K103384 · Ortho Development · LPH · Mar 16, 2011 · Orthopedic
Device Facts
Record ID
K103384
Device Name
ESCALADE ACETABULAR CUP SYSTEM
Applicant
Ortho Development
Product Code
LPH · Orthopedic
Decision Date
Mar 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Escalade Acetabular Cup System is intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.
Device Story
Modular acetabular cup system for total hip replacement; replaces natural hip articular surface. Components include titanium alloy (Ti-6Al-4V ELI) hemispherical shells with plasma spray coating (Ti-CP) for biological fixation, cross-linked UHMWPE liners, bone screws, apical screw hole covers, and cobalt chromium (Co-28Cr-6Mo) femoral heads. Used in surgical settings by orthopedic surgeons. System provides stable, cementless fixation to bone. Liners available in neutral and 10° hooded configurations for femoral heads 26mm-40mm. Benefits include increased patient mobility and pain reduction. Clinical decision-making supported by implant stability and wear reduction properties demonstrated in bench testing.
Clinical Evidence
Bench testing only. In-vitro hip simulator study (ISO 14242-1) compared Escalade UHMWPE liners to TriPlus control liners over 5 million cycles. Escalade liners demonstrated 88% reduction in wear rate (1.5 ± 0.6 mg/MC) compared to TriPlus control (12.8 ± 1.6 mg/MC). Additional mechanical testing included interconnection strength, impingement, range of motion, and fatigue testing.
Indicated for total hip arthroplasty with uncemented, biological fixation. Patient population includes individuals with impaired hip joints due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, failed prior hip surgery, femoral neck/head fractures, avascular necrosis, or congenital dysplasia/structural abnormalities with sufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters. The characters are 'K103384'. The characters appear to be written with a dark ink or marker on a white background. The handwriting is somewhat uneven, but the characters are generally legible.
### 510(k) Summary
MAR 1 6 2011
| NAME OF SPONSOR: | Ortho Development Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) CONTACT: | Tom Haueter<br>Regulatory Affairs Manager<br>Telephone: (801) 553-9991<br>Facsimile: (801) 553-9993<br>Email: thaueter@orthodevelopment.com |
| DATE PREPARED: | November 15, 2010 |
| PROPRIETARY NAME: | Escalade Acetabular Cup System |
| COMMON NAME: | Acetabular Cup Prosthesis |
| CLASSIFICATION: | 21 CFR 888.3358, Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis, Class II device<br>21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis |
| DEVICE PRODUCT CODE: | LPH, LZO |
| PREDICATE DEVICES: | TriPlus Acetabular Cup System, Ortho Development Corp.<br>(K042565) |
| | Pinnacle Acetabular Cup System, DePuy Orthopedics, Inc.<br>(K062148, K000306) |
## DEVICE DESCRIPTIONS:
The Escalade Acetabular Cup System is a modular system intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. The system consists of acetabular shells, liners, bone screws, apical screw hole cover, and femoral heads.
Acetabular shells are hemispherical in shape and are designed for press-fit, cementless surgical applications. The shells are machined from titanium alloy (Ti-6Al-4V ELI) and are plasma spray coated (Ti-CP) for biological fixation. Escalade liners are manufactured from extensively cross-linked UHMWPE and are available in neutral face and 10° hooded options. Liners are configured for use with femoral head diameters ranging from 26mm to 40mm. The Ø6.5mm cancellous bone screws and the apical screw hole cover are manufactured from titanium alloy (Ti-6Al-4V ELI). The 36mm and 40mm diameter femoral heads made of cobalt chromium alloy (Co-28Cr-6Mo) are being added to the currently offered femoral head sizes in conjunction with this system.
