POWDER FREE NITRILE EXAMINATION GLOVE, BLUE (Colored).
K142862 · Careplus (M) Sdn Bhd · LZA · Jan 16, 2015 · General Hospital
Device Facts
Record ID
K142862
Device Name
POWDER FREE NITRILE EXAMINATION GLOVE, BLUE (Colored).
Applicant
Careplus (M) Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable nitrile examination glove; blue colored; powder-free. Manufactured via on-line polymer-coating and chlorination process to modify surface characteristics, ensuring tack-free finish without dusting powder. Used in clinical settings by healthcare professionals to provide a barrier against contamination between patient and examiner. Benefits users sensitive to powdered gloves or those requiring a powder-free environment. Device is non-sterile.
Clinical Evidence
Bench testing only. Device performance verified against ASTM D6319-10 (dimensions, physical properties), ASTM D5151-06 (water tightness), and ASTM D6124-06 (powder residue). Biocompatibility testing conducted per ISO 10993-10 (dermal sensitization) and 16 CFR Part 1500 (primary skin irritation), confirming the material is not a sensitizer or irritant.
Technological Characteristics
Material: Nitrile. Form factor: Disposable examination glove. Surface treatment: On-line polymer-coating and chlorination. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10, 16 CFR Part 1500. Non-sterile.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination Glove Blue and White Colored, Nonsterile (K123469)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
Careplus (M) SDN BHD Mr. Lim Kwee Shyan Managing Director Lot 120 & 121 Jalan Senawang 3 Senawang Industrial Estate 70450 Seremban Negeri Sembilan Darul Khusus MALAYSIA
Re: K142862
Trade/Device Name: Powder Free Nitrile Examination Glove, Blue (Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: I Product Code: LZA Dated: December 1, 2014 Received: December 3, 2014
Dear Mr. Shyan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K142862
Device Name: Powder Free Nitrile Examination Glove, Blue (Colored)
Indications for Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo with a white square background. Inside the square, there is a gray vertical rectangle on the left side and a red semi-circle on the right side. The semi-circle is positioned such that it forms the letter 'P' when combined with the gray rectangle. There is a black shadow underneath the square.
### AREPLUS (M) SDN BHD (Company No. 212677-K)
ng Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus. Malaysia. Fax: 60-6-6772780 Tel: 60-6-6772781
E-mail: info@careplus.com
Quality and Relationship You Can Trust
## 510(K) SUMMARY
510(k) No.: K142862
- CAREPLUS (M) SDN BHD 1.0 Applicant: Address: Lot 120 & 121 , Jalan Senawang Senawang Industrial Estate,
70450 Seremban, Negeri Sembilan Darul Khusus. Malaysia.
- Phone No.: 60-6-6772781 Fax No. 60-6-6772780
- 2.0 Lim Kwee Shvan Contact Person: Phone No.: 60-6-6772781 Fax No. 60-6-6772780
- Preparation Date: 16th January, 2015 3.0
- 4.0 Device Information
Device Name: POWDER FREE NITRILE EXAMINATION GLOVE, BLUE(COLORED) Common Name: POWDER FREE NITRILE EXAMINATION GLOVE Classification Name: Patient Examination Gloves
#### 5.0 Claim of Equivalence
The device is a class I latex patient examination gloves 21 CFR 880.6250, Patient Examination Glove, LZA which is made powder-free by a process of on-line polymercoating and chlorination which is substantially equivalent in safety and effectiveness to the predicate device Powder Free Nitrile Examination Glove Blue and White Colored, Nonsterile 510(k) number K123469, product code LZA.
- 6.0 Device Description
It is the powder-free variation of the class I latex patient examination gloves made by online polymer-coating and on-line chlorination process. The process modifies the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder. It is particularly suitable to users who prefer a powder-free work environment or who may be sensitive to the powdered version of the same gloves.
- 7.0 Intended Use of Device
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo that features a square with a thick black border. Inside the square, there is a gray vertical rectangle on the left side and a red semi-circle on the right side. The semi-circle is positioned so that its flat edge aligns with the right edge of the gray rectangle, creating a shape that resembles the letter 'P'.
