Merge PACS™ is a Picture Archiving and Communication System (PACS) for multi-modality (CT, MR, PT, US, MG, BTO, CR, DR/DX, NM, XA, RF, secondary capture (SC), and other DICOM-compliant modalities) image processing and display, diagnostic reading and reporting, communication, printing, and storage of medical imaging studies and other patient data. Intended clinical users include radiologists, orthopedic and other surgeons, referring physicians, technologists, and other qualified medical professionals. Data can be received directly from acquisition modalities, CAD systems, and other image processing systems, or indirectly via importing. Data that is not DICOM-compliant, such as photos, can be converted to DICOM format by Merge PACS. Merge PACS provides image manipulation tools to enable users to view and compare images such as: linkinq, MPR, MIP, 3D image fusion/registration of CT, MR, and PET; as well as CVR (Color Volume Rendering), measurements (linear distances, angles, areas, SUV, etc.), and annotations (for example, outline and label regions of interest, label spinal vertebrae). The Real Time Worklist (RTWL) displays the real-time status of radiology activity and provides customizable workflow management capabilities. Communication of critical results is facilitated and documented through optional, configurable components. The Patient Dashboard provides a composite view of patient data, both imaging and non-imaging. The optional Reach component provides clinicians with secure, proactive communication and access to clinical reports and images. Multi-tier patient identity matching provides a comprehensive view even when dealing with multiple disparate patient identities. Order and report information generated by the HIS/RIS and report creation systems are received and displayed via the transmission of HL7 messaging. Lossless (reversible) and lossy (irreversible) image compression are supported for viewing, storage and communication. Merge PACS displays full fidelity DICOM images for use in the diagnostic interpretation of mammography using MG or BTO images. Thick slab MIP presentation can be applied to BTO images. Lossy compressed images and digitized screen film images must not be used for primary diagnosis of mammography studies, and only display monitors that have regulatory clearance for mammography interpretation should be used for the interpretation of mammography studies.
Device Story
Merge PACS is a software-based diagnostic workstation for viewing, reporting, and managing medical images. It ingests DICOM-compliant data from modalities (CT, MR, PET, etc.) and non-DICOM data (photos). It provides tools for image manipulation, including MPR, MIP, 3D image fusion/registration, and SUV measurements for PET. It operates on off-the-shelf hardware, supporting standalone or enterprise-integrated workflows (HIS/RIS/EMR). Clinicians use the system to view images, perform measurements, and document findings. It facilitates clinical decision-making by providing a composite patient view and workflow management. Benefits include efficient multi-modality image review and standardized quantitative analysis for tumor activity assessment.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via in-house non-clinical software testing and compliance with design controls. Performance of image fusion and SUV measurements verified against RSNA/QIBA guidelines using Digital Reference Objects (DRO).
Technological Characteristics
Software-based diagnostic workstation; runs on off-the-shelf computer hardware. Supports DICOM, HL7, JPEG compression. Features include 64-bit viewer, image fusion (rigid transformation), and SUV calculation. Connectivity via network integration with VNA, RIS, and HIS. No physical materials; software-only device.
Indications for Use
Indicated for multi-modality medical image processing, display, diagnostic reading, reporting, communication, printing, and storage. Intended for use by radiologists, surgeons, referring physicians, and technologists. No specific disease state or patient population contraindications specified.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
December 8, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Merge Healthcare Incorporated % Ms. Carol Nakagawa Senior Manager of Regulatory Affairs 900 Walnut Ridge Drive HARTLAND WI 53029
Re: K173475
Trade/Device Name: Merge PACS™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 2, 2017 Received: November 9, 2017
Dear Ms. Nakagawa:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K173475
Device Name
Merge PACS(TM)
#### Indications for Use (Describe)
Merge PACS™ is a Picture Archiving and Communication System (PACS) for multi-modality (CT, MR, PT, US, MG, BTO, CR, DR/DX, NM, XA, RF, secondary capture (SC), and other DICOM-compliant modalities) image processing and reporting, communication, printing, and storage of medical imaging studies and other patient data. Intended clinical users include radiologists, orthopedic and other surgeons, referring physicians, technologists, and other qualified medical professionals.
