K173951 · Ortho Development Corporation · LPH · May 30, 2018 · Orthopedic
Device Facts
Record ID
K173951
Device Name
Legend Acetabular Liners
Applicant
Ortho Development Corporation
Product Code
LPH · Orthopedic
Decision Date
May 30, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of: Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis. Previously failed hip surgery. Fractures of the femoral neck or head. Avascular necrosis of the femoral head. Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to prosthesis.
Device Story
Legend Acetabular Liners are modular orthopedic implants used in total hip arthroplasty. Designed as a line extension to the previously cleared acetabular cup system (K103384), these liners replace the natural articular surface of the hip joint. The device is a hemispherical component made of highly cross-linked polyethylene, featuring positive locking, liner stabilization, and anti-rotation mechanisms. It is intended for uncemented, press-fit implantation by orthopedic surgeons in a clinical/hospital setting. The device functions as a bearing surface within the acetabular shell to facilitate joint articulation. Benefits include restoration of hip joint function for patients with degenerative or traumatic joint conditions. No software or algorithmic components are involved.
Clinical Evidence
No clinical studies were performed. Evidence is based on non-clinical bench testing, including impingement performance per ASTM F-2582 and bacterial endotoxin testing per ANSI/AAMI ST72:2011. All tests met predetermined acceptance criteria.
Indicated for patients undergoing total hip arthroplasty requiring uncemented, press-fit fixation due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, failed prior hip surgery, femoral neck/head fractures, avascular necrosis, or congenital dysplasia/structural abnormalities with sufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized symbol. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 30, 2018
Ortho Development Corporation Drew Weaver Director of Quality Assurance and Regulatory Affairs 12187 South Business Park Drive Draper, Utah 84020
Re: K173951
Trade/Device Name: Legend® Acetabular Liners Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: April 30, 2018 Received: May 1, 2018
Dear Drew Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173951
Device Name Legend® Acetabular Liner
#### Indications for Use (Describe)
This device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:
- · Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis.
- · Previously failed hip surgery.
- · Fractures of the femoral neck or head.
- · Avascular necrosis of the femoral head.
- · Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to prosthesis.
| Type of Use (Select one or both, as applicable) | Depository Institution (31 CFR 221 Subpart D) Securities Intermediary (31 CFR 221 Subpart C) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
12187 So. Business Park Drive Draper, Utah 84020 801-553-9991/ fax 553-9993 Orthodevelopment.com
K173951 Page 1/2
Image /page/3/Picture/2 description: The image shows the logo for "Ortho Development". The logo is purple and features a stylized circle with curved lines emanating from it on the left. To the right of the circle is the word "ORTHO" in capital letters, and below that is the word "DEVELOPMENT" in smaller letters.
# Section 5 510(k) Summary
| Name of Sponsor: | Ortho Development Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Drew Weaver<br>Director of Quality Assurance and Regulatory Affairs<br>Telephone: (801) 619-3419<br>Facsimile: (801) 553-9993<br>Email: DWeaver@orthodevelopment.com |
| Date Prepared: | May 24, 2018 |
| Proprietary Name: | Legend® Acetabular Liner |
| Common Name: | Acetabular Liner Prosthesis |
| Classification: | 21 CFR 888.3358, Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis<br>21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis |
| Device Class: | Class II device |
| Device Product Code: | LPH, LZO |
| Predicate Devices: | K103384 – Acetabular Cup System<br>Ortho Development Corporation |
### 5.1 Device Description
The proposed subject device is a line extension to the acetabular cup system previously cleared in K103384. The predicate is a modular system intended for the replacement of the natural articular surface of the hip joint in total hip replacement surgery. The system consists of acetabular shells, liners, bone screws, apical screw hole cover, and femoral heads.
The subject and predicate acetabular liners are both made of highly cross-linked polyethylene and are hemispherical in shape with positive locking, liner stabilization, and anti-rotation features.
The Legend Acetabular Liners (subject device) differ from the predicate device, cleared in K103384, through the addition of new sizes. Other components of the system have not been modified and are not a part of this submission other than by reference.
{4}------------------------------------------------
## 5.2 Indications for Use
This device is intended for use in total hip arthroplasty. The device is intended for uncemented, press-fit use only in cases of:
- Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic 1. arthritis.
- 2. Previously failed hip surgery.
- 3. Fractures of the femoral neck or head.
- 4. Avascular necrosis of the femoral head.
- Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to 5. properly seat the prosthesis.
### 5.3 Basis for Substantial Equivalence:
The Legend Acetabular Liners have the same technological characteristics as the predicate devices. These include:
- Indications for use
- Intended Use
- Fundamental design
- Material
- Manufacturer
- Manufacturing Processes
- Sterilization
- Packaging
The following non-clinical tests were conducted on various configurations of the Legend Acetabular Liners and met the predetermined acceptance criteria:
- lmpingement Performance per ASTM F-2582
- Bacterial endotoxin testing using LAL pyrogen testing methodology per ANSI/AAMI ST72:2011
The fundamental scientific technology of the Legend Acetabular Liner is the same as the previously cleared predicate device. The mechanical test results demonstrate that the Legend Acetabular Liners are as safe, as effective, and perform as well as or better than the legally marketed device predicate. No clinical studies were performed.
Based on similarities in intended use, design, manufacturing methods, sterilization, packaging, and the mechanical test results, the Legend Acetabular Liner is substantially equivalent to the predicate device that was cleared under K103384.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.