PRESS DUO elite, PRESS DUO elite AG

K173450 · Nemoto Kyorindo Co., Ltd. · DXT · Feb 7, 2018 · Cardiovascular

Device Facts

Record IDK173450
Device NamePRESS DUO elite, PRESS DUO elite AG
ApplicantNemoto Kyorindo Co., Ltd.
Product CodeDXT · Cardiovascular
Decision DateFeb 7, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K173450 · Feb 7, 2018PRESS DUO elite, PRESS DUO elite AGNemoto Kyorindo Co., Ltd.Japanese market postmarket performance dataThe sponsor utilized approximately 5 years of real-world postmarket performance data from the Japanese market to justify the sample size used in bench testing and to substantiate the device's performance and reliability.postmarket_experience; market_history

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Japanese Market Performance History; Retrospective market performance review; Follow-up/Duration: Approximately 5 years; Study Period: Prior to 2017Patients in the Japanese marketNot applicable for this studyDevice performance and reliability

Indications for Use

The PRESS DUO elite is intended to be used for injecting contrast media and common flushing solution into the vascular system in Angiographic and Computed Tomography (CT) procedures whereas the PRESS DUO elite AG is intended to be used for injecting contrast media and common flushing solution into the vascular system for Angiographic procedures only.

Device Story

Contrast delivery system for angiographic and CT suites; comprises console, powerhead, and main control unit. Operator programs injection parameters (volume, flow rate, pressure) via touchscreen console. Powerhead drives lead screw ram to push syringe plunger, expelling contrast/saline. System supports single/multiple phase injections; provides real-time injection feedback and pressure monitoring. Used by clinicians in angiographic suites. Output informs diagnostic imaging quality. Benefits include precise, controlled delivery of contrast media to enhance vascular visualization. PRESS DUO elite AG is functionally identical but restricted to angiographic mode via software.

Clinical Evidence

Bench testing only. Performance verification included accuracy of flow rate, volume, pressure, and manual control speeds. EMC, electrical, mechanical, and thermal safety testing performed on prototypes. Usability and human factors analysis conducted; no critical tasks identified, and risks mitigated to ALARP.

Technological Characteristics

Pedestal, ceiling, or table-mount configuration. 24V DC powerhead with screw-driven ram. 150ml single-use disposable syringe (ISO594-2 compliant). Touchscreen console interface. Pressure limit 50-1200 PSI (Angio) or 50-300 PSI (CT). Connectivity via communications interface cable between console and main control unit. Software-controlled injection protocols.

