CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A/B Typing Kit, CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A Subtyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza B Lineage Genotyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza A/H5 Subtyping Kit
K172091 · Centers for Disease Control and Prevention · OZE · Aug 9, 2017 · Microbiology
Device Facts
Record ID
K172091
Device Name
CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A/B Typing Kit, CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A Subtyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza B Lineage Genotyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza A/H5 Subtyping Kit
Applicant
Centers for Disease Control and Prevention
Product Code
OZE · Microbiology
Decision Date
Aug 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172091 · Aug 9, 2017
CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A/B Typing Kit, CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A Subtyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza B Lineage Genotyping Kit, CDC Human Influenza Virus Real-time RT-PCR, Influenza A/H5 Subtyping Kit
Centers for Disease Control and Prevention
Retrospective clinical specimens
Retrospective clinical specimens were used to evaluate the performance of new nucleic acid extraction instrument platforms (QIAGEN EZ1 Advanced XL and Roche MagNA Pure 96) compared to a previously FDA-cleared extraction method.
Retrospective Clinical Study; Retrospective comparative study; Study Period: 2011-2012 and 2013-2014 influenza seasons
Clinical specimens previously determined to be positive or negative for influenza A(H3) virus; Sample Size: 60 specimens (30 positive, 30 negative) per extraction platform
Roche MagNA Pure Compact instrument using the RNA Isolation Kit
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)
Indications for Use
The Influenza A/B Typing Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information: - For qualitative detection of influenza virus type A or B viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture; - To provide epidemiologic information for surveillance of circulating influenza viruses. Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens. All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees. (Note: Similar statements apply to the Influenza A Subtyping Kit (VER 2), Influenza B Lineage Genotyping Kit, and Influenza A/H5 Subtyping Kit (VER 3) as described in the document).
Device Story
The CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel consists of four kits using rRT-PCR to detect and characterize influenza viral RNA. Input samples include upper and lower respiratory tract specimens. Nucleic acids are extracted using validated automated platforms (QIAGEN EZ1 Advanced XL or Roche MagNA Pure 96). Extracted RNA is amplified on the Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument using specific primers and fluorescently labeled hydrolysis probes targeting conserved regions of viral genes (M, NS, NP, HA). The instrument monitors fluorescence intensity during PCR cycles to detect target amplification. Results are used by trained laboratory personnel in conjunction with clinical and epidemiological data to aid in diagnosis and surveillance. The device provides qualitative detection and subtyping/lineage information, assisting healthcare providers in patient management and public health monitoring.
Clinical Evidence
Retrospective clinical study using 60 specimens (30 positive for influenza A(H3), 30 negative) compared the new extraction methods (Roche MagNA Pure 96, QIAGEN EZ1 Advanced XL) against the cleared Roche MagNA Pure Compact method. Results showed 100% agreement (95% CI: 88.7-100.0) for all platforms tested.
Technological Characteristics
Real-time RT-PCR assay. Instrumentation: Applied Biosystems (ABI) 7500 Fast Dx. Nucleic acid extraction platforms: QIAGEN EZ1 Advanced XL (EZ1 DSP Virus Kit, EZ1 RNA Tissue Mini Kit) and Roche MagNA Pure 96 (DNA and Viral NA Small Volume Kit). Analyte: Viral RNA. Detection: Fluorescently labeled hydrolysis probes. Energy source: Electrical (PCR instrument).
Indications for Use
Indicated for qualitative detection, subtyping, or genotyping of influenza A and B viruses in human upper and lower respiratory tract specimens or viral culture from patients with signs/symptoms of respiratory infection. Not for sole basis of treatment decisions. Requires training by CDC instructors.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K172091
B. Purpose for Submission:
Device modification to add three additional options of nucleic acids extraction instrument and extraction method combinations (QIAGEN EZ1 DSP Virus Kit on the QIAGEN EZ1 Advanced XL instrument, QIAGEN EZ1 RNA Tissue Mini Kit on the QIAGEN EZ1 Advanced XL instrument, and Roche DNA and Viral NA Small Volume Kit on the Roche MagNA Pure 96 instrument) that are acceptable for use with the four previously FDA-cleared diagnostic kits, the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza A/B Typing Kit, Influenza A Subtyping Kit (VER 2), Influenza B Lineage Genotyping Kit, and Influenza A/H5 Subtyping Kit (VER 3).
