The IU Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
Device Story
IU Implant System consists of endosseous dental implants (fixtures) and various abutments; fixtures surgically placed in upper or lower jaw bone to support prosthetic devices; restores patient chewing function. Fixtures available in triangular or square screw thread designs; made of pure titanium (ASTM F67); surface treated via sandblasting and acid etching (SLA). Abutments made of titanium alloy (ASTM F136). System used by dental professionals in clinical settings. Fixtures supplied sterile (gamma radiation); abutments supplied non-sterile (require steam sterilization). Output is a stable foundation for dental prostheses; clinical decision-making based on anatomical situation and prosthetic needs. Benefits include restoration of oral function and support for artificial teeth.
Clinical Evidence
No clinical data were included in this submission. Evidence is based on non-clinical bench testing, including biocompatibility (ISO 10993-1, ISO 7405), sterilization validation (ISO 11137, ISO 17665), shelf-life testing (ASTM F1980), and risk analysis (ISO 14971).
Technological Characteristics
Endosseous root-form dental implants. Materials: Pure titanium (ASTM F67) for fixtures; titanium alloy (ASTM F136) for abutments. Surface: Sandblasted and acid-etched (SLA). Connection: Internal hex. Sterilization: Gamma radiation (fixtures), steam sterilization (abutments). Dimensions: Various diameters and lengths provided. No software or electronic components.
Indications for Use
Indicated for patients requiring dental implants in the upper or lower jaw arches to support prosthetic devices (e.g., artificial teeth) and restore chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below. The logos are placed side-by-side.
WARANTEC Company, Limited % Peter Chung Representative Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K172345
Trade/Device Name: IU Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 21, 2018 Received: May 24, 2018
#### Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
June 18, 2018
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K172345
Device Name IU Implant System
#### Indications for Use (Describe)
The IU Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary [as required by 807.92(c)]
# I. SUBMITTER
- a) Company: WARANTEC Co., Ltd.
- b) Address: #411, #412, 474, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13229, Korea
- Tel : +82-31-778-6328 c)
- Fax : +82-70-4324-6328 d)
- President of Company: Mr. Dong-Hwa Jang e)
- f) Contact Person : Peter Chung, 412-687-3976
- Contact Person Address: 300, Atwood Street, Pittsburgh, PA, 15213, USA g)
- Date Prepared: June 16, 2018 h)
# II. DEVICE
- IU Implant System a) Trade Name :
- Common Name : Endosseous Dental Implant System b)
- c) Classification Name : implant, endosseous, root-form
- Product Code : DZE (primary), NHA (secondary) d)
- Regulation Number : 872.3640 e)
- f) Class of device : Class II
- Panel : Dental g)
## III.PREDICATE DEVICE
- Primary Predicate Device: a) Luna Dental Implant System / SHINHUNG MST Co., Ltd. / K123155
- Reference Device: b) SuperLine / DENTIUM CO., LTD. / K160965 AnyOne 100 Internal Implant System / MegaGenImplant Co., Ltd. / K123988 Dentium Implantium® & SuperLine® Prosthetics / DENTIUM CO., LTD. / K141457 Dentium Implantium & SuperLine Prosthetics / DENTIUM CO., LTD. / K160828
## IV. DEVICE DESCRIPTION
The IU Implant System is an integrated system of endosseous dental implants with corresponding various abutments (cover screw, two piece abutment, solid abutment, abutment screw, ball abutment, retained abutment, temporary abutment, and multi abutment).
The IU Implant System is a dental implant made of titanium metal intended to be
{5}------------------------------------------------
surgically placed in the bone of the upper or lower jaw arches.
