The Sterile Disposable Syringes are intended for the injection of contrast media or saline, they shall be used with an U.S legally marketed angiographic injector.
Device Story
Sterile disposable syringes designed for use with power-driven angiographic injectors to deliver contrast media or saline. Device components include syringe barrel, piston, piston seat, protective cap, connecting tubes (straight, spiral, or T-shape), J-shape tubes, and spikes. Used in clinical settings by healthcare professionals. Syringes are single-use, EO-sterilized, and available in various volumes (50ml to 200ml). Operation involves manual loading of contrast/saline via spike or J-tube, followed by power-driven injection via the angiographic injector. Output is the controlled delivery of fluid into the patient's vascular system. Benefits include reliable, sterile fluid delivery for diagnostic imaging procedures.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 7886-2 (cleanliness, acidity/alkalinity, extractable metals, lubricant, capacity, scale, barrel, plunger, nozzle, performance), ISO 594-1/2 (gauging, leakage, separation force, stress cracking, torque, assembly), ASTM F88/F1929 (packaging), and ISO 10993-7 (EO/ECH residue). Biocompatibility verified via cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, hemolysis, pyrogen, and complement activity testing.
Technological Characteristics
Materials: Polycarbonate (PC), Polypropylene (PP), Natural Rubber, Polyethylene (PE), Acrylonitrile-butadiene-styrene (ABS). Sensing/Actuation: Mechanical, power-driven by external angiographic injector. Dimensions: 50ml to 200ml volumes. Connectivity: None (mechanical interface). Sterilization: Ethylene Oxide (EO). Standards: ISO 7886-2, ISO 594-1, ISO 594-2, ISO 10993, ASTM F88, ASTM F1929, ASTM F756, USP <85>.
Indications for Use
Indicated for injection of contrast media or saline using a legally marketed angiographic injector.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 6, 2018
Shenyang MasTech Medical Device Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 CN
Re: K172694
Trade/Device Name: Sterile Disposable Syringes Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: Class II Product Code: DXT Dated: May 31, 2018 Received: June 5, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael John -S 2018.07.06 11:58:42 -04'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K172694
Device Name Sterile Disposable Syringes
Indications for Use (Describe)
The Sterile Disposable Syringes are intended for the injection of contrast media or saline, they shall be used with an U.S legally marketed angiographic injector.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------|
| | <div> <input type="checkbox"/> </div> |
| | <div> <input type="checkbox"/> </div> |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Exhibit #10 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K172694
- 1. Date of Preparation: 05/29/2018
- 2. Sponsor Identification
Shenyang MasTech Medical Device Co., Ltd. Zhaojia Village, Qianjin Street, Dadong District, Shenyang 110165, China
Establishment Registration Number: Not yet registered
Contact Person: Liu Han Position: General Manager Tel: +86-24-24331351 Fax: +86-24-24326101 Email: sy@mastech-medical.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
#### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Sterile Disposable Syringes Common Name: Disposable angiographic syringe
#### Regulatory Information
Classification Name: Angiographic injector and syringe; Classification: II; Product Code: DXT Regulation Number: CFR 870.1650 Review Panel: Cardiovascular
#### Intended Use Statement:
The Sterile Disposable Syringes are intended for the injection of contrast media or saline, they shall be used with an US legally marketed angiographic injector.
#### Device Description
The proposed devices are available in packs, which may include different configuration of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S legally marketed angiography injector, compatibilities are shown in Table 1.
| Model (Syringe) | Volume (ml) | Type | K Number of Compatible Angiographic Injector |
|-------------------|-------------|-------------|------------------------------------------------------|
| DSA-150-MARK | 150ml | Single shot | K903390 |
| DSA-150-LF | 150ml | Single shot | K860204 |
| DSA-130-NE | 125ml | Single shot | K092896 |
| CT/DSA-150-LF | 150ml | Single shot | K963071, K873687 |
| CT-200-MARK | 200ml | Single shot | K993728, K991557, K982814, K934086, K924116, K913624 |
| CT-200-MARK-E | 200ml | Single shot | K071378, K063029, K011160 |
| CT-200/200-MARK-E | 200/200ml | Dual shoot | K071378, K063029, K031571 |
| CT-200-LF | 200ml | Single shot | K912944 |
| CT-100-NE | 100ml | Single shot | K133189, K071691, K062168, K052633 |
| CT-200-NE | 200ml | Single shot | K133189, K071691, K062168, |
Table 1 Compatibility between Syringes and Injectors
{5}------------------------------------------------
| | | | K052633 |
|--------------|-----------|-------------|------------------------------------------------------|
| CT/MRI-60-NE | 50ml | Single shot | K133189, K091734 (MRI),<br>K071691, K062168, K052633 |
| MRI-60-LF | 60/60ml | Dual shoot | K073592, K984088 |
| MRI-60-MARK | 65/65ml | Dual shoot | K011991, K935668 |
| MRI-100-MARK | 65/115ml | Dual shoot | K042784, K033247, K012950 |
| MRI-100-EZEM | 100/100ml | Dual shoot | K020892 |
The connecting tubes, which is used to connect the syringe and the catheter. The tubes are available in three configurations, which are straight tube, Type T tube and spiral tube. The different between these tubes is tube shape, which is available in straight shape for straight tube, spiral shape for spiral tube and T shape for T tube. The Type T tube is available in a various specifications to connect different model of syringe.
J shape tube, which is used to draw contrast media/ saline into the syringe barrel.
Spikes, which are used to draw contrast media/saline into the syringe barrel.