Data can be received directly from acquisities. CAD systems, and other image processing systems, or indirectly via importing. Data that is not DICOM-compliant, such as photos, can be converted to DICOM format by Merge PACS.
Merge PACS provides inage manipulation tools to view and compare images such as: linking. MPR, MP, 3D image fusion! reqistration of CT, MR, and PET; as well as CVR (Color Volume Rendering), measurements (linear distances, angles, areas, SUV, etc.), and annotations (for example, outline and label regions of interest, label spinal vertebrae).
The Real Time Worklist (RTWL) displays the real-time status of radiology activity and provides customizable workflow management capabilities. Communication of critical results is facilitated and documented through optional, components.
The Patient Dashboard provides a composite view of patient data, both imaging. The optional Reach component provides clinicians with secure, proactive communication and images. Multi-tier patient identity matching provides a comprehensive view even when dealing with multiple disparate patient identities.
Order and report information generated by the HS/R/S and report creation systems are received and displayed via the transmission of HL7 messaging.
Lossless (reversible) and lossy (irreversible) image compression are supported for viewing, storage and communication. Merge PACS displays full fidelity DICOM images for use in the diagnostic interpretation of mammography using MG or BTO images. Thick slab MIP presentation can be applied to BTO images.
Lossy compressed images and digitized screen film images must not be used for primary diagnosis of mammography studies, and only display monitors that have regulatory clearance for mammography interpretation of mammography studies.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in a combination of blue and orange colors, with the top half of each letter in orange and the bottom half in blue. Below the word "MERGE", the text "An IBM Company" is written in a smaller, blue font. The logo is simple and modern, reflecting the company's focus on technology and innovation.
# 510(k) Summary
#### l. SUBMITTER
Merge Healthcare Incorporated 900 Walnut Ridge Drive Hartland, Wisconsin 53029 U.S.A.
Tel: 262.367.0700 Fax: 262.367.0717
Contact Person: Carol Nakagawa, RAC Sr. Manager of Requlatory Affairs Merge Healthcare, an IBM Company Tel.: 905.364.8027 Fax: 905.364.8100
November 3, 2017 Date Prepared:
#### ll. DEVICE
Merge PACS™ Proprietary Name of Device:
- Picture Archiving and Communications Common or Usual Name: System (PACS)
- Radiological Image Processing System (21 Classification Name: CFR 892.2050)
- Regulatory Class: Class II
- LLZ Product Code:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in large, bold letters, with the first half of the word in blue and the second half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller, blue font. The logo is simple and modern, and it clearly identifies the company as a subsidiary of IBM.
# III. PREDICATE DEVICES
| Trade Name | 510(k) Submitter | 510(k) Number |
|----------------------------------------------|------------------------------------------|-----------------------------------------|
| Primary Predicate Device:<br>AMICAS PACS 6.0 | AMICAS, Inc. | K082144<br>Class II<br>Product Code LLZ |
| Synapse PACS | Fujifilm Medical Systems<br>U.S.A., Inc. | K160108<br>Class II<br>Product Code LLZ |
#### DEVICE DESCRIPTION IV.
Merge PACS™, a software medical device, is a standards-based medical imaging diagnostic workstation that serves as an adjunct to assist the clinician to view, read, and report their findings. Merge PACS processes and displays medical images from DICOM-compliant modalities. The device is designed to enable efficient workflows by maintaining clinicians' worklists and retrieving and managing studies for reading, reporting, communication, and storage.
Merge PACS software runs on off-the-shelf computer hardware and can be configured to operate standalone or to integrate with vendor-neutral imaging archives (VNAs) such as iConnect Enterprise Archive (iCEA) for image storage, and with radiological and hospital information systems (RIS and HIS) and medical record systems (EMR, EHR, etc.).
Merge PACS can be accessed from within the hospital or enterprise, or from remote locations via web-based access. Images viewed on mobile devices must not be used for diagnostic review.