Indications for Use

Indicated for humans undergoing angiographic and/or CT procedures requiring intravascular injection of contrast media and flushing solutions.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below. February 7, 2018 Nemoto Kyorindo Co., Ltd. Jim Knipfer General Manager, Overseas Department Profit Building 5F 2-27-3 Hongo, Bunkyo-ku Tokyo, Japan 113-0033 Re: K173450 Trade/Device Name: PRESS DUO elite, PRESS DUO elite AG Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: October 31, 2017 Received: November 6, 2017 Dear Jim Knipfer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K173450 Device Name PRESS DUO elite, PRESS DUO elite AG #### Indications for Use (Describe) The PRESS DUO elite is intended to be used for injecting contrast media and common flushing solution into the vascular system in Angiographic and Computed Tomography (CT) procedures whereas the PRESS DUO elite AG is intended to be used for injecting contrast media and common flushing solution into the vascular system for Angiographic procedures only. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## NEMOTO KYORINDO CO., LTD 2-27-20 Hongo Bunkvo-ku. Tokyo 113-0033 Japan TELEPHONE +81- 3-5842-8571 FACSIMILE +81- 3-5805-0551 #### K173450 ## PRESS DUO elite and PRESS DUO elite AG Contrast Delivery System 510(k) Summary | Contact Name: | Jim Knipfer | |----------------------|------------------------------------------------------------------------| | | General Manager, Overseas Department | | | Nemoto Kyorindo Co., Ltd. | | | Profit Building 5F | | | 2-27-3 Hongo, Bunkyo-ku | | | Tokyo, Japan 113-0033 | | Phone: | +81-3-5842-8571 | | FAX: | +81-3-5805-0551 | | Date Prepared: | November 2, 2017 | | Proprietary Name(s): | PRESS DUO elite, PRESS DUO elite AG | | Common Name: | CT/Angiography Contrast Media Delivery System | | Classification Name: | Injector and Syringe, Angiographic | | Predicate Device(s): | Angiomat Illumena Injector, K963071 (Primary predicate) | | | Dual Shot alpha 7, K133189 (Secondary predicate) | | | Rempress, K092896 (Secondary predicate) | | | Medrad Mark 7 Arterion Injection System, K132928 (Secondary predicate) | ## Indications for Use / Intended Use: The PRESS DUO elite is intended to be used for injecting contrast media and common flushing solution into the vascular system in Anqiographic and Computed Tomography (CT) procedures whereas the PRESS DUO elite AG is intended to be used for injecting contrast media and common flushing solution into the vascular system for Angiographic procedures only. ## Description: The main components of the PRESS DUO elite and PRESS DUO elite AG are the Console, Powerhead and Main Control Unit. The basic confiqurations of the PRESS DUO elite and PRESS DUO elite AG are a pedestal, ceiling mount or table mount configuration. With any configuration, the three main components are most often contained in the angiographic suite and normally near the patient. The parameters of the injection, such as volume, flow rate and pressure are programmed by the graphical user interface {4}------------------------------------------------ ## NEMOTO KYORINDO CO., LT 2-27-20 Hongo Bunkvo-ku. Tokyo 113-0033 Japan TELEPHONE +81- 3-5842-8571 +81- 3-5805-0551 FACSIMILE with touchscreen input. The system is designed to deliver a variety of injection protocols. It can deliver single or multiple phase injections. The Console is powered via 24 volts DC which is derived from a remote AC to DC converter (like that used with laptop computers) and communicates with Main Control Unit communications interface cable. After the injection protocol has been set, the Powerhead performs the injection by driving the lead screw ram. The ram pushes the push-rod of the syringe which expels fluid from the barrel of the syringe. The only difference between the PRESS DUO elite and PRESS DUO elite AG is the availability of the CT Mode. The PRESS DUO elite provides both an Angiographic Mode and CT Mode of operation whereas the PRESS DUO elite AG has the CT Mode of operation turned off in the software and cannot be turned on by the user. All other software, hardware and mechanical aspects of the PRESS DUO elite and PRESS DUO elite AG are identical. The following paragraphs provide more details for each of these main components. POWERHEAD: The Powerhead provides a means for accepting and identifying a given syringe and then applying a force to the plunger of the syringe via its screw driven ram which will eject the contrast or common flushing solution from the barrel of the syringe. The Powerhead is in constant communications with the Main Control Unit for determining injection protocol and to monitor all Powerhead functions and injection sequences. The injection is communicated to the user Console to provide immediate feedback of injection operation. As highlighted in the Description section of this document there are 3 configurations for the PRESS DUO elite and PRESS DUO elite AG. These configurations refer to where the Powerhead component would be located 1) pedestal stand, 2) ceiling