K162078 · Smith & Nephew, Inc. · HRS · Nov 18, 2016 · Orthopedic
Device Facts
Record ID
K162078
Device Name
EVOS Small Fragment Plating System
Applicant
Smith & Nephew, Inc.
Product Code
HRS · Orthopedic
Decision Date
Nov 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The EVOS Small Fragment Plating System is indicated for adult and pediatric patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
Device Story
EVOS Small Fragment Plating System consists of metallic bone plates and screws for fracture fixation. System includes 2.7mm and 3.5mm locking/non-locking straight plates, 3.5mm locking/non-locking screws, and 4.7mm non-locking osteopenia screws. Used by orthopedic surgeons in clinical settings for internal fixation of small and long bone fractures. Devices provide mechanical stabilization of bone fragments to facilitate healing. Surgeon selects appropriate plate/screw configuration based on fracture site and bone quality. Output is physical stabilization of bone; clinical benefit is restored skeletal alignment and support during fracture healing.
Clinical Evidence
No clinical data. Evidence consists of bench testing: FEA for stress concentration, four-point bend fatigue testing, static torsional performance (ASTM F543), axial pull-out strength (ASTM F543), and static cantilever bending of locking mechanisms. Packaging and bacterial endotoxin testing (ANSI/AAMI ST72) were also performed.
Technological Characteristics
Metallic bone plates and screws. Locking plates feature threaded or variable angle holes; non-locking plates have non-threaded holes. Screws feature hex drive. Materials and threadforms are similar to existing Smith & Nephew predicates. Testing performed per ASTM F543. Sterilization validated per FDA guidance and ANSI/AAMI ST72.
Indications for Use
Indicated for adult and pediatric patients, including those with osteopenic bone, requiring fixation of small and long bone fractures (e.g., tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, clavicle).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Synthes Small Fragment Dynamic Compression Locking System (K000684)
Smith & Nephew Locking Bone Plate System (K033669)
PERI-LOC Periarticular Locked Plating System – B Plate Locking Bone Plates and Screws (K062216)
PERI-LOC Periarticular Locked Plating System Hexalobular Bone Screws (K082516)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2016
Smith & Nephew, Inc. Samantha Staubach Regulatory Affairs Specialist 1450 Brooks Road Memphis. Tennessee 38119
Re: K162078
Trade/Device Name: EVOS Small Fragment Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 27, 2016 Received: October 28, 2016
Dear Samantha Staubach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# K162078 page 1 of 1
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
| 510(k) Number (if known) | K162078 |
|--------------------------|---------|
|--------------------------|---------|
Device Name EVOS Small Fragment Plating System
## Indications for Use (Describe)
045
The EVOS Small Fragment Plating System is indicated for adult and pediatric patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
| Type of Use (Select one or both, as applicable) | <table><tr><td><span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# >< We are smith&nephew
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary: | October 27, 2016 |
| Name of Device: | EVOS Small Fragment Plating System |
| Common Name: | Bone Plates and Screws |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br><br>21 CFR 888.3040 Smooth or threaded metallic bone fixation<br>fastener |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | HRS/HWC |
### Predicates
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|----------------------|---------------------------------------------------------------------------------------------|----------------------|-----------------------|
| Smith & Nephew, Inc. | Smith & Nephew Bone Plate<br>System (primary predicate,<br>now branded as TC-100) | K993106 | December 9,<br>1999 |
| Synthes (USA) | Synthes Small Fragment<br>Dynamic Compression Locking<br>System | K000684 | April 28, 2000 |
| Smith & Nephew, Inc. | Smith & Nephew Locking Bone<br>Plate System (now branded as<br>PERI-LOC) | K033669 | December 10,<br>2003 |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System – B Plate Locking<br>Bone Plates and Screws | K062216 | September 15,<br>2006 |
| Smith & Nephew, Inc. | PERI-LOC Periarticular Locked<br>Plating System Hexalobular Bone<br>Screws | K082516 | September 17,<br>2008 |
# Device Description
Subject of this premarket notification is an extension to the EVOS family of plates and screws, the EVOS Small Fragment Plating System. This system features similarities to existing Smith & Nephew small fragment plates (TC-100 Small Bone Plating System, PERI-LOC Plating System) and also shares some instruments and implants from the existing EVOS MINI Plating System.
{4}------------------------------------------------
It is comprised of a variety of locking and non-locking 2.7mm and 3.5mm straight plates as well as 3.5mm locking and non-locking screws and 4.7mm non-locking osteopenia screws.
#### Indications for Use
The EVOS Small Fragment Plating System is indicated for adult and pediatric patients, as well as patients with osteopenic bone. It is indicated for fixation of small and long bone fractures, including, but not limited to, those of the tibia, fibula, femur, humerus, ulna, radius, pelvis, acetabulum, metacarpals, metatarsals, and clavicle.
#### Technological Characteristics
Device comparisons described in this premarket notification demonstrated that the proposed devices are substantially equivalent to legally marketed predicates with respect to intended use, indications, and performance characteristics. The subject devices include locking and non-locking plates. Screw holes in the locking plates can either be threaded holes or variable angle holes and the non-locking plates have non-threaded holes. The subject screws feature a hex drive and are similar to existing Smith & Nephew predicate screws with respect to threadform and major and minor diameter.
#### Summary of Pre-Clinical Testing
- Finite element analysis (FEA) was conducted on the proposed plate designs to determine the worst case plates as well as the high stress region of the plate to be used for subsequent mechanical testing. The acceptance criteria were met in that the subject bone plates exhibited similar or superior structural strength compared to the existing predicates. Results of the FEA demonstrated that the plates identified for mechanical testing were the appropriate bone plates because they possessed the highest stress concentrations.
- Four point bend fatigue testing was conducted on the worst-case designs of the . proposed bone plates, as identified through FEA. Results of the testing concluded that the bending fatigue performance achieved by the proposed bone plates met the acceptance criteria in that they were found to be similar to the bending fatigue performance of a previously cleared predicate.
- . Torque to failure testing for the proposed 3.5mm and 4.7mm bone screws was conducted following the guidelines of ASTM F543. The acceptance criterion was met in that the static torsional performance of the EVOS screws was found to be similar to the static torsional performance of a previously cleared predicate device.
- Axial pull-out testing was conducted on the proposed 4.7mm osteopenia screws . following the guidelines of ASTM F543. The acceptance criterion was met in that the subject screws that were tested showed similar or superior (higher) pull-out strength compared to the predicates.
- . Static cantilever bending performance of the locking mechanism for the 3.5mm and 2.7mm screws was evaluated at a fixed angle through the threaded locking mechanism and variable angle locking holes and the results met the acceptance criteria in that they were similar or superior (higher) than the predicates.
{5}------------------------------------------------
- . Engineering rationales were leveraged for several proposed devices due to the similarity in design when compared against the predicates.
- . Packaging testing was conducted for the larger pouch/carton design and the results of this testing showed that the product will not be damaged during shipment and will adequately maintain sterility post shipment.
- Bacterial endotoxin testing was completed and met the acceptable endotoxin limits . as stated in the FDA Guidance , "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72.
## Conclusion
This Traditional 510(k) premarket notification is being submitted to request clearance for the EVOS Small Fragment Plating System Straight Plates and Screws. Based on similarities to the predicate components and a review of the mechanical testing performed, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.