The registry data was used to support the safety and efficacy of adding 'Barrett's Esophagus with high grade dysplasia' as a specific indication for use for the truFreeze System.
Post-market registry; Barrett's Esophagus; Real-world clinical practice
Patients with Barrett's Esophagus (BE) and Barrett's Esophagus with high grade dysplasia (BE-HGD); Sample Size: 111 (Safety Population); 46 (Efficacy Population)
Not applicable for this study
Complete Eradication of Dysplasia (CE-D), Adverse Event rates
Indications for Use
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia) and malignant lesions.
Device Story
Cryosurgical tool applying medical-grade liquid nitrogen to ablation sites via small, low-pressure, open-tipped catheters. System comprises mobile console with touch panel computer (TPC), cryogen/suction/electronics modules, and disposable spray kits (active or passive venting). Physician operates console via dual foot pedal and touch panel to manage cryogen level, pressure, cooling, defrost, and suction. Console monitors parameters; provides safety features including pressure relief valves and emergency stop. Used in clinical settings to ablate unwanted tissue; output is controlled cryogenic spray. Clinical benefit includes effective removal of Barrett's Esophagus with high grade dysplasia, reducing lesion burden.
Clinical Evidence
Prospective post-market registry study (NCT01802203) evaluated 111 patients (safety population) and 46 patients with Barrett's Esophagus with high grade dysplasia (efficacy population). Primary endpoint: complete eradication of dysplasia (CE-D). Results: 40/46 (87.0%) achieved CE-D. Mean procedure sessions: 1.89. Adverse events: 0.9% stricture rate and 0.9% pancreatitis rate in efficacy population. Data supports safety and efficacy for the expanded indication.
Technological Characteristics
Cryosurgical unit; liquid nitrogen delivery via open-tipped catheter. Console includes TPC, cryogen/suction/electronics modules. Mobile cart form factor. Connectivity: standalone. Software-managed cryogen sensing, filling, pressure, cooling, defrost, and suction. Biocompatible, sterile, single-use catheters.
Indications for Use
Indicated for patients requiring ablation of benign or malignant lesions in dermatology, gynecology, and general surgery, including Barrett's Esophagus with high grade dysplasia.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Coldplay CryoBalloon™ Focal Ablation System (K161202)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's commitment to health, human services, and science. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
CSA Medical, Inc. Ms. Sherrie Coval-Goldsmith. M.S. VP of Regulatory & Quality 91 Hartwell Avenue Lexington, Massachusetts 02421
Re: K163244
Trade/Device Name: Trufreeze® System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: April 25, 2017 Received: April 27, 2017
Dear Ms. Coval-Goldsmith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163244
Device Name truFreeze® System
Indications for Use (Describe)
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (e.g. Barrett's Esophagus with high grade dysplasia) and malignant lesions.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K163244
#### 510(k) SUMMARY
Applicant Establishment Registration Number Contact Person
Summary Date Proprietary Name Classification Classification Name Regulation Number Classification Predicate Device Reference Devices
CSA Medical, Inc. 3010140265 Sherrie Coval-Goldsmith, M.S. Vice President RA/QA CSA Medical 91 Hartwell Ave Lexington, Ma 02421 Phone:781-538-7447 Fax:781-538-4730 sgoldsmith@csamedical.com
May 19, 2017 truFreeze® System Class II Cryosurgical Unit, Cryogenic Surgical Device 21 CFR 878.4350 Product Code GEH K162695 (truFreeze® System) K152329 (Coldplay CryoBalloon™ Focal Ablation System), K083737 (Barrx System), and K 161202 ((Coldplay CryoBalloon™ Focal Ablation System)
#### Device Description
The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the ablation area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit.
#### Console:
The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart. Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction. The console is the same as previously cleared for the predicate truFreeze System (K162695). Therefore, there are no new issues of safety or effectiveness raised.
#### Disposable spray kit:
There are 2 types of spray kits available. One kit is available for active venting procedures and one is available for passive venting procedures. Both active and passive venting kits are provided in a carton of five (5) individually packaged sterile, single-use catheters with introducers in individual pouches. Additionally, the active venting kit is provided with includes a carton of five (5) individually packaged sterile, single-use CDTs with associated tubing in individual pouches. Each carton within a spray kit contains the instructions for use. The disposable kit is the same as previously cleared for the predicate
{4}------------------------------------------------
truFreeze System (K162695). Therefore, there are no new issues of safety or effectiveness raised.
