XinBo Electrode (Model: Neck Therapy Pro - I, Neck Therapy Pro - II, Neck Therapy Pro - III, Neck Therapy Pro - IV)
Applicant
Guangzhou Xinbo Electronic Co., Ltd.
Product Code
GXY · Neurology
Decision Date
Apr 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
XinBo Electrodes (Models: Neck Therapy Pro - I, Neck Therapy Pro - II, Neck Therapy Pro - III, Neck Therapy Pro - IV) are cutaneous electrodes to be used with legally marketed TENS stimulating devices for which stimulation of the back of the neck and upper back is appropriate. They will deliver the stimulation signals generated by the stimulator to the back of the neck, upper back and shoulder surface areas with which they are in contact.
Device Story
Cutaneous electrodes designed for use with external TENS stimulators; deliver electrical stimulation signals to back of neck, upper back, and shoulder areas. Device consists of conductive pads (silver-plated nylon or conductive silicone rubber) that interface with skin; connects to stimulator via lead wires with snap/socket connectors. Used in home or clinical settings by patients or providers. Low resistance (<7 ohms/inch) ensures uniform current distribution. Benefits include targeted pain relief via TENS therapy. No active electronic components; passive interface device.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1), hot spot verification, and usability study (IEC 60601-1-6, IEC 62366). No clinical data presented.
Technological Characteristics
Materials: Silver-plated nylon (silver fiber knitted fabric), cotton (crystal super cashmere, mutispandex), conductive silicone rubber, PU foam. Impedance: <7 ohms/inch. Connectivity: Lead wires with female socket or male snap connectors. Standards: ISO 10993-5/10, IEC 60601-1, IEC 60601-1-6, IEC 62366.
Indications for Use
Indicated for use with legally marketed TENS stimulating devices to deliver electrical stimulation to the back of the neck, upper back, and shoulder surface areas.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 19, 2018
Guangzhou Xinbo Electronic Co., Ltd. % Ms. Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, Guandong, 510006 Cn
Re: K172887
Trade/Device Name: XinBo Electrode (Model: Neck Therapy Pro - I. Neck Therapy Pro - II. Neck Therapy Pro - III, Neck Therapy Pro - IV) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 14, 2018 Received: January 19, 2018
Dear Ms. Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172887
Device Name XinBo Electrodes
#### Indications for Use (Describe)
XinBo Electrodes (Model: Neck Therapy Pro - II; Neck Therapy Pro - II; Neck Therapy Pro - III, Neck Therapy Pro - IV) are cutaneous electrodes to be used with legally marketed TENS stimulating devices for which stimulation of the back of the neck and upper back is appropriate. They will deliver the stimulation signals generated by the back of the neck, upper back and shoulder surface areas with which they are in contact.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------|
| | Indoor Use (Applicable to OFF!® Insect Repellents Only) |
| | Outdoor Use (Applicable to OFF!® Insect Repellents Only) |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
510(k) Owner's Name: Guangzhou Xinbo Electronic Co., Ltd. Establishment Registration Number: K172887 Address: 2nd Building, Juntuo Industry Park, XinqyeDadao, Nancun Town, Panyu District, Guangzhou City, Guangdong Province, China Contact Person: Li Huifang (Manager) Email: drtvsammy@hotmail.com
#### Application Correspondent:
Contact Person: Ms. Cecilia Ceng / Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com
# 2. Subject Device Information
Trade Name: XinBo Electrodes Common Name: Cutaneous Electrode Review Panel: Neurology Model: Neck Therapy Pro - I, Neck Therapy Pro - II, Neck Therapy Pro - III, Neck Therapy Pro - IV Regulation Number: 882.1320
# 3. Predicate Device Information
| Sponsor | Neurotron Medical, Inc | DRTV Asia Ltd. | AMPCAXE, LLC |
|----------------------|------------------------|-------------------------------|------------------------------------|
| Device Name | Theraknit Garments | Dr-Ho's Foot Pad<br>Electrode | AMPCARE 50709<br>Series Electrodes |
| 510(k) Number | K053214 | K151693 | K121483 |
| Product Code | GXY | GXY | GXY |
| Regulation<br>Number | 882.1320 | 882.1320 | 882.1320 |
| Regulation Class | II | II | II |
#### 4. Device Description
The electrodes can be use together with any legally marketed TENS stimulating devices for which stimulation of the back of the neck and upper back is appropriate. For the silver coated Nylon yarn electrodes, they can be used dry or wet when in contact with the skin. The entire surface of each
{4}------------------------------------------------
electrodes is very conductive having a resistance of less than 7 ohms per inch; For the model "Neck Therapy Pro - III", conductive silicone rubber having a resistance of less than 7 ohms per inch, these low resistance provides low current density with uniform current distribution.
