truFreeze System

K172041 · CSA Medical, Inc. · GEH · Jul 27, 2017 · General, Plastic Surgery

Device Facts

Record IDK172041
Device NametruFreeze System
ApplicantCSA Medical, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateJul 27, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia) and malignant lesions.

Device Story

The truFreeze System is a cryosurgical device used to ablate benign and malignant lesions. It consists of a mobile console with a touch panel computer, cryogen/suction/electronics modules, and disposable spray kits. The system delivers medical-grade liquid nitrogen to the target tissue via a low-pressure, open-tipped catheter. During active venting procedures, a CryoDecompression Tube (CDT) is used to evacuate gas from the patient. The console manages cryogen levels, pressure, cooling, defrost, and suction. Physicians operate the system using a dual foot pedal and touch panel interface. The device provides visual feedback on the GUI regarding system status, including potential obstructions. Modifications to the CDT (decreased durometer and bump extrusion) improve flexural strength and kink resistance, providing physicians more space and freedom of movement during procedures while maintaining gas egress performance. The device is intended for use in clinical settings by trained healthcare professionals.

Clinical Evidence

No clinical data. The submission relies on bench testing, including CDT outer diameter measurement, bend testing, kink resistance testing, gas egress flow rate testing, CDT deformation under suction, biocompatibility, and accelerated aging. All tests passed.

Technological Characteristics

Cryosurgical unit using liquid nitrogen. Console includes touch panel computer, cryogen/suction/electronics modules. CDT materials: Pebax shaft. CDT dimensions: 16 Fr (internal) to 20 Fr (proximal external). Connectivity: Standalone mobile cart. Sterilization: Sterile, single-use catheters. Software: Off-the-shelf controller for cryogen management, pressure, cooling, defrost, suction, and data management.