- న్. Identification of Predicate Device
510(k) Number: K072696 Product Name: ANT Angiographic Syringes Manufacturer: Shenzhen Ant Hi-Tech Industrial Co., Ltd.
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
| Cleanliness | Clause 5 of ISO 7886-2:1996 |
|----------------------------------|------------------------------|
| Limits for acidity or alkalinity | Clause 6 of ISO 7886-2:1996 |
| Limits for extractable metals | Clause 7 of ISO 7886-2:1996 |
| Lubricant | Clause 8 of ISO 7886-2:1996 |
| Tolerance on graduated capacity | Clause 9 of ISO 7886-2:1996 |
| Graduated scale | Clause 10 of ISO 7886-2:1996 |
| Barrel | Clause 11 of ISO 7886-2:1996 |
| Piston/ plunger assembly | Clause 12 of ISO 7886-2:1996 |
| Nozzle | Clause 13 of ISO 7886-2:1996 |
| Performance | Clause 14 of ISO 7886-2:1996 |
{6}------------------------------------------------
| Gauging | Clause 4.1 of ISO 594-1:1986 |
|--------------------------|------------------------------|
| Liquid leakage | Clause 4.2 of ISO 594-1:1986 |
| Air leakage | Clause 4.3 of ISO 594-1:1986 |
| Separation force | Clause 4.4 of ISO 594-1:1986 |
| Stress cracking | Clause 4.5 of ISO 594-1:1986 |
| | |
| Gauging | Clause 4.1 of ISO 594-2:1998 |
| Leakage | Clause 4.2 of ISO 594-2:1998 |
| Separation force | Clause 4.3 of ISO 594-2:1998 |
| Unscrewing torque | Clause 4.4 of ISO 594-2:1998 |
| Ease of assembly | Clause 4.5 of ISO 594-2:1998 |
| Resistance to overriding | Clause 4.6 of ISO 594-2:1998 |
| Stress cracking | Clause 4.7 of ISO 594-2:1998 |
Sterile Barrier Packaging Testing performed on the proposed device:
| Seal strength | ASTM F88/F88-09 |
|-----------------|-----------------|
| Dye penetration | ASTM F1929-12 |
Sterilization and Shelf Life Testing performed on the proposed device:
| EO residue | ISO 10993-7:2008 |
|--------------------------|-----------------------------------------------|
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP 38-NF 33 <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package Test |
| | were performed on aging samples to verify the |
| | claimed shelf life of the device |
#### Biocompatibility Testing:
The patient-contact materials of proposed devices are identified and the proposed devices were tested for Cytotoxicity (ISO 10993-5), Skin Sensitization (ISO 10993-10), Intracutaneous Reactivity (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), Hemolytic Property (ASTM F756), Pyrogen (ISO 10993-11) and Complement Activity (ISO 10993-4).
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
#### Substantially Equivalent (SE) Comparison 8.
Table 1 Comparison of Technology Characteristics
| Item | Proposed Device | | | Predicate Device<br>K072696 | | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------------------------------------------|
| Product Code | DXT | | | DXT | | |
| Regulation<br>Number | CFR 870.1650 | | | CFR 870.1650 | | |
| Intended Use | The Sterile Disposable Syringes are intended for the injection of contrast<br>media or saline, they shall be used with an U.S legally marketed<br>angiographic injector. | | | ANT Angiographic Syringes are syringes for the injection of contras<br>media or saline. This syringe is for single use with U.S leally marketec<br>angiographic injection. | | |
| Operation mode | Power-driven operation, single use | | | Power-driven operation, single use | | |
| Configuration<br>and Material | Angiographic<br>Syringe | Configuration | Material | Angiographic<br>Syringe | Configuration | Material |
| | | Barrel | Polycarbonate (PC) or<br>Polypropylene (PP) | | Barrel | Polypropylene (PP) or<br>Polyethylene terephthalate (PET) |
| | | Piston | Natural Rubber | | Piston | Unknown |
| | | Piston Seat | Polycarbonate (PC) or<br>Polypropylene (PP) | | Piston Seat | Unknown |
| | | Protective Cap | Polypropylene (PP) | | Protective Cap | Unknown |
| | | Connecting<br>tube | Tubing | | Polyvinyl Chloride (PVC) | Connecting<br>tube |
| | | | Check Valve | | Polycarbonate (PC) | |
| J shape tube | | Polyethylene (PE) | J shape tube | | Unknown | |
| Spike | | Acrylonitrile-butadiene-styrene (ABS) | Spike | | Unknown | |
| Sterility | EO Sterilized | | | EO Sterilized | | |
| Single Use | Yes | | | Yes | | |
S001
{8}------------------------------------------------
| Biocompatibility | Cytotoxicity | No cytotoxicity. | Cytotoxicity | No cytotoxicity. |
|------------------|------------------------------------------------|-------------------------------|------------------------------------------------|-------------------------------|
| | Irritation | No intracutaneous reactivity. | Irritation | No intracutaneous reactivity. |
| | Sensitization | No skin sensitization | Sensitization | No skin sensitization |
| | Systemic Toxicity | No systemic toxicity | Systemic Toxicity | No systemic toxicity |
| | Hemolysis | No hemolysis | Hemolysis | No hemolysis |
| | Pyrogen | No pyrogen | Pyrogen | No pyrogen |
| | Complement Activation | No complement | Complement Activation | No complement |
| Performance | Comply with ISO 7886-2, ISO594-1 and ISO 594-2 | | Comply with ISO 7886-2, ISO594-1 and ISO 594-2 | |
Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
S001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.