POWDER FREE NITRILE EXAMINATION GLOVES, BLUE (COLORED)
K172015 · Careglove Global Sdn. Bhd. · LZA · Sep 26, 2017 · General Hospital
Device Facts
Record ID
K172015
Device Name
POWDER FREE NITRILE EXAMINATION GLOVES, BLUE (COLORED)
Applicant
Careglove Global Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Sep 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, ambidextrous nitrile examination glove; beaded cuff; blue color. Manufactured via on-line polymer-coating and mild chlorination to ensure tack-free surface without dusting/donning powder. Used in clinical settings by healthcare professionals to provide barrier protection against contamination between patient and examiner. Device is single-use; provides physical barrier; does not involve electronic processing or clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only, including physical property testing (tensile strength, elongation), water leak testing, and biocompatibility testing (primary skin irritation and dermal sensitization).
Indicated for use as a disposable medical device worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Careplus Powder Free Nitrile Examination Gloves, Blue (Colored) (K142862)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2017
Careglove Global Sdn Bhd Lim Shyan Managing Director Lot 17479, Lrg Senawang 3/2, Off Jln Senawang 3, Senawang In Seremban, 70450 My
Re: K172015
Trade/Device Name: Powder Free Nitrile Examination Gloves, Blue (colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: Julv 3, 2017 Received: July 3, 2017
Dear Lim Shyan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
### Tara A. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K172015
Device Name
POWDER FREE NITRILE EXAMINATION GLOVES, BLUE (COLORED)
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "CARE GLOVE". The word "CARE" is written in large, bold, blue letters. Below "CARE" is the word "GLOVE" in a light green color with a blue circle in the middle of the word with a hand in the center. Below the word "GLOVE" is the text "CAREPLUS-DESCARPACK" in blue.
### AREGLOVE GLOBAL
Lot 17479, Lorong Senawang 3/2, Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Malaysia. Tel: 60-6-6782377, 60-6-6788377 Fax: 60-6-6785377 Email: info@careglove.com
### K172015 510(K) SUMMARY
#### CAREGLOVE GLOBAL SDN BHD Applicant:
- Location Lot 17479, Lorong Senawang 2/3 Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia.
Phone No. (60) 6 6782377 Fax No. (60) 6 6785377
Contact Person:Lim Kwee Shyan
- 31st August, 2017 Summary Preparation Date:
### Device Information
| Trade Name: | POWDER FREE NITRILE EXAMINATION GLOVES, BLUE (COLORED) |
|----------------------|--------------------------------------------------------|
| Common Name: | POWDER FREE NITRILE EXAMINATION GLOVES |
| Classification Name: | Patient Examination Gloves |
| Regulatory Class: | |
| Product Code: | LZA |
| Regulation: | 21 CFR 880.6250 |
| | |
| Predicate Device | |
| Device Name: | CAREPLUS POWDER FREE NITRILE EXAMINATION GLOVES, BLUE (COLORED) |
|----------------------|-----------------------------------------------------------------|
| Product Code: | LZA |
| Classification Name: | Patient Examination Gloves |
| 510K Number: | K142862 |
| Regulatory Class: | I |
#### Device Description
It is the powder-free variation of the class I latex patient examination gloves made by on-line polymer-coating and mild on-line chlorination process modifies the surface characteristics and causes it to remain tackfree without the use of any dusting or donning powder.
#### Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo with the words "CARE GLOVE" stacked on top of each other. The word "CARE" is in blue, and the word "GLOVE" is in light green. There is a blue circle with a light green hand in the middle of the word "GLOVE". Below the words "CARE GLOVE" is the word "CAREPLUS-DESCARPACK" in blue.
CAREGLOVE GLOBAL SDI
Lot 17479, Lorong Senawang 3/2, Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Malaysia. Tel: 60-6-6782377, 60-6-6788377 Fax: 60-6-6785377 Email: info@careglove.com
#### Summary of the Technological Characteristic
The Powder Free Nitrile Examination Gloves – Blue are summarized with the following technological characteristic compare to ASTM D6310 or equivalent standards.
| Characteristic | Subject Device | Predicate Device | Comparison<br>Analysis |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Powder Free Nitrile<br>Examination Gloves, Blue<br>(Colored) | Careplus Powder Free Nitrile<br>Examination Glove, Blue<br>(Colored) | Different |
| 510(k) Reference | K172015 | K142862 | N/A |
| Product Code | LZA | LZA | Same |
| Regulatory Class | I | I | Same |
| Intended Use | Intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner | Intended for medical purposes<br>that is worn on the examiner's<br>hand to prevent contamination<br>between patient and examiner | Same |
| Design | Powder Free, Non-Sterile,<br>Ambidextrous, Beaded Cuff | Powder Free, Non-Sterile,<br>Ambidextrous, Beaded Cuff | Same |
| Indications for<br>Use | A patient examination glove<br>is a disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner | A patient examination glove is<br>a disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>to prevent contamination<br>between patient and examiner | Same |
| Construction | Ambidextrous, Polymer<br>Coated or Chlorinated,<br>Powder Free Nitrile | Ambidextrous, Polymer<br>Coated or Chlorinated, Powder<br>Free Nitrile | Same |
| Color Description | Blue | Blue | Same |
| Material | Nitrile | Nitrile | Same |
| Single Use | Yes | Yes | Same |
| Packaging | Packed in Dispenser Boxes | Packed in Dispenser Boxes | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for a company called "CARE GLOVE." The word "CARE" is written in large, bold, blue letters on top. Below it, the word "GLOVE" is written in a lighter green color. There is a blue circle with a light green hand in the middle of the word "GLOVE." Underneath the words "CARE GLOVE" is the text "CAREPLUS-DESCARPACK" in smaller, blue letters.
