K152339 · Owen Mumford, Ltd. · FMI · Jun 24, 2016 · General Hospital
Device Facts
Record ID
K152339
Device Name
Unifine Pentips/ Unifine Pentips Plus
Applicant
Owen Mumford, Ltd.
Product Code
FMI · General Hospital
Decision Date
Jun 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Unifine Pentips range of pen needles are intended for use with pen injector devices for the subcutaneous injection of drugs, including insulin.
Device Story
Sterile, single-use pen needles for subcutaneous drug delivery via pen injectors; used by patients, caregivers, and clinicians. Assembly includes double-ended cannula, hub, shield, and container. User screws needle onto pen injector; non-patient end penetrates cartridge septum; patient end delivers medication. Unifine Pentips Plus includes integrated needle remover chamber in primary container to facilitate safe disposal and prevent needlesticks. Lubricated with silicone to aid penetration. Available in various lengths (4-12mm) and gauges (29G-32G). Benefits include ease of use, compatibility with standard pen injectors, and enhanced safety features for disposal in the 'Plus' variant.
Clinical Evidence
No clinical data required. Substantial equivalence established via bench testing, including ISO 11608-2:2012 compliance (dimensions, flow rate, dose accuracy, bond strength, needle points) and biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity).
Technological Characteristics
Cannula: Stainless steel; Lubricant: Silicone; Hub: Polypropylene; Shield/Container: High-density polyethylene; Sterility seal: Paper laminate. Manual operation. Sterilization: Gamma radiation or Ethylene Oxide (SAL 10^-6). Dimensions: 4-12mm length, 29G-32G gauge. Complies with ISO 11608-2:2012.
Indications for Use
Indicated for subcutaneous injection of drugs, including insulin, using pen injector devices. Suitable for use by consumers, caregivers, and healthcare professionals.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2016
Owen Mumford Ltd % Ms. Patty Cronan Owen Mumford USA, Inc. 1755 West Oak Commons Court Marietta, Georgia 30062
Re: K152339
Trade/Device Name: Unifine® Pentips® and Unifine® Pentips® Plus Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: June 17, 2016 Received: June 20, 2016
Dear Ms. Cronan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152339
Device Name Unifine Pentips/ Unifine Pentips Plus
Indications for Use (Describe)
The Unifine Pentips range of pen needles are intended for use with pen injector devices for the subcutaneous injection of drugs, including insulin.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5.0
# 510(k) SUMMARY: 510k Number: K152339
# 1. Submitter
| Prepared by: | Patty Cronan - Quality Manager<br>Quality Manager<br>Owen Mumford USA Inc.<br>Tel. (770) 977-2226<br>Fax (770) 977-2866 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared for: | Owner/ Operator<br>Owen Mumford Ltd<br>Brook Hill<br>Woodstock<br>Oxfordshire<br>OX20 1TU<br>United Kingdom<br>Establishment Registration Number: 8021764 |
| Contact Person: | Patty Cronan - Quality Manager<br>Quality Manager<br>Owen Mumford USA Inc.<br>Tel. (770) 977-2226<br>Fax (770) 977-2866<br>Email: p.cronan@owenmumfordinc.com |
Date Prepared: October 16, 2015
# 2. Device
Name of Device: Unifine® Pentips® & Unifine® Pentips® Plus Common Name: Pen Needles Classification Name: Needle, Hypodermic, Single Lumen (21 CFR 880.5570) Regulatory Class: II Product Code: FMI
# 3. Predicate Devices
Primary Predicate Device 1 Name: Unifine® Pentips®, K973899
Reference Predicate Device 2 Name: Artsana InsuPen® Insulin Pen Needles, K051783
Reference Predicate Device 3 Name: BD Pen Needles, K051899
Reference Predicate Device 4 Name: BD 32G x 4mm Pen Needle, K100005
# 4. Description Of The Device
Unifine® Pentips® and Unifine® Pentips® Plus are sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors. The pen needles are used by consumers, caregivers, and healthcare professionals. They are available in a variety of lengths (4mm, 5mm, 6mm, 8mm and12mm) and gauges (29G, 31G and 32G).