The focus of this premarket notification is on the addition of the ability to "fuse" images for viewing (image fusion) and on the ability to measure Standardized Uptake Values (SUV) on PET (Positron Emission Tomography) images.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in large, bold letters, with the first half of the word in blue and the second half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size.
# Image Fusion
The primary use for image fusion is to view PET and CT images overlaid on top of each other in addition to the previously available ability to view them side-by side in different viewports, or sequentially one after the other.
The ability to view images overlaid on top of each other in one viewport does not change the intended use of Merge PACS, which is intended to be used for multi-modality image viewinq in 2D and 3D. The fusion of the display of two images does not change the fundamental image processing that is used. The original images are always available and each can be faded completely in or out in the fused display. The unfused image is the default display.
PET and CT images from modern scanners are acquired with a common DICOM frame of reference. The spatial processing of the PET and CT image objects uses the same algorithms (i.e., having a common DICOM frame of reference provides the basis from which the images can either be crossreferenced with localizer lines or registered with each other). The DICOM frame of reference, pixel spacinq, slice spacinq, orientation vector and other data already provide the technological basis for the display of localizer or cross reference lines and the 3D cursor tool. These are core functions of most PACS viewers including those that support 3D operations (such as Merge PACS under its previous 510(k) clearance) and are applied now to image registration for image fusion.
Alignment is done as a simple rigid transformation allowing for translation and rotation about each of the axes of the 3D image volumes. This is similar to the existing technology used to provide manual linking of studies (e.q., to enable side-by-side simultaneous scrolling of two studies in different viewports) or to allow re-linking with an offset for common-frame-ofreference series or studies, as was previously cleared for Merge PACS.
Merge PACS does not enable warping or non-rigid transformation of the images, nor automation of the realignment or registration adjustment process.
The original images are always available and the most common viewing paradigm is to view the fused (corrected PET and CT) images alongside the unfused corrected PET images.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in large, bold letters, with the top half of each letter in blue and the bottom half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size and blue color.
# Standardized Uptake Values (SUV)
SUVs (Standardized Uptake Values) are calculated for FDG PET studies to judge the biological activity of a tumor and its response to therapy.
The SUV calculation can be performed for individual pixels or on a Region Of Interest (ROI) as an adjunct to viewing the relative intensities of different areas of the PET images.
SUV measurements can be made in Merge PACS discretely via a point / pixel measurement tool, or via SUVmean measurements via a planar elliptical ROI measurement tool. These raw measurements can be normalized according to any of the four common normalizations:
- · Body Weight (BW)
- · Lean Body Mass (LBM)
- Body Surface Area (BSA) .
- Ideal Body Weight (IBW) .
Merge PACS does not provide automated or semi-automated segmentation of PET lesion volumes. Merge PACS provides manual single point SUV values and SUVs for manually defined ROI. The user determines where they want to measure SUV on the image and decides how to interpret the information, for example whether activity is related to tumor activity or to other processes such as infection or inflammation. Merge PACS does not provide a diagnosis or treatment recommendation.
Merge PACS SUV measurements and image registration meet the RSNA (Radiological Society of North America)/ QIBA (Quantitative Imaging Biomarkers Alliance) guidelines using the DRO (Digital Reference Object) that they developed in conjunction with their guideline document.
The SUV measurements in Merge PACS are based on the existing pixel value and ROI tools of AMICAS PACS 6.0. The RSNA/ QIBA testing and qualification process for SUV measurements is being implemented to have an independent means of assessing the measurement and registration accuracy.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in a combination of blue and orange colors. The letters "M, E, R" are in blue, while the letters "G, E" are in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size.
#### INTENDED USE / INDICATIONS FOR USE V.
Merge PACS™ is a Picture Archiving and Communication System (PACS) for multi-modality (CT, MR, PT, US, MG, BTO, CR, DR/DX, NM, XA, RF, secondary capture (SC), and other DICOM-compliant modalities) image processing and display, diagnostic reading and reporting, communication, printing, and storage of medical imaging studies and other patient data.