C. Measurand:
Influenza A, Influenza B, seasonal human influenza A(H3), seasonal human influenza A(H1)pdm09, B/Victoria genetic lineage of human influenza B, B/Yamagata genetic lineage of human influenza B, and influenza A subtype A(H5) (Asian lineage) viral RNA target sequences.
D. Type of Test:
Real-time RT-PCR (rRT-PCR) assays.
E. Applicant:
Centers for Disease Control and Prevention.
F. Proprietary and Established Names:
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A/B Typing Kit, Influenza A Subtyping Kit (VER 2), Influenza B Lineage Genotyping Kit, and Influenza A/H5 Subtyping Kit (VER 3).
G. Regulatory Information:
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A/B Typing Kit
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OZE | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
{1}
2
| NSU | Class II | 21 CFR 862.2570
Instrumentation for Clinical Multiplex Test Systems | Clinical Chemistry (75) |
| --- | --- | --- | --- |
| OOI | Class II | 21 CFR 862.2570
Instrumentation for Clinical Multiplex Test Systems | Clinical Chemistry (75) |
## CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A Subtyping Kit (VER 2)
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OZE | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
| OEP | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
| OQW | Class II | 21 CFR 866.3332
Reagents for Detection
of Specific Novel
Influenza A Viruses | Microbiology (83) |
| NSU | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
| OOI | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
## CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza B Lineage Genotyping Kit
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OZE | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
| NSU | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
{2}
3
| OOI | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
| --- | --- | --- | --- |
## CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A/H5 Subtyping Kit (VER 3)
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OZE | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
| OEP | Class II | 21 CFR 866.3980
Respiratory Viral Panel
Multiplex Nucleic Acid
Assay | Microbiology (83) |
| NXD | Class II | 21 CFR 866.3332
Reagents for Detection
of Specific Novel
Influenza A Viruses | Microbiology (83) |
| NSU | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
| OOI | Class II | 21 CFR 862.2570
Instrumentation for
Clinical Multiplex Test
Systems | Clinical Chemistry (75) |
## H. Intended Use:
### 1. Intended use(s):
## CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A/B Typing Kit:
The Influenza A/B Typing Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
- For qualitative detection of influenza virus type A or B viral
{3}
RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
- To provide epidemiologic information for surveillance of circulating influenza viruses.
Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
# CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A Subtyping Kit (VER 2):
The Influenza A Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast
4
{4}
Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
- For determination of the subtype of seasonal human influenza A viruses as seasonal A(H3), and/or A(H1)pdm09 from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) and lower respiratory tract specimens (including bronchoalveolar lavage [BAL], bronchial wash [BW], tracheal aspirate [TA], sputum, and lung tissue) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
- To provide epidemiologic information for surveillance of circulating influenza viruses.
Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
5
{5}
6
# CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza B Lineage Genotyping Kit:
The Influenza B Lineage Genotyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
- For the determination of the genetic lineage of human influenza B viruses as B/Victoria or B/Yamagata lineage from viral RNA in upper respiratory tract clinical specimens (including nasopharyngeal swabs [NPS], nasal swabs [NS], throat swabs [TS], nasal aspirates [NA], nasal washes [NW] and dual nasopharyngeal/throat swabs [NPS/TS]) from human patients with signs and symptoms of respiratory infection and/or from viral culture;
- To provide epidemiologic information for surveillance of circulating influenza viruses.
Performance characteristics for influenza B lineage genotyping were established during a season when influenza B/Victoria and B/Yamagata lineages were found in approximately equal proportion.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
# CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel - Influenza A/H5 Subtyping Kit (VER 3):
The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:
{6}
- For the presumptive identification of virus in patients who may be infected with influenza A subtype A(H5) (Asian lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors;
- To provide epidemiologic information for surveillance of circulating influenza viruses.
Performance characteristics for influenza were established during a season when seasonal influenza viruses A(H1N1) and A(H3N2) were the predominant influenza A viruses in circulation and during a season when the A(H1N1)pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.
