K173435 · 3M Company · FRC · Feb 9, 2018 · General Hospital
Device Facts
Record ID
K173435
Device Name
3M Attest Rapid Readout Biological Indicator
Applicant
3M Company
Product Code
FRC · General Hospital
Decision Date
Feb 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the following systems: AMSCO® V-PRO™ 1 Low Temperature Sterilization System (Lumen cycle), AMSCO® V-PRO™ 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles). AMSCO® V-PRO™ max Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), AMSCO® V-PRO™ 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) and in STERRAD® NX (Standard and Advanced cycles), STERRAD® 100NX (Standard, Flex, Express and Duo cycles) systems, STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles) and STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express and Duo cycles).
Device Story
Self-contained biological indicator (BI) for monitoring vaporized hydrogen peroxide sterilization; polycarbonate sleeve contains Geobacillus stearothermophilus spores on polyethylene terephthalate carrier and growth media ampoule. Post-sterilization, user activates BI by depressing cap to crush ampoule, mixing media with spores. BI incubated in 3M Attest Auto-reader 490H; reader detects fluorescence from enzyme activity indicating sterilization failure. Chemical indicator on cap provides visual confirmation of exposure to sterilant (blue to pink). Used in clinical settings to verify sterilization efficacy; provides rapid 24-minute readout. Benefits include reliable, timely confirmation of sterilization process success, ensuring patient safety by preventing use of improperly sterilized instruments.
Clinical Evidence
No clinical data included. Effectiveness demonstrated via bench testing, including full and half-cycle performance verification in specified sterilization systems, chemical indicator color change, and verification of spore population characteristics, per ANSI/AAMI/ISO 11138-1:2006/(R) 2010.
Technological Characteristics
Self-contained BI; polycarbonate sleeve; polyethylene terephthalate spore carrier; Geobacillus stearothermophilus spores (≥1x10^6); H2O2-sensitive chemical indicator ink; 60±2°C incubation; 24-minute fluorescence readout; manual activation; no electronic components in BI; compatible with 3M Attest Auto-reader 490H.
Indications for Use
Indicated for routine monitoring of vaporized hydrogen peroxide sterilization processes in specified AMSCO V-PRO and STERRAD sterilization systems using the 3M Attest Auto-reader 490H.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles a caduceus, with three human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2018
3M Health Care Ms. Nadia Battah Regulatory Affairs Associate 3M Center, Bldg. 275-5W-06 St. Paul. MN 55144-1000
Re: K173435
Trade/Device Name: 3M™ Attest™ Rapid Readout Biological Indicator 1295 Regulation Number: 21 CFR 880.2800 Regulation Name: Biological Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: January 05, 2018 Received: January 08, 2018
Dear Ms. Nadia Battah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K173435
Device Name
3MTM Attest™ Rapid Readout Biological Indicator 1295
#### Indications for Use (Describe)
Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3MTM Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the following systems: AMSCO® V-PRO™ 1 Low Temperature Sterilization System (Lumen cycle), AMSCO® V-PRO™ 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles). AMSCO® V-PRO™ max Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), AMSCO® V-PRO™ 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) and in STERRAD® NX (Standard and Advanced cycles), STERRAD® 100NX (Standard, Flex, Express and Duo cycles) systems, STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles) and STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express and Duo cycles).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and consists of the number "3" and the letter "M". The number and letter are in a bold, sans-serif font. The logo is simple and recognizable.
# 510(k) Summary for 3MTM Attest™ Rapid Readout Biological Indicator 1295
#### Sponsor Information:
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact: Nadia Battah Regulatory Affairs Associate Phone Number: (651) 733-0929 Fax Number: (651) 737-5320
Date of Summary: February 6, 2018
Submission Number: K173435
{4}------------------------------------------------
#### Device Name and Classification:
| Common or Usual Name: | Biological Indicator |
|------------------------|---------------------------------------------------------|
| Proprietary Name: | 3M TM Attest TM Rapid Readout Biological Indicator 1295 |
| Classification Name: | Indicator, Biological Sterilization Process |
| Device Classification: | Class II, 21 CFR $ \S $ 880.2800(a) |
| Product Code: | FRC |
#### Predicate Device:
- . K171003 - 3MTM Attest™ Rapid Readout Biological Indicator 1295
#### Description of Device:
The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is a self-contained biological indicator specifically designed for rapid and reliable routine monitoring of vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3M™ Attest™ Auto-reader 490H. The 1295 BI is a single-use device composed of a polycarbonate sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. A chemical process indicator printed with stripes which change from blue toward pink upon exposure to vaporized hydrogen peroxide is located on the top of the cap. The detection of fluorescence upon incubation of the 1295 BI in the 490H Auto-reader indicates a sterilization failure.
#### Indications for Use
Use the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M™ Attest™ Auto reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the following systems: AMSCO® V-PROIM 1 Low Temperature Sterilization System (Lumen cycle), AMSCO® V-PRO™ 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles), AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), AMSCO® V-PRO™ 60 Low Temperature Sterilization System (Lumen and Flexible cycles) and in STERRAD® 100S, STERRAD® NX (Standard and Advanced cycles), STERRAD® 100NX (Standard, Flex, Express and Duo cycles) systems, STERRAD® NX with ALLClear™ Technology (Standard and Advanced cycles) and STERRAD® 100NX with ALLClear™ Technology (Standard, Flex, Express and Duo cycles).
