Unifine SafeControl 5mm x 30G, Unifine SafeControl 8mm x 30G
Applicant
Owen Mumford, Ltd.
Product Code
FMI · General Hospital
Decision Date
Jul 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unifine SafeControl™ range of pen needles are intended for use with multi-dose injection devices for the subcutaneous injection of FDA approved drugs, including insulin.
Device Story
Unifine SafeControl is a sterile, single-use, 30-gauge safety pen needle (5mm-8mm lengths) for subcutaneous drug delivery. It attaches to multi-dose injection pens. The device features an integrated safety guard that covers the needle after use to prevent sharps injuries. Operation is manual; the user attaches the needle to the pen, performs the injection, and the safety mechanism engages. It is used in clinical or home settings by patients, caregivers, or clinicians. The device provides a physical barrier to reduce accidental needlestick risks. It is supplied sterile via gamma radiation.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 11608-2 (torque, dose accuracy, flow rate, needle retention/dislocation) and ISO 23908 (sharps injury prevention). Biocompatibility testing conducted per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity).
Technological Characteristics
30G stainless steel (AISI 304) cannula with silicone coating. Hub/carrier: polypropylene. Safety guard: polystyrol. Adhesive: Dymax 1180-M-SV04. Sterilization: Gamma radiation (20-40kGy, SAL 10^-6). Complies with ISO 11608-2, ISO 11607-1, ISO 23908, and ASTM F1140-07.
Indications for Use
Indicated for subcutaneous injection of FDA-approved drugs (including insulin) using multi-dose injection devices. Intended for use by self-administering patients, caregivers, and healthcare professionals.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Unifine Pentips and Unifine Pentips Plus (K152339)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
July 11, 2018
Owen Mumford Ltd % Patty Cronan Quality Manager Owen Mumford USA Inc 1755 West Oak Commons CT Marietta, Georgia 30062
Re: K173881
Trade/Device Name: Unifine SafeControl Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: May 24, 2018 Received: May 25, 2018
Dear Patty Cronan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Digitally signed by Alan M. Alan M Stevens -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, Stevens 0.9.2342.19200300.100.1.1=1300 189211, cn=Alan M. Stevens -S Date: 2018.07.11 07:16:33 -04'00' for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K173881
Device Name Unifine SafeControl™M
Indications for Use (Describe)
The Unifine SafeControl™ range of pen needles are intended for use with multi-dose injection devices for the subcutaneous injection of FDA approved drugs, including insulin.
| Type of Use ( <i>Select one or both, as applicable</i> ) | <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="text-decoration: overline;">☑</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff(@fda.hhs.gov
H
{3}------------------------------------------------
# SECTION 5.0
# 510(k) SUMMARY
## 1.Submitter
| Prepared by: | Watheq Al-Hakam<br>Regulatory Affairs Officer - Projects<br>Owen Mumford Ltd.<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared for: | Owner/ Operator<br>Owen Mumford Ltd<br>Brook Hill<br>Woodstock<br>Oxfordshire<br>OX20 1TU<br>United Kingdom<br>Establishment Registration Number: 3003348846 |
| Contact Person: | Darren Mansell<br>Regulatory Affairs Manager<br>Owen Mumford Ltd,<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466<br>Email: darren.mansell@owenmumford.com |
| Contact Person: | Paul Smith<br>Head of Quality<br>Owen Mumford Ltd,<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466<br>Email: paul.smith@owenmumford.com |
| Contact Person: | Watheq Al-Hakam<br>Regulatory Affairs Officer - Projects<br>Owen Mumford Ltd,<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466<br>Email: watheq.al-hakam@owenmumford.com |
| Date Prepared: | 15 DECEMBER 2017 |
{4}------------------------------------------------
## 2.Device
Name of Device: Unifine SafeControl™ Common Name: Pen Needles Classification Name: Needle, Hypodermic, Single Lumen (21 CFR 880.5570) Regulatory Class: II Product Code: FMI
## 3.Predicate Devices
Predicate Device Name: Unifine Pentips and Unifine Pentips Plus, 510k number K152339
### 4.Description of The Device
The Unifine SafeControl™ safety pen needle is a sterile, single-use, disposable device intended for use with multi-dose Injector devices for the subcutaneous injection of FDA approved drugs, including insulin. The device is designed for prescription and over-the-counter use and to be used by self-administering patients, care-givers and healthcare professionals. The safety pen needle is a 30-gauge needle available in sizes between 5mm and 8mm lengths.
The target population includes male and female right or left-handed self-administering patients, care givers and healthcare professionals. The frequency of use and intended patient population is dependent on given treatment regime.
The pen needle assembly consists of a cannula attached to a needle carrier assembled into a plastic moulded needle hub and safety guard, a primary container that houses the entire assembly and a sterility seal that covers the assembly inside the primary container. The entire device is packaged and labeled as a sterile single-use device.