suspension system or 3) table rail mounted. CONSOLE: The Console is the main user interface for the PRESS DUO elite and PRESS DUO elite AG. It provides a color graphical user interface with an overlying touchscreen for easy injection monitoring and injection setup. At the completion the injection results are clearly displayed to the user. The Console provides a connection for a remote Handswitch that can be used to start or stop an injection. The Console resides normally in the angiographic suite and is typically located near the patient. MAIN CONTROL UNIT: The Main Control Unit is the interface between the Console and the Powerhead. The Main Control Unit receives the user input data from the Console and converts to the necessary control information then relays to the Powerhead. The Main Control unit also provides the power necessary to operate the Powerhead. The Main Control Unit is located inside the angiographic suite and communicates with the Console via a communications link. SYRINGES AND ACCESSORIES: The PRESS DUO elite and PRESS DUO elite AG injection system have only {5}------------------------------------------------ # NEMOTO KYORINDO CO., LTD 2-27-20 Hongo Bunkyo-ku, Tokyo 113-0033 Japar TELEPHONE +81- 3-5842-8571 FACSIMILE +81- 3-5805-0551 one compatible syringe which is a 150ml single use disposable syringe as defined by 510(k) K090487, manufactured by Coeur, Inc. Other consumable accessories such as tubing, valves, stop cocks, are required to have luer connections compliant to ISO594-2 and shall have a pressure rating consistent with the pressure limit set-point on the PRESS DUO elite or PRESS DUO elite AG or a maximum of 1200PSI. ## SUBSTANTIAL EQUIVALENCE A comparison chart shown in Table 1 compares the technological characteristics of the PRESS DUO elite and PRES DUO elite AG contrast delivery systems to the predicate devices the Liebel-Flarseim's Angiomat Illumena, Nemoto's Dual Shot alpha 7, Nemoto's Rempress and Medrad Mark 7 Arterion. The PRESS DUO elite and PRESS DUO elite AG is substantially equivalent to the predicate devices. {6}------------------------------------------------ # NEMOTO KYORINDO CO., LTD. 2-27-20 Hongo Bunkyo-ku, Tokyo 113-0033 Japan | Characteristic / Feature | Angiomat Illumena<br>(Primary Predicate)<br>K963071 | Dual Shot alpha 7<br>(Secondary Predicate)<br>K133189 | REMPRESS<br>(Secondary Predicate)<br>K092896 | Medrad Mark 7 Arterion<br>(Secondary Predicate)<br>K132928 | PRESS DUO elite<br>PRESS DUO elite AG<br>(K173450) | Device Comparison<br>Result | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Multi-phasic Injections | 4 phases per protocol | 5 phases per protocol | 1 phase per protocol | 4 phases per protocol | 2 phase per protocol | Same | | Single or Dual Syringe | Single | Dual | Single | Single | Dual | Same | | Angio and CT Modes | Angio, CT | CT | Angio | Angio | Angio, CT | Same | | Infusion Mode | Yes | No | Yes | No | Yes | Same | | Protocol Storage | 45 protocols (Angio Mode)<br>45 protocol (CT Mode) | 400 protocols | 20 protocols | 40 protocols | 240 protocols | Inside Specification | | X-ray/Scan Delay | 0 - 300 secs. | 0 – 300 secs | 0 - 99 secs | 0 - 99.9 secs | 0 - 99.9 secs | Inside Specification | | Inject Delay | 0 - 300 secs. | No | 0 - 99 secs | 0 - 99.9 secs | 0 - 99.9 secs | Inside Specification | | Inter-phase Delay | 0 - 300 secs. | 0 – 300 secs | None | Not indicated | None | Same | | Syringe System | 150ml or 125ml | 100ml or 200ml | 150ml | 150ml | 150ml | Same | | Volume Remaining Display | LED display on Powerhead | Optional Powerhead display | LED display on Powerhead | Color LCD | LED display on Powerhead | Same | | Filling Rate | 3 to 25ml/sec (manual) | 0.5ml/sec, 1.5ml/sec and<br>8.0ml/sec | 0.5 – 2.5ml/sec | 1 - 10ml/sec | 0.5 – 2.5ml/sec | Same | | Rise Time | 0 to 10 seconds | Off = 0 secs, On = 2 secs | 0 - 9.9 seconds | 0 - 9.9 secs | 0 - 9.9 seconds | Same | | Arming Modes | Single or Multi | Single | Single or Multi | Single or Multi | Single or Multi | Same | | Variable Jog Speeds | Yes | Yes | Yes | Yes | Yes | Same | | Body Weight Protocol | No | Yes | No | No | No | Same | | Quick Return | No | Yes (see Filling Rate) | No | Not Indicated | Yes (see Filling Rate) | Same | | Quick Purge | No | Yes | No | Not Indicated | Yes | Same | | Remote Start | Yes | Yes | Yes | Yes | Yes | Same | | Pressure Graph | Not indicated | Yes | Yes | No | Yes | Same | | Syringe Heater | Yes | Yes | Yes | Yes | Yes | Same | | Interlocking with Angio | Yes | No | Yes | Yes | Yes | Same | | Interlocking with CT | Not indicated | Yes | No | No | No | Same | | Injection Abort | Yes | Yes | Yes | Yes | Yes | Same | | Powerhead Indicator Lights | Yes | Yes | Yes | Yes | Yes | Same | | Power-On Self-Test | Yes | Yes | Yes | Yes | Yes | Same | | Selectable Inject Patterns | Yes | Yes | Yes | Yes | Yes | Same | | Variable Flow Rate | No | No | Yes - optional | Yes - optional | No | Same | | Flow Rate | 0.1 to 40ml/sec for<br>125/150ml syringe (Angio<br>Mode)<br>0.1 to 10.0ml/sec in (CT<br>Mode)<br>0.1 to 99ml/min