K172556 · Actuated Medical, Inc. · KNT · Jun 29, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K172556
Device Name
TubeClear System
Applicant
Actuated Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Jun 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TubeClear System is intended to clear occlusions/clogs in Feeding and Decompression Tubes.
Device Story
TubeClear System consists of reusable Control Box and single-use Clearing Stem; operator manually inserts Clearing Stem into occluded feeding/decompression tube; Control Box actuates internal Clearing Stem Wire; wire tip mechanically clears occlusion to restore tube patency; used by certified/licensed healthcare practitioners in clinical settings; restores tube function; benefits patient by avoiding tube replacement.
Clinical Evidence
Bench testing only. Water ingress testing performed on modified O-ring to support substantial equivalence.
Technological Characteristics
System comprises reusable Control Box and single-use Clearing Stem. Actuation via Control Box Model 101. Modified O-ring material. Sixteen Clearing Stem models for various tube sizes/lengths. ISO 13485:2003 certified; ISO 14971:2007 compliant.
Indications for Use
Indicated for clearing occlusions/clogs in feeding and decompression tubes in adult patients. Compatible with specific nasoenteral, nasogastric, gastrostomy, and jejunostomy tubes ranging from 6-18 Fr and 20-140 cm in length, composed of PVC or polyurethane.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" in a larger font size, with "ADMINISTRATION" written below in a smaller font size.
June 29, 2018
Actuated Medical, Inc. Michael T. Britton, M.S. Regulatory Affairs & Compliance Leader 310 Rolling Ridge Drive Bellefonte, PA 16823
Re: K172556
Trade/Device Name: TubeClear System Regulation Number: 21 CFR& 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: Mav 31, 2018 Received: June 1, 2018
Dear Michael T. Britton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the FDA logo on the left side. On the right side, the text "Jeffrey W. Co" is present, followed by the date "2018.06.29 18" and the time zone "-04'00'". The FDA logo is in light gray, and the text is in black.
Jeffrev W. Cooper -S 18.06.29 18:35:13
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172556
Device Name TubeClear System
### Indications for Use (Describe)
TubeClear System Clearing Stem Models are indicated for use ONLY and SOLELY in clearing occlusions/clogs in Feeding and Decompression Tubes in adult patients that have the following Tube type and size (French and length):
· TC-0608: for Nasoenteral and Nasogastric Tubes composed of Polyvinyl Chloride (PVC) and Polyurethane that are size 6 - 8 Fr and have a length of 38 - 140 cm (15 - 55 in);
· TC-0812: for Nasoenteral and Nasogastric Tubes composed of Polyviny1 Chloride (PVC) and Polyurethane that are size 8 - 12 Fr and have a length of 38 - 140 cm (15 - 55 in);
· TC-1018: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 91 - 140 cm (36 - 55 in); and Gastrostomy and Jejunostomy Tubes that are size 10 - 18 Fr and have a length of 20 - 36 cm (8 - 14 in);
- · NE-1036: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in);
- NE-1042: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in);
- NE-1043: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in);
- NE-1045: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in);
- · NE-1048: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in);
- NE-1050: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in);
- NE-1055: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in);
- · G-1008: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in);
- · G-1009: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in);
- · G-1010: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 25 cm (10 in);
- · G-1011: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in);
- · G-1012: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in); and
- · G-1014: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K172556 Page 1 of 3
Image /page/3/Picture/1 description: The image shows the logo for Actuated Medical. The logo has the word "ACTUATED" in gray and "MEDICAL" in orange. Below the logo is the text "Innovative motion + Positive outcomes" in gray.
310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355.0003 + f (814) 355.1532 ActuatedMedical.com
