K172576 · BioHorizons Implant Systems, Inc. · DZE · Nov 22, 2017 · Dental
Device Facts
Record ID
K172576
Device Name
BioHorizons Tapered Short Implants
Applicant
BioHorizons Implant Systems, Inc.
Product Code
DZE · Dental
Decision Date
Nov 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172576 · Nov 22, 2017
BioHorizons Tapered Short Implants
BioHorizons Implant Systems, Inc.
Published clinical literature (Nevins et al., 2008) containing human histologic data
Human histologic evidence from published literature is used to support the functional claims of the Laser-Lok surface technology, specifically regarding connective tissue attachment, inhibition of epithelial downgrowth, and crestal bone retention.
Human histology; Connective tissue attachment; Laser-Lok surface; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Nevins et al. Human Histologic Evidence of a Connective Tissue Attachment to a Dental Implant; Human histologic study
Human subjects
Not applicable for this study
Connective tissue attachment, epithelial cell downgrowth inhibition, crestal bone retention
Indications for Use
BioHorizons Tapered Short Implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The implants may be restored using delayed loading, or with a terminal or intermediate abutment for fixed or removable bridgework, and for overdentures.
Device Story
BioHorizons Tapered Short Implants are machined titanium alloy (ASTM F136) screw-form endosseous dental implants; available in 4.6mm and 5.8mm diameters and 6.0mm and 7.5mm lengths. Devices feature Resorbable Blast Texture (RBT) on threaded surfaces and Laser-Lok microchannels on the collar to inhibit epithelial downgrowth and promote connective tissue attachment. Used by dental clinicians in clinical settings for tooth replacement or prosthetic retention. Implants are surgically placed in the mandible or maxilla; restored via delayed loading protocols. Benefits include artificial root structure for dental restoration. Device is non-active, non-mechanical, and provided sterile via gamma radiation.
Clinical Evidence
Bench testing only. Mechanical fatigue testing performed per ISO 14801 and FDA guidance. Axial pullout testing performed per ASTM F543. Surface area and bone-to-implant contact (BIC) comparative analysis performed. Sterility validated per ANSI/AAMI/ISO 11137-1. Endotoxin testing performed per USP <85>. Shelf life validated per ASTM standards for packaging. Biocompatibility supported by historical data for ASTM F136 titanium alloy.
Indicated for patients requiring single tooth replacement, fixed bridgework, or dental retention in the mandible or maxilla. Suitable for use with delayed loading protocols or with terminal/intermediate abutments for fixed/removable bridgework and overdentures.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
BioHorizons Tapered Internal Plus Implants (K121787)
Bicon, LLC The Bicon 5.0 x 5.0mm and 6.0 x 5.0mm Dental Implant (K073368)
Bicon, LLC 4.5 x 6.0mm and 6.0 x 6.0mm Dental Implant (K050712)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
BioHorizons Implant Systems, Inc. Michael Davis Director, Regulatory Affairs 2300 Riverchase Center Birmingham, Alabama 35244
Re: K172576 Trade/Device Name: BioHorizons Tapered Short Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: August 25, 2017 Received: August 28, 2017
Dear Michael Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
November 22, 2017
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172576
#### Device Name BioHorizons Tapered Short Implants
#### Indications for Use (Describe)
BioHorizons Tapered Short Implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The implants may be restored using delayed loading, or with a terminal or intermediate abutment for fixed or removable bridgework, and for overdentures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary 21 CFR 807.92
#### Submitter's Name & Address
| Manufacturer: | BioHorizons Implant Systems, Inc.<br>2300 Riverchase Center<br>Birmingham, AL 35244<br>Phone (205) 967-7880<br>Fax (205) 870-0304 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Official contact: | Michael Davis, Director, Regulatory Affairs |
| Date prepared: | November 21, 2017 |
### Name of the Device
| Trade Name: | BioHorizons Tapered Short Implants |
|------------------------|------------------------------------|
| Common or Usual Name: | Screw-type dental implant |
| Classification Name: | Endosseous dental implant |
| Classification Number: | Class II (21 CFR 872.3640) |
| Primary Product Code : | DZE |
## Predicate Devices
Primary Predicate Device:
1. K121787, BioHorizons Tapered Internal Plus Implants, September 5, 2012.
Reference Predicate Devices:
- 1. K150571, Biomet 3i T3 Short Implants, November 20, 2015.