Testing with the influenza H5a and H5b primer and probe sets should not be performed unless the patient meets the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspect A(H5) specimens. The definitive identification of influenza A(H5) (Asian lineage) either directly from patient specimens or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.
Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted unless a BSL 3E facility is available to receive and culture specimens.
All users, analysts, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.
2. Indication(s) for use:
Same as Intended Use(s)
{7}
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument
I. Device Description:
The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel is used in real-time RT-PCR (rRT-PCR) assays on the Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR system. The panel is configured in four separate kits. Each kit consists of oligonucleotide primers, fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection and characterization of influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). Oligonucleotide primers and probes for detection of influenza A, influenza B, and 2009 influenza A (swine origin) were selected from highly conserved regions of the matrix (M), non-structural (NS), and nucleoprotein (NP) genes, respectively. Oligonucleotide primers and probes for characterization and differentiation of influenza A(H3) and A(H1)pdm09 viruses and genetic lineages of influenza B were selected from highly conserved regions of their HA genes. Oligonucleotide primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens are also included in the panel.
J. Substantial Equivalence Information:
1. Predicate device name(s):
CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel –
Influenza A/B Typing Kit
Influenza A Subtyping Kit (VER 2)
Influenza B Lineage Genotyping Kit
Influenza A/H5 Subtyping Kit (VER 3)
2. Predicate 510(k) number(s):
K133869
K161556
K140857
K153148
3. Comparison with predicate:
{8}
| Similarities or Differences | | |
| --- | --- | --- |
| Item | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A/B Typing Kit (K133869) | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A/B Typing Kit (K172091) |
| Specimen Types | Upper respiratory tract and lower respiratory tract specimens | Same |
| Organisms Detected | Influenza A viruses (animal and human) and influenza B viruses | Same |
| Analyte | RNA | Same |
| Technological Principles | Real-time RT-PCR | Same |
| Instrumentation | Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR system | Same |
| Enzyme Master Mix | Invitrogen SuperScript III Platinum One-Step Quantitative RT-PCR Kit (with or without ROX) or Quanta BioSciences qScript One-Step qRT-PCR Kit, Low ROX | Same |
| Nucleic Acid Extraction | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche |
{9}
| Similarities or Differences | | |
| --- | --- | --- |
| Item | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A Subtyping Kit (VER 2) (K161556) | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A Subtyping Kit (VER 2) (K172091) |
| Specimen Types | Upper respiratory tract and lower respiratory tract specimens | Same |
| Organisms Detected | Influenza A viruses (animal and human), swine-origin influenza A viruses, influenza A subtypes: seasonal A(H3) and A(H1)pdm09 | Same |
| Analyte | RNA | Same |
| Technological Principles | Real-time RT-PCR | Same |
| Instrumentation | Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR system | Same |
| Enzyme Master Mix | Invitrogen SuperScript III Platinum One-Step Quantitative RT-PCR Kit (with or without ROX) or Quanta BioSciences qScript One-Step qRT-PCR Kit, Low ROX | Same |
| Nucleic Acid Extraction | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche |
{10}
| Similarities or Differences | | |
| --- | --- | --- |
| Item | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza B Lineage Genotyping Kit (K140857) | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza B Lineage Genotyping Kit (K172091) |
| Specimen Types | Upper respiratory tract and lower respiratory tract specimens | Same |
| Organisms Detected | Influenza B viruses, lineages B/Victoria and B/Yamagada | Same |
| Analyte | RNA | Same |