The purpose of this 510(k) Premarket Notification is to obtain a Prescription Use and Over-The-Counter Use clearance for the Unifine SafeControl™ safety pen needle device. The intended use for the Unifine SafeControl™ remains the same as the predicate device.
Figure 5.1: Exploded view of the Unifine SafeControl™ safety pen needle assembly
Image /page/4/Figure/10 description: The image shows an exploded view of a medical device, likely a needle or syringe, and its packaging. The components are arranged in a line, starting with the primary container or packaging on the left, followed by a safety guard, a needle carrier with cannula, a hub, and finally, a sterility seal or packaging on the right. The labels indicate the function of each component, emphasizing the packaging and safety features of the device.
{5}------------------------------------------------
| Component | Base Material | Patient Contact | Duration |
|----------------------------------|-------------------------------------------------------|-----------------------------------------------------|------------------------|
| Hub | Polypropylene | Surface (Intact Skin) | A - Limited (≤24 hour) |
| Safety Guard | Polystyrol | Surface (Intact Skin) | A - Limited (≤24 hour) |
| Needle Carrier | Polypropylene | Surface (Intact Skin) | A - Limited (≤24 hour) |
| Cannula 30G | STAINLESS STEEL -<br>AISI 304 coated with<br>silicone | Externally<br>Communicating<br>(Bloodpath Indirect) | A - Limited (≤24 hour) |
| Adhesive (Glue) | DYMAX 1180-M-SV04 | Surface (Intact Skin) | A - Limited (≤24 hour) |
| Primary Packaging | | | |
| Primary Container<br>(Packaging) | Polypropylene | Surface (Intact Skin) | A - Limited (≤24 hour) |
| Foil seal (Packaging) | Paper Laminate | Surface (Intact Skin) | A - Limited (≤24 hour) |
## Table 5.1: Biocompatibility Device Categorization for Submission Device Components
## 5.Indications for Use
The Unifine SafeControl™range of pen needles are intended for use with multi-dose injection devices for the subcutaneous injection of FDA approved drugs, including insulin.
## 6. Technological Characteristics
The Unifine SafeControl™ safety pen needles are substantially equivalent to the predicate device with an additional safety feature allowing the exposed needle to be covered after use.
A comparison of the intended use and technological characteristics is summarized in the table below.
{6}------------------------------------------------
Table 5.2: Comparison of Device Characteristics between Submission Devices and Predicate Device
| Device Characteristic | Predicate Device: Unifine Pentips &<br>Unifine Pentips Plus - K152339 | Submission Device - Unifine<br>SafeControl™ |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Unifine Pentips range of pen<br>needles are intended for use with<br>multi-dose injection devices for the<br>subcutaneous injection of FDA<br>approved drugs, including insulin | Unchanged from the predicate<br>device |
| Operating principle | Manual | Manual with an additional safety<br>feature allowing the exposed<br>needle to be covered after use. |
| Design/ construction | Needle assembly (cannula, needle<br>hub packaged in a needle shield and<br>primary container) | Device unchanged from the<br>predicate device with the<br>exception of an addition of a<br>safety feature (safety guard)<br>integrated in the device. The<br>needle carrier was integrated in<br>the needle hub in the predicate<br>device. |
| Components and Materials | Unifine Pentips by Owen Mumford:<br>Device:<br>• Cannula - Stainless Steel<br>• Lubricant - Silicone<br>• Adhesive – Medical Grade Adhesive<br>• Needle Hub – Polypropylene<br>Packaging:<br>• Needle Shield - High Density<br>Polyethylene<br>• Outer Container - High Density<br>Polyethylene<br>• Sterility Seal - Paper Laminate<br>Unifine Pentips Plus by Owen<br>Mumford:<br>Device:<br>• Cannula - Stainless Steel<br>• Lubricant - Silicone<br>• Adhesive – Medical Grade Adhesive<br>• Needle Hub – Polypropylene<br>Packaging:<br>• Needle Shield - High Density<br>Polyethylene<br>• Primary Container - Polypropylene<br>• Sterility Seal - Paper Laminate | Unchanged from the predicate<br>device with the exception of a<br>safety feature (safety guard)<br>component:<br>• Safety guard -<br>Polystyrol<br>The needle carrier was<br>integrated in the needle hub in<br>the predicate device, the colour<br>additive is changed. |
| Package | • Plastic outer container<br>• Sterility Seal<br>• Shelf box | Unchanged from the predicate<br>device |
{7}------------------------------------------------
| Needle<br>Specification | | Needle Length | Unifine Pentips by Owen Mumford:<br>4mm for 32G; 5mm, 6mm and 8mm<br>for 31G; 12mm for 29G<br>Unifine Pentips Plus by Owen<br>Mumford: 4mm for 32G; 5mm, 6mm<br>and 8mm for 31G; 12mm for 29G | 5mm for 30G<br>8mm for 30G |
|-------------------------|--|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| | | Effective Gauge | 29G, 31G and 32G | 30G |
| | | Tip<br>Configuration | Patient-side: Tri-bevel (3) | Unchanged from the predicate<br>device |
| Sterilization | | | Unifine Pentips by Owen Mumford:<br>Gamma Radiation (validated to<br>achieve SAL 10-6) | Unchanged from the predicate<br>device, Gamma Radiation<br>(validated to achieve SAL 10-6). |
| | | | Unifine Pentips Plus by Owen<br>Mumford: Gamma Radiation (validated<br>to achieve SAL 10-6) | |
## 7.Performance Data
## Non-clinical performance data:
To support this submission, design verification testing was performed to demonstrate that the device operates safely and effectively. The following table shows a summary of the performance tests to standards.