- 2. K073368, Bicon, LLC The Bicon 5.0 x 5.0mm and 6.0 x 5.0mm Dental Implant, October 10. 2008.
- 3. K050712, Bicon, LLC 4.5 x 6.0mm and 6.0 x 6.0mm Dental Implant, May 16, 2005.
### Device Description
The BioHorizons Tapered Short Implants are machined titanium, screw-form endosseous dental implants supplied in 4.6mm and 5.8mm diameters. The 4.6mm diameter implant includes a 3.5mm prosthetic platform, while the 5.8mm diameter implant includes a 4.5mm prosthetic platform. The implants are provided in 6.0mm and 7.5mm lengths across both diameters. Implant material is titanium alloy as specified in ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Allov (UNS R56401) for Surgical Implant Applications.
The devices are further processed by roughening the threaded surface with Resorbable Blast Texture (RBT) media (tricalcium phosphate) and by micro-machining grooves, known as Laser-Lok® microchannels, on the implant collar. The product is packaged using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10°, validated in compliance with ANSI/AAMI/ISO 11137-1 Sterilization of healthcare products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
### Indications for Use
BioHorizons Tapered Short Implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The implants may be restored using delayed loading, or with a terminal or intermediate abutment for fixed or removable bridgework, and for overdentures.
{4}------------------------------------------------
The main difference between the Indications for Use statements between the subject device and the primary predicate device is with regard to immediate versus delayed loading protocols. The subject device is intended for delayed loading, whereas the primary predicate device was cleared for immediate restoration 1) with a temporary prosthesis that is not in functional occlusion, or 2) when splinted together for multiple tooth replacement, or when stabilized with an overdenture supported by multiple implants. Both devices are indicated for fixed or removeable bridgework and for overdentures. No other substantive differences exist between the Indications for Use statements between the subject and primary predicate devices.
# Technological Characteristics
The fundamental scientific technology of the BioHorizons Tapered Short endosseous dental implant devices subject to this 510(k) is substantially equivalent to the referenced primary predicate device. The threaded portion of the implants is RBT-blasted, and Laser-Lok microchannels are applied to the implant collar.
Laser-Lok is a surface feature in which patterns of micro-machined grooves are applied to the collar of a dental implant, providing a roughened surface to establish a physical, connective tissue attachment. This tissue connection:
- 1) is functionally oriented,
- 2) inhibits epithelial cell downgrowth, and
- 3) enables crestal bone adjacent to the implant to attach and be retained.
The preceding claims 1-3 are supported in the published literature by Nevins et al. Human Histologic Evidence of a Connective Tissue Attachment to a Dental Implant, International Journal of Periodontics & Restorative Dentistry 2008; 28:111-121.
All materials, suppliers, processing, packaging and sterilization methods remain the same as for the primary predicate device, BioHorizons Tapered Internal Plus Implants (K121787). The Laser-Lok feature is substantially equivalent to that cleared for the BioHorizons Tapered Internal Plus Implants. The BioHorizons Tapered Short Implants are substantially equivalent to the features of the predicate implant devices because of the similarities in design, materials and intended use. Refer to Table 1, Summary Table of Substantial Equivalence, immediately following.
### Summary of Testing
Mechanical fatigue testing of the worst-case (smallest prosthetic platform/smallest implant body diameter combined with the shortest implant length) 4.6mm x 6.0mm Tapered Short Implant was performed in accordance with the Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. May 12, 2004 and ISO 14801. The load-bearing features of the implant-abutment connection were tested in conjunction with angled prosthetic abutments representative of the worst-case scenario. The results of the fatigue load testing demonstrate that the subject devices are substantially equivalent to the predicate devices. Additionally, surface area and bone-to-implant contact (BIC) comparative analysis was performed in accordance with BioHorizons internally developed test protocol, followed by axial pullout testing in accordance with ASTM F543. Analysis compared the shortest length worst-case subject implants across both offered implant diameters to the shortest length worst-case primary predicate implants across equivalent implant diameters. Both evaluations demonstrated that the subject device provides increased external surface area, increased bone-to-implant contact and increased axial pullout strength as compared to the primary predicate device.
