K171845 · Emba Medical Limited · KRD · Aug 18, 2017 · Cardiovascular
Device Facts
Record ID
K171845
Device Name
EMBA Hourglass Peripheral Embolization Device
Applicant
Emba Medical Limited
Product Code
KRD · Cardiovascular
Decision Date
Aug 18, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EMBA Peripheral Embolization Device (PED) is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature in 6-8mm blood vessels. The device is not indicated for use in blood vessels subject to repetitive motion, such as extremity or pulmonary vessels.
Device Story
Implantable, self-expanding, covered nitinol embolic device; preloaded in catheter-based delivery system. Used in peripheral vasculature to obstruct/reduce blood flow; triggers clotting cascade. Physician-operated under fluoroscopic guidance. Delivery system tracked to target site; device released by retracting outer body to allow expansion and detachment via interlock mechanism. Benefits patient by providing mechanical embolization. Used in clinical settings.
Clinical Evidence
Clinical study conducted on 51 subjects (n=51) to evaluate safety and effectiveness endpoints. Additional evidence includes extensive in-vitro and in-vivo performance testing: kink radius, trackability, deployment accuracy, migration resistance, occlusion effectiveness, MR compatibility, corrosion, fatigue, and delivery system integrity.
Technological Characteristics
Laser-cut nitinol structure with polymer covering; multi-lobed/hourglass shape. Self-expanding. Delivered via catheter-based system. Terminally sterilized by ethylene oxide (EtO). Biocompatibility per ISO 10993.
Indications for Use
Indicated for obstruction or reduction of blood flow in peripheral vasculature (6-8mm vessels). Contraindicated for vessels < 6mm and vessels subject to repetitive motion (e.g., extremity or pulmonary vessels).
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Reverse Medical® MVP® Micro Vascular Plug System (K141313)
{0}------------------------------------------------
## 510(k) Summary
| Date Prepared: | July 31, 2017 |
|-----------------|---------------------------------------------------------------------------------------------------|
| 510(k) SPONSOR: | EMBA Medical Limited<br>3451 Commerce Parkway<br>Miramar, FL 33025<br>Phone Number: 954-874-1016 |
| | |
#### EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) TRADE NAME:
#### COMMON NAME: Vascular Embolization Device
| Product | Regulation and<br>Description | Product Code | Device | Device Class |
|--------------------------------------------|----------------------------------------------------------|--------------|------------------------------------|--------------|
| Peripheral<br>Embolization<br>Device (PED) | 21 CFR<br>870.3300<br>Vascular<br>Embolization<br>Device | KRD | Vascular<br>Embolization<br>Device | II |
## PREDICATE DEVICES:
Reverse Medical® MVP® Micro Vascular Plug System (K141313) Amplatzer® Vascular Plug 4 (K113658)
## DEVICE DESCRIPTION:
The EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) consists of a covered, implantable, self-expanding structure (Embolic Device) preloaded in a catheter-based Delivery System. The Embolic Device is intended to be deployed to the target site in the vasculature under fluoroscopic guidance. The product is shipped sterile and labeled for single use only.
The configuration of the EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) is listed below.
| Reference Number | Embolic Device<br>Diameter<br>(mm) | Embolic Device<br>Length<br>(mm) | Embolic Device<br>Maximum Covered<br>Length<br>(mm) |
|------------------|------------------------------------|----------------------------------|-----------------------------------------------------|
| PEP-001 | 10 | 20 | 10 |
{1}------------------------------------------------
#### INDICATIONS FOR USE:
The EMBA Peripheral Embolization Device (PED) is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature in 6-8mm blood vessels. The device is not indicated for use in blood vessels subject to repetitive motion, such as extremity or pulmonary vessels.
#### CONTRAINDICATIONS
The device should not be implanted in vessels < 6 mm due to risk of vessel perforation or device fracture.
The device is not indicated for use in blood vessels subject to repetitive motion, such as extremity or pulmonary vessels due to risk of device fracture.
#### BASIS FOR SUBSTANTIAL EQUIVALENCE:
The EMBA" HOURGLASS" Peripheral Embolization Device (PED) and the predicate devices, Reverse Medical® MVP® Micro Vascular Plug System and Amplatzer® Vascular Plug 4 are intended for embolization/obstruction of blood flow in the peripheral vasculature - arterial and venous. Compared to the predicate devices, the subject device is not indicated for use in blood vessels < 6 mm or subject to repetitive motion: however, these differences do not introduce a new intended use. Additionally, the combination of device labeling and totality of data support that the EMBA Hourglass does not raise new questions of safety and effectiveness.