| Technological Principles | Real-time RT-PCR | Same |
| Instrumentation | Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR system | Same |
| Enzyme Master Mix | Invitrogen SuperScript III Platinum One-Step Quantitative RT-PCR Kit (with or without ROX) or Quanta BioSciences qScript One-Step qRT-PCR Kit, Low ROX | Same |
| Nucleic Acid Extraction | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche |
11
{11}
| Similarities or Differences | | |
| --- | --- | --- |
| Item | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A/H5 Subtyping Kit (VER 3) (K153148) | CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel – Influenza A/H5 Subtyping Kit (VER 3) (K172091) |
| Specimen Types | Upper respiratory tract and lower respiratory tract specimens | Same |
| Organisms Detected | Influenza A viruses (animal and human), and influenza A subtype A(H5) (Asian lineage) | Same |
| Analyte | RNA | Same |
| Technological Principles | Real-time RT-PCR | Same |
| Instrumentation | Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR system | Same |
| Enzyme Master Mix | Invitrogen SuperScript III Platinum One-Step Quantitative RT-PCR Kit (with or without ROX) or Quanta BioSciences qScript One-Step qRT-PCR Kit, Low ROX | Same |
| Nucleic Acid Extraction | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche | • QIAamp DSP Viral RNA Mini Kit, QIAGEN
• MagNA Pure Compact – Nucleic Acid Isolation Kit I, Roche
• MagNA Pure Compact – RNA Isolation Kit, Roche
• MagNA Pure LC – Total Nucleic Acid Kit, Roche
• QIAcube – QIAamp DSP Viral RNA Mini Kit, QIAGEN
• NucliSENS easyMAG, bioMérieux
• EZ1 Advanced XL – EZ1 DSP Virus Kit and EZ1 RNA Tissue Mini Kit, QIAGEN
• MagNA Pure 96 - DNA and Viral NA Small Volume Kit, Roche |
{12}
K. Standard/Guidance Document Referenced (if applicable):
NA
L. Test Principle:
The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel includes sets of oligonucleotide primers and dual-labeled hydrolysis probes to be used in real-time RT-PCR assays on the Applied Biosystems (ABI) 7500 Fast Dx Real-time PCR instrument. The targeted regions of influenza viral RNA are transcribed into complimentary DNA (cDNA) and amplified by the polymerase chain reaction (PCR). The fluorescently labeled probes anneal to amplified DNA fragments and the fluorescent signal intensity is monitored by the ABI 7500 Fast Dx Real-time PCR instrument during each PCR cycle. Amplification of target is recorded as increase of fluorescence over time in comparison to background signal.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Using the CDC Influenza A Subtyping Kit as the representative assays from the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel, studies were conducted to assess the reproducibility of the two additional nucleic acids extraction instrument platforms to be used with the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel test kits. The Roche MagNA Pure 96 was evaluated with the Roche DNA and Viral NA Small Volume Kit. The QIAGEN EZ1 Advanced XL was evaluated with the QIAGEN EZ1 DSP Virus Kit and the QIAGEN EZ1 RNA Tissue Mini Kit.
A blinded panel of contrived samples containing a background of beta-propiolactone (BPL) treated A549 cells in viral transport medium (VTM) was assembled by adding a BPL treated influenza A(H3N2) virus, A/Hong Kong/4801/2014. The samples included a moderate positive sample, a low positive sample near the established assay LoD for the CDC Influenza A Subtyping Kit, and a negative sample consisting of background A549 cells and VTM.
Three separate testing sites were selected for each extraction instrument platform. The sample panel was tested five times by two analysts at each site over five days. Analysts performed extractions with the investigational nucleic acids extraction instrument platform and the relevant extraction method(s),
13
{13}
and tested the extracted nucleic acids with the InfA, H3, and RP assays from the CDC Influenza A Subtyping Kit using Invitrogen SuperScript and utilizing the Applied Biosystems 7500 Fast Dx (ABI 7500 Fast Dx) real-time PCR system.
The results for the reproducibility studies of each nucleic acids extraction instrument platform and the relevant extraction method(s) are summarized in Table 1 to Table 3 below.