There are two types of Fixtures, triangular screw threads design and square screw threads design. The IU Implant System includes the each fixtures is available in various diameters (lengths) 3.6 mm (8.5 mm, 10.0 mm, 11.5 mm, 13.0 mm) & 4.1 mm, 4.6 mm, 5.1 mm, 5.6 mm, 6.1 mm (7.5 mm, 8.5 mm, 10.0 mm, 11.5 mm, 13.0 mm) of triangular screw threads design and 4.3 mm, 4.8 mm, 5.3 mm (7.0 mm, 8.5 mm, 10.0 mm, 11.5 mm, 13.0 mm) & 6.3 mm (7.0 mm, 8.5 mm, 10.0 mm, 11.5 mm) of square screw threads design according to the anatomical situation. Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated.
Abutment is device made of titanium alloy and it is intended for use to make permanent prostheses and/or temporary prosthesis. The abutments are provided non-sterile and should be sterilized before use.
#### V. INDICATIONS FOR USE
The IU Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.
#### VI. SUBSTANIAL EQUIVALNECE COMPARISON
The IU Implant System is similar designs and dimensions, and has the same material, intended use, surface treatment and technological characteristics as the identified primary predicate device (K123155) and reference devices (K160965/K123988/ K141457). When compared with predicate device, no new questions of substantial equivalence have been raised for the IU Implant System.
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| 510(k) Number | K172345 | K123155 | K160965 | K123988 |
| Manufacturer | WARANTEC Co., Ltd. | SHINHUNG MST Co., Ltd. | DENTIUM CO., LTD. | MegaGenImplant Co., Ltd. |
| Common Name | Endosseous Dental Implant System | Endosseous Dental Implant System | Endosseous Dental Implant System | Endosseous Dental Implant System |
| Trade Name | IU Implant System | Luna Dental Implant System | SuperLine | AnyOneTM Internal Implant System |
| Device class | Class II | Class II | Class II | Class II |
| Product Code(s) | DZE | DZE | DZE | DZE |
| Indications for Use/<br>Intended Use | The IU Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to | The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or | SuperLine is indicated for use in surgical and restorative applications for placement in the bone of the upper or | The AnyOneTM Internal Implant System is intended to be surgically placed in the maxillary or |
| | provide support for | lower jaw arches to | lower jaw to provide | mandibular molar |
| | prosthetic devices, | provide support for | support for prosthetic | areas for the purpose |
| | such as artificial teeth, | prosthetic devices, | devices, such as | providing prosthetic |
| | and to restore the | such as artificial teeth, | artificial teeth, in order | support for dental |
| | patient's chewing | and to restore the | to restore the patient's | restorations (Crown, |
| | function. | patient's chewing | chewing function. | bridges, and |
| | | function. | SuperLine is indicated | overdentures) in |
| | | | also for immediate | partially or fully |
| | | | loading when good | edentulous individuals. |
| | | | primary stability is | It is used to restore a |
| | | | achieved and with | patient's chewing |
| | | | appropriate occlusal | function. Smaller |
| | | | loading. | implants (less than |
| | | | | 06.0 mm) are |
| | | | | dedicated for |
| | | | | immediate loading |
| | | | | when good primary |
| | | | | stability is achieved |
| | | | | and with appropriate |
| | | | | occlusal loading. |
| | | | | Larger implants are |
| | | | | dedicated for the molar |
| | | | | region and are |
| | | | | indicated for delayed |
| | | | | loading. |
