TUBECLEAR MODEL 101 (WITH NEW CLEARING STEM NE AND G MODELS)
K131052 · Actuated Medical, Inc. · KNT · Aug 16, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K131052
Device Name
TUBECLEAR MODEL 101 (WITH NEW CLEARING STEM NE AND G MODELS)
Applicant
Actuated Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Aug 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr. The Clearing Stem Models are indicated for use as follows: NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in). NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in). NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in). NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in). NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in). NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in). NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in). G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in). G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in). G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 25 cm (10 in). G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in). G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in). G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
Device Story
TubeClear system consists of reusable Control Box (Model 101) and single-use Clearing Stems. Licensed healthcare practitioners manually insert Clearing Stem into patient's existing feeding/decompression tube. Control Box motor provides electromechanical linear reciprocating motion to internal wire within Clearing Stem. Reciprocating motion transfers to distal wire tip, which mechanically clears occlusions within tube. Used in hospitals, long-term care, or home settings. Restores tube patency, avoiding need for tube replacement. No direct patient contact by device; sterile conditions not required.
Clinical Evidence
No clinical data presented. Bench testing only. Efficacy evaluated via simulated use testing comparing proposed models to predicates. Safety testing evaluated tube damage, heating, and movement. Electrical safety tested to IEC 60601-1 3rd Edition.
Technological Characteristics
System: Reusable Control Box (Model 101) and single-use Clearing Stems. Clearing Stem: PTFE sheath, 304V stainless steel wire (Lunderquist style mandrel). Wire features stiff core, flexible tip, and brazed coil. Actuation: Electromechanical linear reciprocation. Dimensions: 10-18 Fr compatibility; lengths 20-140 cm. Sterilization: Not required. Connectivity: None.
Indications for Use
Indicated for clearing occlusions in feeding and decompression tubes (10-18 Fr) in adult patients. Specific models indicated for various tube lengths (20-140 cm) and types (nasoenteral, nasogastric, gastrostomy, jejunostomy).
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Actuated Medical. The logo consists of a stylized medical symbol to the left of the word "ACTUATEDMEDICAL" in all caps. Below the company name is the tagline "Innovative motion · Positive outcomes".
pg 1 of 7 310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355.0003 + f (814) 355 1532 + w ActuatedMedical com
#### Certified
+ ISO 13485 & 14971
- + 9001 Compliant
+ WBF
#### SECTION 5: 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR807.92.
510(k) Number: K131052
## 5.1: APPLICANT INFORMATION
| Date Prepared: | August 9, 2013 | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------|-------------|
| Name and Address: | Actuated Medical, Inc.<br>310 Rolling Ridge Drive<br>Bellefonte, PA 16823<br>Ph: 814-355-0003<br>Fax: 814-355-1532 | AUG 16 2013 |
| Contact Person: | Victoria A. Kellogg, Ph.D., CRNP, M.B.A., Esq.<br>Director Medical Affairs<br>Ph: 814-355-0003 x118<br>Fax: 814-355-1532 | |
Email: Victoria.kellogg@actuatedmedical.com
# 5.2: DEVICE INFORMATION
| Classification: | KNT |
|----------------------|----------------------------------------------------------------|
| Trade Name: | TubeClear |
| Common Name: | In Patient Tube Clearing System |
| Classification Name: | Tubes, Gastrointestinal and Accessories<br>21 C.F.R. §876.5980 |
{1}------------------------------------------------
## 5.3: PREDICATE DEVICE
| 510(K)<br>Number | Trade or Proprietary<br>Name | Manufacturer | Comment |
|------------------|------------------------------|------------------------|-----------------------------------------------------------|
| K121571 | TubeClear | Actuated Medical, Inc. | Original and<br>Traditional<br>510K<br>Cleared<br>6/18/12 |
| K123659 | TubeClear | Actuated Medical, Inc. | Special<br>510K1<br>Cleared<br>12/20/12 |
The legally marketed device to which substantial equivalence is being claimed:
The Predicate Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NG-1036 and NG-1043.
## 5.4: DEVICE DESCRIPTION
The Proposed Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NE-1036, NE-1042, NE-1043, NE-1045, NE-1048, NE-1050, NE-1055, G-1008, G-1009, G-1010, G-1012, and G-1014. The Control Box is reusable and the Clearing Stems are single use. The Operator manually inserts the Clearing Stem into the feeding and/or decompression tube (i.e., Tube) and directs the Clearing Stem's progression along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear The linear reciprocating motion is transferred to the reciprocating motion. proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Mire. The motion at the Wire Tip mechanically acts on the occlusion and restores Tube patency.