{1}------------------------------------------------
The following marketing claim will be made for Escalade Acetabular Liners:
Ortho Development Corp Escalade™ UHMWPE liners reduce wear by 88% compared to previously cleared TriPlus UHMWPE liners when used with CoCrMo femoral heads. An in-vitro hip simulator study was conducted using the following polyethylene liners: Escalade 36mm ID liners for a 52mm OD shell, Escalade 40mm ID liners for a 56mm OD shell, and TriPlus 32mm ID liners for a 52mm OD shell as a control. The bearing surface thickness was 6.7mm for the Escalade liners and 5.8mm for the TriPlus liners. Test and control liners were manufactured by Ortho Development Corp. The Escalade Liners were made from compression molded GUR 1020 material which was crosslinked, and ETO sterilized prior to the study. TriPlus liners were machined from compression molded GUR 1020 material and sterilized by gamma irradiation in an inert gas package. Test and control liners were tested for 5 million cycles on an AMTI 12 station hip simulator with a standard walking gait cycle as specified by ISO 14242-1 with a 3.0 kN peak load. The femoral head components were CoCrMo heads manufactured by Ortho Development. The lubricant was 94% bovine serum (Sigma-Aldrich) with 10.7mMol EDTA and penicillin-streptomycin solution (Fisher Scientific). The protein concentration of the lubricant was 74 g/L. The steady state wear rates determined from the study were 1.5 ± 0.6 mg/MC for both the 36mm and the 40mm Escalade liners. The steady state wear rate for the TriPlus control liners was 12.8 ±1.6 mg/MC. The average total wear for 5 million cycles was 6.6 mg for the 40mm liners, 6.7 mg for the 36mm liners, and 67.9 mg for the 32mm control liners. Reduced wear claims are based upon the results of in-vitro hip wear simulator tests which have not been shown to quantitatively predict clinical wear performance. Also, a reduction in total polyethylene wear volume or wear rate alone may not result in an improved clinical outcome as wear particle size and morphology are also critical factors in the evaluation of the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not evaluated as part of this wear claim.
#### INTENDED USE:
The Escalade Acetabular Cup System is intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.
#### INDICATIONS FOR USE:
The Escalade Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Escalade Acetabular Cup System is compatible with the Ovation Hip Stem System (indicated for uncemented, biological fixation of the stem) and the Encompass Hip Stem System (indicated for both cemented and uncemented, biological fixation of the stem).
Total hip arthroplasty is indicated for the following conditions:
- Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic 1. arthritis.
- Previously failed hip surgery. 2.
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{2}------------------------------------------------
- 3. Fractures of the femoral neck or head.
- 4. Avascular necrosis of the femoral head.
- 5. Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
#### BASIS OF SUBSTANTIAL EQUIVALENCE:
Ortho Development believes that the Escalade Acetabular Cup System is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, design, materials, size range, manufacturing methods, packaging, liner material and mechanical performance.
Testing that supports the substantial equivalence of the devices includes construct wear, liner/shell interconnection strength testing (assembly, push-out, torque-out), static and dynamic impingement, range of motion, shell deformation testing, and offset liner rim fatigue testing. In addition, porous plasma spray coating and polyethylene liner material characterization was performed.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ortho Development Corporation % Mr. Thomas D. Haueter Regulatory Affairs Manager 12187 South Business Park Drive Draper, Utah 84020
## MAR 1 6 2511
Re: K103384
Trade/Device Name: Escalade Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: March 08, 2011 Received: March 11, 2011
Dear Mr. Haueter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Mr. Thomas D. Haueter
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97): For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Aj. VS. n 2
fer
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K10384
# Indication for Use Form Ortho Development Escalade 510(k)
#### 510(k) Number: K103384
Device Name: Escalade Acetabular Cup System
Indications for Use:
The Escalade Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Escalade Acetabular Cup System is compatible with the Ovation Hip Stem System (indicated for uncemented, biological fixation of the stem) and the Encompass Hip Stem System (indicated for both cemented and uncemented, biological fixation of the stem).
Total hip arthroplasty is indicated for the following conditions:
- Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic 1. arthritis.
- Previously failed hip surgery. 2.
- 3. Fractures of the femoral neck or head.
- Avascular necrosis of the femoral head. ধ ..
বাংলাদেশে পারে বিশ্ববিদ্যালয়ের প্রতিষ্ঠান করে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পা
- 5. Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
M.melkersm
(Division Sign-Off) (Division Sign pical. Orthopedic, Divisionative Devices
510(k) Number K103384
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.