The pen needle assembly consists of a double-ended cannula, a needle hub, a needle shield, primary container and sterility seal that covers the assembly inside the primary container.
{4}------------------------------------------------
The hub has internal threads, which allows the pen needle to be screwed onto a pen injector device. The Non Patient (NP) end of the cannula (inside the hub) penetrates the rubber septum of the cartridge in the pen injector device. The Patient end and NP end of the cannula are lubricated using a silicone based lubricant to aid in injection and penetration into the rubber septum. The inner needle shield is injection molded and assembled over the Patient end of the needle to protect the needle from damage and accidental needle-sticks. The pen needle assembly is then inserted into an iniection molded outer container and sealed with a protective peel-away label to ensure a sterility barrier and tamper evidence. The outer container can be used to remove the hub with cannula from the pen injector device after the injection is completed. The peel-away label is pre-printed with information, which includes the lot number, expiry date, and size (length) and gauge of the needle. The individual needle assemblies are then packaged in bags and/or cartons, and placed into shippers with appropriate labeling. The shipper cases are then placed on pallets for sterilization.
Unifine Pentips Plus pen needles differ from Unifine Pentips pen needles because the primary container consists of two 'chamber contains a new, unused pen needle and the second chamber is empty. The empty chamber serves as a built-in pen needle remover, and may prevent accidental needlesticks. Users can insert their used needle into the chamber, then unscrew from the injection pen. The primary container is designed to hold the used pen needle until it can be disposed of in a suitable sharps container. The actual pen needle is identical between Unifine Pentips and Unifine Pentips Plus in both the physical appearance and functionality.
The purpose of this 510(k) Premarket Notification is to obtain Over-the-Counter (OTC) clearance for Unifine Pentips and Unifine Pentips Plus, to add additional needle lengths of 4mm, 5mm & 6mm, and to document some material changes. Though the Indications for Use are worded slightly different, the intended use for Unifine Pentips and Unifine Pentips Plus is substantially equivalent to the primary predicate and multiple predicate devices.
# 5. Indications for Use
The Unifine Pentips range of pen needles are intended for use with pen injector devices for the subcutaneous injection of drugs, including insulin.
# 6. Technological Characteristics
A comparison of the intended use and technological characteristics is summarized in the table below. Table 6.1. For the table, the BD Predicates, K051899 & K100005, were combined in the third column as the only difference between the two is the length and gauge of the needle. K051899 included needle lengths of 5mm, 8mm & 12.7mm; and gauge sizes of 31G, 30G & 29G. K100005 covers length of 4mm, and gauge size of 32G.
The changes for Owen Mumford included in this submission are the addition of the lengths (4mm, 5mm & 6mm) and gauge sizes (31G & 32G) to the contract manufactured Unifine Pentip product line. This submission also includes the full range of sizes and gauges selfmanufactured by Owen Mumford, including lengths (4mm, 5mm, 8mm &12mm) and qauge sizes (32G, 31G & 29G).
The full range of Unifine Pentips and Unifine Pentips Plus pen needles are substantially equivalent to intended use, device description, appearance, function, principle of operation, and basic composition to the predicate devices. Though the sizes and gauges are not exactly the same as the predicate device sizes and gauges, the range of Unifine Pentips and Unifine Pentips Plus (4mm - 12mm, 32G - 29G) fall within the range of the predicate device sizes and gauges (4mm - 12.7mm, 32G - 29G). The sizes included in this submission do not represent a difference in product technology or product performance versus the predicate devices.