Table 1: Reproducibility Study Summary Results - QIAGEN EZ1 Advanced XL, EZ1 DSP Virus Kit
| Panel Sample | Assay | Site 1 | | | Site 2 | | | Site 3 | | | Agreement Total | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | | |
| A(H3) Moderate | InfA | 10/10 | 25.46 | 2.36 | 10/10 | 26.45 | 2.80 | 10/10 | 28.46 | 4.12 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 27.15 | 2.92 | 10/10 | 28.22 | 3.25 | 10/10 | 28.61 | 4.40 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 22.45 | 1.44 | 10/10 | 23.47 | 2.42 | 10/10 | 24.50 | 5.68 | 30/30 | 100.0 (88.6-100.0) |
| A(H3) Low | InfA | 10/10 | 28.96 | 3.41 | 10/10 | 30.35 | 1.87 | 10/10 | 32.21 | 2.55 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 30.56 | 2.60 | 10/10 | 31.70 | 1.96 | 10/10 | 32.63 | 3.18 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 22.26 | 1.08 | 10/10 | 23.53 | 1.12 | 10/10 | 24.58 | 4.63 | 30/30 | 100.0 (88.6-100.0) |
| Negative | InfA | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 24.80 | 1.49 | 10/10 | 25.35 | 3.91 | 10/10 | 26.93 | 4.83 | 30/30 | 100.0 (88.6-100.0) |
Table 2: Reproducibility Study Summary Results - QIAGEN EZ1 Advanced XL, EZ1 RNA Tissue Mini Kit
| Panel Sample | Assay | Site 1 | | | Site 2 | | | Site 3 | | | Agreement Total | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | | |
| A(H3) Moderate | InfA | 10/10 | 25.97 | 3.48 | 10/10 | 26.40 | 4.06 | 10/10 | 28.59 | 2.01 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 27.40 | 4.13 | 10/10 | 27.56 | 2.35 | 10/10 | 28.66 | 0.79 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 22.08 | 3.31 | 10/10 | 23.52 | 1.56 | 10/10 | 24.77 | 2.14 | 30/30 | 100.0 (88.6-100.0) |
| A(H3) Low | InfA | 10/10 | 30.07 | 2.04 | 10/10 | 30.30 | 3.26 | 10/10 | 32.37 | 2.44 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 31.02 | 3.14 | 10/10 | 30.94 | 2.06 | 10/10 | 32.59 | 1.15 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 21.88 | 2.37 | 10/10 | 23.29 | 2.32 | 10/10 | 24.30 | 2.93 | 30/30 | 100.0 (88.6-100.0) |
| Negative | InfA | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 24.36 | 2.57 | 10/10 | 25.32 | 1.55 | 10/10 | 27.38 | 1.85 | 30/30 | 100.0 (88.6-100.0) |
Table 3: Reproducibility Study Summary Results - Roche MagNA Pure 96, DNA and Viral NA Small Volume Kit
| Panel Sample | Assay | Site 1 | | | Site 2 | | | Site 3 | | | Agreement Total | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | Agreement | Ave. Ct | % CV | | |
| A(H3) Moderate | InfA | 10/10 | 31.29 | 3.97 | 10/10 | 30.05 | 3.37 | 10/10 | 27.61 | 3.93 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 32.00 | 3.36 | 10/10 | 31.92 | 3.32 | 10/10 | 28.44 | 4.05 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 27.16 | 4.07 | 10/10 | 24.25 | 1.66 | 10/10 | 23.35 | 3.48 | 30/30 | 100.0 (88.6-100.0) |
| A(H3) Low | InfA | 10/10 | 34.91 | 2.97 | 10/10 | 33.49 | 2.86 | 10/10 | 31.40 | 3.07 | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 35.45 | 2.54 | 10/10 | 35.35 | 2.75 | 10/10 | 32.37 | 3.30 | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 26.97 | 3.85 | 10/10 | 24.22 | 0.98 | 10/10 | 23.11 | 2.70 | 30/30 | 100.0 (88.6-100.0) |
| Negative | InfA | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | H3 | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 10/10 | 0.00 | n/a | 30/30 | 100.0 (88.6-100.0) |
| | RP | 10/10 | 29.40 | 3.66 | 10/10 | 26.64 | 1.36 | 10/10 | 25.58 | 2.91 | 30/30 | 100.0 (88.6-100.0) |
Each nucleic acids extraction instrument platform and the relevant extraction
{14}
method demonstrated acceptable reproducibility with 100% agreement across different sites, analysts, and days.
b. Linearity/assay reportable range:
Not applicable, qualitative tests.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable, aspects of the tests other than nucleic acids extraction options are not modified from the previously FDA-cleared versions.
d. Detection limit:
The LoD performance equivalency between a previously FDA-cleared nucleic acids extraction method and either the Roche MagNA Pure 96 or the QIAGEN EZ1 Advanced XL nucleic acids extraction instrument platform and the relevant extraction method was evaluated by testing 5-fold serial dilutions of a characterized influenza A(H3N2) virus, A/Hong Kong/4801/2014, of known embryo infectious dose 50% titer (EID₅₀), using the CDC Influenza A Subtyping Kit as the representative assays from the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel.