### CAREGLOVE GLOBAL SDN BHD
Lot 17479, Lorong Senawang 3/2, Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Malaysia. Tel: 60-6-6782377, 60-6-6788377 Fax: 60-6-6785377 Email: info@careglove.com
| Dimension | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|------|
| Length (size: XSmall), mm<br>Length (size: Small), mm<br>Length (size: Medium), mm<br>Length (size: Large), mm<br>Length (size: XLarge), mm | Meet 220mm min<br>Meet 220mm min<br>Meet 230mm min<br>Meet 230mm min<br>Meet 230mm min | Meet 230mm min<br>Meet 220mm min<br>Meet 230mm min<br>Meet 230mm min<br>Meet 230mm min | Same |
| Thickness (palm), mm<br>Thickness (finger), mm | Meet 0.05mm min<br>Meet 0.05mm min | Meet 0.05mm min<br>Meet 0.05mm min | |
| Width (size: XSmall), mm<br>Width (size: Small), mm<br>Width (size: Medium), mm<br>Width (size: Large), mm<br>Width (size: XLarge), mm | Meet 70 ± 10 mm<br>Meet 80 ± 10 mm<br>Meet 95 ± 10 mm<br>Meet 110 ± 10 mm<br>Meet 120 ± 10 mm | Meet 70 ± 10 mm<br>Meet 80 ± 10 mm<br>Meet 95 ± 10 mm<br>Meet 111 ± 10 mm<br>Meet 120 ± 10 mm | |
| Physical Properties | | | |
| (Before Ageing)<br>i) Tensile Strength (MPa)<br>ii) Ultimate Elongation (%) | Meet 14MPa min.<br>Meet 500% min | Meet 14MPa min.<br>Meet 500% min | Same |
| (After Aging)<br>i) Tensile Strength (MPa)<br>ii) Ultimate Elongation (%) | Meets 14MPa min<br>Meet 400% min. | Meets 14MPa min<br>Meet 400% min. | |
| Water Leak Test,<br>1000 ml<br>Before Aging, AQL<br>After Aging, AQL | Meet AQL 1.5<br>Meet AQL 2.5 | Meet AQL 1.5<br>Meet AQL 2.5 | Same |
| Residual Powder Content<br>Residual Powder Content,<br>mg/glove | Meet 2mg/glove max. | Meet 2mg/glove max | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for a company called "CARE GLOVE". The word "CARE" is in blue, and the word "GLOVE" is in light green. There is a blue circle with a light green hand in the middle of the word "GLOVE". The words "CAREPLUS-DESCARPACK" are in blue at the bottom of the logo.
# CAREGLOVE GLOBAL SDN BHD
Lot 17479, Lorong Senawang 3/2, Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Malaysia. Tel: 60-6-6782377, 60-6-6788377 Fax: 60-6-6785377 Email: info@careglove.com
| Biocompatibility Test | | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------|
| i) Primary Skin Irritation Test | Passes<br>i) Primary Skin Irritation Test.<br>Conclusion: Under the conditions of this study the test material did not cause an irritant response | Passes<br>i)Primary Skin Irritation Test.<br>Conclusion: Under the conditions of this study the test material did not cause an irritant response. | Same |
| ii)Skin Sensitization Test | ii)Dermal Sensitization Test.<br>Conclusion: Under the conditions of this study, the test material did not produce a skin sensitization effect | ii)Dermal Sensitization Test.<br>Conclusion: Under the conditions of this study, the test material did not produce a skin sensitization effect | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "CARE GLOVE". The word "CARE" is in large, bold, blue letters on top. Below it, the word "GLOVE" is in a light green color. The "O" in "GLOVE" is replaced by a blue circle with a light green handprint inside. Underneath the word "GLOVE" is the text "CAREPLUS-DESCARPACK" in small, bold, blue letters.
## CAREGLOVE GLOBAL SI
Lot 17479, Lorong Senawang 3/2, Off Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Malaysia. Tel: 60-6-6782377, 60-6-6788377 Fax: 60-6-6785377 Email: info@careglove.com
#### Summary of Clinical Testing
Not applicable
Substantial Equivalence Conclusions.
The subject device is a safe, as effective, and performs as well as or better than the legally marketed predicate device, K142862 (Careplus Powder Free Nitrile Examination Gloves, Blue (Colored)).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.