The 3i T3® Short Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.
Device Story
3i T3® Short Implants are endosseous dental implants designed for surgical placement in the jaw to support prosthetic restorations. The device consists of a screw-type, parallel-walled titanium body with an external hex connection for mating with restorative components. It features a roughened apex and OSSEOTITE® coronal surface, with optional nano-scale discrete crystalline deposition (DCD®) calcium phosphate treatment. Used by dental surgeons in clinical settings, the implant provides a foundation for single teeth, fixed or removable bridgework, or overdentures. The device is intended for delayed loading protocols. Clinical benefit is derived from providing a stable anchor for dental prosthetics in patients with varying anatomy requiring shorter implant lengths. MR compatibility is established for 1.5T and 3.0T environments.
Clinical Evidence
Bench testing only. Fatigue testing performed per ISO 14807:2007. Comparative non-clinical testing against Bicon predicates included implant surface area analysis, bone-to-implant contact analysis, and pullout strength analysis. MR compatibility testing conducted per ASTM F2052-06e, F2213-06, F2119-07, and F2182-11a for 1.5T and 3.0T environments. No clinical data presented.
Technological Characteristics
Commercially pure titanium; screw-type design; parallel-walled body; external hex connection; OSSEOTITE® surface; optional DCD® CaP surface treatment. Dimensions: 5.0mm/6.0mm diameters and 5.0mm/6.0mm lengths. MR Conditional (1.5T/3.0T). Standards: ISO 14807:2007 (fatigue), ASTM F2052-06e, F2213-06, F2119-07, F2182-11a (MR safety).
Indications for Use
Indicated for surgical placement in upper or lower jaw for prosthetic attachment in single tooth restorations, partially or fully edentulous spans, or to retain overdentures in patients requiring dental implants. Utilizes delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure in profile, with three faces overlapping to symbolize the department's focus on people and health. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2015
Biomet 3i Mr. Chris McKee Regulatory Affairs Manager 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K150571
Trade/Device Name: 3i T3® Short Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: October 15, 2015 Received: October 16, 2015
Dear Mr. McKee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature and the name Tina Kiang. The signature is a series of loops and lines, and the name is written in a simple, sans-serif font. Below the name and signature is the letter S with a dash in front of it.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
к150571 510(k) Number (if known):_
Device Name: 3i T3® Short Implants
## Indications for Use:
The 3i T3® Short Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 3i T3® Short Dental Implants 510(k) Summary November 20, 2015
| I. | Company: | BIOMET 3i™<br>4555 Riverside Drive<br>Palm Beach Gardens, Florida 33410<br>(561) 776-6923 | |
|----|-------------------------------|--------------------------------------------------------------------------------------------|--|
| | 1 - - 1 - - 1 - - 1 - - 1 - - | 011- 1 / . 17 | |
Contact: Chris McKee Regulatory Affairs/Regulatory Compliance Leader
- Proprietary Trade Name: 3i T3® Short Implants II.
- III. Classification Name: Implant, Endosseous, Root-Form (21 CFR 872.3640)
- IV. Classification: Class II
- V. Product Code(s): DZE
#### VI. Reason for 510(k)
The purpose of this 510(k) was to expand BIOMET 37's dental implant product offering by adding dental implants with a multi-level surface topography like the existing 3i T3® Ex Hex Dental Implants but with shorter lengths called the 3i T3® Short Implants. In addition, MR compatibility information is being added to the device labeling for Biomet 3i dental implants (inclusive of the subject devices) and restorative devices.
#### VII. Product Description
The 3i T3® Short Implants are manufactured from commercially pure titanium and feature a roughened apex and traditional OSSEOTITE® coronal surface. In addition, the implants are offered with or without a nano-scale discrete crystalline deposition (DCD®) calcium phosphate (CaP) surface treatment. The dental implants are basic screw-type designs available in parallel walled body geometries with an external hex connection for mating with associated Biomet 3t" external connection restorative components. The implants are offered in 5.0mm and 6.0 mm diameters and 5.0mm and 6.0mm lengths to accommodate varying patient anatomy. Size appropriate cover screws are provided with each implant.
#### VIII. Indications
The 3i T3" Short Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.
#### IX. Summary of the Technological Characteristics
The subject devices provide additional implant options for connection to existing Biomet 3i™ external hex restorative components. The primary change from the predicate 3i T3" devices in K133049 is the shorter length. The external connection is the same as the other predicate external connection dental implants. The materials, implant body designs, and surface treatments are identical as the listed primary predicate device. The size offerings are also the same as the secondary predicate device.
{4}------------------------------------------------
Unlike the primary predicate device, the subject devices are not indicated for immediate loading. The removal of this indication for the subject devices does not alter their intended use described within Section VIII above, as it is only related to the timing in which the implant can be placed into occlusal loading after implantation. The remaining indications for delayed loading are identical to the primary predicate device. Furthermore, the intended use and remaining indications are equivalent to the secondary predicate device which is similar in size. Based on these equivalencies along with those demonstrated through non-clinical testing, the removal of immediate loading from the indications of the subject devices does not raise any new questions of safety or effectiveness.
### X. Identification of Legally Marketed Devices The design features, materials, and indications for use of the subject devices are substantially equivalent to Biomet 3i predicate device noted below:
- 3i T3® External Hex Dental Implants (K133049 SE 1/8/2014) ●
The size of the subject devices are substantially equivalent or similar to the secondary predicate device noted below:
- Bicon Short Implant (5.0mm and 6.0mm, K073368 SE 10/10/2008, K062044 SE 11/21/2006 and K042637 SE 11/15/2004)
#### XI. Discussion of the Non-Clinical Testing
Non-clinical testing was performed on the worst case subject devices in the form of fatigue testing in accordance with ISO 14807:2007 and FDA guidance. The results were compared to the previously listed predicate devices in K133049. The subject devices met the pre-determined acceptance criteria.
Additional non-clinical testing was performed on the worst case subject devices and the worst case predicate Bicon devices in the form of implant surface area analysis, bone to implant contact analysis and pullout strength analysis. The results of the head testing were compared and it was determined that the subject devices were substantially equivalent to the predicate Bicon devices in all three tests.
Because the subject devices are manufactured out of identical materials and by identical processes, the primary predicate device biocompatibility data was leveraged.
MR compatibility testing was conducted on the worst case cleared Biomet 3i device constructs utilizing multiple material configurations. Testing included RF heating, MR artifacts, displacement force, and torque under static magnetic fields of 1.5 Tesla (1.5T) and 3.0 Tesla (3.0T) in accordance with FDA Guidance Document [Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment], ASTM F2052-06e, ASTM F2213-06, ASTM F2119-07, and ASTM F2182-11a. The results of the testing demonstrated that the Biomet 3i dental implants (inclusive of the subject devices) and restorative devices are considered to be MR Conditional in both 1.5T and 3.0T MR environments. This has been reflected in the device labeling.
#### XII. Conclusions
The subject devices have demonstrated substantial equivalence to the previously listed predicate devices in that they utilize identical materials and fundamental designs, have the same intended use, and performed equivalently in the aforementioned non-clinical testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.