Per FDA Guidance on "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," BioHorizons conducted baseline
{5}------------------------------------------------
validation testing of representative worst-case BioHorizons endosseous dental implant devices using the Limulus Amebocyte Lysate (LAL) Method - Inhibition/Enhancement Assay (per USP Chapter <85>) for bacterial endotoxins. Testing demonstrated that EU/Device was less than 0.050 for all lots tested, which is below the established acceptance criteria of 20 EU/Device for medical devices not intended to contact cerebrospinal fluid.
Sterilization validation was performed in accordance with ANSI/AAMI/ISO 11137-1. Validation testing was conducted on representative worst-case BioHorizons dental implant devices and demonstrates (1) that a minimum gamma radiation sterilization dose of 25kGy is substantiated for BioHorizons dental implant products, and (2) that BioHorizons dental implant products sterilized with a minimum 25 kGy gamma radiation dose may be released based on data from dosimeters without the need for sterility testing of each log.
Shelf life testing was performed in accordance with applicable ASTM standards for evaluating seal strength of flexible barrier materials and porous medical packaging common to all BioHorizons sterile-packaged dental implants. Test results established the shelf life to be five years provided the sterile seal is not breached. The subject device is a titanium alloy, nonmechanical, non-active device, therefore, degradation in performance characteristics is not likely over the established shelf life period. Special storage conditions are not required for the device, as ambient storage conditions are not expected to adversely affect the safety or efficacy of the device.
Additional data relied on from BioHorizons previous dental implant device submissions to demonstrate substantial equivalence to the predicate devices includes evaluation of the RBT surface treatment process as is applied to all BioHorizons dental implant devices. Evaluation included Scanning Electron Microscopy (SEM) at 100X, 1000X and 5000X demonstrating effective removal of all residual HA particulates, as well as Energy Dispersive X-Ray (EDX) spectroscopy demonstrating that there is no effect on the biocompatibility of the underlying titanium alloy material from RBT processing.
Finally, BioHorizons Tapered Short Implants meet the chemical requirements of ASTM F136. This grade of Titanium is commonly used in surgical implant applications thus no special biocompatibility testing was conducted for the proposed devices. Historical biocompatibility testing conducted on representative BioHorizons dental implant devices that used the same ASTM F136 titanium alloy material that is used exclusively in the manufacture of all BioHorizons dental implant devices included cytotoxicity, irritation and sensitization testing. Test results concluded that the test articles were non-cytotoxic, non-irritating and negative for evidence of dermal sensitization under the test conditions employed.
# Conclusion
The data presented in this submission demonstrates that the proposed devices are substantially equivalent with respect to performance and intended use. The proposed devices perform as well as the legally marketed predicate devices. Furthermore, the proposed devices do not pose any new or increased risks as compared to the legally marketed predicate devices.
