The NCompass Nitinol Stone Extractors are intended for extraction of stones or debris during biliary surgical procedures.
Device Story
NCompass Nitinol Stone Extractor is a single-use, sterile, disposable device for biliary stone/debris retrieval. Device consists of a nitinol basket (12 or 16-wire bulb shape) and a 2.4 Fr shaft diameter sheath. Operated by a physician during biliary surgical procedures; device is deployed from the sheath into the anatomical duct to capture stones. Visualization is achieved via fluoroscopy (NC3 model) or choledochoscope imaging (NCT4 model). The device includes a handle and Touhy-Borst adapter. By facilitating mechanical stone extraction, the device assists in clearing biliary obstructions, potentially reducing patient complications associated with retained stones.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), dimensional/compatibility evaluation, radiopacity, multiple stone retrieval testing, and various tensile tests (basket formation, coil, sheath, and connection points) on accelerated aged samples. All pre-determined acceptance criteria were met.
Technological Characteristics
Nitinol (nickel-titanium alloy) basket; 12 or 16-wire bulb configuration. Sheath composed of braided stainless steel tubing coated with PTFE, FEP, and Polyimide. 2.4 Fr shaft diameter. Radiopaque markers (stainless steel on NC3 tip). Sterilized via Ethylene Oxide (SAL 10^-6). Single-use. Mechanical actuation via handle.
Indications for Use
Indicated for extraction of stones or debris during biliary surgical procedures in patients requiring such intervention.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 7, 2017
Cook Incorporated Colin Jacob Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, Indiana 47402
Re: K173009
> Trade/Device Name: NCompass Nitinol Stone Extractors Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: LOR Dated: September 27, 2017 Received: September 27, 2017
Dear Colin Jacob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173009
#### Device Name NCompass Nitinol Stone Extractors
Indications for Use (Describe)
The NCompass Nitinol Stone Extractors are intended for extraction of stones or debris during biliary surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
### NCompass Nitinol Stone Extractors (21 CFR §876.5010) Date Prepared: September 27, 2017
## Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|-------------------------------------------------------------------------------|
| Contact: | Colin Jacob |
| Applicant Address: | Cook Incorporated<br>P.O. Box 489<br>750 Daniels Way<br>Bloomington, IN 47402 |
| Contact Phone Number: | (812) 335-3575 x 104965 |
| Contact Fax Number: | (812) 332-0281 |
Contact Fax Number:
## Device Information:
| Trade Name: | NCompass Nitinol Stone Extractor |
|----------------------------|----------------------------------|
| Common Name: | Dislodger, Stone, Biliary |
| Classification Name/Panel: | Gastroenterology/Urology |
| Regulation: | 21 CFR § 876.5010 |
| Regulation Name: | Biliary Catheter and Accessories |
| Product Code: | LQR |
# Predicate Device:
- 트 Wittich Nitinol Stone Basket (K902944)
# Device Description:
The NCompass Nitinol Stone Extractor is a single-use, sterile, disposable device. The extractor is a stone retrieval device consisting of nitinol basket with a 1.5 cm wide diameter when fully expanded and a 2.4 Fr shaft diameter sheath. The device is packaged with a Touhy-Borst adapter accessory.
# Indications for Use:
The NCompass Nitinol Stone Extractors are intended for extraction of stones or debris during biliary surgical procedures.
# Comparison to Predicate:
The NCompass Nitinol Stone Extractors and the predicate device, Wittich Nitinol Stone Basket (K902944), have the same intended use; they also share similar indications for use and principle
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, sans-serif, uppercase letters in the top portion of the logo. Below that, the word "MEDICAL" is in white, sans-serif, uppercase letters in a smaller font size.
of operation. The differences in technological characteristics do not raise different questions of safety and/or effectiveness when compared to the predicate device.
### Substantial Equivalence Comparison
| | | PREDICATE DEVICE | SUBJECT DEVICE |
|------------------------------------------|--------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | | Wittich Nitinol Stone Basket<br>(K902944) | NCompass Nitinol Stone Extractor |
| Regulation Number | | 21 CFR § 876.5010 | IDENTICAL TO PREDICATE |
| Product Code | | LQR | IDENTICAL TO PREDICATE |
| Classification | | II | IDENTICAL TO PREDICATE |
| Indications for Use | | For nonoperative removal of stones<br>from the biliary tract, renal pelvis and<br>ureter | For extraction of stones or debris during<br>biliary surgical procedures. |
| One-time Use | | Yes | IDENTICAL TO PREDICATE |
| Duration of Use | | Limited (≤ 24 hours) | IDENTICAL TO PREDICATE |
| Principle of Operation | | Basket deployment from sheath into an<br>anatomical duct | IDENTICAL TO PREDICATE |
| Imaging Technique to Visualize<br>Device | | Fluoroscopy | NC3 model: IDENTICAL TO<br>PREDICATE<br>NCT4 model: Choledochoscope Imaging |
| Basket | Fluoroscopic Radiopaque<br>Material | Tungsten/Rhenium band<br>(located on sheath tip) | Stainless Steel<br>(located only on NC3 basket tip) |
| | Tip Material | None (tipless) | NC3 model: Stainless steel<br>NCT4 model: None (tipless) |
| | Wire Material | Nitinol (nickel and<br>titanium alloy) | IDENTICAL TO PREDICATE |
| | Wire Configuration | 6-wire bulb shape | 12-wire and 16-wire bulb shape |
| | Length (cm) | 4.5 | NC3 model: 1.9-2.3<br>NCT4 model: 2.1-2.5 |
| | Outer Diameter when<br>Deployed (cm) | 2 | 1.5 |
| Catheter / Sheath | Material | Radiopaque tetrafluoroethylene | Braided stainless steel tubing coated<br>with Polytetrafluoroethylene (PTFE),<br>fluorinated ethylene propylene (FEP),<br>and Polyimide |
| | Length (cm) | 24 | 115 |
| | Outer Diameter (Fr) | 12 | 2.4 |
| Additional Components | | Dilator, introducer sheath | Handle, Touhy-Borst Adapter |
| Packaging | | Polyethylene-Polyester/Tyvek | IDENTICAL TO PREDICATE |
| Sterilization Method | | Ethylene Oxide | IDENTICAL TO PREDICATE |
| Sterility Assurance Level (SAL) | | 10-6 | IDENTICAL TO PREDICATE |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font on a darker red background.
## Technological Characteristics:
The subject devices, NCompass Nitinol Stone Extractors, were subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests are listed below:
- . Biocompatibility - Per ISO 10993-1, testing for cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity were performed to ensure the biocompatibility of the subject device set. Test results indicated that all materials are biocompatible.
- Dimensional and Compatibility Evaluation (Accelerated Aged) 트
- Radiopacity Evaluation (Accelerated Aged) .
- . Multiple Stone Retrieval Testing (Zero Time)
- Tensile Test: Basket Formation (Accelerated Aged) 트
- . Tensile Test: Basket Coil (Accelerated Aged)
- . Tensile Test: Basket Sheath (Accelerated Aged)
- 트 Tensile Test: Basket Sheath to Handle Connection (Accelerated Aged)
- 트 Tensile Test: Basket Assembly to Shaft Connection (Accelerated Aged)
- Tensile Test: Cannula to Basket Coil Connection (Accelerated Aged) 트
#### Conclusion:
For these tests, all pre-determined acceptance criteria were met. The results of these tests show that the NCompass Nitinol Stone Extractors meet the design input requirements based on the intended use. Furthermore, these results support the conclusion that the NCompass Nitinol Stone Extractors do not raise new questions of safety or effectiveness and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.