Table 6.1: Comparison of Device Characteristics between Submission Devices and Predicate Devices
{5}------------------------------------------------
| Device<br>Characteristic | Submission<br>Device: Unifine<br>Pentips & Unifine<br>Pentips Plus<br>(K152339) | Primary Predicate:<br>Unifine Pentips &<br>Unifine Pentips<br>Plus (K973899) | Reference<br>Predicate: Artsana<br>InsuPen® Insulin<br>Pen Needles<br>(K051783) | Reference<br>Predicate: BD Pen<br>Needles (K051899<br>& K100005) | |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Intended Use | The Unifine Pentips<br>range of pen<br>needles are<br>intended for use<br>with pen injector<br>devices for the<br>subcutaneous<br>injection of drugs,<br>including insulin. | The Unifine Pentip<br>is single use,<br>disposable<br>hypodermic single<br>lumen needle<br>designed for use<br>with multi-dose<br>injection devices<br>that use prefilled<br>cartridges.<br>The protective<br>paper is removed<br>from the pentip<br>which is then<br>screwed onto the<br>device cartridge<br>housing and then<br>used in the<br>prescribed manner.<br>One of the main<br>uses of the pentip<br>is for use with pen<br>injectors for the<br>delivery of insulin<br>which is supplied in<br>the form of prefilled<br>cartridges. The<br>patient places a<br>fresh pentip needle<br>onto the pen<br>injector, dials a<br>dose, inserts the<br>pentip into the skin<br>and delivers the<br>dose. | Artsana disposable<br>sterile pen needles<br>are used with<br>insulin pen injector<br>devices for the<br>subcutaneous<br>injection of insulin<br>in the treatment of<br>diabetes. | BD Pen Needle is<br>intended for use<br>with pen injector<br>device for<br>subcutaneous<br>injection of drugs,<br>including insulin<br>and exenatide. | |
| Prescription or<br>OTC | OTC | Prescription | Prescription | OTC | |
| Operation<br>Principle | Manual | Manual | Manual | Manual | |
| Design/<br>Construction | Needle assembly<br>(cannula, needle<br>hub, needle shield,<br>outer container) | Needle assembly<br>(cannula, needle<br>hub, needle shield,<br>outer container) | Needle assembly<br>(cannula, needle<br>hub, needle shield,<br>outer container) | Needle assembly<br>(cannula, needle<br>hub, needle shield,<br>outer container) | |
| Materials | Unifine Pentips by<br>Artsana S.p.A.:<br>• Cannula -<br>Stainless Steel<br>• Lubricant -<br>Silicone | Unifine Pentips by<br>Artsana S.p.Α.:<br>• Cannula -<br>Stainless Steel<br>• Lubricant -<br>Silicone | • Cannula –<br>Stainless Steel<br>• Lubricant –<br>Silicone<br>• Adhesive -<br>Medical Grade<br>Adhesive | • Cannula -<br>Stainless Steel<br>• Lubricant -<br>Silicone Based<br>• Outer Container<br>– Polypropylene | |
| Device<br>Characteristic | Submission<br>Device: Unifine<br>Pentips & Unifine<br>Pentips Plus<br>(K152339) | Primary Predicate:<br>Unifine Pentips &<br>Unifine Pentips<br>Plus (K973899) | Reference<br>Predicate: Artsana<br>InsuPen® Insulin<br>Pen Needles<br>(K051783) | Reference<br>Predicate: BD Pen<br>Needles (K051899<br>& K100005) | |
| | Adhesive<br>• Needle Hub -<br>Polypropylene<br>• Needle Shield -<br>High Density<br>Polyethylene<br>• Outer Container<br>- High Density<br>Polyethylene<br>• Sterility Seal -<br>Paper Laminate<br><i>Unifine Pentips by<br/>Owen Mumford:</i><br>• Cannula -<br>Stainless Steel<br>• Lubricant -<br>Silicone<br>• Adhesive -<br>Medical Grade<br>Adhesive<br>• Needle Hub –<br>Polypropylene<br>• Needle Shield -<br>High Density<br>Polyethylene<br>• Outer Container<br>- High Density<br>Polyethylene<br>• Sterility Seal -<br>Paper Laminate<br><i>Unifine Pentips<br/>Plus by Owen<br/>Mumford:</i><br>• Cannula -<br>Stainless Steel<br>• Lubricant -<br>Silicone<br>• Adhesive -<br>Medical Grade<br>Adhesive<br>• Needle Hub -<br>Polypropylene<br>• Needle Shield -<br>High Density<br>Polyethylene<br>• Outer Container<br>- Polypropylene<br>• Sterility Seal -<br>Paper Laminate | Adhesive<br>• Needle Hub -<br>Polypropylene<br>• Needle Shield -<br>High Density<br>Polyethylene<br>• Outer Container<br>- High Density<br>Polyethylene<br>• Sterility Seal -<br>Paper Laminate | Polypropylene<br>• Needle Shield -<br>High Density<br>Polyethylene<br>• Outer Container<br>- High Density<br>Polyethylene<br>• Sterility Seal -<br>Paper Laminate | available for these<br>devices. | |
| Package | • Plastic outer<br>container<br>• Sterility Seal | • Plastic outer<br>container<br>• Sterility Seal | • Plastic outer<br>container<br>• Sterility Seal | • Plastic outer<br>container<br>• Sterility Seal | |
| | • Shelf box | • Shelf box | • Shelf box | • Shelf box | |
| Device<br>Characteristic | Submission<br>Device: Unifine<br>Pentips & Unifine<br>Pentips Plus<br>(K152339) | Primary Predicate:<br>Unifine Pentips &<br>Unifine Pentips<br>Plus (K973899) | Reference<br>Predicate: Artsana<br>InsuPen® Insulin<br>Pen Needles<br>(K051783) | Reference<br>Predicate: BD Pen<br>Needles (K051899<br>& K100005) | |
| Needle Specifications | Needle<br>Length | Unifine Pentips<br>manufactured for<br>Owen Mumford<br>Ltd: 4mm for 32G;<br>5mm, 6mm, 8mm<br>for 31G; 12mm for<br>29G<br>Unifine Pentips by<br>Owen Mumford:<br>4mm for 32G;<br>5mm, 6mm and<br>8mm for 31G;<br>12mm for 29G<br>Unifine Pentips<br>Plus by Owen<br>Mumford: 4mm for<br>32G; 5mm, 6mm<br>and 8mm for 31G;<br>12mm for 29G | Unifine Pentips<br>manufactured for<br>Owen Mumford<br>Ltd: 8mm for 31G;<br>12mm for 29G | 4mm for 33G,<br>4mm, 6mm and<br>8mm for 32G, 8mm<br>for 30G, 5mm,<br>6mm and 8mm for<br>31G, 12mm for<br>29G | 4mm for 32G, 5mm<br>and 8mm for 31G<br>and 12.7mm for<br>29G |
| | Effective<br>Gauge | 29G, 31G and 32G | 29G and 31G | 29G, 30G, 31G,<br>32G and 33G | 29G, 31G and 32G |
| | Tip<br>Configura-<br>tion | Patient-side: Tri-<br>bevel (3) | Patient-side: Tri-<br>bevel (3) | Patient-side: Tri-<br>bevel (3) | Patient-side: 5<br>bevel |
| Sterilization | Unifine Pentips<br>manufactured for<br>Owen Mumford<br>Ltd: Ethylene Oxide<br>(validated to<br>achieve SAL 10-6)<br>Unifine Pentips by<br>Owen Mumford:<br>Gamma Radiation<br>(validated to<br>achieve SAL 10-6)<br>Unifine Pentips<br>Plus by Owen<br>Mumford: Gamma<br>Radiation(validated<br>to achieve SAL 10-6) | Unifine Pentips<br>manufactured for<br>Owen Mumford<br>Ltd: Ethylene Oxide<br>(validated to<br>achieve SAL 10-6) | Ethylene Oxide<br>(validated to<br>achieve SAL 10-6) | Gamma Radiation | |
Medical Grade
Medical Grade
• Needle Hub –
information publicly
{6}------------------------------------------------
{7}------------------------------------------------
Though there are minor differences in the Indications for Use between the Predicate Device, Reference Predicate Devices and the Submission Device, the intended use is substantially equivalent. The differences are that the Predicate Device uses terminology of "for use with multi-dose injection devices that use pre-filled cartridges" and the Submission Device uses terminology of "pen injector devices for the subcutaneous injection of drugs, including insulin". Multi-dose injection pens used with pre-filled cartridges were the injectors available at the time of the original 510(k). Since that time injector pens that are disposable that include the pre-filled cartridge have become prevalent in the market. Both injectors operate using the same principle, a pen needle is screwed onto the end and the drug is then injected.
