K173293 · Smith & Nephew, Inc. · HRS · Jan 8, 2018 · Orthopedic
Device Facts
Record ID
K173293
Device Name
EVOS Small Fragment Upper Extremity Plates
Applicant
Smith & Nephew, Inc.
Product Code
HRS · Orthopedic
Decision Date
Jan 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
Device Story
EVOS Small Fragment Upper Extremity Plates are metallic bone fixation implants designed for surgical stabilization of fractures. Plates feature 2.7mm and 3.5mm holes, including threaded, non-threaded, and variable-angle locking configurations. Used by orthopedic surgeons in clinical settings, these plates are implanted to provide structural support during bone healing. The system is an extension of existing Smith & Nephew plating lines, utilizing compatible bone screws and instruments. Clinical benefit includes anatomical fracture fixation, particularly for upper extremity applications. Device performance is validated through finite element analysis (FEA) and mechanical bending/construct fatigue testing to ensure structural integrity comparable to predicate systems.
Clinical Evidence
Bench testing only. Evidence includes finite element analysis (FEA) to identify worst-case designs, followed by mechanical bending and construct fatigue testing. Results demonstrated performance similar or superior to predicate plates. Packaging verification and bacterial endotoxin testing (per ANSI/AAMI ST72) were performed to confirm sterility maintenance and safety.
Technological Characteristics
Metallic bone fixation plates with 2.7mm/3.5mm hole diameters. Features include threaded, non-threaded, and variable-angle locking holes. Designed for compatibility with existing EVOS system screws and instruments. Sterilization validated per FDA guidance and ANSI/AAMI ST72 standards.
Indications for Use
Indicated for adult and pediatric patients, including those with osteopenic bone, requiring fixation of small and long bone fractures (e.g., tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, clavicle).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
January 8, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smith & Nephew, Inc. Samantha Staubach Regulatory Affairs Specialist II 1450 Brooks Road Memphis, Tennessee 38116
### Re: K173293
Trade/Device Name: EVOS Small Fragment Upper Extremity Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: December 19, 2017 Received: December 20, 2017
Dear Ms. Staubach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# >< smith&nephew
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-----------------------------------------|---------------------------------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020<br>See PRA Statement below. |
**Indications for Use**
| 510(k) Number (if known) | K173293 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | EVOS Small Fragment Upper Extremity Plates |
| Indications for Use (Describe) | The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle. |
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="2" x2="13" y1="2" y2="13"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="13" x2="2" y1="2" y2="13"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="15" width="15"> <rect height="15" style="fill:white;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
## >< We are smith&nephew
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | October 12, 2017 |
| Name of Device: | Samantha Staubach<br>Regulatory Affairs Specialist II<br>T 901-399-6132<br>F 901-566-7596<br>EVOS Small Fragment Upper Extremity Plates |
| Common Name: | Bone Plates |
| | |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | HRS |
#### Predicates
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------|
| Smith & Nephew, Inc. | Smith & Nephew Bone Plate<br>System (primary predicate,<br>now branded as TC-100) | K993106 | December 9,<br>1999 |
| Smith & Nephew, Inc. | Smith & Nephew Locking Bone<br>Plate System (now branded as<br>PERI-LOC) | K033669 | December 10,<br>2003 |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System - B Plate Locking<br>Bone Plates and Screws (now<br>branded as PERI-LOC VLP) | K062216 | September 15,<br>2006 |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System -VLP Locking<br>Bone Plates and Screws | K071563 | August 8, 2007 |
| Smith & Nephew, Inc. | EVOS Small Fragment Plating<br>System Straight Plates and<br>Screws | K162078 | November 18,<br>2016 |
| Smith & Nephew, Inc. | EVOS Small Fragment Lower<br>Extremity Plates | K170457 | June 14, 2017 |
| Smith & Nephew, Inc. | EVOS Small Fragment Plating<br>System | K170887 | April 24, 2017 |
{4}------------------------------------------------
#### Device Description
Subject of this premarket notification is an extension of the EVOS Small Fragment Plating System, the EVOS Small Fragment Upper Extremity Plates. The subject plates feature similarities to existing Smith & Nephew small fragment plating systems (TC-100 Small Bone Plating System, PERI-LOC/PERI-LOC VLP Plating Systems) and also share some instruments and compatible implants from the existing EVOS MINI Plating System and EVOS Small Fragment Plating System. When compared against EVOS Small Fragment Plates already cleared for market, the subject plates use the same hole features and are designed to work with the same bone screws. The EVOS Small Fragment Upper Extremity Plates are available in a variety of plate designs for specific anatomical areas. These plate designs include plates with 2.7mm holes or 3.5mm holes only, or with a combination of 2.7mm and 3.5mm holes. Plate designs may include threaded holes, non-threaded holes, and variable-angle locking holes.
#### Indications for Use
The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
#### Technological Characteristics
Device comparisons described in this premarket notification demonstrated that the proposed devices are substantially equivalent to legally marketed predicates with respect to intended use, indications, and performance characteristics. The subject devices include plates designs with a variety of hole features, which are available on EVOS Small Fragment Plates previously cleared for market. The subject plates differ from the EVOS Small Fragment Plates previously cleared in that the plate shapes are more anatomically specific for upper extremity fractures.
#### Summary of Pre-Clinical Testing
- . Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates based upon structural strength for further mechanical testing. Plates were separated into groups for evaluation based upon similar designs or anatomical application.
- . Bending fatigue performance or construct fatigue performance was evaluated for the worst case plate designs identified through FEA. Results of the testing determined that the subject EVOS plates performed similar or superior to the predicate plates tested, when evaluated under the same conditions.
- Engineering rationales were leveraged to justify the bending strength of some EVOS . plates based on previously conducted FEA, previously conducted mechanical testing, and hand calculations.
{5}------------------------------------------------
- . Packaging verification testing was conducted for the proposed packaging configurations and the results of this testing demonstrated that the product will not be damaged during shipment and will adequately maintain sterility post shipment.
- · Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance , "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72.
#### Conclusion
This Traditional 510(k) premarket notification is being submitted to request clearance for the EVOS Small Fragment Plating System Upper Extremity Plates. Based on similarities to the predicate plating systems and a review of the mechanical testing performed, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.