{6}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Predicate Devices | | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BioHorizons Implant Systems, Inc.<br>Tapered Short Implants<br>K172576 | BioHorizons Implant Systems, Inc.<br>Tapered Internal Plus Implants<br>K121787 | Biomet 3i<br>3i T3 Short Implants<br>K150571 | Bicon LLC<br>The Bicon 5.0 x 5.0mm and 6.0 x<br>5.0mm Dental Implant<br>K073368 | Bicon LLC<br>4.5 x 6.0mm and 6.0 x 6.0mm<br>Dental Implant<br>K050712 |
| Intended Use | BioHorizons Tapered Short Implants are<br>intended for use in the mandible or<br>maxilla as an artificial root structure for<br>single tooth replacement or for fixed<br>bridgework and dental retention. The<br>implants may be restored using delayed<br>loading, or with a terminal or<br>intermediate abutment for fixed or<br>removable bridgework, and for<br>overdentures. | BioHorizons Tapered Internal Plus<br>Implants are intended for use in the<br>mandible or maxilla for use as an<br>artificial root structure for single tooth<br>replacement or for fixed bridgework and<br>dental retention.<br>BioHorizons Tapered Internal Plus<br>Implants may be restored immediately<br>1) with a temporary prosthesis that is<br>not in functional occlusion, or<br>2) when splinted together for multiple<br>tooth replacement, or when<br>stabilized with an overdenture<br>supported by multiple implants. | The 3i T3® Short Implants are<br>intended for surgical placement in the<br>upper or lower jaw to provide a means<br>for prosthetic attachment in single<br>tooth restorations and in partially or<br>fully edentulous spans with multiple<br>single teeth utilizing delayed loading,<br>or with a terminal or intermediary<br>abutment for fixed or removable<br>bridgework, and to retain<br>overdentures. | The 5.0 x 5.0mm and the 6.0 x 5.0mm<br>implants are designed as a one stage<br>or two stage surgical procedure<br>implant for use in edentulous sites in<br>the mandible or maxilla for support of<br>a complete denture prosthesis, a<br>terminal or intermediate abutment for<br>fixed bridgework, partial dentures, or a<br>single tooth replacement. | The 4.5 x 6.0mm and the 6.0 x<br>6.0mm implants are designed for use<br>in edentulous sites in the mandible or<br>maxilla for support of a complete<br>denture prosthesis, a terminal or<br>intermediate abutment for fixed<br>bridgework, partial dentures, or a<br>single tooth replacement. |
| Design | | | | | |
| Implant shape | Tapered | Tapered | Parallel wall | Parallel wall | Parallel wall |
| Implant body diameter | 4.6mm, 5.8mm | 3.8mm, 4.6mm, 5.8mm | 5.0mm, 6.0mm | 5.0mm, 6.0mm | 4.5mm, 6.0mm |
| Implant length | 6.0mm, 7.5mm | 7.5mm (except 3.8mm body), 9mm,<br>10.5mm, 12mm, 15mm | 5.0mm, 6.0mm | 5.0mm | 6.0mm |
| Outer thread | Buttress | Buttress | V-thread | Fin design (tapped into place, not<br>screwed) | Fin design (tapped into place, not<br>screwed) |
| Surface | Implant - RBT<br>Collar - Laser-Lok | Implant - RBT<br>Collar - Laser-Lok | T3 (blasted, acid etched) or<br>T3 DCD (blasted, acid etched and<br>calcium phosphate applied to surface) | Integra-Ti (grit blasted, acid etched) or<br>Integra-CP (hydroxylapatite surface) | Integra-Ti (grit blasted, acid etched)<br>or<br>Integra-CP (hydroxylapatite surface) |
| Connection | Internal Hex | Internal Hex | Flat to flat butt joint | Tapered connection, non-indexing | Tapered connection, non-indexing |
| Internal thread | Spiralock UNF 1-72 | Spiralock UNF 1-72 | Not known | Not known | Not known |
| Prosthetic platform | 3.5mm, 4.5mm | 3.0mm, 3.5mm, 4.5mm | Not known | Not known | Not known |
| Material and Manufacturing | | | | | |
| Implant Material | Ti-6Al-4V (ASTM F136) | Ti-6Al-4V (ASTM F136) | Grade 4 pure titanium | Ti-6Al-4V | Ti-6Al-4V |
| Manufacturing process | Machined by BioHorizons or A-level<br>supplier, surface treated with micro-<br>machined grooves (Laser-Lok) and RBT | Machined by BioHorizons or A-level<br>supplier, surface treated with micro-<br>machined grooves (Laser-Lok) and<br>RBT | Not known - proprietary | Not known - proprietary | Not known - proprietary |
| Packaging | Tyvek-lidded blister tray (primary<br>package), placed inside a tamper-evident<br>outer box (secondary package) | Tyvek-lidded blister tray (primary<br>package), placed inside a tamper-<br>evident outer box (secondary package) | Not known | Not known | Not known |
| Sterilization | 25-40 kGy gamma irradiation dose range | 25-40 kGy gamma irradiation dose<br>range | Not known | Not known | Not known |
#### Table 1. Summary Table of Substantial Equivalence
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.