<sup>1</sup> Special 510K submitted to address a change to the Clearing Stem. The change was the removal of pad-printed markings from the Clearing Stem Sheath.
{2}------------------------------------------------
Thirteen (13) new Clearing Stem Models are proposed. The additional Clearing Stem Models expand the specific indications for use. 5.5: INTENDED USE
The Proposed Clearing Stem Models are substantially equivalent to the Predicate Clearing Stem Models in regards to intended use and therapeutic effect. The intended use for both the Proposed and Predicate Clearing Stems is to clear occlusions / clogs from feeding and decompression tubes (i.e., Tubes). The therapeutic effect is restoring patency to the Tube and alleviating the need for Tube replacement.
All Models of Clearing Stem are disposable, single use devices. All Models are intended for use by Licensed Healthcare Practitioners. 'All Models are intended for use in hospitals, in long term care facilities, and in homes (serviced by licensed healthcare practitioners). Because all Clearing Stem Models enter a Tube that is within the Patient, no Model makes direct contact with the Patient. All models of Clearing Stems are placed into a Tube that is in the Patient's gastro-intestinal (GI) system. Sterile conditions are not required for the intended purpose.
## 5.6: INDICATIONS FOR USE
TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.
The Clearing Stem Models are indicated for use as follows:
- NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
- NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
- NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
- NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
- NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).
{3}------------------------------------------------
- NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
- NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
- G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
- G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
- G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 -- 18 Fr and have a length of 25 cm (10 in).
- G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).
- G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
- G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
#### 5.6.1: Specific Indications for Use of Predicate Clearing Stem Models
Clearing Stem Model NG-1036 is indicated for use in Nasogastric tubes that are of size 10 - 14 French and have a length of 91 - 108 cm (36 - 42 in).
Clearing Stem Model NG-1043 is indicated for use in Nasogastric tubes that are of size 10 - 14 French and have a length of 109 - 127 cm (43 - 50 in).
The specific indications for use vary from Clearing Stem Model to Clearing Stem Model; however, the differences are not critical to the intended therapeutic use of the device, which is restoring patency to the Tube and alleviating the need for Tube replacement. The differences do not affect the safety and effectiveness of the Proposed Clearing Stem Models.
{4}------------------------------------------------
Address Macked Coro
K131052 pg 5 of 7
## 5.7: TECHNOLOGICAL CHARACTERISTICS
The technology is the same for the Proposed and Predicate Clearing Stems. The operating principle, Operator interaction with the Device, and the mode of actuation of the Clearing Stem Wire is identical (i.e., Control Box Model 101). Both the Proposed and Predicate Clearing Stems are manually inserted and advanced into the Patient's Tube. Both the Proposed and Predicate Clearing Stem Models physically interact with the occlusion material inside a Tube. Both the Proposed and Predicate Clearing Stem Models mechanically clear occlusions. The Wire inside the Sheath is linearly actuated by the Motor inside the Control Box.
The design features of the Proposed Clearing Stem Models and the Predicate Clearing Stem Models are very similar. For both the Proposed and Predicate Clearing Stem Models, the Sheath is the same material (i.e., PTFE extruded tubing) and length. Each Model has a Collar specifically positioned to limit the depth that the Clearing Stem can be inserted into the Tube. The Collar position determines the Working Length of the Clearing Stem and is specific to each Model. The main difference between the Proposed Clearing Stem Models and the Predicate Clearing Stem Models is the Wire inside the Sheath of the Clearing Stem; specifically the Wire Tip. The Proposed Clearing Stem Design incorporates a more flexible Wire Tip than the Predicate Clearing Stem Design. The Proposed Wire Tip Design is safe and effective for the expanded indications for use.
The Wire used in the Proposed Clearing Stems is 203.340 cm (80.055 in) in length; 304V stainless steel; and a Lunderquist Style Mandrel Guide Wire. It has a stiff, fixed Core and a flexible Tip. Beginning at 109.22 cm (43.000 in) from the distal end, a flat Coil Wire is wound over the core wire creating an outer diameter of 0.089 ± 0.008 cm (0.035 ± 0.003 in). Ten (10) cm (3.937 in) from the distal end, the core wire gradually tapers over a 5.000 ± 0.500 cm (1.969 ± 0.197 in) length to 0.015 ± 0.013 cm (0.006 ± 0.005 in). The distal 5.000 ± 0.500 cm (1.969 ± 0.197 in) of the core wire are a constant 0.015 ± 0.013 cm (0.006 ± 0.005 in) in diameter. The Coil Wire is brazed onto the Core wire. The distal end of the wire is rounded and smooth with a radius of .046 cm (0.018 in).