## Labeling (Intended Use/Indications for Use and Instructions for Use Document)
The Intended Use/Indications for Use statement is identical in its Intended Use as the predicate device with the exception of adding a specific indication for use as an example of benign tissue that can be ablated with truFreeze. Clinical data has been generated to substantiate a change in label claim. The proposed modification adds Barrett's Esophagus with high grade dysplasia into the Intended Use/Indications for Use statement as an example of benign tissue to be ablated.
## Indications for Use
The truFreeze System is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign (eg. Barrett's Esophagus with high grade dysplasia) and malignant lesions.
## Technical and Operational Characteristics
The truFreeze System is identical in design, operational and technological characteristics as the predicate device and supports that no new safety concerns are being raised by change in intended use/indications for use statement and thus raises no new issues of safety or effectiveness.
## Summary of Testing
The truFreeze System was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility testing, software testing, animal testing, biocompatibility and sterilization testing. Data was previously submitted to support that the use of liquid nitrogen is cold enough to ablate benign or malignant lesions (K133258) and reference is being made to that data. Clinical data are summarized below.
## Rationale For Substantial Equivalence:
The Intended Use/Indications for Use statement and technological characteristics of the truFreeze System and the predicate device were compared. The Intended Use/Indications for Use statement of the two devices had equivalent general claims but a specific example of a benign lesion that can be ablated was added to the indications for use statement. Clinical data are presented to support the safety and efficacy of adding Barrett's Esophagus with high grade dysplasia as an example of specific benign that can be ablated.
A clinical summary of data collected from a post market registry study (ClinicalTrials.gov Identifier: NCT01802203) supports this submission. This post-marketing registry prospectively collected data on patients treated with the FDA-cleared cryotherapy device (truFreeze® System). The instructions for use document guided the dosimetry used for ablation procedures.
Similar to any clinical trial, physicians offered "all comers" with BE to participate in the registry if they met all inclusion criteria and none of the exclusion criteria. Patients were sequentially and consecutively
{5}------------------------------------------------
enrolled into the registry at participating trial.
The registry had pre-defined inclusion / exclusion criteria, which are consistent with the instructions for use of the cleared device. CSA Medical applied 3 additional exclusion criteria that were sequentially applied to the population of BE patients. These exclusion criteria included items 3-5 in table 1 below. This yielded patients for the efficacy population.
Two medical societies cite endoscopic radiofrequency ablation (RFA) as a recognized ablation method for BE HGD. 1-2 A rate of 81% CE-D is cited in the literature with a lower rate of 62% also being cited.1-
The clinical data for the patients with Barrett's Esophagus high grade dysplasia are provided in Tables 1-5 below.
| Inclusion Criteria | 1. Able to read, comprehend, and complete the informed consent form.<br>2. Aged 18 or older.<br>3. Receiving spray cryotherapy procedure with the truFreeze device for<br>the first time for removal of unwanted tissues including malignant or pre-<br>malignant conditions of the aerodigestive system. |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Exclusion Criteria | 1. Contraindication to spray cryotherapy.<br>2. Prior treatment with spray cryotherapy. Previous or concurrent<br>treatment using other mucosal therapies such as endoscopic mucosal<br>resection or radiofrequency ablation is acceptable.<br>3. Patients not having Barrett's Esophagus with high grade dysplasia.<br>4. Patients without a pre-therapy and post-therapy histology.<br>5. Patients who received other therapies (such as endoscopic mucosal<br>resection) during the ablation therapy period. |
| Safety Population | 111 Patients |
| Efficacy population | 46 patients |
| Device Used | truFreeze System |
Table 1 Summary of Populations analyzed
## Table 2 Demographics
| | BE Patients<br>(Safety Population)<br>(N=111) | BE HGD Patients<br>(Efficacy Population)<br>(N=46) |
|----------------------------|-----------------------------------------------|----------------------------------------------------|
| Age Mean (yrs) | 67.8 (10.8) | 67.4 (9.1) |
| Range (yrs) | 24 - 89 | 50 - 86 |
| Female | 29 (26.1%) | 15 (32.6%) |
| Male | 82 (73.9%) | 31 (67.4%) |
