AMPCARE 50709 Series of cutaneous electrodes are intended to be used to apply electrical stimulation current to the patient's skin. Example electrical stimulation current applications of these electrodes are: a) Transcutaneous Electrical Nerve Stimulation (TENS) for pain relief, b) Electrical muscle stimulation (EMS) for neck muscle stimulation, c) Functional electrical stimulation (FES), d) Galvanic stimulation, e) Microcurrent electrical nerve stimulation (MENS), f) Interferential (IF) stimulation, and g) Neuromuscular electrical stimulation (NMES), including for muscle re-education by application of external stimulation to the muscles necessary for pharyngeal contraction.
Device Story
AMPCARE 50709 Series are non-sterile, self-adhering, reusable cutaneous electrodes; available with pin/socket or snap connectors. Electrodes interface with external electrical stimulation devices to deliver current to patient skin. Used in clinical or home settings under prescription. Construction includes non-conductive top layer, conductive center layer, and adhesive bottom layer. Plastic liner removed prior to application. Electrodes facilitate various stimulation modalities (TENS, EMS, FES, etc.) for therapeutic purposes like pain relief or muscle re-education. Benefit derived from uniform current distribution, preventing 'hot spots' and patient discomfort during stimulation.
Clinical Evidence
Bench testing only. No clinical data. Testing included dispersion testing to verify uniform current distribution and absence of 'hot spots' compared to Columbia (K080386) and Uni-Patch (K915333) electrodes. Electrical impedance and impedance uniformity testing performed to demonstrate equivalence to predicate devices.
Technological Characteristics
Three-layer construction: non-conductive top layer, conductive center layer, adhesive bottom layer. Pin/socket or snap connectors; pin/socket version compatible with 21 CFR 898 lead wires. Non-sterile, self-adhering, reusable. No active electronic components; passive electrode.
Indications for Use
Indicated for patients requiring electrical stimulation therapy, including TENS for pain, EMS for neck muscles, FES, galvanic stimulation, MENS, IF stimulation, and NMES for muscle re-education (including pharyngeal contraction).
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
Predicate Devices
Labeltape Meditect, Inc. (Uni-Patch / Covidien) (K915333)
SelectiveMed Components, Inc. (Guardian) (K083756)
Submission Summary (Full Text)
{0}------------------------------------------------
# AMPCARE
OCT 15 2012
## 5.510(k) SUMMARY - K121483
| Submitter: | AMPCARE, LLC<br>1201 Richardson Drive, Suite 280<br>Richardson, Texas 75080 | |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Russell Campbell<br>CEO / President<br>TEL: 682-561-2444<br>FAX: 817-923-1377<br>rcampbell@ampcarellc.com | |
| Date Prepared: | May 17, 2012 revised September 13, 2012 | |
| Trade Name: | Reusable Cutaneous Electrode | |
| Common Name: | Cutaneous Electrode | |
| Classification Name: | GXY 882.1320 | Electrode, Cutaneous |
| Predicate Devices: | K915333 GXY<br>K080386 GXY<br>K852267 GXY<br>K083756 GXY | Labeltape Meditect, Inc. (Uni-Patch / Covidien)<br>Columbia Scientific Development, LLC<br>Axelgaard Manufacturing Company, Ltd.<br>SelectiveMed Components, Inc. (Guardian) |
| Device Description: | The new AMPCARE 50709 Series Electrodes device is a family of cutaneous<br>electrodes with various shapes and sizes, which are similar in design and<br>construction to several predicate cutaneous electrodes. AMPCARE electrodes are<br>non-sterile, self-adhering, and are for multiple use by a single patient, available with | |
AMPCARE 50709 Series of cutaneous electrodes are intended to be used to apply Statement of electrical stimulation current to the patient's skin. Example electrical stimulation Intended Use: current applications of these electrodes are: a) Transcutaneous Electrical Nerve Stimulation (TENS) for pain relief, b) Electrical muscle stimulation (EMS) for neck muscle stimulation, c) Functional electrical stimulation (FES), d) Galvanic stimulation, e) Microcurrent electrical nerve stimulation (MENS), f) Interferential (IF) stimulation, and g) Neuromuscular electrical stimulation (NMES), including for muscle re-education by application of external stimulation to the muscles necessary for pharyngeal contraction.
either pin/socket or snap connections.
The new AMPCARE device is designed in accordance with the general design Summary of approach of the predicate devices referenced above. Specifically, each electrode in Technological the AMPCARE 50709 Series is constructed with a non-conductive top layer, Characteristics: conductive center layer and adhesive bottom layer. The plastic liner at the bottom of the electrode is peeled away just prior to placement on the patient, consistent with the four predicate devices. The pin/socket connector version of the new device will be provided with a lead wire that is 21 CFR 898 compatible, containing a female recessed socket for electrical connection.
{1}------------------------------------------------
AMPCARE
Summary of Non-Clinical Testing: Dispersion testing has demonstrated that the new AMPCARE device has uniform current distribution with no evidence of "hot spots" that could cause patient discomfort or injury. The Columbia (K080386) and Uni-Patch (K915333) electrodes were included in the dispersion testing for comparison purposes. Additional comparison testing to predicate device electrodes was performed to demonstrate suitability of use for each example of electrical stimulation current applications given in the Indications for Use. Both electrical impedance testing and electrical impedance uniformity testing was performed on AMPCARE electrodes as well as predicate device electrodes to demonstrate substantial equivalence.
Conclusion:
AMPCARE considers the AMPCARE 50709 Series Electrodes to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, nonclinical test results, and established medical use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a bird-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 15 2012
AMPCARE, LLC % Ms. Diane Rutherford Submissions Manager Ken Block Consulting 1201 Richardson Drive, Suite 280 Richardson, TX 75080
Re: K121483
Trade/Device Name: AMPCARE 50709 Series Electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Electrode, cutaneous Regulatory Class: Class II Product Code: GXY Dated: September 13, 2012 Received: September 14, 2012
Dear Ms. Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
### Page 2- Ms. Diane Rutherford
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
12
Sincerely yours,
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE
#### 510(k) Number: K121483
Device Name: AMPCARE 50709 Series Electrodes
Indications for Use:
AMPCARE 50709 Series of cutaneous electrodes are intended to be used to apply electrical stimulation current to the patient's skin. Example electrical stimulation current applications of these electrodes are:
a) Transcutaneous Electrical Nerve Stimulation (TENS) for pain relief
b) Electrical muscle stimulation (EMS) for neck muscle stimulation.
- c) Functional electrical stimulation (FES)
- d) Galvanic stimulation.
- e) Microcurrent electrical nerve stimulation (MENS).
- Interferential (IF) stimulation. 0
- g) Neuromuscular electrical stimulation (NMES), including for muscle re-education by application of external stimulation to the muscles necessary for pharyngeal contraction.
Prescription Use X (21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDHR, Office of Device Evaluation (ODE)
Callaway
(Divisjon Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121483
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.