Intended clinical users include radiologists, orthopedic and other surgeons, referring physicians, technologists, and other qualified medical professionals. Data can be received directly from acquisition modalities, CAD systems, and other image processing systems, or indirectly via importing. Data that is not DICOM-compliant, such as photos, can be converted to DICOM format by Merge PACS.
Merge PACS provides image manipulation tools to enable users to view and compare images such as: linkinq, MPR, MIP, 3D image fusion/registration of CT, MR, and PET; as well as CVR (Color Volume Rendering), measurements (linear distances, angles, areas, SUV, etc.), and annotations (for example, outline and label regions of interest, label spinal vertebrae).
The Real Time Worklist (RTWL) displays the real-time status of radiology activity and provides customizable workflow management capabilities. Communication of critical results is facilitated and documented through optional, configurable components.
The Patient Dashboard provides a composite view of patient data, both imaging and non-imaging. The optional Reach component provides clinicians with secure, proactive communication and access to clinical reports and images. Multi-tier patient identity matching provides a comprehensive view even when dealing with multiple disparate patient identities.
Order and report information generated by the HIS/RIS and report creation systems are received and displayed via the transmission of HL7 messaging. Lossless (reversible) and lossy (irreversible) image compression are supported for viewing, storage and communication.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in large, bold letters, with the first half of the word in blue and the second half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size. The logo is simple and modern, with a clear emphasis on the company name.
Merge PACS displays full fidelity DICOM images for use in the diagnostic interpretation of mammography using MG or BTO images. Thick slab MIP presentation can be applied to BTO images.
Lossy compressed images and digitized screen film images must not be used for primary diagnosis of mammography studies, and only display monitors that have regulatory clearance for mammography interpretation should be used for the interpretation of mammography studies.
#### COMPARISON WITH PREDICATES VI.
Merge PACS and the predicate devices are software products and physical technological characteristics do not apply. The comparison of Merge PACS to the predicate devices is based primarily on the functionality and technological design of their software features.
Merge PACS, AMICAS PACS 6.0, and Synapse PACS devices have the same intended use. Each is a multi-modality diagnostic workstation for viewing, reading, and reporting on medical images from various imaging modalities. They are all "tool type" devices that have the same general indications for use that do not specify a disease, condition, or population.
Each device has essentially the same basic image processing, display, storage, and communication features, based on the same or similar technology as the primary predicate device. The availability of additional specific features may vary between devices. Bundling of subsets of features or toolsets makes the workflow more suitable for clinicians with different specializations, such as mammography or orthopedics, however the image processing technology behind each feature is essentially the same regardless of how the features are bundled.
It is our opinion that Merge PACS is substantially equivalent to AMICAS PACS 6.0, a multi-modality diagnostic workstation that represents an earlier version of Merge PACS and shares the same basic technology and intended uses as Merge PACS.
It is also our opinion that Merge PACS is substantially equivalent to Synapse PACS, which is a multi-modality diagnostic workstation. Synapse PACS has
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MERGE, an IBM company. The word "MERGE" is written in a large, sans-serif font, with the letters "M", "E", "R", and "G" in blue and the letter "E" in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller, sans-serif font.
essentially the same basic image processing, display, storage, and communication features as Merge PACS. Synapse PACS supports 3D image fusion and is capable of performing SUV measurements.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MERGE, an IBM company. The word "MERGE" is written in a bold, sans-serif font, with the letters "M", "E", "R", and "G" in blue and the letter "E" in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller, sans-serif font.
