K172639 · Laseroptek Co., Ltd. · GEX · Nov 3, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172639
Device Name
PALLAS 308/311 Solid-State UV Laser System
Applicant
Laseroptek Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
UVB phototherapy for psoriasis, vitiligo, atopic dermatitis, and leukoderma of affected skin.
Device Story
PALLAS 308/311 is a self-contained solid-state UV laser system; provides targeted UV-B energy to skin lesions while sparing healthy tissue. System comprises a protective console containing the laser source and a handpiece connected via an articulating arm. Operated by clinicians in a clinical setting; parameters (timing/dosing) controlled via console display; activation via foot-switch. Delivers monochromatic UV-B light at 308 nm or 311 nm. Flat-top, non-Gaussian beam profile. Benefits patients by enabling localized phototherapy for dermatological conditions.
Clinical Evidence
Bench testing only. Performance data includes measurement of wavelength, average output power, and energy fluence. Compliance verified against IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser equipment), IEC 60825-1 (laser safety), and ISO 14971 (risk management). Biocompatibility supported by use of aluminum alloy 6061, previously used in other medical devices.
Technological Characteristics
Solid-state UV laser; 308 nm or 311 nm wavelength. Aluminum alloy 6061 patient-contact parts. Flat-top, non-Gaussian beam profile. Max power 20W. Pulse duration 15-20 ns. Foot-switch activation. Complies with 21 CFR 1040.10/1040.11, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, and ISO 14971.
Indications for Use
Indicated for UVB phototherapy of psoriasis, vitiligo, atopic dermatitis, and leukoderma of affected skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
XTrac Excimer Laser System, Model AL7000 (K020847)
TheraLight, Inc.'s UV120-2 UVA/UVB Phototherapy System (K024020)
Laseroptek Co. Ltd.'s Helios III Q-Switched Nd: Y AG Laser System (K152856)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 3, 2017
Laseroptek Co. Ltd % Mr. Dave Yungvirt Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041
Re: K172639
Trade/Device Name: PALLAS 308/311 Solid-State UV Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 25, 2017 Received: October 26, 2017
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -ડિ
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172639
Device Name
PALLAS 308/311 Solid-State UV Laser System
Indications for Use (Describe)
The indications for use are UVB phototherapy of psoriasis, vitiligo, atopic dermatitis, and leukoderma of affected skin.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary
## 1. General Information
| Applicant/Submitter: | Laseroptek Co., Ltd. |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | #203 & #204 Hyundai I Valley 31<br>Galmachi-Ro, Jungwon-Gu<br>Seongnam-Si, Gyeonggido, 13212<br>Rep. of Korea (South Korea)<br>Tel) +82.31.8023.5150<br>Fax) +82.31.8023.5151 |
| Contact Person: | Mina Joo, BT Solutions, Inc. |
| Address: | Unit 502, 148 Yuksamro, Gangnam-gu, Seoul,<br>Republic of Korea<br>Tel) +82.2.538.9140<br>Email) smanager@btsolutions.co.kr |
| Preparation Date: | August-30-2017 |
### 2. Device Name and Code
| Device Trade Name: | PALLAS 308/311 Solid-State UV Laser System | |
|----------------------|----------------------------------------------------------------------------------------|--|
| Common Name: | UV Laser | |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology | |
| Product Code: | GEX | |
| Regulation Number: | 878.4810 | |
| Classification: | Class II | |
| Review Panel: | General & Plastic Surgery (ODE) | |
### 3. Technical Characteristics in Comparison to Predicate Devices
PALLAS 308/311 Solid-State UV Laser System is substantially equivalent to the following legally marketed predicate devices
| | Predicate Device | Predicate Device | Proposed Device |
|---------------|---------------------------------------------|--------------------------------|--------------------------------------|
| 510(K) Number | K020847 | K062963 | Not Available |
| Manufacturer | PhotoMedex, Inc. | RA Medical Systems, Inc. | Laseroptek Co. Ltd. |
| Device Name | XTrac Excimer Laser<br>System, Model AL7000 | PHAROS Excimer Laser<br>EX-308 | PALLAS Sold-State UV<br>Laser System |
{4}------------------------------------------------
#### PALLAS 308/311 Solid-State UV Laser System
| Clearance Date: | 16 May 2002 | April 3, 2007 | N/A |
|--------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Classification /<br>Regulation | Class 2 / 878.4810 | Class 2 / 878.4810 | Class 2 / 878.4810 |
| Product Code | GEX | GEX | GEX |
| Intended Use | UVB phototherapy for<br>psoriasis, vitiligo, atopic<br>dermatitis, and leukoderma. | UVB phototherapy for<br>psoriasis, vitiligo, atopic<br>dermatitis, and leukoderma. | UVB phototherapy for<br>psoriasis, vitiligo, atopic<br>dermatitis, and leukoderma. |
| Mode of<br>Operation | Monochromatic UVB<br>Phototherapy | Monochromatic UVB<br>Phototherapy | Monochromatic UVB<br>Phototherapy |