| Implant design | Root-form endosseous | Implant, Endosseous, | Root-form endosseous | Root-form endosseous |
| | dental implants | Root-Form | dental implants | dental implants |
| Shape | a)<br>b) | | | |
| Connection | 2.1, 2.5 Internal Hex | 2.1, 2.5 Internal Hex | Internal Hex | Internal Hex |
| Diameter of fixture | a) 3.6, 4.1, 4.6, 5.1, | 3.7, 4.2 – 5.7 mm | 3.6, 4.0, 4.4, 4.9, 6.0, | Internal type |
| | 5.6, 6.1 mm | | 7.0 mm | 3.9, 4.3, 5.3, 6.3, 7.3 |
| | b) 4.3, 4.8, 5.3, 6.3 | | | mm (for normal |
| | mm | | | thread) |
| | | | | 4.8, 5.8, 6.8, 7.8, 8.3 |
| | | | | mm (for deep thread) |
| | | | | 4.8, 5.3, 6.3, 7.3 mm |
| | | | | (for special length) |
| Length of fixture | a) 7.5, 8.5, 10.0, 11.5, | 8.5 — 15.0 mm | 7.0, 8.0, 10.0, 12.0, | Internal type |
| | 13.0 mm | | 14.0 mm | 7.0, 8.0, 9.5, 11.0, |
| | (cf. 3.6 mm diameter | | | 12.5, 14.5 mm (for |
| | does not include a 7.5 | | | normal and deep |
| | mm length.) | | | thread) |
| | b) 7.0, 8.5, 10.0, 11.5, | | | 7.0 mm (for special |
| | 13.0 mm | | | length) |
| | (cf. 6.3 mm diameter | | | |
| | does not include a 13.0 | | | |
| | mm length.) | | | |
| Material | Pure titanium (ASTM<br>F67) | Pure titanium (ASTM<br>F67) | Pure titanium (ASTM<br>F67) | Pure titanium (ASTM<br>F67) |
| Physical testing<br>performed | Not tested/no<br>angulation submitted<br>for clearance. | Test according to ISO<br>14801 | Test according to ISO<br>14801 | Test according to ISO<br>14801 |
| Biocompatibility | Biocompatible<br>according to ISO<br>10993-1 | Biocompatible<br>according to ISO<br>10993-1 | Biocompatible<br>according to ISO<br>10993-1 | Biocompatible<br>according to ISO<br>10993-1 |
| Surface treatment | S.L.A | S.L.A | S.L.A | S.L.A |
| Sterilization method | Gamma radiation | Gamma radiation | Gamma radiation | Gamma sterilization |
| Shelf-life | 7years | - | 8years | - |
Table 1 - Comparison of Endosseous Implant Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
# Table 2 - Comparison of Endosseous Abutments Characteristics
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K172345 | K123155 | K141457 | K123988 |
| Manufacturer | WARANTEC Co., Ltd. | SHINHUNG MST<br>Co., Ltd. | DENTIUM CO., LTD. | MegaGenImplant Co.,<br>Ltd. |
| Common Name | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System |
| Trade Name | IU Implant System | Luna Dental Implant<br>System | Dentium Implantium®<br>& SuperLine®<br>Prosthetics | AnyOne™ Internal<br>Implant System |
| Device class | Class II | Class II | Class II | Class II |
| Product Code(s) | NHA | NHA | NHA | NHA |
| Indications for Use | The IU Implant<br>System is intended to<br>be surgically placed in<br>the bone of the upper<br>or lower jaw arches to<br>provide support for<br>prosthetic devices,<br>such as artificial teeth,<br>and to restore the<br>patient's chewing<br>function. | The Luna Dental<br>Implant System is<br>intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices,<br>such as artificial teeth,<br>and to restore the<br>patient's chewing<br>function. | Dentium Implantium®<br>& SuperLine®<br>Prosthetics are<br>intended for use as an<br>aid in prosthetic<br>rehabilitation. | The AnyOne ™<br>Internal Implant<br>System is intended to<br>be surgically placed in<br>the maxillary or<br>mandibular molar<br>areas for the purpose<br>providing prosthetic<br>support for dental<br>restorations (Crown,<br>bridges, and<br>overdentures) in<br>partially or fully<br>edentulous individuals.<br>It is used to restore a<br>patient's chewing<br>function. Smaller |
{8}------------------------------------------------
