K171942 · Edan Instruments, Inc. · DPS · Jan 24, 2018 · Cardiovascular
Device Facts
Record ID
K171942
Device Name
Electrocardiograph
Applicant
Edan Instruments, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jan 24, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and pediatric patients (beginning at birth through 21 years of age) through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Device Story
The SE-12 series electrocardiograph acquires 12-lead ECG signals simultaneously from adult and pediatric patients. The device features a 32-bit processor, high-resolution thermal recorder, and internal memory. It operates via mains or battery power. The system offers manual, auto, rhythm, R-R analysis, and VCG (on Express models) modes. SE-12 Express includes stress ECG functionality for diagnosing coronary heart disease and assessing treatment effects. Used in hospitals/clinics by physicians and trained healthcare professionals. The device displays ECG waveforms and parameters on an integrated LCD screen. It provides automated measurements and interpretive statements as an advisory tool for clinicians to support heart disease diagnosis. The device is intended to aid clinical decision-making, not replace professional judgment.
Clinical Evidence
No clinical data. Evidence consists of bench testing per IEC 60601-2-25:2011, biocompatibility testing (cytotoxicity, skin sensitization, skin irritation per ISO 10993-1), electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2), and software verification/validation.
Technological Characteristics
12-lead ECG system; 24-bit A/D resolution; 16,000 Hz sampling frequency; thermal dot-matrix recorder (200 dpi amplitude, 1000 dpi time axis); LCD display (320x240 or 800x600); Type CF applied part; Class I with internal battery; connectivity via mains/battery; software-based interpretation; biocompatible materials (surface contact <24h).
Indications for Use
Indicated for adult and pediatric patients (birth through 21 years) for the acquisition of ECG signals via body surface electrodes to assist in the analysis and diagnosis of heart disease in hospital or healthcare facility settings.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Edan Instruments, Inc., PC ECG, model SE-1515 (K152427)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 24, 2018
Edan Instruments, Inc. % Doug Worth Sr. Dir. US RA/QA Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94089
Re: K171942
Trade/Device Name: Electrocardiograph with models SE-12, SE-12 Express, SE-1200 and SE-1200 Express Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 21, 2017 Received: December 26, 2017
Dear Doug Worth:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MA Willel em
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
# K171942
#### Device Name
Electrocardiograph: SE-12, SE-12 Express, SE-1200, SE-1200 Express
Indications for Use (Describe)
The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and pediatric patients (beginning at birth through 21 years of age) through body surface ECG electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and disease. However, the interreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community,<br>Kengzi Sub-District, Pingshan District, Shenzhen, 518122 P.R.China.<br>Tel.: (0755) 28340011<br>Fax: +1 (408) 418-4059 |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Alice Yang |
| Date prepared: | June 19, 2017 |
| 2. Device name<br>and classification: | Device Name: Electrocardiograph |
| | Model: SE-12, SE-1200, SE-12 Express, SE-1200 Express |
| | Classification Name: |
| | 870.2340 Electrocardiograph<br>Product code: DPS |
| | Regulatory Class: Class II |
| 3.Premarket<br>Notification Class<br>III Certification<br>and Summary | Not applicable, the subject device is Class II. |
| 4. Predicate<br>Device(s): | 1. Edan Instruments, Inc., Electrocardiograph, models SE-12, SE-12 Express,<br>SE-1200, and SE-1200 Express cleared under K160876 (Primary)<br>2. Edan Instruments, Inc., PC ECG, model SE-1515 cleared under<br>K152427(Reference) |