#### 5. Intended Use / Indications for Use
XinBo Electrodes (Models: Neck Therapy Pro - I, Neck Therapy Pro - II, Neck Therapy Pro - III, Neck Therapy Pro - IV) are cutaneous electrodes to be used with legally marketed TENS stimulating devices for which stimulation of the back of the neck and upper back is appropriate. They will deliver the stimulation signals generated by the stimulator to the back of the neck, upper back and shoulder surface areas with which they are in contact.
#### 6. Test Summary
XinBo Electrodes have been evaluated the safety and performance by lab bench testing as following:
- Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards �
- The electrode wire is compliance with 21 CFR 898 by IEC 60601-1 (version 3.1, clause 8.5.2.3) � evaluation.
- � Performance test (hot spots verification test) according to manufacturer self-requirements.
- � Usability study according to IEC 60601-1-6, IEC 62366, and "FDA Guidance for Applying Human Factors and Usability Enqineering to Medical Devices"
# 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of subject device are substantially equivalent to the predicate devices.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Remark |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name<br>and Model | XinBo Electrodes<br>(Models: Neck<br>Therapy Pro - I, Neck<br>Therapy Pro - II, Neck<br>Therapy Pro - III, Neck<br>Therapy Pro - IV) | Theraknit Garments | Dr-Ho's Foot Pad<br>Electrode | AMPCARE 50709<br>Series Electrodes | -- |
| 510(k) Number | K172887 | K053214 | K151693 | K121483 | -- |
| Product Code | GXY | GXY | GXY | GXY | SE |
| OTC or Rx-<br>only | OTC | Rx-only | OTC | Rx-only | SE |
| Intended Use /<br>Indications for<br>Use | XinBo Electrodes<br>(Models: Neck<br>Therapy Pro - I, Neck<br>Therapy Pro - II, Neck<br>Therapy Pro - III, Neck<br>Therapy Pro - IV) are<br>cutaneous electrodes | The TheraKnit<br>Garment Electrodes<br>are cutaneous<br>electrodes to be used<br>with legally marketed<br>TENS stimulating<br>device. The knitted | The Cutaneous<br>Electrodes, "Dr-Ho's<br>Foot Pad Electrodes",<br>are intended to be<br>used with legally<br>marketed electrical<br>stimulating devices | AMPCA RE 50709<br>Series of cutaneous<br>electrodes are<br>intended to be used to<br>apply electrical<br>stimulation current to<br>the patient's skin. | SE |
| to be used with legally<br>marketed TENS<br>stimulating devices for<br>which stimulation of<br>the back of the neck<br>and upper back is<br>appropriate. They will<br>deliver the stimulation<br>signals generated by<br>the stimulator to the<br>back of the neck,<br>upper back and<br>shoulder surface<br>areas with which they<br>are in contact. | garment electrodes<br>will deliver the<br>stimulation signals<br>generated by the<br>stimulator to the body<br>surface with which<br>they are in contact.<br>These body parts can<br>include hand (glove),<br>feet (socks), elbow or<br>knee (sleeve), arm,<br>leg, shoulder, back<br>(pads). | such as<br>transcutaneous<br>electrical nerve<br>stimulators or<br>powered muscle<br>stimulators. The<br>cutaneous electrodes<br>will deliver the<br>stimulation signals<br>generated by the<br>stimulator to the<br>bottom of the feet<br>which they are in<br>contact with. | Example electrical<br>stimulation current<br>applications of these<br>electrodes are:<br>a) Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS)<br>for pain relief<br>b) Electrical muscle<br>stimulation (EMS) for<br>neck muscle<br>stimulation.<br>c) Functional<br>electrical stimulation<br>(FES)<br>d) Galvanic<br>stimulation<br>e) Microcurrent<br>electrical nerve<br>stimulation (MENS)<br>J) Interferential (IF)<br>stimulation<br>g) Neuromuscular<br>electrical stimulation<br>(NMES), including for<br>muscle re-education<br>by application of<br>external stimulation to<br>the muscles<br>necessary for<br>pharyngeal<br>contraction. | | |