Virus dilutions were prepared using a suspension of BPL treated A549 cells in VTM as diluent. Triplicate samples of each dilution were extracted separately with the previously FDA-cleared Roche MagNA Pure Compact RNA Isolation Kit, as well as the investigational nucleic acids extraction instrument platform and the relevant extraction method. The Roche MagNA Pure 96 was evaluated with the Roche DNA and Viral NA Small Volume Kit. The QIAGEN EZ1 Advanced XL was evaluated with the QIAGEN EZ1DSP Virus Kit and the QIAGEN EZ1 RNA Tissue Mini Kit.
Extracted RNA was tested with the InfA and H3 assays from the CDC Influenza A Subtyping Kit using Invitrogen SuperScript III Platinum One-Step RT-PCR System (Invitrogen SuperScript™) and utilizing the ABI 7500 Fast Dx real-time PCR system.
The pre-established acceptance criteria for demonstrating LoD equivalence between the previously FDA-cleared and the investigational extraction methods was defined as a demonstration of 100% positivity (3 out of 3 replicates) at either the same endpoint concentration or within one 5-fold dilution. The results of the study are summarized in Table 4 to Table 6 below.
15
{15}
Table 4: LoD Equivalency Determination - QIAGEN EZ1 Advanced XL, EZ1 DSP Virus Kit
| Titer (EID50/mL)1 | Roche MagNA Pure Compact - RNA Isolation Kit (Comparator) | | QIAGEN EZ1 Advanced XL - EZ1 DSP Virus Kit | |
| --- | --- | --- | --- | --- |
| | InfA | H3 | InfA | H3 |
| 103.2 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 102.3 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.8 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.1 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 100.4 | 1/3 (+) | 0/3 (+) | 1/3 (+) | 0/3 (+) |
$^{1}$ EID $_{50}$ = Egg Infectious Dose 50%
Table 5: LoD Equivalency Determination - QIAGEN EZ1 Advanced XL, EZ1 RNA Tissue Mini Kit
| Titer (EID50/mL)1 | Roche MagNA Pure Compact - RNA Isolation Kit (Comparator) | | QIAGEN EZ1 Advanced XL - EZ1 RNA Tissue Mini Kit | |
| --- | --- | --- | --- | --- |
| | InfA | H3 | InfA | H3 |
| 103.2 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 102.3 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.8 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.1 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 100.4 | 1/3 (+) | 0/3 (+) | 2/3 (+) | 2/3 (+) |
$^{1}$ EID $_{50}$ = Egg Infectious Dose 50%
{16}
Table 6: LoD Equivalency Determination - Roche MagNA Pure 96, DNA and Viral NA Small Volume Kit
| Titer (EID50/mL)1 | Roche MagNA Pure Compact - RNA Isolation Kit (Comparator) | | Roche MagNA Pure 96 - DNA and Viral NA Small Volume Kit | |
| --- | --- | --- | --- | --- |
| | InfA | H3 | InfA | H3 |
| 103.2 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 102.3 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.8 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 101.1 | 3/3 (+) | 3/3 (+) | 3/3 (+) | 3/3 (+) |
| 100.4 | 1/3 (+) | 0/3 (+) | 2/3 (+) | 2/3 (+) |
$^{1}$ EID $_{50}$ = Egg Infectious Dose 50%
Each of the three investigational nucleic acids extraction instrument platform and extraction method combinations demonstrated an equivalent endpoint LoD concentration when compared to the previously FDA-cleared nucleic acids extraction instrument platform and extraction method combination (comparator).
# e. Analytical specificity:
Not applicable, aspects of the tests other than nucleic acids extraction options are not modified from the previously FDA-cleared versions.
