K152427 · Edan Instruments, Inc. · DPS · Feb 5, 2016 · Cardiovascular
Device Facts
Record ID
K152427
Device Name
PC ECG
Applicant
Edan Instruments, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Feb 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
SE-1515 PC ECG is intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. The SE-1515 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the SE-1515 PC ECG can help users to analyze and diagnose heart diseases. However, the ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
Device Story
PC-based ECG system; acquires, processes, and stores ECG signals; supports 12, 16, and 18-lead configurations via wired (DE18, DE15, DP12) or wireless (DX12) sampling boxes. Wireless communication uses Bluetooth. For stress testing, integrates with external BP monitors and treadmills/ergometers. Used in hospitals/clinics by physicians and trained healthcare professionals. Software provides ECG measurements and interpretive statements on an advisory basis. Clinicians review output to assist in diagnosis and treatment decisions. Benefits include standardized ECG data acquisition and automated analysis to support clinical heart disease assessment.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), performance testing (IEC 60601-2-25), biocompatibility (ISO 10993-1), and software verification/validation.
Technological Characteristics
PC-based ECG system; wired/wireless sampling boxes; Bluetooth connectivity for wireless modules. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-25 (performance), ISO 10993-1 (biocompatibility). Features vector loop analysis (FRANK leads). Software: Moderate level of concern.
Indications for Use
Indicated for adult and pediatric patients undergoing stress exercise or resting ECG tests in hospitals and healthcare facilities.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2016
Edan Instruments. Inc. % Doug Worth Sr. Dir. US RA/QA Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94086
Re: K152427
Trade/Device Name: PC ECG, model SE-1515 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 24, 2015 Received: December 28, 2015
Dear Doug Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152427
Device Name PC ECG, model SE-1515
#### Indications for Use (Describe)
SE-1515 PC ECG is intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. The SE-1515 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the SE-1515 PC ECG can help users to analyze and diagnose heart diseases. However, the ECG with measurements is offered to clinicians on an advisory basis only. It is mainly used in the ECG Outpatient Department and Physical Examination Department.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>Shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86(0755) 26858736<br>Fax: +1 (408) 418-4059 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:<br>Preparing date: | Queena Chen<br>December 24, 2015 |
| 2. Device name and<br>classification: | Device Name: PC ECG<br>Model: SE-1515<br>Classification Name/ Product code:<br>870.2340 Electrocardiograph/DPS<br>Regulatory Class: Class II |
| 3. Predicate Device(s): | 1) EDAN Instrument, Inc. SE-1010, K131819 (Primary)<br>2) Philips Medical Systems, Philips DXL 12/16-lead ECG<br>Algorithm, K132068 (Reference)<br>3) Philips Medical Systems, Philips TC70, K113144(Reference) |
| 4. Reason for<br>Submission | New product. |
| 5. Pre-Submission,<br>IDE | Not applicable, there is no prior submission. |
| 6. Device Description: | SE-1515 is a PC-based ECG designed to acquire, process and<br>store ECG signals and analyze and diagnose heart disease.<br>SE-1515 consists of analysis software and sampling boxes, and<br>has four configurations: Wired 18-lead ECG system, Wired<br>16-lead ECG system, Wired 12-lead ECG system and Wireless<br>12-lead ECG system.<br><br>Wired 18-lead ECG system consists of analysis software and<br>wired DE18 sampling box.<br><br>Wired 16-lead ECG system consists of analysis software and<br>wired DE15 sampling box. Wired 12-lead ECG system consists<br>of analysis software and wired DP12 sampling box. Wireless |
| | 12-lead ECG system contains analysis software, wireless DX12<br>sampling boxes (DX12 transmitter and DX12 receiver). The<br>communication between DX12 transmitter and DX12 receiver<br>is based on Bluetooth technology.<br>For stress exercise testing, the user should connect a BP<br>monitor and a treadmill or an ergo meter to the PC. |
| 7. Intended Use: | SE-1515 PC ECG is intended to acquire, process and store ECG<br>signals from adult and pediatric patients undergoing stress<br>exercise test or resting test. The SE-1515 PC ECG is intended<br>to be used only in hospitals and healthcare facilities by doctors<br>and trained healthcare professionals. The cardiogram recorded<br>by the SE-1515 PC ECG can help users to analyze and diagnose<br>heart diseases. However, the ECG with measurements and<br>interpretive statements is offered to clinicians on an advisory<br>basis only. |
{4}------------------------------------------------
## 8. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters items, which are shown in the following tables in details.
