study and Clinical Study
No animal study or clinical study was conducted.
### Statement of Substantial Equivalence
The Cirona 6300 Disposable Deep Vein Thrombosis Prevention System is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.