| Segment length (cm) <= 3 | 65 (59%) | 30 (65%) |
| Segment length (cm) 3-6 | 17 (15%) | 8 (17%) |
| Segment length (cm) >=6 | 21 (19%) | 5 (11%) |
| Segment length (cm)Unknown | 8 (7%) | 3 (7%) |
{6}------------------------------------------------
| | BE Patients<br>(Safety Population)<br>(N=111) | BE HGD Patients<br>(Efficacy Population)<br>(N=46) |
|-----------------------------------------|-----------------------------------------------|----------------------------------------------------|
| Previous EMR, no. (%) | 30 (28%) | 17 (37%) |
| Mean time since EMR (months), mean (SD) | 6.9 (11.8)<br>N=28<br>(2 missing data) | 7.0 (12.7)<br>N=17 |
Table 3 . Procedure Information to Achieve Best Response
| | 46 BE-HGD Patients in Efficacy<br>Population |
|--------------------------------------------------------------|----------------------------------------------|
| Patients receiving 1 procedure session to best<br>response | 19 (41%) |
| Patients receiving 2 procedure sessions to best<br>response | 17 (37%) |
| Patients receiving 3 procedure sessions to best<br>response | 6 (13%) |
| Patients receiving >3 procedure sessions to best<br>response | 4 (9%) |
| Mean Procedure Sessions (SD) | 1.89 (0.95) |
| Mean Cycles per procedure (SD) | 2.23 (0.52) |
| Mean Time (sec) per cycle (SD) (seconds) | 21.17 (2.94) |
Table 4 Procedure or Device Related Adverse Event Rate per Procedure Session
| | BE Patients<br>Safety Population Procedures<br>(258 procedures) | BE HGD Patients<br>Efficacy Population Procedures<br>(113 procedures) |
|---------------|-----------------------------------------------------------------|-----------------------------------------------------------------------|
| Stricture | 3 (1.2%) | 1 (0.9%) |
| Pancreatitis* | 1 (0.4%) | 1 (0.9%) |
* This is the only serious adverse event reported. The physician stated the relationship of truFreeze treatment and development of pancreatitis was highly unlikely. However, a possible causal relation with the device could not be ruled out because of the temporal association.
{7}------------------------------------------------
# Table 5 Barrett's segment length for CE-D Results
| | Response<br>CE-D |
|------------------------------------|------------------|
| Efficacy Population | 46 |
| N Responders | 40 (87.0%) |
| Response rate by BE segment length | |
| Unknown | 2 (4.3%) |
| <= 3cm | 27 (58.7%) |
| 3-6cm | 7 (15.2%) |
| >=6cm | 4 (8.7%) |
# Table 6 Safety Data
| | truFreeze Per Patient Rate |
|---------------------|----------------------------|
| | Safety Population (N=111) |
| Stricture | 3 (2.7%) |
| Abdominal Pain | 0 (0%) |
| Pancreatitis | 1 (0.9%) |
| Chest Pain | 0 (0%) |
| GI Hemorrhage | 0 (0%) |
| Mucosal Lacerations | 0 (0%) |
# Table 7 Efficacy Data (N=46)
| | CE-D Patients (%) |
|-----------|-------------------------------|
| truFreeze | 40 (87% BE-HGD patients only) |
{8}------------------------------------------------
## Discussion
The data show that the majority of BE-HGD patients were successfully treated with the truFreeze device. Among the 46 BE-HGD patients in the efficacy population, 40 achieved a response of CE-D of 87.0 %. This success rate is comparable to the success rate for the Barrx System and the Coldplay CryoBalloon™ Focal Ablation System among all phenotypes of BE patients, ranging from no dysplasia to early adenocarcinoma and across 3 different ablation doses.
The safety rate of the trufreeze device is also substantially equivalent to these 2 reference devices.
## Conclusion
In summary, endoscopic ablation of Barrett's Esophagus with high grade dysplasia with the truFreeze device does not raise new questions of safety or efficacy as compared to the reference devices or the predicate device and warrant adding Barrett's Esophagus with high grade dysplasia as an example of benign tissues that can be ablated, to the truFreeze label. This indication was supported by clinical data obtained from the post market registry study (ClinicalTrials.gov Identifier: NCT01802203).
#### References
- ASGE Standards of Practice Committee. The role of endoscopy in Barrett's esophagus and other 1. premalignant conditions of the esophagus. Gastrointestinal Endoscopy 2012; 76 (6): 1087 (available at https://www.asge.org/docs/default-source/education/practice_guidelines/doc-therole-of-endoscopy-in-barretts- esophagus-and-other-premalignant-conditions-of-theesophagus.pdf?sfvrsn=8).
- Nicholas J. Shaheen , MD, MPH, FACG , Gary W. Falk , MD, MS, FACG, Prasad G. lyer , MD, MSc, 2. FACG andLauren Gerson , MD, MSc, FACG. ACG Clinical Guideline: Diagnosis and Management of Barrett's Esophagus. Am J Gastroenterol advance online publication, 3 November 2015; doi: 10.1038/ajg.2015.322 (available at https://gi.org/guideline/diagnosis-and-management-of-barretts-esophagus/).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.