| Item | SE-1515 | SE-1010 | Remark |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | Current Submission | K131819 | ---- |
| Indications for Use | | | |
| Intended use | SE-1515 PC ECG is intended to acquire process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. The SE-1515 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the SE-1515 PC ECG can help users to analyze and diagnose heart diseases. However, the ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. It is mainly used in the ECG | SE-1010 is a PC-based diagnostic tool intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-1010 can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinicians on an | Different |
| | Outpatient Department and<br>Physical Examination<br>Department. | advisory basis only. | |
| High/Low Pass<br>Filters | DFT Filter:<br>0.01 Hz/0.05 Hz/0.32 Hz/0.67 Hz<br>AC Filter: 50 Hz/60 Hz/Off<br>EMG Filter:<br>25 Hz/35 Hz/45 Hz/Off<br>Low Pass Filter:<br>300 Hz/270 Hz/150 Hz/100 Hz/75<br>Hz | DFT Filter: weak/strong<br>AC Filter: 50 Hz/60 Hz/Off<br>Low Pass/EMG Filter:<br>25 Hz/35 Hz/45 Hz/75 Hz/100<br>Hz/150 Hz | Different |
| Sampling Box | | | |
| ---- | DE15 | DP12 | ---- |
| Classification | | | |
| Anti-electric-sho<br>ck type | Class II | Class II | Same |
| Anti-electric-sho<br>ck degree | Type CF with defibrillation-proof | Type CF with<br>defibrillation-proof | Same |
| Lead number | 16-lead | 12-lead | Different |
| Degree of<br>protection<br>against harmful<br>ingress of water | Sealed equipment without<br>waterproof | Sealed equipment without<br>waterproof | Same |
| Degree of safety<br>of application in<br>the presence of<br>flammable gas | Equipment not suitable for use in<br>the presence of flammable gas | Equipment not suitable for use<br>in the presence of flammable<br>gas | Same |
| Working mode | Continuous operation | Continuous operation | Same |
| Processing | | | |
| Sampling Rate | 16 kHz /channel | 1 kHz /channel | Different |
| Resolution | 0.1575 µV/LSB @ 1 kHz | 2.52 µV/LSB @ 1 kHz | Different |
| Time Constant | >3.2 s | ≥3.2 s | Same |
| Frequency<br>Response | 0.01 Hz~300 Hz | 0.05 Hz ~ 150 Hz | Different |
| Input Impedance | ≥100 MΩ | >50 ΜΩ | Different |
| Dynamic Range | AC Differential ± 5 mV, DC<br>offset ±600 mV | AC Differential ± 5 mV, DC<br>offset ±600 mV | Same |
| System Noise | ≤12.5μ Vp-p | ≤12.5μ Vp-p | Same |
| Common Mode<br>Rejection | ≥140 dB(AC ON)<br>≥123 dB(AC OFF) | >110 dB | Different |
| Pace Detect | ±750 uV~±700 mV @ 50 us ~2.0<br>ms | ±2 to ±700 mV @ 0.1 ms to<br>2.0 ms | Different |
| ESD Sensitivity | +6 kV contact | +6 kV contact | Same |
| | ±8 kV air | ±8 kV air | |
| Heart Rate Meter | 30 BPM ~300 BPM, ±1 BPM<br>DE18 | 30 BPM ~300 BPM, ±1 BPM<br>DP12 | Same |
| Classification | | | |
| Anti-electric-sho<br>ck type | Class II | Class II | Same |
| Anti-electric-sho<br>ck degree | Type CF with defibrillation-proof | Type CF with<br>defibrillation-proof | Same |
| Lead number | 18-lead | 12-lead | Different |
| Degree of<br>protection<br>against harmful<br>ingress of water | Sealed equipment without<br>waterproof | Sealed equipment without<br>waterproof | Same |
| Degree of safety<br>of application in<br>the presence of<br>flammable gas | Equipment not suitable for use in<br>the presence of flammable gas | Equipment not suitable for use<br>in the presence of flammable<br>gas | Same |
| Working mode | Continuous operation | Continuous operation | Same |
| Processing | | | |
| Sampling Rate | 16 kHz /channel | 1 kHz /channel | Different |
| Resolution | 0.1575 µV/LSB @ 1 kHz | 2.52 µV/LSB @ 1 kHz | Different |
| Time Constant | ≥3.2 s | ≥3.2 s | Same |
| Frequency<br>Response | 0.01 Hz~300 Hz | 0.05 Hz ~ 150 Hz | Different |
| Input Impedance | ≥100 ΜΩ | >50 ΜΩ | Different |
| Dynamic Range | AC Differential ± 5 mV, DC<br>offset ±600 mV | AC Differential ± 5 mV, DC<br>offset ±600 mV | Same |
| System Noise | ≤12.5μ Vp-p | ≤12.5μ Vp-p | Same |
| Common Mode<br>Rejection | ≥140 dB (AC ON)<br>≥123 dB (AC OFF) | >110 dB | Different |
| Pace Detect | ±750 uV~±700 mV @ 50 us ~2.0<br>ms | ±2 to ±700 mV @ 0.1 ms to<br>2.0 ms | Different |
| ESD Sensitivity | ±6 kV contact<br>±8 kV air | ±6 kV contact<br>±8 kV air | Same |
| Heart Rate Meter | 30 BPM ~300 BPM, ±1 BPM | 30 BPM~300 BPM, ±1 BPM | Same |
| DP12 Sampling Box | | | |
| Classification | | | |
| Anti-electric-sho<br>ck type | Class II | Class II | Same |
| Anti-electric-sho<br>ck degree | Type CF with defibrillation-proof | Type CF with<br>defibrillation-proof | Same |
| Lead number | 12-lead | 12-lead | Same |
| | | | |
| protection<br>against harmful<br>ingress of water | protection against ingress of water | protection against ingress of water | |
| Degree of safety<br>of application in<br>the presence of<br>flammable gas | Not suitable for use in the<br>presence of a flammable<br>anesthetic mixture with air or with<br>oxygen or nitrous oxide | Not suitable for use in the<br>presence of a flammable<br>anesthetic mixture with air or<br>with oxygen or nitrous oxide | Same |
| Working mode | Continuous operation | Continuous operation | Same |
| <b>Processing</b> | | | |
| Sampling Rate | 1,000 /sec/channel | 1,000 /sec/channel | Same |
| Resolution | 2.52 $ μ $ V/LSB @ 1 kHz | 2.52 $ μ $ V/LSB @ 1 kHz | Same |
| Time Constant | $ ≥ $ 3.2 s…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.