## Bench Testing:
Table 5.3: Summary of the performance tests to standards:
| Test | Standard/ Requirement | Results |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|----------------|
| Visual Inspection | ISO 11608-2:2012 Section 12.2.2 | Meets standard |
| Torque to attach to pen | ISO 11608-2:2012 Section 11.4.1.5 | Meets standard |
| Torque to remove from pen | ISO 11608-2:2012 Section 11.4.3 | Meets standard |
| Dose accuracy | ISO 11608-2:2012 Section 11.4.2 | Meets standard |
| Flow Rate | ISO 11608-2:2012 Section 4.3 | Meets standard |
| Force to over-ride safety feature | ISO 11608-5:2012 Section 5.1.11.2 | Meets standard |
| Needle Retention | ISO 11608-2:2012 Section 9 | Meets standard |
| Needle Dislocation | ISO 11608-2:2012 Section 4.8 | Meets standard |
| Sterility - Seal integrity | ISO 11607-1:2007 Section 6.3 | Meets standard |
| Internal pressure test (Burst<br>Testing) | ASTM-F-1140-07:2007<br>ASTM 4169 2016 Section 16.2 | Meets standard |
| NIS Compatibility | ISO 11608-2:2012 Section 4.9 | Meets standard |
| Simulated Clinical use<br>testing (sharps safety<br>testing) | ISO 23908:2011 &<br>Guidance for Industry and FDA Staff-<br>Medical Devices with Sharps<br>Injury Prevention Features | Meets standard |
{8}------------------------------------------------
The objective of all the bench testing conducted was to verify that the submission device met the pre-determined specifications, in order to support the conclusion that it is fit for purpose and is substantially equivalent. Where applicable, the testing for the submission device was conducted in accordance with ISO 11608-2:2012 Needle-based injection systems for medical use – Requirements and test methods – Part 2: Needles. Written protocols specified the scope, objectives, specifications, test equipment, test methods and acceptance criteria. Additional design verification testing was performed that was not required by ISO 11608- 2:2012. The type of Bench testing are summarized below.
The following are the tests conducted:
- Visual Inspection
- Torque to attach to pen
- Torque to remove from pen
- Dose accuracy
- Glide Force
- Penetration Force
- · Force to remove foil
- · Flow Rate
- · Force to activate safety feature
- · Force to over-ride safety feature
- Needle Retention
- · Needle Dislocation
- · Sterility Seal integrity
- · Internal pressure test
- · Force to remove from Primary Container
- Torque to Break Thread Bond
#### Biocompatibility:
In accordance with ISO 10993-1, the Unifine SafeControl™ safety pen needle is classified as: Externally Communicating Device, Blood Path Indirect, Limited (≤24 hour) use. The biocompatibility evaluation for finished devices' blood/body contacting parts were conducted in accordance with the FDA quidance document "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The tests used for the biocompatibility evaluation of the contact materials are as follows:
- Cytotoxicity
- . Sensitization
- Irritation ●
- Systemic Toxicity
- Hemocompatibility ●
- Pyrogenicity ●
In addition to the tests set out in ISO 10993-1; the submission included evaluation of Pyrogenicity (ISO 10993-11 Annex F) and the whole device evaluation (ISO 10993-1 section 7). Pyrogenicity testing shall be considered for any components which come into internal contact with the user, classed as externally communicating. Full device evaluation includes a review of all the materials, consumables, manufacturing processes and environments by an expert toxicologist to ensure the device in its final form is biocompatible.
Results of the testing demonstrate the materials are biocompatible.
{9}------------------------------------------------
## Sterilization:
Unifine SafeControl™ safety pen needles manufactured by Owen Mumford in the UK are sterilized by gamma radiation. A summary of sterilization and shelf life details for all variants of device are as follows:
#### Sterilization Method: Gamma radiation
#### Radiation dose range: 20-40kGy
#### Description of the method used to validate the sterilization cycle:
Sterilization validation (Dose mapping) was conducted in accordance with Section 9 of ISO 11137: "Sterilization of Healthcare Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices."
#### Shelf life
Shelf life studies were conducted per ISO 11608-2, ISO 23908, and ASTM F1980-16 to support an initial 2-year shelf life, and protocols were submitted to support an increase to a 5-year shelf life upon successful completion of studies.
### 8.Conclusion
In summary, the differences between the Unifine SafeControl™ and the predicates do not raise different questions of safety and effectiveness and the performance data shows that the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.