| Body Parts to<br>Use on | Back of the neck,<br>upper back and<br>shoulder | Hand, feet, elbow or<br>knee, arm, leg,<br>shoulder, back | Bottom of the feet | Back of the neck | SE |
| Design<br>(Shape) | Model A: Neck<br>Therapy Pro - I<br>Model B: Neck<br>Therapy Pro - II<br>Model C: Neck<br>Therapy Pro - III<br>Model D: Neck<br>Therapy Pro - IV | Electrode A: Glove<br>Style<br>Electrode B: Socks<br>Style<br>Electrode C: Sleeve<br>Style<br>Electrode D: Pads<br>Style | Electrode A: foot-shaped Electrode | | SE<br>Note 1 |
| Dimensions | Model A:<br>For the whole product:<br>260 x 297.8 mm<br>For only the electronic<br>pads part: 4<br>electrodes on each<br>models, two of them<br>are: 65 mm in<br>diameter, the other<br>two of them are<br>irregularly: 75 mm in<br>diameter and about<br>105 mm long<br>Model B:<br>For the whole product: | | | | SE<br>Note 1 |
| | | | | | |
| | 208 x115 x110 mm<br>For only the electronic<br>pads part: 2<br>electrodes on each<br>models, each<br>electrode is: 50 mm in<br>diameter<br>Model C:<br>For the whole product:<br>264.2 x 298.5 mm<br>For each the<br>electronic pads part: 6<br>electrodes on each<br>models, each<br>electrode is: 48 x26.2<br>mm²<br>Model D:<br>For the whole product:<br>300x 450 x 110mm<br>For each electronic<br>pads part: 4<br>electrodes on each<br>models, each<br>electrode is: 45 mm in<br>diameter | | | | |
| Impedance | For Model A: 7 ohms | 7 ohms resistance per inch | <7 ohms per inch | <7 ohms per inch | SE |
| Parameters | per inch<br>For Model B: 7 ohms<br>per inch<br>For Model C: Less<br>than 7 ohms per inch<br>For Model D: 7 ohms<br>resistance per inch | | | | |
| Patient<br>Contacting<br>Material | For Model A: Silver<br>plated nylon (Silver<br>Fibre Knitted Fabric),<br>cotton (Crystal Super<br>Cashmere)<br>For Electrode B: Silver<br>plated nylon (Silver<br>Fibre Knitted Fabric),<br>cotton (Crystal Super<br>Cashmere and<br>Mutispandex)<br>For Electrode C:<br>Conductive silicone<br>rubber, PU Foam<br>For Electrode D: Silver<br>plated nylon (Silver<br>Fibre Knitted Fabric),<br>cotton (Crystal Super<br>Cashmere) | Silver plated nylon | Conductive silicone<br>rubber | -- | SE<br>Note 1 |
| Lead Wire | Lead wire with female | Lead wire with female | Connector Lead wire | -- | SE |
| Type and | socket, or snap button | socket, or snap button | .080" female socket | | |
| Characteristics | with male snap connector | with male snap connector | connector | | |
| Operating<br>Environment | Temperature: 5~40°C<br>Humidity: ≤ 80%RH<br>Atmospheric Pressure:<br>86~106kPa | Temperature: 5~40°C<br>Humidity: ≤ 80%RH<br>Atmospheric<br>Pressure: 86~106kPa | Temperature: 5~40°C<br>Humidity: ≤ 80%RH<br>Atmospheric<br>Pressure: 86~106kPa | -- | SE |
| Storage<br>Environment | Temperature: -20 ~<br>55°C<br>Humidity: ≤ 95% RH<br>Atmospheric Pressure:<br>50~106 kPa | Temperature: -20 ~<br>55°C<br>Humidity: ≤ 95% RH<br>Atmospheric<br>Pressure: 50~106 kPa | Temperature: -20 ~<br>55°C<br>Humidity: ≤ 95% RH<br>Atmospheric<br>Pressure: 50~106 kPa | -- | SE |
| Biocompatibility | Complied with ISO<br>10993-5, ISO 10993-<br>10 | Complied with ISO<br>10993-5, ISO 10993-<br>10 | Complied with ISO<br>10993-5, ISO 10993-<br>10 | Complied with ISO<br>10993-5, ISO 10993-<br>10 | SE |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
# Comparison in Detail(s):
#### Note 1:
Although the "Design (shape)", "Dimensions", and "Patient Contacting Material" are a little different from the predicate devices, they all comply with ISO 10993 requirements. So the differences will not raise any safety or effectiveness issue.
# Final Conclusion:
The subject device XinBo Electrodes (Models: Neck Therapy Pro - I, Neck Therapy Pro - II, Neck Therapy Pro -III, Neck Therapy Pro -IV) are Substantial Equivalent to the predicate devices.
# 8. Date of the summary prepared: April 20, 2018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.