# VII. The following table compares features/technologies of changes
| Category<br>New Features: | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5,<br>2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3,<br>2016 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | |
| Image Display | PET/CT/MR Image Fusion | Yes | No | Yes<br>Rigid 3D fusion<br>(MIP/MPR) of CT,<br>MR, PET, SPECT<br>Non-Rigid 3D fusion<br>(MIP/MPR) of CT,<br>MR, PET, SPECT<br>2D Multi-modality<br>fusion |
| Image Display | PET/CT Fusion enhancements:<br>Rotation of MIP images, manual<br>entry of missing SUV attributes<br>(height, weight, BSA) | Yes | No | Has PET/CT fusion<br>and SUV<br>No significant<br>difference |
| Measurement | SUV Calculation (PET) | Yes | No | Yes |
| Non-Significant<br>Changes: | | | | |
| Category | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5, 2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3, 2016 |
| Image Display | Enhanced mammography tools:<br>Stacked scrolling, Dual<br>magnifier, Binocular, Hotlight | Yes | No | Supports<br>mammography<br>reading<br><br>No significant<br>difference |
| Image Display | BTO – Breast Tomosynthesis<br>Objects (MG Tomosynthesis) | Yes | No | Yes |
| Study Management | Real Time Study List with<br>Saved Searches | Yes | No | Not known |
| NA | Access to Teaching Worklists<br>from iConnect Access | Yes | No | Not a clinically<br>significant feature<br>Similar, sends<br>images and notes to<br>teaching archives |
| Compliance | Credentialing, Licensure, ME<br>filtering | Yes | No | Not known<br>Not a clinically<br>significant feature |
| Study Management | Real-Time Study List (RTSL)<br>Enhancements: Quick filter<br>and worklist folders fly-out,<br>Reading pools, Saved searches,<br>Custom fields, Configurable | Yes | No | Not a clinically<br>significant feature<br>Similar, e.g. has<br>custom work folders<br>and other<br>configurable study |
| Category | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5,<br>2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3,<br>2016 |
| | columns, Due in time display,<br>Column filters and sorting | | | management<br>features<br>Not a clinically<br>significant feature |
| Storage | Use iConnect VNA as the Long<br>Term Archive (iConnect VNA is<br>now known as iConnect<br>Enterprise Archive | Yes | No | Similar, uses own<br>VNA<br>No significant<br>difference |
| Storage | PACS EA (Enterprise Archive)<br>interoperation | Yes | No | Yes |
| Interoperability | Improved DICOM alignment<br>between PACS and EA | Yes | No | Not known |
| Study Management | Automatic archive retrieval | Yes | No | Yes |
| Interoperability,<br>Security | Tight integration between PACS<br>and EA: common security<br>model, viewer loads directly<br>from EA, seamless integration<br>between PACS and EA storage,<br>image ingestion from PACS or<br>from EA | Yes | No | Similar<br>No significant<br>difference |
| Workflow | Composite worklists deliver<br>efficient high volume reading | Yes | No | Not known |
| Category | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5, 2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3, 2016 |
| Storage | Instant access to archive with<br>iConnect Enterprise Archive | Yes | No | Not a clinically<br>significant feature<br>Similar, access to an<br>archive (VNA)<br>No significant<br>difference |
| Compliance | Maintain compliance and<br>manage control with physician<br>credentialing and licensure | Yes | No | Not known<br>Not a clinically<br>significant feature |
| Technology | 64-Bit Viewer: effectively<br>removes study size limits and<br>improves performance | Yes | No | Yes |
| Collaboration | Instant Messaging | Yes | No | Not known<br>Not a clinically<br>significant feature |
| Image Manipulation | Enhanced skin line detection for<br>mammography | Yes | No | Similar, breast<br>bounds image<br>alignment<br>No significant<br>difference |
| Annotation | Enhancements to DICOM<br>overlays | Yes | No | Similar, has<br>modality specific<br>overlays |
| Category | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5,<br>2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3,<br>2016 |
| | | | | No significant<br>difference |
| Image Manipulation | Enhanced Window/Level for NM<br>/ PT | Yes | No | Similar, supports NM<br>and PT<br>No significant<br>difference |
| Measurement | Orthopedic specific tools | Yes | Yes<br>(Cobb Angle) | Yes |
| Print | DICOM Printing, including True<br>Size | Yes | No | Yes |
| Export | Key image collage: save key<br>image layout as SC for export<br>to non KO aware PACS | Yes | No | Not known<br>Not a clinically<br>significant feature |
| Technology | Infrastructural enhancements:<br>worklist over port 80 -<br>simplifies deployments, full 64-<br>bit support on server, replace<br>IIS with Apache, Jboss7, Java 7<br>(functionality, security and<br>performance) | Yes | No | Similar, 64-bit<br>server support<br>No significant<br>difference |