## Operating As ...
- + ISO 13485:2003 Certified
- + ISO 14971:2007 Compliant
- + Women-Owned Certified
# SECTION 5: 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
510(k) Number: K172556
#### 5.1 Applicant Information
| Date Prepared: | June 22, 2018 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address: | Actuated Medical, Inc.<br>310 Rolling Ridge Drive<br>Bellefonte, PA 16823<br>Ph: (814) 355-0003<br>Fx: (814) 355-1523 |
| Contact Person: | Michael T. Britton<br>Regulatory Affairs and Compliance Leader<br>Ph: (814) 355-0003 x118<br>Fx: (814) 355-1523<br>Email: michael.britton@actuatedmedical.com |
#### 5.2 Device Information
| Classification: | 876.5980, product code KNT | |
|----------------------|-----------------------------------------|--|
| Trade Name | TubeClear System | |
| Common Name: | In Patient Tube Clearing System | |
| Classification Name: | Gastrointestinal tubes and accessories. | |
#### 5.3 Predicate Devices
The legally marketed devices to which substantial equivalence is being claimed are as follows:
| 510(k) Number | Trade Name | Manufacturer |
|---------------|------------------|------------------------|
| K163092 | TubeClear System | Actuated Medical, Inc. |
| K131052 | TubeClear System | Actuated Medical, Inc. |
Image /page/3/Picture/17 description: The image is an abstract design featuring a combination of geometric and organic elements. The background consists of a checkered pattern in varying shades of orange, creating a textured effect. Overlaying this pattern is a series of wavy lines, also in orange, that meander across the image, adding a sense of movement and fluidity. The overall composition is visually engaging, with the interplay of shapes and colors creating a dynamic and modern aesthetic.
{4}------------------------------------------------
#### 5.4 Device Description
The TubeClear System is comprised of a reusable Control Box and a single use Clearing Stem. The Clearing Stem is connected to the Control Box. The Operator then manually inserts the Clearing Stem into the Feeding and Decompression Tube (i.e., Tube) and directs the progression of the Clearing Stem inside the Tube. The Control Box actuates the Clearing Stem Wire. The Clearing Stem Wire Tip mechanically clears the occlusion to restore Tube patency. Control Box Model 101 is used to actuate all Clearing Stem models. Sixteen (16) Clearing Stem Models are proposed to accommodate the different types, sizes, and materials of Tubes.
#### 5.5 Intended Use
The TubeClear System is intended to clear occlusions/clogs in Feeding and Decompression Tubes.
#### 5.6 Indications for Use of the Proposed TubeClear System
TubeClear System Clearing Stem Models are indicated for use ONLY and SOLELY in clearing occlusions/clogs in Feeding and Decompression Tubes in adult patients that have the following Tube type and size (French and length):
· TC-0608: for Nasoenteral and Nasogastric Tubes composed of Polyvinyl Chloride (PVC) and Polyurethane that are size 6 - 8 Fr and have a length of 38 -140 cm (15 - 55 in);
· TC-0812: for Nasoenteral and Nasogastric Tubes composed of Polyvinyl Chloride (PVC) and Polyurethane that are size 8 - 12 Fr and have a length of 38 - 140 cm (15 - 55 in):
· TC-1018: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 91 - 140 cm (36 - 55 in); and Gastrostomy and Jejunostomy Tubes that are size 10 - 18 Fr and have a length of 20 - 36 cm (8 - 14 in);
• NE-1036: for Nasoenteral and Nasogastric Tubes that are size 10 -18 Fr and have a length of 91 cm (36 in);
• NE-1042: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 107 cm (42 in);
• NE-1043: for Nasoenteral and Nasogastric Tubes that are size 10 -18 Fr and have a length of 109 cm (43 in);
• NE-1045: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 114 cm (45 in);
• NE-1048: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 122 cm (48 in);
{5}------------------------------------------------
· NE-1050: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 127 cm (50 in);
· NE-1055: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 140 cm (55 in);
• G-1008: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 20 cm (8 in);
• G-1009: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 23 cm (9 in);
• G-1010: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 25 cm (10 in):
• G-1011: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 28 cm (11 in);
• G-1012: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 30 cm (12 in); and
• G-1014: for Gastrostomy and Jejunostomy Tubes that are size 10 -18 Fr and have a length of 36 cm (14 in).
#### 5.7 Technological Characteristics
The TubeClear System referenced in this 510(k)-submission is technologically identical to the predicate devices. Changes include modified material of the Control Box O-Ring, combining the Clearing Stem models' labeling from each predicate device into a single Instructions for Use (IFU), and expanding the user population for the NE/G models from Licensed Healthcare Practitioners to Certified and Licensed Healthcare Practitioners to be consistent with the TC models.
#### 5.8 Performance Data
Performance testing on the material changes demonstrate substantial equivalence to the predicate devices. Performance bench testing used to support substantial equivalence was water ingress testing of the o-ring.
#### 5.9 Conclusions
The scientific data demonstrates that the Proposed Clearing Stem Models are substantially equivalent to the Predicate Clearing Stem Models.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.