The Wire used in the Predicate Clearing Stems is 203 cm (80 in) in length; flexible twisted 1x7 strand 304V stainless steel. The Wire diameter is 0.069 ± 0.008 cm (0.027 ± 0.003 in). The ends of the Wire are fused and rounded.
{5}------------------------------------------------
Actuato Madizal Com
K131052 pg 6 of 7
## 5.8: NON-CLINICAL PERFORMANCE DATA
As the same Control Box (i.e., Model 101) will be used, testing focused predominantly on the Clearing Stem. Testing for the Control Box Model 101 consisted of upgrading electrical safety testing to IEC 60601-1 30 Edition; Medical Electrical Equipment, Part 1: General requirements for basic safety and essential performance. Testing was performed by Intertek Group (Boxborough, MA) and was successfully completed in December, 2012.
Bench Testing of the Clearing Stem focused in three areas: Technical, Efficacy, Technical testing included verification of product specifications, and Safety. transportation vibration, and shelf life. Efficacy testing included usability and effectiveness. Safety testing evaluated the potential for Tube damage, Tube heating and Tube movement during use of TubeClear.
## 5.8.1: Technical Testing
Verification confirmed that the Clearing Stem Models met product specifications passing all acceptance criteria. Transportation testing successfully passed acceptance criteria for US highway truck. Shelf life was tested via accelerated life testing and passed 24 month storage. Additional testing for longer shelf life is ongoing.
## 5.8.2: Efficacy Testing
Usability testing by licensed healthcare practitioners was successfully completed. Evaluation of effectiveness during simulated use was performed for the Proposed Clearing Stem Models and compared to the effectiveness of the Predicate Clearing Stem Models. They were found to be of equivalent effectiveness.
## 5.8.3: Safety Testing
Safety testing in terms of Tube damage (e.g., scratches, nicks, tears, abrasions, punctures) to the inner tube surface after device operation was found to be equivalent for the Proposed and Predicate Devices. Heating during device operation was negligible and equivalent for the Proposed and Predicate Devices, thus the Devices have equivalent safety in terms of heating. Tube movement was measured and found to be of the same magnitude, thus equivalent for both
{6}------------------------------------------------
Actuata Madical com
K131052 pg 7 of 7
Devices. The potential for Tube damage, Tube heating and Tube movement during use of TubeClear was found to be equivalent for the Proposed and Predicate Clearing Stems.
## 5.9: CLINICAL PERFORMANCE DATA
No clinical data was collected; therefore, no clinical data is presented in this submission.
## 5.10: CONCLUSIONS
The scientific data demonstrates that the Proposed Clearing Stem Models are substantially equivalent to the Predicate Clearing Stem Models. The 510(k) Substantial Equivalence Decision Making Process Flow Chart was used by Actuated Medical, Inc. to determine Substantial Equivalence.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16, 2013
Actuated Medical, Inc. % Jeffrey Rongero Senior Project Engineer UL LLC 12 Laboratory Drive Research Triangle Park, NC 27709-3995
Re: K131052
Trade/Device Name: TubeClear™ Regulation Number: 21 CFR §876.5980 Regulation Name: Tubes, Gastrointestinal and Accessories Regulatory Class: Class II Product Code: KNT, FPD Dated: July 31, 2013 Received: August 1, 2013
Dear Jeffrey Rongero,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{8}------------------------------------------------
Page 2 - Jeffrey Rongero
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## 4.0 Indications for Use
510(k) Number (if known): K131052
Device Name: TubeClear
#### Indications for Use Statement
TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.
The Clearing Stem Models are indicated for use as follows:
- NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
- NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
- NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
- NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
- NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).
- NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
- NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
- G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
- G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
- G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 25 cm (10 in).
- G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).
Page 1 of 2
4-1
{10}------------------------------------------------
- G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
- G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
Prescription Use ____X__ (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Benjamin R. Fisher -S 2013.08.16 15:02:28 -04'00'
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K131052 510(k) Number
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.