| 5. Reason for<br>Submission | Modification for previous cleared product Electrocardiograph, models<br>SE-12, SE-1200, SE-12 Express, SE-1200 Express. |
| 6. Pre-Submission,<br>IDE | Not applicable, there is no prior submission. |
{4}------------------------------------------------
| 7. Device<br>Description: | The SE-12/12 Express/1200/1200 Express Electrocardiograph gathers<br>ECG signals of 12 leads simultaneously. It displays the operation menu,<br>ECG parameters as well as electrocardiograms, and be powered by the<br>mains supply or battery. And the system has advanced performance and<br>high reliability due to high resolution thermal recorder, 32-bit processor<br>and a large-capacity memorizer. Design of the system took much<br>consideration on ergonomics so the size is suitable for clinic and<br>hospital uses. |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | There are four selectable modes in the system, including manual, auto,<br>rhythm, R-R analysis or VCG, and VCG is configured with SE-12<br>Express & SE-1200 Express. |
| | SE-12 and SE-1200 share the same single-color LCD screen of which<br>the resolution is 320×240 dot; and LCD screen of SE-12 Express and<br>SE-1200 Express is 800×600 multicolor LCD screen. |
| | Moreover, SE-12 Express is configured with stress ECG function, which<br>will allow to diagnose concealed coronary heart disease and atypical<br>angina pectoris, prescribe the workload for patients with myocardial<br>infarction before leaving hospital, and assess the effect of the treatment. |
| 8. Intended Use: | The intended use of SE-12 series electrocardiograph is to acquire ECG<br>signals from adult and pediatric patients (beginning at birth through 21<br>years of age) through body surface ECG electrodes. The<br>electrocardiograph is only intended to be used in hospitals or healthcare<br>facilities by doctors and trained healthcare professionals. The<br>cardiogram recorded by the electrocardiograph can help users to analyze<br>and diagnose heart disease. However, the interpreted ECG with<br>measurements and interpretive statements is offered to clinicians on an<br>advisory basis only. |
# 9. Predicate Device Comparison
Comparison to the predicate devices, the subject device has the same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following tables:
| Item | Predicatedevice<br>(SE-12, SE-12Express,<br>SE-1200, andSE-1200Express) | Proposed device<br>(SE-12, SE-12Express, SE-1200,<br>andSE-1200Express) | Comparison<br>Result |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K# | K160876 | Current Submission | — |
| Intended Use | The intended use of the<br>12-channel electrocardiograph is<br>to acquire ECG signals from<br>adult and pediatric patients<br>through body surface ECG<br>electrodes. The<br>electrocardiograph is only | The intended use of SE-12 series<br>electrocardiograph is to acquire<br>ECG signals from adult and<br>pediatric patient (beginning at<br>birth through 21 years of age)<br>through body surface ECG<br>electrodes. The | Different |
| | | | |
| intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. | | electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. | |
| Safety Specifications | | | |
| Safety Standards | IEC 60601-1:2005/A1:2012<br>EN 60601-1:2006/A1:2013<br>IEC 60601-1-2:2007<br>EN 60601-1-2:2007/AC:2010<br>IEC/EN 60601-2-25 | IEC 60601-1:2005/A1:2012<br>EN 60601-1:2006/A1:2013<br>IEC 60601-1-2:2007<br>EN 60601-1-2:2007/AC:2010<br>IEC/EN 60601-2-25 | Same |
| Anti-electric-shock type: | Class I with internal power supply | Class I with internal power supply | Same |
| Anti-electric-shock degree: | Type CF | Type CF | Same |
| Degree of protection against harmful ingress of water: | Ordinary equipment (Sealed equipment without liquid proof) | Ordinary equipment (Sealed equipment without liquid proof) | Same |
