The Strange Bile Duct Stone Exploration Set is intended to be used for cholangiography and bile duct stone retrieval using fluoroscopy. The device is indicated for adults only.
Device Story
Set includes NCompass Nitinol Tipped Stone Extractor, radiopaque polyurethane cholangiography catheter, introducer catheter with needle stylet, and PTFE-coated wire guide. Used for biliary stone retrieval and cholangiography; operated by physicians under fluoroscopic guidance. Device components deployed into biliary duct; basket mechanism captures stones. Provides visualization and mechanical retrieval capability. Benefits include combined procedural utility for stone exploration and imaging in single sterile kit.
Clinical Evidence
Bench testing only. Dimensional and compatibility evaluations performed on accelerated aged samples; all acceptance criteria met.
Technological Characteristics
Components: Nitinol wire, stainless steel, polyurethane, PTFE, polyimide. Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: 115cm extractor, 7.5Fr cholangiography catheter, 10Fr introducer. Single-use, sterile. No software or electronic components.
Indications for Use
Indicated for adults requiring cholangiography and bile duct stone retrieval via fluoroscopy.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2017
Cook Incorporated Colin Jacob Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47402
Re: K173687
> Trade/Device Name: Strange Bile Duct Stone Exploration Set Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: II Product Code: LOR, GCD, GBZ, OCY, GCB Dated: November 30, 2017 Received: December 1, 2017
Dear Colin Jacob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the Strange Bile Duct Stone Exploration Set have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173687
Device Name Strange Bile Duct Stone Exploration Set
Indications for Use (Describe)
The Strange Bile Duct Stone Exploration Set is intended to be used for cholangiography and bile duct stone retrieval using fluoroscopy. The device is indicated for adults only.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K173687
Page 1 of 4
COOK
MEDICAL
750 DANIELS WAY, P.C
BLOOMINGTON, IN 47402-C
PHONE: 812.339.2235 TOLL FREE: 800
WWW.COOKMED
#### 510(k) Summary
#### Strange Bile Duct Stone Exploration Set (21 CFR §807.92) Date Prepared: November 30, 2017
#### Submitted Bv:
Applicant: Contact: Applicant Address: Cook Incorporated Colin Jacob Cook Incorporated 750 Daniels Way Bloomington, IN 47402 (812) 335-3575 x 104965 (812) 332-0281
#### Device Information:
Contact Fax Number:
Contact Phone Number:
Trade Name: Panel: Regulation: Regulation Name: Primary Product Code: Secondary Product Codes:
Strange Bile Duct Stone Exploration Set Gastroenterology/Urology 21 CFR § 876.5010 Biliary catheter and accessories LOR GCD, GBZ, OCY, GCB
## Predicate Devices:
The predicate device is the NCompass Nitinol Stone Extractor (K173009).
## Device Description:
The Strange Bile Duct Stone Exploration Set is provided sterile for single-use only. The set consists of:
- . a 115 centimeter long NCompass Nitinol Tipped Stone Extractor with Tuohy Borst adapter
- a 7.5 French, double lumen, 40 centimeter long radiopaque polyurethane cholangiography catheter with a preloaded inner catheter;
- . a 10 French, 12 centimeter long introducer catheter with preloaded needle stylet; and
- a 125 centimeter long, polytetrafluoroethylene coated wire guide.
## Indications for Use:
The Strange Bile Duct Stone Exploration Set is intended to be used for cholangiography and bile duct stone retrieval using fluoroscopy. The device is indicated for adults only.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, bold, white letters on a red square. Below the red square is a smaller red rectangle with the word "MEDICAL" in white letters. The logo is simple and clean, with a focus on the company name.
COOK
INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A..
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
#### Comparison to Predicates:
The Strange Bile Duct Exploration Set and the predicate device, the NCompass Nitinol Stone Extractor (K173009), are substantially equivalent in that these devices have the same intended use and principles of operation, and similar indications for use. The differences between the subject device and the predicate device include:
- . Additional components.
- " Assist in cholangiography via the additional components.
A comparison table on the following page elaborates on the features of the subject device and the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters at the top. Below "COOK" is the word "MEDICAL" in smaller, white letters, set against a darker red background.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
#### Substantial Equivalence Comparison
| | | PREDICATE DEVICE | SUBJECT DEVICE |
|---------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| | | NCompass Nitinol Stone<br>Extractor<br>(K173009) | Strange Bile Duct Stone<br>Exploration Set |
| | Regulation Number | 21 CFR § 876.5010 | IDENTICAL TO PREDICATE |
| | Primary Product Code | LQR | IDENTICAL TO PREDICATE |
| | Classification | II | IDENTICAL TO PREDICATE |
| | Indications for Use | For extraction of stones or debris<br>during biliary surgical procedures | For cholangiography and bile duct<br>stone retrieval using fluoroscopy |
| | One-time Use | Yes | IDENTICAL TO PREDICATE |
| | Duration of Use | Limited (≤ 24 hours) | IDENTICAL TO PREDICATE |
| | Principles of Operation | Basket deployment from sheath into<br>the biliary duct location | Basket deployment from sheath into<br>the biliary duct location. In addition<br>to cholangiography of the duct. |
| | Imaging Technique to Visualize<br>Device | Fluoroscopy | IDENTICAL TO PREDICATE |
| | Tip Material | Stainless steel | IDENTICAL TO PREDICATE |
| | Wire Material | Nitinol | IDENTICAL TO PREDICATE |
| Basket | Wire Configuration | 12-wire bulb configuration | IDENTICAL TO PREDICATE |
| | Length (cm) | 1.9-2.3 | IDENTICAL TO PREDICATE |
| | Outer Diameter when<br>Deployed (cm) | 1.5 | IDENTICAL TO PREDICATE |
| Catheter | Material | Braided stainless steel tubing coated<br>with polytetrafluoroethylene<br>(PTFE), fluorinated ethylene<br>propylene (FEP), and polyimide | IDENTICAL TO PREDICATE |
| | Length (cm) | 115 | IDENTICAL TO PREDICATE |
| | Outer Diameter (Fr) | 2.4 | IDENTICAL TO PREDICATE |
| | Adapter | Tuohy Borst adapter | IDENTICAL TO PREDICATE |
| | Additional Components | None | Cholangiography catheter, wire<br>guide, introducer catheter with<br>preloaded needle stylet |
| Packaging | | Polyethylene-Polyester/Tyvek | IDENTICAL TO PREDICATE |
| Sterilization Method | | Ethylene Oxide | IDENTICAL TO PREDICATE |
| Sterility Assurance Level (SAL) | | 10-6 | IDENTICAL TO PREDICATE |
# Technological Characteristics:
The subject device, Strange Bile Duct Stone Exploration Set, was subjected to applicable nonclinical testing to assure reliable design and performance under the testing parameters. Testing includes:
- י Dimensional and Compatibility Evaluation of subject device set (Accelerated Aged)
All pre-determined acceptance criteria were met.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font, also on a red background.
WWW.COOKMEDICAL.COM
#### Conclusion:
The Strange Bile Duct Stone Exploration Set meets the design input requirements based on the intended use. Furthermore, the results and comparison supports the conclusion that the Strange Bile Duct Stone Exploration Set does not raise new questions of safety or effectiveness and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.