| Wavelength | 308 nm | 308 nm | 308 nm or 311 nm |
| Pulse<br>Characteristics: | | | |
| Maximum Pulse<br>Duration | 30 ns | 20 to 50 ns | 15 to 20 ns |
| Output Energy<br>per Pulse | 15 mJ | Up to 15 mJ Max | Up to 4.2 mJ |
| Fluence per<br>pulse (mJ/cm²) | 2 to 3.8 mJ/cm² | 2 to 5 mJ/cm², 3 mJ/cm² | 3.5 - 4.4 mJ/cm² |
| Physical<br>Characteristics: | | | |
| Activation | Via foot-switch | Via foot-switch | Via foot-switch |
| Beam Profile | Flat-Top, Non-Gaussian | Flat-Top, Non-Gaussian | Flat-Top, Non-Gaussian |
| Electrical<br>Requirements | AC 230 V, 50/60 Hz | AC 230 V, 50/60 Hz | AC 230 V, 50/60 Hz |
| Maximum Power | 20W | 20 W | 20W |
#### 510(k) Summary
Reference Devices: TheraLight, Inc.'s UV120-2 UVA/UVB Phototherapy System . (K024020) *
> *This reference device is referred to technical aspects, of which Narrow Band (NB) UVB lamps are used and emit UVB pulsed energy at the nominal center wavelength of 311 nm the same to those of the PALLAS 308/311 Solid-State UV Laser System.
- Laseroptek Co. Ltd.'s Helios III Q-Switched Nd: Y AG Laser o System (K152856) **
**This reference device is referred, of which hand piece's material are the same to those of the PALLAS 308/311 Solid-State UV Laser System.
### 4. Device Description
Laseroptek Co. Ltd.'s PALLAS 308/311 Solid-State UV laser system is a self-contained ultraviolet laser light source and optical energy delivery system that provides targeted energy to the treatment site while avoiding exposure to non-affected tissue. The light source is contained within the protective console. The complete system also includes a hand piece connected to the console via an articulating arm. Timing and dosing parameters and the user interface are controlled through a
{5}------------------------------------------------
#### 510(k) Summary
display on the console. The delivery system allows UV-B light to pass through the hand piece to selectively treat skin legions without exposure to the unwanted skin.
### 5. Indications / Intended Use
UVB phototherapy for psoriasis, vitiligo, atopic dermatitis, and leukoderma of affected skin.
### 6. Performance Data
Non-clinical tests: Measurement of wavelength, average output power, and energy fluence (in units of mW/cm2) of treatment were performed. Testing conducted on the PALLAS 308/311 Solid-State UV Laser System shows that it refers to the relevant mandatory performance standards for laser products 21 CFR 1040.10 and 1040.11. Other performance, such as electromagnetic compliance, etc, were tested using following consensus standards:
- PALLAS 308/311 Solid-State UV Laser System is tested and evaluated according to IEC -60601-1:2005 + CORR.1 (2006) + CORR. 2 (2007) + AM1 (2012) or IEC 60601-1:2012. All the results presented here demonstrated general requirements for basic safety and essential performance.
- -Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2: 2007. All the results presented here demonstrated the requirements and tests for electromagnetic disturbances.
- PALLAS 308/311 Solid-State UV Laser System is tested and evaluated according to FDA-recognized consensus standard IEC 60601-2-22: 2007 (Third Edition) + A1:2012. All the results presented here demonstrated the particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- -Safety of laser products is evaluated according to FDA-recognized consensus standard IEC 60825-1: 2007. All the results presented here demonstrated the equipment classification and requirements.
- -Risk management was recorded according to the FDA-recognized consensus standard ISO 14971: 2012. All the results presented here demonstrated the application of risk management to medical devices.
- Usability was documented according to the FDA-recognized consensus standard ISO -14971: 2012. All the results presented here demonstrated the application of risk management to medical devices.
The portion of the device that touches patient body is made of aluminum alloy 6061, which have been used for other medical devices without any biocompatibility risk.
### 7. Substantial Equivalence
The intended use of the PALLAS 308/311 UV Laser System is within the scope of the predicate devices. PALLAS 308/311 UV Laser System, from both a design and clinical perspective, uses similar or identical technology as the cited predicate devices and has the same intended uses. Based upon the predicted overall performance characteristics for the PALLAS 308/311 UV Laser System, Laseroptek Co. Ltd. believes that no significant differences exist in usage of its underlying
{6}------------------------------------------------
technological principles between PALLAS 308/311 UV Laser System and the cited predicate devices.
### 8. Conclusions
On the basis of the information provided in this Summary, Laseroptek Co., Ltd. Believes that PALLAS 308/311 UV Laser System is substantially equivalent to legally commercialized predicate devices for the purposes of this 510 (k) submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.