# CAREPLUS (M) SDN BHD (company No. 212677-K)
Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia. Tel: 60-6-6772781 Fax: 60-6-6772780
E-mail: info@careplus.com
Quality and Relationship You Can Trust
#### 8.0 1)Comparison To Predicate Device
| No. | Characteristic | Standards | Related Defects | Predicate Device<br>(Powder Free Nitrile<br>Examination Glove,<br>Blue and White<br>Colored K123469) | Powder-Free Nitrile<br>Examination Glove<br>(Blue Color) |
|-----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Water Tight Test,<br>1000 ml | ASTM D6319-10<br>ASTM D5151-06 | G- I, AQL<br>2.5<br>(FDA GII,<br>AQL 2.5) | Meets ASTM D6319-10<br>Meets ASTM D5151-06 | Meets ASTM D6319-10<br>Meets ASTM D5151-06 |
| 2 | Physical Properties<br>(Before Ageing)<br>i) Tensile Strength(Mpa)<br>ii)Ultimate Elongation(%)<br>(After Ageing)<br>i) Tensile Strength(Mpa)<br>ii) Ultimate Elongation(%) | ASTM D6319-10 | Min. 14<br>Min. 500<br><br>Min. 14<br>Min. 400 | Meets ASTM D6319-10<br>Meets ASTM D6319-10<br><br>Meets ASTM D6319-10<br>Meets ASTM D6319-10 | Meets ASTM D6319-10<br>Meets ASTM D6319-10<br><br>Meets ASTM D6319-10<br>Meets ASTM D6319-10 |
| 3 | Powder Content | ASTM D6319-10<br>ASTM D6124-06 | Max.<br>2mg/glove | Meets ASTM D6319-10<br>Meets ASTM D6124-06 | Meets ASTM D6319-10<br>Meets ASTM D6124-06 |
| 4 | Biocompatibility Test<br><br>i) Primary Skin Irritation<br>Test<br><br>ii)Dermal Sensitization<br>Test | ISO 10993-10<br><br>As per 16CFR<br>Part 1500 | No Animal<br>Irritation<br><br>No Animal<br>Irritation | Passes<br>i) Primary Skin<br>Irritation Test.<br>Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>cause an irritant<br>response.<br><br>ii)Dermal Sensitization<br>Test.<br>Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>produce a skin<br>sensitization effect. | Passes<br>i) Primary Skin Irritation Test.<br>Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>cause an irritant<br>response.<br><br>ii)Dermal Sensitization<br>Test.<br>Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>produce a skin<br>sensitization effect. |
| 5 | Color Extraction | | | Blue | Blue |
### Conclusions:
Careplus (M) Sdn Bhd, Powder Free Nitrile Examination Glove, Blue (colored) is similar with Wear Safe (M) Sdn Bhd, Powder Free Nitrile Examination Glove, Blue and White Colored, K123469.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo that features a square divided into two distinct sections. On the left side, there is a gray vertical rectangle. On the right side, there is a red semi-circle. The logo has a simple and modern design, using only two colors and basic geometric shapes.
# CAREPLUS (M) SDN BHD (company No. 212677-K)
Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia. Tel: 60-6-6772781 Fax: 60-6-6772780
E-mail: info@careplus.com
Quality and Relationship You Can Trust
## 2) Summary of Technological Characteristics of the Device
Powder Free Nitrile Examination Glove, Blue(Colored) possesses the following technological characteristic (as compared to ASTM or equivalent standards):
| Parameter | Standard References | Device Performance |
|------------------------|-------------------------------------------------------------|----------------------------------|
| Dimensions | ASTM D6319-10 | Meet Specification |
| Physical Properties | ASTM D6319-10 | Meet Specification |
| Freedom from pin-holes | ASTM D5151-06 | Meet Specification |
| Powder Free Residue | ASTM D6124-06 | Meet Specification |
| | ASTM D6319-10 | Meet Specification |
| Biocompatability Test | Dermal Sensitization<br>Test<br>(ISO 10993-10:2010) | Not a contact skin<br>sensitizer |
| | Primary Skin Irritation<br>Test<br>(As per 16CFR Part 1500) | Not a primary skin<br>irritant |
#### 9.0 Conclusion
Powder Free Nitrile Examination Glove, Blue(Colored) performs according to the glove performance standards referenced in Section 8.0 above. The device is substantially equivalent to current marketed devices per Section 5.0.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.