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K172345 | K123155 | K141457 | K123988<br>implants (less than<br>06.0 mm) are<br>dedicated for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with appropriate<br>occlusal loading.<br>Larger implants are<br>dedicated for the molar<br>region and are<br>indicated for delayed<br>loading. |
| Cover Screw | | | | |
| Shape | Image: Cover Screw Shape 1 | Image: Cover Screw Shape 2 | Image: Cover Screw Shape 3 | Image: Cover Screw Shape 4 |
| Material | Pure titanium (ASTM<br>F67) | Pure titanium (ASTM<br>F67) | Ti-6Al-4V-ELI<br>(ASTM F136) | CP4 Titanium (ASTM<br>F67) |
| Diameter | 2.82, 3.32 mm | - | 3.6 mm | - |
| Abutment Screw | | | | |
| Shape | Image: Abutment Screw Shape 1 | Image: Abutment Screw Shape 2 | Image: Abutment Screw Shape 3 | Image: Abutment Screw Shape 4 |
| Material | Titanium alloy (ASTM<br>F136) | Titanium alloy (ASTM<br>F136) | Titanium alloy (ASTM<br>F136) | Ti-6Al-4V-ELI |
| Diameter | 2.2 mm | - | 1.8 ~ 2.3 mm | 3.8 ~ 10.0 mm |
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| 510(k) Number | K172345 | K123155 | K160828 | K123988 |
| Manufacturer | WARANTEC Co., Ltd. | SHINHUNG MST<br>Co., Ltd. | DENTIUM CO., LTD. | MegaGenImplant Co.,<br>Ltd. |
| Common Name | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System | Endosseous Dental<br>Implant System |
| Trade Name | IU Implant System | Luna Dental Implant<br>System | Dentium Implantium®<br>& SuperLine®<br>Prosthetics | AnyOneTM Internal<br>Implant System |
| Device class | Class II | Class II | Class II | Class II |
| Product Code(s) | NHA | NHA | NHA | NHA |
| Indications for Use | The IU Implant<br>System is intended to<br>be surgically placed in | The Luna Dental<br>Implant System is<br>intended to be | Dentium Implantium®<br>& SuperLine®<br>Prosthetics are | The AnyOne TM<br>Internal Implant<br>System is intended to |
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| 510(k) Number | K172345 | K123155 | K160828 | K123988 |
| | the bone of the upper<br>or lower jaw arches to<br>provide support for<br>prosthetic devices,<br>such as artificial teeth,<br>and to restore the<br>patient's chewing<br>function. | surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices,<br>such as artificial teeth,<br>and to restore the<br>patient's chewing<br>function. | intended for use as an<br>aid in prosthetic<br>rehabilitation. | be surgically placed in<br>the maxillary or<br>mandibular molar<br>areas for the purpose<br>providing prosthetic<br>support for dental<br>restorations (Crown,<br>bridges, and<br>overdentures) in<br>partially or fully<br>edentulous individuals.<br>It is used to restore a<br>patient's chewing<br>function. Smaller<br>implants (less than<br>06.0 mm) are<br>dedicated for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with appropriate<br>occlusal loading.<br>Larger implants are<br>dedicated for the molar<br>region and are<br>indicated for delayed<br>loading. |
| Temporary Abutment | | | | |
| Shape | Image: (Hex) (Non-Hex) | Image: (Hex) (Non-Hex) | Image: (Hex) (Non-Hex) | Image: Ti-6A1-4V-ELI |
| Material | Titanium alloy (ASTM<br>F136) | Titanium alloy (ASTM<br>F136) | Pure Titanium Grade 5<br>(ASTM F67) | Ti-6A1-4V-ELI |
| Diameter | 3.9 mm | 4.0 ~ 6.0 mm | 4.5 mm | 3.8 ~ 10.0 mm |
| Two Piece Abutment | | | | |
| Shape | Image: (Hex) | Image: (Hex) (Non-Hex) | Image: (Hex) (Non-Hex) | Image: (Hex) (Non-Hex) |
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| 510(k) Number | K172345 | K123155 | K160828 | K123988 |
| | (Hex) (Non-Hex) | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.