| Image Display | Mammo tomosynthesis position<br>indicator | Yes | No | Similar, displays MG<br>Tomosynthesis<br>images |
| Category | Feature | Subject 510(k):<br>Merge PACS | Primary Predicate:<br>AMICAS PACS 6.0<br>510(k) K082144<br>Cleared Sept. 5,<br>2008 | Predicate:<br>Fuji Synapse PACS<br>K160108<br>Cleared Feb. 3,<br>2016 |
| | | | | No significant<br>difference |
| Image Display | Tomosynthesis improvements:<br>creation of nominal Axial,<br>Sagittal, and Coronal views<br>oriented to the slice stack | Yes | No | Similar, displays MG<br>Tomosynthesis<br>images |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MERGE, an IBM company. The word "MERGE" is written in a combination of blue and orange colors, with the letters "M", "R", and "G" in blue and the letters "E" in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size and blue color. The logo is simple and modern, with a clear emphasis on the company name and its affiliation with IBM.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Merge Healthcare, an IBM company. The word "MERGE" is written in large, bold letters, with the first half of the word in blue and the second half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size. The logo is simple and modern, with a focus on the company's name and its affiliation with IBM.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for MERGE, an IBM company. The word "MERGE" is written in a large, bold font, with the letters "M", "R", and "G" in blue and the letters "E" in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font. The logo is simple and modern, and the use of blue and orange gives it a professional look.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Merge Healthcare, an IBM company. The word "MERGE" is written in large, bold letters, with the first half of the word in blue and the second half in orange. Below the word "MERGE" is the text "An IBM Company" in a smaller font size. The logo is simple and modern, with a focus on the company's name and its affiliation with IBM.
{15}------------------------------------------------
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Merge, an IBM company. The word "MERGE" is written in a combination of blue and orange colors. The letters "M", "E", "R", and "G" are blue, while the letter "E" is orange. Below the word "MERGE", the text "An IBM Company" is written in a smaller font size.
## VIII. PERFORMANCE DATA
Mandatory performance specifications or special controls applicable to diagnostic workstation devices or to image fusion or SUV features specifically, have not been established by the FDA.
Merge PACS and the primary predicate device AMICAS PACS 6.0 comply with the following voluntary standards and quidelines:
- · DICOM Digital Imaging and Communications in Medicine, for the format of medical images and data
- · JPEG Joint Photographic Experts Group, for image compression standards
- FDA Guidance, "Guidance for the Submission of Premarket Notifications for Medical Image Management Devices", July 27, 2000
The determination of substantial equivalence is based on a comparison of features and software technology and on tests to confirm the correct behavior or result for each feature or measurement.
## Non-Clinical Tests
In-house non-clinical testing of the features of Merge PACS, a software device, was performed in compliance with design control processes.
The results of software testing confirmed that the Merge PACS device with image fusion and SUV measurement features performs as expected, and documented evidence was provided in this submission.
## Clinical Tests
Clinical studies were not required to demonstrate the safety and effectiveness of Merge PACS.
### Conclusion
The intended uses and indications for use of Merqe PACS are, in our opinion, substantially equivalent to the primary predicate device AMICAS PACS 6.0 as well as to the predicate device Synapse PACS.
{17}------------------------------------------------
The majority of features and functionalities, including image fusion and SUV measurements, were compared and found to be similar or the same as each other. We believe that any differences in features and their underlying technologies between the subject and predicate devices are minor and not siqnificant since all comply with the same fundamental standards such as DICOM, and the resulting displayed images and measurements would be essentially equivalent.
The results of non-clinical testing and compliance with design controls demonstrate that the subject device, Merge PACS, is safe and effective, and performs as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.