The operating principle of the EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) and the predicate devices, Reverse Medical® MVP® Micro Vascular Plug System, and Amplatzer® Vascular Plug 4 are equivalent. All the devices are delivered through a Delivery System that is introduced into the vasculature using standard techniques. The Delivery System is advanced to the site of interest by the physician using fluoroscopic guidance. Once in position, the Embolic Device is delivered by manipulating the Delivery System to mechanically release the Embolic Device. The Amplatzer® Vascular Plug 4 and Reverse Medical® MVP® Micro Vascular Plug System are mounted on a pusher and tracked through a guide catheter to the desired location. The EMBA™ HOURGLASS™ Peripheral Embolization Device is mounted on a pusher, inside an outer body, which is tracked to the desired location. The Amplatzer® Vascular Plug 4 and Reverse Medical® MVP® Micro Vascular Plug System are deployed by rotating the delivery wire to unscrew the proximal end of the embolic device from the wire. The EMBA™ HOURGLASS™ Peripheral Embolization Device is released by retracting the outer body until the embolic device is exposed, thus allowing the proximal portion of the device to expand, releasing the device from the interlock mechanism. Although there are technological differences in the delivery system and modes of detachment, these differences do not raise any different questions of safety and effectiveness. All of the Embolic Devices consist of a self-expanding material which, when deployed, expand to obstruct the blood flow in the vessel thus triggering the clotting cascade.
The EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) and the predicate devices, Reverse Medical® MVP® Micro Vascular Plug System, and Amplatzer® Vascular Plug 4 have similar technological characteristics and materials of construction. The Reverse Medical® MVP® Micro Vascular Plug System Embolic Device is laser cut nitinol in an oblong shape tapering at each end. The device includes a polymer covering. The Amplatzer® Vascular Plug 4 is a braided wire construction in a multi-lobed shape. The EMBA" HOURGLASS"
{2}------------------------------------------------
Peripheral Embolization Device (PED) is a laser cut nitinol in a multi-lobed or hourglass shape including a polymer covering. These characteristics are present in the other predicate devices. Design features of the embolic device of the EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) and the predicate devices are very similar and are considered substantially equivalent: Delivery System length, system compatibility, Embolic Device shape, and Embolic Device size, Embolic Device materials of construction, and indicated vessel sizes. Although there are slight technological differences in the embolic device designs, these differences do not raise any different questions of safety and effectiveness.
The subject device, EMBA™ HOURGLASS™ Peripheral Embolization Device (PED), passed appropriate ISO 10993 testing to demonstrate biocompatibility.
The subject device, EMBA™ HOURGLASS™ Peripheral Embolization Device (PED), and the cited predicate devices are provided packaged sterile and all of the devices are terminally sterilized by ethylene oxide (EtO).
Comparative performance testing was conducted to demonstrate that the technological differences the EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) and the predicate devices do not raise concerns of substantial equivalence. Testing was conducted in accordance with the FDA guidance document of "Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices." Testing conducted consisted of Kink Radius, Trackability, Partial Deployment and Retraction, Deployment Accuracy, Chronic Outward Force, Embolic Device Length and Diameter, Foreshortening, Contrast Flow, Migration Resistance, Occlusion Effectiveness, MR Compatibility, Corrosion, Fatigue Testing, Delivery System Integrity. Shelf Life, and Animal Testing.
Human Use Clinical Testing has been conducted on the EMBA™ HOURGLASS™ Peripheral Embolization Device (PED) (n=51) to evaluate safety and effectiveness endpoints.
## CONCLUSIONS:
Review of the in-vitro and in-vivo performance testing as well as the comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, EMBA™ HOURGLASS™ Peripheral Embolization Device (PED), is substantially equivalent to the cited predicate devices, Reverse Medical® MVP® Micro Vascular Plug System (K141313) and Amplatzer® Vascular Plug 4 (K113658). Any differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
EMBA Medical Limited % Randy Prebula Partner Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K171845
Trade/Device Name: EMBA Hourglass Peripheral Embolization Device Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: June 20, 2017 Received: June 20, 2017
Dear Randy Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{4}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.