Use the 3M™ Attest™ Rapid Readout Biological Indicator in conjunction with the 3M™ Attest™ Auto reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the Amsco® V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), and in STERRAD® 100S, STERRAD® NX® (Standard and Advanced cycles) and STERRAD® 100NX® (Standard, Flex, Express, and Duo cycles) systems.
Device Story
System comprises 3M Attest Rapid Readout Biological Indicator 1295 (self-contained BI) and 3M Attest Auto-reader 490H. BI contains Geobacillus stearothermophilus spores and media ampoule in polycarbonate sleeve; cap features chemical process indicator (blue-to-pink). User places BI in sterilizer; post-cycle, BI is activated and inserted into Auto-reader. Auto-reader incubates BI at 60°C; detects fluorescence via enzymatic activity. Fluorescence indicates sterilization failure. Software update reduces final readout time from 4 hours to 24 minutes. Used in clinical settings for sterilization quality assurance; provides rapid feedback to healthcare staff regarding sterilization efficacy; ensures patient safety by confirming successful sterilization of instruments.
Clinical Evidence
No clinical data included. Effectiveness demonstrated via bench testing: positive control, survival/kill time (survival ≥ 5s, kill = 7 min), and reduced incubation time validation (24-minute readout vs 7-day conventional).
Indicated for routine monitoring of vaporized hydrogen peroxide sterilization processes in specified Amsco and STERRAD low-temperature sterilization systems.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2017
3M Company c/0 Hilary Hovde Regulatory Affairs Associate 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144
Re: K171003
Trade/Device Name: 3M Attest Rapid Readout Biological Indicator 1295, 3M Attest Autoreader 490H Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: June 29, 2017 Received: June 30, 2017
Dear Hilary Hovde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Tara A. Ryan -S
for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171003
Device Name
3M™ Attest™ Rapid Readout Biological Indicator 1295 and 3M™ Attest™ Auto-reader 490H
Indications for Use (Describe)
Use the 3M™ Attest™ Rapid Readout Biological Indicator in conjunction with the 3M™ Attest™ Auto reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the Amsco® V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), and in STERRAD® 100S, STERRAD® NX® (Standard and Advanced cycles) and STERRAD® 100NX® (Standard, Flex, Express, and Duo cycles) systems.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is in red and features a large, bold "3M" in a sans-serif font. The letters are closely spaced and appear to be slightly rounded. The background is white, providing a strong contrast to the red letters.
# 510(k) Summary for 3M™ Attest™ Rapid Readout Biological Indicator 1295 and 3M™ Attest™ Auto-reader 490H
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
> Contact: Hilary B. Hovde Regulatory Affairs Associate Phone Number: (651) 736-0364 FAX Number: (651) 737-5320
Submission Date: July 20, 2017
K171003
{4}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k)] 3MTM Attest™ Rapid Readout Biological Indicator 1295 and 3MTM Attest™ Auto-reader 490H
#### Device Name and Classification:
| Common or Usual Name: | Biological Indicator |
|------------------------|-------------------------------------------------------------------------------------|
| Proprietary Name: | 3M™ Attest™ Rapid Readout Biological Indicator 1295<br>3M™ Attest™ Auto-Reader 490H |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR § 880.2800(a)) |
| Device Classification: | Class II |
| Product Code: | FRC |
#### Predicate Device:
- Intended use Predicate for system K160546 3M™ Attest™ Rapid Readout . Biological Indicator 1295
- . 3MTM AttestTM Auto-reader 490H, K140392
#### Description of Device:
The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is a self-contained biological indicator specifically designed for rapid and reliable routine monitoring of vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3M™ Attest™ Auto-reader 490H. The 1295 BI is a single-use device composed of a polycarbonate sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. A chemical process indicator printed with stripes which change from blue toward pink upon exposure to vaporized hydrogen peroxide is located on the top of the cap. The detection of fluorescence upon incubation of the 1295 BI in the 490H Auto-reader indicates a sterilization failure.
#### Nonclinical Comparison to the Predicate Device
This submission is addressing a software change to the 3MTM Attest™ Auto-reader 490H to reduce the final fluorescent readout from 4 hours to 24 minutes. The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is the same design as the previously cleared device of the same model number. The device has the same materials and fundamental scientific technology. The only change to the performance specifications is the change in final fluorescent readout from 4 hours to 24 minutes.
{5}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k)] 3M™ Attest™ Rapid Readout Biological Indicator 1295 and 3M™ Attest™ Auto-reader 490H
#### Summary of Nonclinical Testing
The effectiveness of the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M™ Attest™ Auto-reader 490H with a final fluorescent readout of 24 minutes is demonstrated in the following tests:
| Test | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Positive Control | Passed |
| Survival Time = 5 seconds<br>Kill Time = 7 minutes | Passed |
| Reduced Incubation Time<br>Meets FDA's requirements for Reduced Incubation Time with > 97%<br>alignment with the conventional incubation time of 7 days for the<br>following readout time:<br>• 24 minutes | Passed |
## Summary of Clinical Testing
No clinical data was included in this premarket application submission.
## Indications for Use
Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the AMSCO® V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), and in STERRAD® 100S, STERRAD® NX® (Standard and Advanced cycles) and STERRAD® 100NX® (Standard, Flex, Express and Duo cycles) systems.
{6}------------------------------------------------
## Comparison to Predicate Device
| Feature | Submission Device:<br>3MTM AttestTM Rapid Readout<br>Biological Indicator 1295 and<br>3MTM AttestTM Auto-reader 490H | Predicate Device (K160546):<br>3MTM AttestTM Rapid Readout<br>Biological Indicator 1295 and<br>3MTM AttestTM Auto-reader 490H |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Use the 3MTM AttestTM Rapid Readout<br>Biological Indicator 1295 in conjunction<br>with the 3MTM AttestTM Auto-reader<br>490H as a standard method of routine<br>monitoring of vaporized hydrogen<br>peroxide sterilization processes in the<br>Amsco® V-PRO® maX Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen, and Flexible<br>cycles), and in STERRAD® 100S,<br>STERRAD® NX® (Standard and<br>Advanced cycles) and STERRAD®<br>100NX® (Standard, Flex, Express and<br>Duo cycles) systems. | Identical |
| Organism | Geobacillus stearothermophilus<br>traceable to ATCCTM 7953 | Identical |
| Viable spore<br>population | ≥ 1 x 106 | Identical |
| Resistance<br>characteristics | (Tested at 10 mg/L vaporized hydrogen<br>peroxide)<br>D-value<br>Survival/Kill Window<br> | Identical |
| • D-value<br>• Survival/Kill<br>Window | $D_{10 mg/L} ≥ 1 second$<br>Survival Time ≥ 5 seconds<br>Kill Time = 7 minutes | Identical |
| Carrier material | polyethylene terephthalate | Identical |
| Incubation<br>temperature | 60 ± 2°C | Identical |
| Readout time | 24 minute fluorescence result read in the<br>3MTM AttestTM Auto-reader 490H | 4 hour fluorescence result read in the<br>3MTM AttestTM Auto-reader 490H |
| Chemical<br>indicator | H2O2 sensitive ink; changes from blue<br>towards pink | Identical |
| Shelf-life | 24 months | Identical |
## Conclusion
The 3M™ Attest™ Rapid Readout Biological Indicator 1295 and the 3MTM Attest™ Autoreader 490H meet all applicable standards and are substantially equivalent to their predicate devices in terms of their intended use, physical properties and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.