Indications for Use

Indicated for use as a cryosurgical tool in dermatology, gynecology, and general surgery to ablate benign (e.g., Barrett's Esophagus with high grade dysplasia) and malignant lesions.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2017 CSA Medical, Inc. Ms. Sherrie Coval-Goldsmith VP of Regulatory & Quality 91 Hartwell Avenue Lexington, Massachusetts 02421 Re: K172041 Trade/Device Name: truFreeze System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: July 5, 2017 Received: July 6, 2017 Dear Ms. Coval-Goldsmith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172041 Device Name truFreeze ® System Indications for Use (Describe) The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (eg. Barrett's Esophagus with high grade dysplasia) and malignant lesions. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div style="display:inline-block; vertical-align:middle;"><input type="checkbox"/></div> For Exhibition Use Only (21 CFR 201.66 Subject to 201.68) | |----------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:inline-block; vertical-align:middle;"><input type="checkbox"/></div> For Test Consumption (21 CFR 201.64) | | | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K172041 510(k) SUMMARY Applicant Establishment Registration Number Contact Person CSA Medical, Inc. 3010140265 Sherrie Coval-Goldsmith, M.S. Vice President RA/QA CSA Medical 91 Hartwell Ave Lexington, MA 02421 Phone: 781-538-7447 Fax: 781-538-4730 sgoldsmith@csamedical.com Summary Date Proprietary Name Classification Classification Name Regulation Number Classification Predicate Device July 5, 2017 truFreeze® System Class II Cryosurgical Unit, Cryogenic Surgical Device 21 CFR 878.4350 Product Code GEH K163244 (truFreeze® System) ### Device Description The truFreeze System is a cryosurgical tool that applies medical-grade liquid nitrogen to the ablation area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit. ### Console: The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction. ### Disposable Spray Kits: Currently, there are 2 types of spray kits available. One kit is available for active venting procedures and one kit is available for passive venting procedures. Both active and passive venting kits are provided in a carton of five (5) sterile, single-use catheters with introducers in individual pouches. Additionally, the active venting kit includes a carton of five (5) sterile, single-use, dual-lumen Active Venting Cryo Decompression Tubes (CDTs) with associated tubing in individual pouches. Each carton within a spray kit contains Instructions for Use. {4}------------------------------------------------ ## K172041 ### CryoDecrompression Tube (CDT): Proposed changes involve the CryoDecrompression Tube (CDT) only. The CDT is a necessary part of the truFreeze Active Venting Kit to suction gas out of the patient when liquid nitrogen is sprayed. User Feedback suggests that some physicians experience significant space constraints within their procedure room resulting in the need to bend (portion of the tube that is outside the mouth) an excessive amount to get the CDT out of their way due to this space constraint. By moving the tube out of the way, the physician has more freedom to move during the ablation procedure. This excessive bending could result in a pinching of the tube leading to a transient increase in the absence of the absence of the cryospray that may be perceived as a physical obstruction. This is manifested by an optical color display of orange on the GUI. If this pinch created by the excessive bend is minimized by moving the tube, the orange indicator color disappears. Despite this appearance of the orange indicator display on the GUI, the CDT is performing as intended and is evacuating the gas as intended. This has led to some user annoyance. Additionally, User Feedback suggests that the weight of the suction tubing at the location of the console may also weigh down the CDT, depending upon tube position, resulting in the same pinching and orange indicator display described above. Therefore, this Special 510k covers 2 CDT modifications: - 1) decrease the durometer of the current Pebax shaft material to resist excessive bending forces of the CDT and increase the robustness of the CDT - 2) create a bump extrusion to allow for a bump up in the current size of the proximal segment of the CDT (located outside the mouth).The bump does not modify the 16 Fr portion within the body but increases to 20Fr external to the body. This configuration reduces the likelihood of pinching outside of the oral cavity due to greater flexural strength of the 20 Fr section. This design not only resists forces resulting in bending of the CDT but maintains the current gas egress flow performance specification. These modifications address the user annoyance and provide the physician additional space (outside the patient's body) during the cryogenic procedure. Therefore, the minor technological modifications do not change the intended use or indications for use of the device and no new questions of safety or performance are raised. ### Labeling (Intended Use/Indications for Use and Instructions for Use Document) The Intended Use/Indications for Use statements are identical to the predicate device. The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia) and malignant lesions. The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions. This is the same indication for use as previously cleared for the previous version of the truFreeze System (K163244). {5}------------------------------------------------ ### K172041 ### Technical and Operational Characteristics The truFreeze System, with the exception of the proposed CDT changes in durometer and dimension, is identical in design, operational and technological characteristics as the predicate device. The bump extrusion and the decreased durometer of the CDT modifications reduce user annoyance by providing additional space and freedom to move during the cryogenic procedure and eliminate the indicator display that there is an obstruction during the non-spraying portion of the ablation procedure. Therefore, the minor technological changes do not change the intended use or indications for use of the device and no new questions of safety or performance are raised. The fundamental technological characteristics is unchanged between the predicate device and the proposed device. There are no changes to the console or software. The proposed modifications do not alter the principles of operation. The principle of operation is still to destroy unwanted tissue using a cryogenic agent. No change to the energy output or flow rate occurs. Therefore, the proposed modification still permits the device to destroy unwanted tissue. There are no new safety or performance questions raised with this proposed modification, therefore the proposed CDT is substantially equivalent to the CDT described in K163244. | Specification Test associated with proposed modifications | Result | |-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Measure the maximum OD of the CDT | PASS<br>The CDT OD measurement was Less than<br>or equal to 16.5F (0.22 inches) | | Bend Test | PASS<br>All samples maintained structural<br>integrity and no kinks observed. | | Kink Resistance Test | PASS<br>All samples endured a complete flex of<br>the CDT without a fracture | | K172041 | | | Specification Test associated with proposed modifications | Result | | Gas Egress Flow Rate | PASS<br>All samples passed the minimum<br>required LPM requirement at 3 inHg | | CDT OD deformation with suction test | PASS<br>The CDT OD did not deform at the<br>minimum required % at 3 inches of<br>mercury (inHg) of vacuum | | Biocompatibility | All materials were determined to be<br>biocompatible | | Sterility | Sterility adoption was accepted | | Accelerated Aging | Pass<br>A 6 month shelf life is supported. | | Packaging | PASS | ### Summary of Testing {6}------------------------------------------------ The requirements specified have been successfully met, the proposed modifications have completed design verification in accordance with the specifications. The truFreeze System was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility testing, software testing, therefore, those tests {7}------------------------------------------------ #### K172041 are not being repeated. Verification testing to support the proposed catheter modification in this submission was performed. #### Rationale for Substantial Equivalence: The proposed changes do not affect the Instructions for Use of the truFreeze System or alter the fundamental scientific technology of the device compared with the predicate device. They do not introduce any new safety or efficacy questions. The company through its design control activities, has determined that clinical data are not necessary to evaluate safety and effectiveness related to these changes and that the results of design validation do not raise new issues of safety and effectiveness. Consequently, in the company, the truFreeze System, lower durometer material and bump extrusion design, is substantially equivalent to the predicate device. #### Conclusion The proposed modifications do not change the Intended use or indications for use and the principles of operation of the truFreeze® System are unchanged. The modifications are limited a decreased durometer and an added bump extrusion to the CDT. These insignificant modifications do not affect the technological characteristics or intended use of the truFreeze System. This proposed modification is accomplished while maintaining the fundamental technological characteristics of the system. These modifications address user feedback. These differences do not present any new issues of safety or effectiveness. Thus, the proposed truFreeze System and the predicate truFreeze System subject of K163244 are substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...