#### Nonclinical Comparison to the Predicate Device
The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is the same design as the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Attest™ Rapid Readout Biological Indicator 1295 (K171003).
{5}------------------------------------------------
There has been no change to the device's design, materials, physical/chemical characteristics, performance specifications or fundamental scientific technology or to the device's accessory, the 3M Attest™ Auto-reader 490H. The intent of this submission is to expand the indications for use to include use in additional sterilizers.
#### Summary of Nonclinical Testing
Nonclinical testing of the 3M™ Attest™ Rapid Readout Biological Indicator 1295 was conducted in accordance with the FDA Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions, and ANSV/AAMI/ISO 11138-1:2006/(R) 2010 Sterilization of health care products- Biological indicators- Part 1: General requirements (FDA Recognition Number 14-296).
The effectiveness of the 3M™ Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M™ Attest™ Auto-reader 490H with a final fluorescent readout of 24 minutes is demonstrated in the following tests:
| Performance Test | Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Full Cycle Performance Verification – STERRAD® NX with ALLClearTM<br>Technology, STERRAD® 100NX with ALLClearTM Technology, AMSCO®<br>V-PROTM 1, AMSCO® V-PROTM 1 Plus and AMSCO® V-PRO 60 | Pass |
| Half Cycle Performance Verification – STERRAD® NX with ALLClearTM<br>Technology, STERRAD® 100NX with ALLClearTM Technology, AMSCO®<br>V-PROTM 1, AMSCO® V-PROTM 1 Plus and AMSCO® V-PRO 60 | Pass |
| Chemical Indicator (CI) Color Change | Pass |
| Verification of population characteristics | Pass |
#### Summary of Clinical Testing
No clinical data was included in this premarket application submission.
{6}------------------------------------------------
### Comparison to Predicate Device
| | Submission Device: | Predicate Device (K171003): |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | 3MTM AttestTM Rapid Readout | 3MTM AttestTM Rapid Readout |
| | Biological Indicator 1295 | Biological Indicator 1295 |
| Device Models | 1295 | Identical. |
| Intended Use | Monitoring of vaporized hydrogen<br>peroxide sterilization processes. | Identical. |
| Indications for use | Use the 3MTM AttestTM Rapid Readout<br>Biological Indicator 1295 in conjunction<br>with the 3MTM AttestTM Auto reader 490H<br>as a standard method of routine<br>monitoring of vaporized hydrogen<br>peroxide sterilization processes in the<br>following systems: AMSCO® V-PROTM<br>1 Low Temperature Sterilization System<br>(Lumen cycle), AMSCO® V-PROTM 1<br>Plus Low Temperature Sterilization<br>System (Lumen and Non Lumen cycles),<br>AMSCO® V-PROTM maX Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen, and Flexible<br>cycles), AMSCO® V-PROTM 60 Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen and Flexible cycles)<br>and in STERRAD® 100S, STERRAD®<br>NX (Standard and Advanced cycles),<br>STERRAD® 100NX (Standard, Flex,<br>Express and Duo cycles) systems,<br>STERRAD® NX with ALLClearTM<br>Technology (Standard and Advanced<br>cycles) and STERRAD® 100NX with<br>ALLClearTM Technology (Standard, Flex,<br>Express and Duo cycles). | Use the 3MTM AttestTM Rapid<br>Readout Biological Indicator 1295 in<br>conjunction with the 3M AttestTM<br>Auto reader 490H as a standard<br>method of routine monitoring of<br>vaporized hydrogen peroxide<br>sterilization processes in the<br>AMSCO® V-PRO® maX Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen, and Flexible<br>cycles), and in STERRAD® 100S,<br>STERRAD® NX (Standard and<br>Advanced cycles) and STERRAD®<br>100NX (Standard, Flex, Express and<br>Duo cycles) systems. |
| Device Design | The 3M AttestTM 1295 Rapid Readout<br>Biological Indicator (1295 BI) is a self-<br>contained biological indicator that<br>contains greater than 1x106 Geobacillus<br>stearothermophilus spores on a carrier<br>within the device. During activation, the<br>cap is depressed fully onto the sleeve<br>which pushes the growth media ampoule<br>into the ampoule crusher, causing the<br>glass ampoule to break and allowing the<br>liquid growth media to flow down to the<br>spores. After activation, the cap has a<br>surface that seals against the sleeve which<br>closes the sterilant entry ports and<br>contains the liquid growth media within<br>the BI. | Identical. |
| Organism | Geobacillus stearothermophilus<br>traceable to ATCCTM 7953 | Identical. |
| Viable spore<br>population | $≥1x106$ | Identical. |
| Resistance<br>characteristics<br>• D-value<br>• Survival/<br>Kill<br>Window | (Tested at 10 mg/L vaporized hydrogen<br>peroxide)<br>$D10 mg/L≥ 1 second$<br>Survival Time $≥$ 5 seconds<br>Kill Time = 7 minutes | Identical. |
| Carrier material | Polyethylene terephthalate | Identical. |
| Incubation<br>temperature | 60 ± 2°C | Identical. |
| Readout time | 24 minute fluorescence result read | Identical. |
| Chemical<br>indicator | H2O2 sensitive ink; changes from blue<br>towards pink | Identical. |
| Shelf-life | 24 months. | Identical. |
{7}------------------------------------------------
#### Conclusion
The results of the studies demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is substantially equivalent to, and is as safe and as effective as the legally marketed predicate device, K171003, Class II (21 CFR 880.2800, Product code FRC).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.