for infusion<br>purposes | 0.1 to 10ml/sec in 0.1ml/sec<br>increments | 0.1 to 25ml/sec with 150ml<br>syringe<br>0.1 to 99ml/min for infusion<br>purposes | 0.1 to 45ml/sec in<br>0.1ml/sec increments<br>0.1 to 59.9ml/min in<br>0.1ml/min (not specified as<br>infusion mode but operates<br>at infusion rates) | 0.1 to 30ml/sec in<br>0.1ml/sec increments<br>(Angio Mode)<br>0.2 to 10.0ml/sec in<br>0.1ml/sec increments (CT<br>mode) | Inside Specification | | Pressure Limit | 75 - 1200 PSI (Angio Mode)<br>75 - 300 (CT Mode) | 25 to 300 PSI | 50 - 1200 PSI | 100 – 1200 PSI | 50-1200 PSI (Angio Mode)<br>50-300 PSI (CT Mode) | Same | | Remote Start Switch | Yes | Yes | Yes | Yes | Yes | Same | | Test Shot | Yes (Re-Arming) | No | Yes (Trace Shot) | Yes (Re-arming) | Yes (Re-Arming) | Same | | Safety Stop Mechanism | Electrical stop when<br>injection parameters are<br>out of specification | Electrical stop when<br>injection parameters are out<br>of specification | Electrical stop and<br>mechanical backup stopper<br>when injection parameters<br>are out of specification | Electrical stop when<br>injection parameters are<br>out of specification | Electrical stop when<br>injection parameters are<br>out of specification | Same | | User Interface Features | | | | | | | | - Fill / Expel Control | Yes | Yes | Yes | Yes | Yes | Same | | - Air Detection | Operator visual inspection | Operator visual inspection | Operator visual inspection | Operator visual inspection | Operator visual inspection | Same | | FACSIMILE<br>+81- 3-5805-0551 | | | | | | | | Status Display | Powerhead lights | Powerhead lights | Powerhead lights | Powerhead lights | Powerhead lights | Same | | Console Display | Color LCD with touchscreen | Color LCD with touchscreen | Color LCD with touchscreen | Color LCD with touchscreen | Color LCD with touchscreen | Same | | Materials | Plastics and metals | Plastic and metals | Plastics and metals | Plastic and metals | Plastic and metals | Same | | Anatomical Sites | Arterial and Venous injections | Arterial and venous injections | Arterial and Venous injections | Arterial and Venous injections | Arterial and Venous injections | Same | | Intended Use Statement | The Angiomat Illumena is<br>designed to inject<br>radiopaque contrast<br>medium into the vascular<br>system for Angiography<br>and CT procedures as<br>prescribed by qualified<br>health care professionals. | The contrast delivery system<br>Dual Shot alpha 7 is an<br>intravascular injection<br>system intended for the<br>administration of ionic and<br>non-ionic contrast media<br>and saline used in<br>conjunction with computed<br>X-ray tomography (CT). | The contrast delivery<br>system REMPRESS is an<br>intravascular injection<br>system intended for the<br>administration of contrast<br>media or flushing solutions<br>used in conjunction with<br>angiographic imaging<br>procedures. | The MEDRAD Mark 7<br>Arterion Injection system is<br>intended to be used<br>specifically for the purposes<br>of injecting contrast<br>medium and common<br>flushing solutions into<br>humans for angiographic<br>studies. | The PRESS DUO elite is<br>intended to be used for<br>injecting contrast media<br>and common flushing<br>solution into the vascular<br>system in Angiographic and<br>Computed Tomography<br>(CT) procedures whereas<br>the PRESS DUO elite AG is<br>intended to be used for<br>injecting contrast media<br>and common flushing<br>solution into the vascular<br>system for Angiographic<br>procedures only. | Same | | Target Population | Humans | Humans | Humans | Humans | Humans | Same | {7}------------------------------------------------ # NEMOTO KYORINDO CO., LTD. 2-27-20 Hongo Bunkyo-ku, Tokyo 113-0033 Japan TELEPHONE +81- 3-5842-8571 {8}------------------------------------------------ ## EMOTO KYORINDO CO., LT 2-27-20 Hongo Bunkvo-ku. Tokvo 113-0033 Japan TELEPHONE +81- 3-5842-8571 +81- 3-5805-0551 FACSIMILE ### SUMMARY OF PERFORMANCE TESTING Performance testing was completed to verify the PRESS DUO elite and PRESS DUO elite AG contrast delivery systems are capable of achieving the specification parameters for the system as outlined in the substantial equivalence chart. Verification of the system's ability to accurately achieve these the substantial equivalence claims. Injection performance was tested to verify accuracy of delivered flow rate, volume, pressure and manual control speeds. The EMC, Electrical, Mechanical and Thermal Safety testing was performed using a prototype device however there were no electrical, safety or structural mechanical differences between the tested device and final device. A ferrite bead shown in the test reports for the prototype device is also included in the final device. One variation between the tested device and the final device is a Footswitch. The Footswitch was included in the testing as this optional component is provided to our customers outside of the USA however it would not be provided to USA customers since the option is not required. Testing was not specifically performed with the PRESS DUO elite AG model however