K160108 · Fujifilm Medical Systems U.S.A, Inc. · LLZ · Feb 3, 2016 · Radiology
Device Facts
Record ID
K160108
Device Name
Synapse PACS
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities. The FUJIFILM Synapse PACS Software enables the display, comparison of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIPMPR/Fusion) are not intended for Mammography use. The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.
Device Story
Synapse PACS is a web-based client/server software system for medical image management; integrates capabilities of previous Synapse Workstation and MPR/Fusion software. Inputs: DICOM-compliant medical images (X-ray, CT, MRI, ultrasound, nuclear medicine, PET, SPECT). Processing: Server-side image processing; web-based UI for display, manipulation, and 3D reconstruction (MIP/MPR/Fusion). Outputs: Visualized medical images on standard off-the-shelf PCs. Used in clinical environments by radiologists and clinicians for diagnostic interpretation. Enables comparison of current/prior studies, measurement, and annotation. Benefits: Centralized, platform-independent access to multi-modal imaging data; facilitates clinical decision-making through advanced visualization and efficient workflow.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level functionality, component, integration, usability, and installation testing. Benchmark performance testing was conducted using actual clinical images to demonstrate accuracy. All tests passed based on predefined requirements.
Technological Characteristics
Web-based client/server architecture; server-side image processing. Client platform: Windows 7/8.1, browser-based (IE 10/11, Chrome). Server platform: Windows Server 2008 R2/2012 R2. Database: Oracle. Connectivity: Networked DICOM-compliant. Programming: Javascript (TypeScript), HTML5, CSS. Security: Windows Authentication. Software class: Moderate concern per FDA guidance.
Indications for Use
Indicated for radiologists and clinicians to process, display, and compare medical images from DICOM-compliant modalities including X-ray, CT, MRI, ultrasound, nuclear medicine, PET, and SPECT. Includes 3D visualization (MIP/MPR) and fusion capabilities. Supports processing and interpretation of mammography images (For Processing/For Presentation). Not intended for mammography use with MIP/MPR/Fusion functions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. Above the profiles is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2016
FUJIFILM Medical Systems U.S.A., Inc. % Jyh-Shyan Lin Senior Manager, Regulatory, Ouality and Clinical Affairs 419 West Avenue STAMFORD CT 06902
Re: K160108 Trade/Device Name: Synapse PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 15, 2016 Received: January 19, 2016
Dear Jyh-Shyan Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160108
Device Name Synapse PACS
#### Indications for Use (Describe)
FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities.
The FUJIFILM Synapse PACS Software enables the display. comparison of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIPMPR/Fusion) are not intended for Mammography use.
The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.
| Type of Use (Select one or both, as applicable) | | | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|--|--|
| <div><span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | | | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## FUJ}FILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse PACS
#### 510(k) Summary
## 510(k) Number: K160108
## Date Prepared:
February 1, 2016
#### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin
## Device Trade Name:
Synapse PACS
## Device Common Names:
Picture Archiving and Communications System (PACS)
## Device Classification Name:
System, Image Processing, Radiological
#### Product Code:
LLZ
Regulation Number: 21 CFR 892.2050
## Device Class:
Class II
## Panel:
Radiology
## Predicate Devices:
- . FUJIFILM Synapse Workstation Software (K112439)
- Synapse MPR Fusion (K113244) .
{4}------------------------------------------------
Synapse PACS
## 1. Description of the Device
The proposed Synapse PACS Software is an implementation that combines capabilities of Synapse Workstation software (K112439) and Synapse MPR/Fusion software (K113244) in a single system, using the current Internet standards for Web clients and servers. The Synapse PACS Software and the predicate devices, i.e., Synapse Workstation Software (K112439) and Synapse MPR Fusion Software (K113244), are picture archiving and communication systems (as defined by 21 CFR 892.2050). Synapse PACS Software is the web based (client/server) application and implementation of the Synapse Workstation Software and the Synapse MPR Fusion Software. The Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with Synapse PACS Software (server). The Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities. Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies (K113244).
In summary, this 510(k) submission introduces the Synapse PACS with the web based (client/server) application and the ability to the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies.
## 2. Indication for Use
FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities.
The FUJIFILM Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIP/MPR/Fusion) are not intended for Mammography use.
The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.