# f. Assay cut-off:
Not applicable, aspects of the tests other than nucleic acids extraction options are not modified from the previously FDA-cleared versions.
# 2. Comparison studies:
a. Method comparison with predicate device:
Refer to the "Clinical studies" section
b. Matrix comparison:
Not applicable
{17}
# 3. Clinical studies:
Performance testing natural clinical specimens using the two additional nucleic acids extraction instrument platforms with the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel test kits, the Roche MagNA Pure 96 and the QIAGEN EZ1 Advanced XL instrument platforms, was evaluated in a study using the CDC Influenza A Subtyping Kit as the representative assays from the CDC Human Influenza Real-Time RT-PCR Diagnostic Panel to demonstrate equivalent performance when testing natural clinical specimens to a nucleic acids extraction instrument platform and extraction method combination that was previously FDA-cleared for use with the CDC Human Influenza Real-Time Diagnostic Panel test kits (the comparator).
The Roche MagNA Pure 96 was evaluated using the Roche DNA and Viral NA Small Volume Kit. The QIAGEN EZ1 Advanced XL was evaluated using the QIAGEN EZ1 DSP Virus Kit and the QIAGEN EZ1 RNA Tissue Mini Kit. The comparator was the Roche MagNA Pure Compact instrument using the RNA Isolation Kit.
The study was conducted internally testing a set of retrospective clinical specimens collected during the 2011-2012 and 2013-2014 influenza seasons. A total of thirty clinical specimens that were previously determined to be positive for influenza A(H3) virus and thirty negative clinical specimens were evaluated with the InfA, H3, and RP assays from the CDC Influenza A Subtyping Kit as representative assays from the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel. Testing was performed using Invitrogen SuperScript and utilizing the ABI 7500 Fast Dx real-time PCR system. The study results are summarized in the Table 7 to Table 9 below.
Table 7: Retrospective Clinical Study Results - QIAGEN EZ1 Advanced XL, EZ1 DSP Virus Kit
| | Roche MagNA Pure Compact – RNA Isolation Kit (Comparator) | | Total |
| --- | --- | --- | --- |
| QIAGEN Advanced XL – EZ1 DSP Virus Kit | Positive | Negative | |
| Positive | 30 | 0 | 30 |
| Negative | 0 | 30 | 30 |
| Total | 30 | 30 | 60 |
| Positive Percent Agreement (PPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
| Negative Percent Agreement (NPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
{18}
Table 8: Retrospective Clinical Study Results - QIAGEN EZ1 Advanced XL, EZ1 RNA Tissue Mini Kit
| | Roche MagNA Pure Compact – RNA Isolation Kit (Comparator) | | Total |
| --- | --- | --- | --- |
| QIAGEN Advanced XL – EZ1 RNA Tissue Mini Kit | Positive | Negative | |
| Positive | 30 | 0 | 30 |
| Negative | 0 | 30 | 30 |
| Total | 30 | 30 | 60 |
| Positive Percent Agreement (PPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
| Negative Percent Agreement (NPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
Table 9: Retrospective Clinical Study Results – Roche MagNA Pure 96, DNA and Viral NA Small Volume Kit
| | Roche MagNA Pure Compact – RNA Isolation Kit (Comparator) | | Total |
| --- | --- | --- | --- |
| Roche MagNA Pure 96 – DNA and Viral NA Small Volume Kit | Positive | Negative | |
| Positive | 30 | 0 | 30 |
| Negative | 0 | 30 | 30 |
| Total | 30 | 30 | 60 |
| Positive Percent Agreement (PPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
| Negative Percent Agreement (NPA) = 100% (30/30), 95% CI: 88.6% to 100% | | | |
Each of the three investigational nucleic acids extraction instrument platform and extraction method combinations demonstrated $100\%$ agreement with the comparator.
# 4. Clinical cut-off:
Not applicable, aspects of the tests other than nucleic acids extraction options are not modified from the previously FDA-cleared versions.
# 5. Expected values/Reference range:
Not applicable, aspects of the tests other than nucleic acids extraction options are not modified from the previously FDA-cleared versions.
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.