New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded)
K172761 · Shanghai New Century Dental Materials Co., Ltd. · EIH · Nov 17, 2017 · Dental
Device Facts
Record ID
K172761
Device Name
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded)
Applicant
Shanghai New Century Dental Materials Co., Ltd.
Product Code
EIH · Dental
Decision Date
Nov 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded) are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
Device Story
Device consists of zirconia powder-based blocks, discs, and rods; used as raw material for CAD/CAM milling of dental restorations (crowns, bridges, inlays, veneers). Material composition: ZrO2, Y2O3, HfO2, Al2O3; pre-shaded versions include trace inorganic pigments. Operated by dental professionals in clinical/laboratory settings; material is milled into final patient-specific geometry then sintered at ~1500 °C. Output is a custom-fit ceramic dental restoration. Benefits include high-strength, biocompatible, aesthetic tooth replacement.
Clinical Evidence
Bench testing only. Performance verified against ISO 6872:2008 (Dentistry: Ceramic Materials) for density and mechanical properties. Biocompatibility assessment conducted per ISO 10993-1.
Technological Characteristics
Materials: Yttria-stabilized Zirconia (TOSOH TZ series). Form factors: blocks, discs, rods. Processing: Sintering at ~1500 °C. Standards: ISO 6872:2008 (performance), ISO 10993-1 (biocompatibility). Non-sterile. No software or electronic components.
Indications for Use
Indicated for dental patients requiring all-ceramic restorations, including full contour crowns, bridges, inlays, and veneers, fabricated via CAD/CAM technology as prescribed by a dentist.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai New Century Dental Materials Co., Ltd. % Mr. Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland, New Jersey 07436
Re: K172761
Trade/Device Name: New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded) Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: September 13, 2017 Received: September 13, 2017
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
November 17, 2017
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172761
Device Name
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded)
Indications for Use (Describe)
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded) are Intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K172761: 510(k) Summary:
### Submitter & Foreign Manufacture Identification
| Name: | Shanghai New Century Dental Materials Co., Ltd. |
|------------|-----------------------------------------------------------------------------|
| Address: | No.461 Tanghang Village, Huating Town, Jiading District<br>Shanghai, China. |
| Telephone: | (086) 21-59954240 |
### Contact Person
Charles Shen Manton Business and Technology Services 37 Winding Ridge, Oakland, New Jersey 07436 Telephone: 608-217-9358 Email: cyshen@aol.com
#### Date of Summary: November 8, 2017
| Device Name: | |
|------------------------|-----------------------------------------------------------------------------|
| Proprietary Name: | New Century All-Ceramic Dental Zirconia Blocks (Un-<br>Shaded & Pre-Shaded) |
| Common Name: | Dental Zirconia Ceramics |
| Classification Name: | Powder, Porcelain |
| Device Classification: | II |
| Regulation Number: | 21 CFR 872.6660 |
| Panel: | Dental |
| Product Code: | EIH |
#### Predicate Device Information:
K140070 Luxer Shaded Zirconia
#### Device Description:
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded) are derived from Zirconia powder that has been processed into their final net shapes. These blanks are then further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/ onlay. The zirconia powder is composed of ZrO2 + Y2O3 + HfO2 + A12O2. The performance of the dental blanks conforms to ISO 6872: 2008, Dentistry: Ceramic Materials
{4}------------------------------------------------
The shading effects of New Century All-Ceramic Dental Zirconia Blocks (Pre-Shaded) are achieved by blending different zirconia powder material, all from the same supplier; have almost identical composition, differing only in the trace amount of inorganic pigments to achieve the shading effects in dental blanks. The performance of the preshaded dental zirconia blanks also conforms to ISO 6872, Dentistry: Ceramic Materials
### Indications for Use:
New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded) are intended for use with CAD/CAM technology to produce all ceramic dental restorations (full contour crowns, bridges, inlays, and Veneers) as prescribed by a dentist.
# Summary of Non-Clinical Performance Testing:
Bench testing was performed per ISO 6872:2008 and using internal testing procedures (for presintered and sintered density) to ensure that the New Century All-Ceramic Dental Zirconia Blocks (Un-Shaded & Pre-Shaded) met the specifications. All tests were verified to meet acceptance criteria. A biocompatibility assessment was conducted in accordance with ISO 10993-1.
# Technological Comparison with Predicate Device
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
| Description | Subject Device | Predicate Device (K140070) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | New Century All-Ceramic Dental<br>Zirconia Blocks (Un-Shaded & Pre-<br>Shaded) are intended for use with<br>CAD/CAM technology to produce all<br>ceramic dental restorations (full contour<br>crowns, bridges, inlays and Veneers) as<br>prescribed by a dentist. | Luxer Shaded Zirconia blanks are<br>Intended for use with CAD/CAM<br>technology to produce all ceramic dental<br>restorations (full contour crowns and<br>bridges) as prescribed by a dentist. |
| Basic Design | Blocks, disc, and rod | Blocks |
| Materials | Yttria stabilized Zirconia from TOSOH TZ<br>Series of zirconia (white, yellow, pink, and<br>gray) | Yttria stabilized Zirconia from TOSOH<br>TZ Series of zirconia (white, yellow,<br>pink, and gray) |
| Processing | Sintering at temperature around 1500 °C | Public information not available |
| Dimension | Various | Various |
| Single Use | Yes | Yes |
### Table 5.1: Comparison of Intended Use, Design, Material, and Processing
{5}------------------------------------------------
| Color | None, and Pre-shaded ( for pre-shaded<br>series) | Pre-shaded |
|---------|--------------------------------------------------|-------------|
| Sterile | Non-sterile | Non-sterile |
Our device is highly similar to the predicate device in terms of indications for use, design, material, and work flow to fabricate dental restorations. Our device is different from the predicate device in the following:
- The subject device has blocks, disc, and rod, while the predicate device only has (1) blocks. This difference does not raise any safety or efficacy concerns for two reasons: First, the initial product shape does not matter because they are going to be milled into the final shape before patient use. Secondly, the dental zirconia blanks are known to have multiple different shapes, including blocks, discs, and rods, such as described in K141724 by Liaoning Upcera Company Limited.
- (2) The sintering temperature is about 1500 ℃ for the subject device, while this information is not publicly available for the predicate device. This difference does not raise any safety or efficacy concerns as the sintering temperature mainly affects the physical and mechanical property of the dental blanks. Both the subject and predicate device have similar physical/mechanical properties that met the requirements of ISO 6872.
- (3) The subject has both non-shaded and pre-shaded blanks, while the predicate device has only the pre-shaded series. This difference does not raise any safety or efficacy concerns because there are many non-shaded zirconia blanks on the market, such as described in K141724 by Liaoning Upcera Company Limited.
### Conclusion
In conclusion, based on the indications for use, technology and non-clinical performance testing, New Century All-Ceramic Dentistry Zirconia Blocks (Un-Shaded and Pre-Shaded) are substantially equivalent to K140070 Luxer Shaded Zirconia.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.