the functionality of the PRESS DUO elite AG can be tested using the PRESS DUO elite model since it can perform both angiographic functions and the CT functions which are not included in the PRESS DUO elite AG. The utilization of the PRESS DUO elite further permits this disabling of the CT Mode thereby making it a PRESS DUO elite AG model. All performance specifications when operating in the Angio Mode are identical between the PRESS DUO elite and PRESS DUO elite AG, therefore separate testing was not necessary to confirm the PRESS DUO elite AG operation. The sample size used for the testing was a single unit of the justification for testing only a single unit is based upon our long history of the PRESS DUO in the Japanese market. The PRESS DUO has been on the Japanese market for approximately 5 years therefore the performance of the device is well understood and substantiated by the existing market. With respect to consumables used during the testing, the only compatible syringe is the 150ml disposable syringe identified in K090487, which is reflected in the User Manual. Therefore, the 150ml syringe was used for all testing. The 150ml connection point is compliant to ISO594-2 therefore compatible with any tubing compliant to ISO594-2 and 1200PSI requirement. The ISO594-2 and 1200PSI requirements are also reflected in the updated User Manual. The extension tubing H-70 used in this testing is compatible with the required specifications. All needles and catheters on the market today that comply with the ISO594-2 specification can be used with the 150ml syringe in K090487 which is compatible with the Protection Cover {9}------------------------------------------------ ## EMOTO KYORINDO CO., LT 2-27-20 Hongo Bunkvo-ku. Tokvo 113-0033 Japa TELEPHONE +81- 3-5842-8571 +81- 3-5805-0551 FACSIMILE that is used with the PRESS DUO elite and PRESS DUO elite AG. To develop injection pressure various needles are used during the performance testing to ensure the injector can be forced to perform a pressure limited injection. A pressure limited injection is required to verify the necessary hardware and software mechanisms are in place and functioning correctly to prevent an over pressure limit situation from presenting itself. Various need to enable the pressure limit condition to occur at different Flow Rates. The 21G and 23G needle sizes are sufficient to force the injector device into pressure limit conditions at various flow rate values. Since the pressure is limited by the injector a smaller orifice size would not cause the pressure to limit to change. For example, the execution of an injection using a 23G needle which is pressure limiting at 500PSI would still pressure limit at 500PSI using a 24G (smaller) needle. The pressure limit is set by the injector device and not by the needle, tubing etc. #### VOLUME Fluid delivery of an injector is characterized by three primary parameters, flow rate, volume and pressure limit. The volume delivered is the most critical parameter in that is has the greatest effect on image quality. During testing, the volume delivered was accurately measured and compared to the volume programmed. Over the range of volumes programmed, the PRESS DUO elite and PRESS DUO elite AG contrast delivery system successfully delivered the volumes within the allowable specifications. #### FLOW RATE The second most critical variable parameter in delivering fluid to a patient is the flow rate. During the test injections, the injection time and volume were recorded then the flow rate calculated. In the event of a pressure limit injection, the volume and time are known not to accurately represent the flow rate therefore the data was omitted from this specific calculation. In all non-pressure limited injections the flow rate measured within the allowable ranges of the specification. NOTE: Pressure limit injections automatically reduce the flow rate from the user set flow rate to maintain a constant pressure, as desired by the operator. #### PRESSURE During all injections, the PRESS DUO elite and PRESS DUO elite AG monitors the pressure to determine if the injection should be pressure limit is set for the syringe used in the injector in order to avoid failure of the syringe. During testing, the actual pressure was measured and compared to the displayed values and product specifications. All pressure readings were within the allowable specifications. #### MANUAL CONTROL SPEEDS The PRESS DUO elite and PRESS DUO elite AG have two speeds for manually moving the rams. By {10}------------------------------------------------ ## NEMOTO KYORINDO CO., LT 1-27-20 Hongo Bunkvo-ku. Tokyo 113-0033 Japa TELEPHONE +81- 3-5842-8571 FACSIMILE +81- 3-5805-0551 depressing the forward or reverse keys on the Powerhead, the rams can be moved at varying speeds. By pressing the forward or reverse key in combination with the accelerator key, the rams can be moved at maximum speed. During testing, speeds were confirmed to be within the allowable specificati…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...