We have tested our pen needles with both a re-useable pen injectors (which we referred to as "multi-dose injection device" in the original 510(k)) and disposable pen injectors. Our pen needles operate as intended on both re-useable pen injectors and disposable pen injectors.
{8}------------------------------------------------
The differences in the Indications for Use do not affect performance of the devices, or the safety and effectiveness of the devices. Therefore, Unifine Pentips and Unifine Pentips Plus are substantially equivalent to the Predicate Device and the Reference Predicate Devices in the device characteristics.
# 7. Performance Data
#### Non-clinical Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
#### Bench testing
Table 7.1: List of Performance Tests to Standards
| Test | Standard | Result |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Dimensions for needles | ISO 11608-2:2012 Section 4.2.2 | Meets standard |
| Needle points | ISO 11608-2:2012 Section 4.5 | Meets standard |
| Freedom from defects | ISO 11608-2:2012 Section 4.6 | Meets standard |
| Lubrication | ISO 11608-2:2012 Section 4.7 | Meets standard |
| Dislocation of measuring point at<br>patient end | ISO 11608-2:2012 Section 4.8 | Meets standard |
| Bond between hub and needle<br>tube | ISO 11608-2:2012 Section 4.4 | Meets standard |
| Ease of assembly/ disassembly | ISO 11608-2:2012 Section 4.10 | Meets standard |
| Sterilization | ISO 11608-2:2012 Section 4.11,<br>ISO 11137-2:2007 (manufactured<br>pen needles in the UK), ISO<br>11135-1:2006(manufactured pen<br>needles in Italy) | Meets standard |
| Flow rate | ISO 11608-2:2012 Section 4.3 | Meets standard |
| Needle dose accuracy | ISO 11608-2:2012 Section 4.9,<br>11.4.2 | Meets standard |
| Needle hub torque removal | ISO 11608-2:2012 Section 4.9,<br>11.4.3 | Meets standard |
Additionally, performance testing other than the above ISO Standards was performed on the devices. The devices comply with the acceptance criteria established based on the specifications of the devices.
Table 7.2: List of Internal Performance Tests
| Performance Test | Results |
|--------------------------------------------------------------|---------------------------|
| Force required to remove pen needle from primary container | Meets acceptance criteria |
| Force required to fit pen needle into primary container (UP) | Meets acceptance criteria |
| Force to remove needle shield from pen needle | Meets acceptance criteria |
The results of these tests demonstrate that the Unifine Pentips and Unifine Pentips Plus devices perform equivalent to the predicate devices when used as intended.
# Biocompatibility
In accordance with ISO 10993-1, Unifine Pentips and Unifine Pentips Plus are classified as: Externally Communicating Device, Blood Path Indirect, Short Term (<24 hours) Use, as the cannula is immediately withdrawn after injection into the body. The biocompatibility evaluation for finished devices' blood/body contacting parts were conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1:
{9}------------------------------------------------
Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- . Cytotoxicity
- . Sensitization
- Irritation ●
- Systemic Toxicity ●
- Hemocompatibility .
- LAL & Rabbit Material Mediated Pyrogen Testing ●
Results of the testing demonstrate the devices are biocompatible.
# Sterilization
The sterility of the devices is assured using a sterilization method validated in accordance with ISO 11137 – Medical Devices – Validation and Routine Control of Radiation Sterilization for Unifine Pentips and Unifine Pentips Plus manufactured by Owen Mumford UK. The sterility of Unifine Pentips manufactured for Owen Mumford by Artsana S.p.A. is assured using a sterilization method validated in accordance with ISO 11135 - Medical Devices -Validation and Routine Control of Ethylene Oxide Sterilization. Through both sterilization methods used, all Unifine Pentips and Unifine Pentips Plus are sterilized to provide a Sterility Assurance Level (SAL) of 10-6.
# 8. Clinical Data
Clinical Data per 807.92(b)(2) was not required to establish substantial equivalence for these devices.
# 9. Conclusions
In summary, the Unifine Pentips and Unifine Pentips Plus devices, subject of this 510(k), are substantially equivalent in their intended use, technology/principle of operation, materials, and performance to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.