| Disinfection/sterilization method: | Refer to the user manual for details | Refer to the user manual for details | Same |
| Degree of safety of application in the presence of flammable gas: | Equipment not suitable for use in the presence of flammable gas | Equipment not suitable for use in the presence of flammable gas | Same |
| Working mode: | Continuous operation | Continuous operation | Same |
| EMC: | CISPR 11 Group 1, Class A | CISPR 11 Group 1, Class A | Same |
| Ingress rating | IPX0 | IPX0 | Same |
| Environment Specifications | | | |
| Temperature | | | |
| Transport & Storage | -20°C (-4°F) ~ +55°C (+131°F) | -20°C (-4°F) ~ +55°C (+131°F) | Same |
| Working | +5°C (+41°F) ~ +40°C (+104°F) | +5°C (+41°F) ~ +40°C (+104°F) | Same |
| Relative Humidity: | | | |
| Transport & Storage | 25%~93%<br>Non-Condensing | 25%~93%<br>Non-Condensing | Same |
| Working | 25% RH ~ 80% RH<br>Non-Condensing | 25% RH ~ 80% RH<br>Non-Condensing | Same |
| Atmospheric Pressure: | | | |
| Transport & Storage | 700 hPa ~1060 hPa | 700 hPa ~1060 hPa | Same |
| Working | 860 hPa ~1060 hPa | 860 hPa ~1060 hPa | Same |
| Power Supply Specifications | | | |
{5}------------------------------------------------
{6}------------------------------------------------
| Operating frequency = 50 Hz / 60<br>Operating frequency = 50 Hz / 60<br>Same<br>Hz<br>Hz<br>Mains Supply:<br>Input Current = 0.9-0.4A<br>Input Current = 0.9-0.4A<br>Same<br>Or Input power = 96VA<br>Or Input power = 96VA<br>Rated voltage = 14.8 V<br>Rated voltage = 14.8 V<br>Same<br>SE-12 Express&SE-1200<br>SE-12 Express&SE-1200<br>Express: Rated capacity =<br>Express: Rated capacity =<br>5000mAh<br>5000mAh<br>Same<br>SE-12& SE-1200: Rated capacity<br>SE-12& SE-1200: Rated capacity<br>= 2500mAh<br>= 2500mAh<br>When the battery is fully<br>When the battery is fully charged,<br>charged, SE-12&SE-1200 can<br>work<br>SE-12& SE-1200can work<br>normally about4hours, and it can<br>normally about 4 hours, and it<br>Built-in Lithium<br>can continually print about 1.5<br>continually print about 1.5 hours<br>Battery Pack:<br>hours in the manual mode or<br>in the manual mode or print<br>print about 300 ECG reports of<br>about300 ECG reports of<br>3×4+1R in the auto mode: SE-12<br>3×4+1Rin the auto mode; SE-12<br>Same<br>Express &SE-1200 Express can<br>Express& SE-1200 Express can<br>work normally about 5 hours, and<br>work<br>it can continually print about 2.5<br>normally about 5 hours, and it<br>can continually print about 2.5<br>hours in the manual mode or print<br>hours in the manual mode or<br>about 350 ECG reports of<br>print about 350 ECG reports of<br>3×4+1Rin the auto mode.<br>3×4+1R in the auto mode.<br>Performance Specifications<br>Recording<br>Thermal dot-matrix recorder<br>Thermal dot-matrix recorder<br>Recorder:<br>Same<br>8 dots per mm / 200 dots per inch<br>8 dots per mm / 200 dots per inch<br>(amplitude axes)<br>(amplitude axes)<br>Printing Density<br>Same<br>40 dots per mm / 1000 dots per<br>40 dots per mm / 1000 dots per<br>inch (time axes, @, 25 mm/s)<br>inch (time axes, (@, 25 mm/s)<br>Folded thermal paper:<br>Folded thermal paper:<br>210mm×295mm×100pages<br>210mm×295mm×100pages<br>Folded thermal paper:<br>Folded thermal paper:<br>Recorder Paper:<br>215mm×280mm×100pages<br>215mm×280mm×100pages<br>Same<br>(Optional)<br>(Optional)<br>Rolled thermal paper:<br>Rolled thermal paper:<br>210mm×30m (Optional)<br>210mm×30m (Optional)<br>Effective Width:<br>210mm<br>210mm<br>Same<br>5mm/s, 6.25mm/s, 10mm/s,<br>5mm/s, 6.25mm/s, 10mm/s,<br>12.5mm/s, 25mm/s, 50mm/s<br>12.5mm/s, 25mm/s, 50mm/s<br>Paper Speed:<br>Same<br>(+3%)<br>(+3%)<br>±5% (x-axis), ±5%(y-axis)<br>Accuracy of data:<br>±5% (x-axis), ±5%(y-axis)<br>Same | | Operating Voltage =100V-240V~ | Operating Voltage =100V-240V~ | Same | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-------------------------------|-------------------------------|------|--|
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | | | | | |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.