{5}------------------------------------------------
# FUJIYFILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse PACS
#### 3. Substantial Equivalence Comparison
#### 3.1 Device features and technological characteristics comparison
## Table 1. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)
| Features | Synapse PACS 5.1.0<br>(This submission) | Synapse Workstation 3.3.0<br>(K112439) |
|----------------------------------------------------------------------------|---------------------------------------------------------|----------------------------------------|
| Product availability | Provided as Software only | Provided as Software only |
| Operating Systems | Windows 7/8.1 | Windows 2000/XP |
| Web Browser | Internet Explorer Version: 10 or 11<br>or Google Chrome | Internet Explorer |
| Database and file access method | Web based (http(s)) | Web based (http(s)) |
| Display of Digital Mammography Images | Yes | Yes |
| Process of Digital Mammography (Non-<br>Processed or For Process) Images | Yes | Yes |
| On-demand access to database and images | Yes | Yes |
| Viewing study lists | Yes | Yes |
| Decompression for compressed images before<br>display | Yes | Yes |
| Display images | Yes | Yes |
| Viewing Reports | Yes | Yes |
| Hanging Protocol | Yes | Yes |
| Spine Labeling | Yes | Yes |
| Reference line display | Yes | Yes |
| Key images identification | Yes | Yes |
| Link multiple series | Yes | Yes |
| Measurement Tools | Yes | Yes |
| Annotation Tools | Yes | Yes |
| Standard Image Manipulation Tools (Window<br>width/level, Zoom, Pan, etc.) | Yes | Yes |
| FCR IPSS for Fuji-CR images | Yes | Yes |
| CT IPSS for CT images | Yes | Yes |
{6}------------------------------------------------
Synapse PACS
| Features | Synapse PACS 5.1.0<br>(This submission) | Synapse MPR Fusion 2.5<br>(K113244) |
|--------------------------------------------------------------------------------|---------------------------------------------------------|-------------------------------------|
| Product availability | Provided as Software only | Provided as Software only |
| Operating Systems | Windows XP/Vista/Windows 7 | Windows 7, Windows 8.1 64bit |
| Web Browser | Internet Explorer Version: 10 or 11 or Google<br>Chrome | Internet Explorer |
| Database and file access method | Web based (http(s)) | Web based (http(s)) |
| Orthogonal and Oblique Multi Planar<br>Reconstruction (MPR) | Yes | Yes |
| Maximum, Average, and Minimum Intensity<br>Projection (MIP, MPVR, MinIP) | Yes | Yes |
| Curved Multi Planar Reconstruction (CPR) | No | Yes |
| Independent Window/Level of upper and lower<br>images in Fusion Study | Yes | Yes |
| Save Preferences by User Profile | No | Yes |
| Save Window/Level Preferences | Yes | Yes |
| Multi-Modal Fusion to include MR/PET/NM/and<br>CT | Yes | Yes |
| Compare Current Study to previous study in<br>Fusion View | Yes | Yes |
| Compare Current Study to Prior Study in<br>Compare Mode | Yes | Yes |
| Manual Fusion Correction with Rigid<br>Registration in Blend and overlay modes | No | Yes |
| Save series to Synapse PACS | Yes | Yes |
{7}------------------------------------------------
## FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse PACS
#### 3.2 Technology Similarities and Differences
| Technology | Synapse PACS 5.1.0<br>(This Submission) | Synapse Workstation 3.3.0<br>(K112439) |
|------------------------------------------|-----------------------------------------|----------------------------------------|
| Image and Data Processing | Server Side | Client Side |
| Technology Platform (Synapse Client) | Platform Independent (Browser based UI) | Windows Active-X |
| Technology Platform (Synapse Server) | Windows Server | Windows Server |
| System Security | Windows Authentication | Windows Authentication |
| Programming Language(s) (Synapse Client) | Javascript (TypeScript), HTML5, CSS | C++, Active-X |
## Table 3. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)
## Table 4. Synapse PACS 5.1.0 vs. Synapse MPR Fusion 2.5 (K113244)
| Technology | Synapse PACS 5.1.0<br>(This Submission) | Synapse MPR Fusion 2.5<br>(K113244) |
|------------------------------------------|-----------------------------------------|-------------------------------------|
| Image and Data Processing | Server Side | Client Side |
| Technology Platform (Synapse Client) | Platform Independent (Browser based UI) | Windows Native Application |
| Technology Platform (Synapse Server) | Windows Server | Windows Server |
| System Security | Windows Authentication | Windows Authentication |
| Programming Language(s) (Synapse Client) | Javascript (TypeScript), HTML5, CSS | C++ |
{8}------------------------------------------------
## FUJ}{F||LM|MEDICAL SYSTEMS U.S.A., Inc.
Synapse PACS
#### 3.3 Device Specifications Comparison
| Specifications | Synapse PACS 5.1.0<br>(This Submission) | Synapse Workstation 3.3.0<br>(K112439) |
|----------------|---------------------------------------------------|----------------------------------------|
| Client O/S | Windows 7/8.1 | Windows 2000/XP |
| Browser | Internet Explorer Version: 10/11 or Google Chrome | Internet Explorer |
| Server O/S | Windows 2008 R2, or 2012 R2 Server OS | Windows 2003, or 2008 Server OS |
| Database | Oracle | Oracle |
## Table 5. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)
## Table 6. Synapse PACS 5.1.0 vs. Synapse MPR Fusion 2.5 (K113244)
| Specifications | Synapse PACS 5.1.0<br>(This Submission) | Synapse MPR Fusion 2.5<br>(K113244) |
|----------------|------------------------------------------------------|-------------------------------------|
| Client O/S | Windows 7/8.1 | Windows 2000/XP |
| Browser | Internet Explorer Version: 10 or 11 or Google Chrome | Internet Explorer |
| Server O/S | Windows 2008 R2, or 2012 Server R2 OS | Windows 2003, or 2008 Server OS |
| Database | Oracle | N/A |
{9}------------------------------------------------
## FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.
Synapse PACS
## 4. Safety Information
Synapse PACS introduces no new safety or efficacy issues other than those already indentified with the predicate device. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse PACS labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.
## 5. Testing and Performance Information
Synapse PACS is tested successfully with reference to its product requirements, as well as design verification and validation documents and traceability matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse PACS software, which is found to be safe and effective and substantially equivalent to the predicate devices.
Testing involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, we conducted benchmark performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.
Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.
## 6. Conclusion
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.