Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves are intended for medical purposes that are worn on the examiners hands to prevent contamination between patient and examiner.
Device Story
Non-sterile, single-use, powder-free nitrile butadiene rubber examination gloves; ambidextrous design with beaded cuff and straight fingers. Inner surface features polyacrylic polymer coating and HydraSoft (glycerine) coating to facilitate donning. Used in clinical settings by healthcare personnel to provide a barrier against contamination between patient and examiner. Device is passive; no electronic or mechanical components. Benefits include protection against cross-contamination and improved donning comfort via HydraSoft coating.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified against ASTM D6319-10 (dimensions, tensile strength, elongation), ASTM D5151-06 (freedom from holes), and ASTM D6124-06 (powder residual). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (skin irritation and sensitization), and ISO 10993-11 (acute systemic toxicity).
Technological Characteristics
Material: Nitrile butadiene rubber. Coating: Polyacrylic polymer and HydraSoft (glycerine). Design: Ambidextrous, beaded cuff, finger-textured. Sizes: XS, S, M, L, XL. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-11. Non-sterile, single-use.
Indications for Use
Indicated for use by medical examiners to prevent contamination between patient and examiner. Intended for use as a disposable, non-sterile, powder-free patient examination glove.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Patient Examination Glove, Blue Colored and White (Non-colored), Non-sterile, Polymer Coated (K123469)
Reference Devices
Encore Sterile Powder-Free Polymer Coated Latex Surgical Glove with Hydrasoft™ Coating (K051793)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 29, 2017
Ansell Healthcare Products LLC Robert Mahler Director, Regulatory Affairs for the Americas 111 Wood Avenue South, Suite 210 Iselin, New Jersey 08830
Re: K171737
Trade/Device Name: Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: December 8, 2017 Received: December 11, 2017
Dear Robert Mahler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171737
#### Device Name
Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Glove
### Indications for Use (Describe)
Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves are intended for medical purposes that are worn on the examiners hands to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CER 801 Subnart D) | Over-The-Counter Use (21 CFR 801 Subnart C) |
| | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary
### 510(k): k171737
### Submitter:
Ansell Healthcare Products LLC. 111 Wood Avenue South, Suite 210 Iselin, NJ 08830 USA
### Contact Person:
Robert Mahler Director, Regulatory Affairs for the Americas Phone: (732) 345-2174 Email: rob.mahler@ansell.com
### Date Prepared:
### 12/8/2017
### Name of the Device:
| Trade Names: | Micro Touch Denta Glove Nitrile Hydrasoft Patient Examination Gloves | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Common Name: | Patient Examination Glove | |
| Classification Name: | Patient Examination Glove | |
| Classification Regulation: | 21 CFR 880.6250 | |
| Device Class: | 1 | |
| Product Code: | LZA | |
| Classification Panel: | General and Plastic Surgery | |
| Legally Marketed Predicate Device: | | |
| Company: | Wear Safe (Malaysia) Sdn Bhd | |
| Trade Name: | Powder Free Nitrile Patient Examination Glove, Blue Colored and White (Non-<br>colored), Non-sterile, Polymer Coated<br>Powder Free Nitrile Patient Examination Gloves, Blue Colored and white (Non-<br>colored), Non-sterile, without Polymer (Chlorinated) | |
| 510(k) Number: | K123469 | |
| Device Class: | Class I | |
| 510(k) Number: | K123469 |
|----------------|---------------------------------------------|
| Device Class: | Class I |
| Product Code: | LZA (Nitrile) |
| Device Name: | Patient Examination Glove (21 CFR 880.6250) |
# Reference Device:
| Company: | Ansell Healthcare LLC |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Encore Sterile Powder-Free Polymer Coated Latex Surgical Glove with Hydrasoft™<br>Coating (containing glycerine) and Protein Labeling Claim (50mg or less) |
| 510(k) Number: | K051793 |
| Device Class: | Class I |
| Product Code: | KGO |
| Device Name: | Surgeon's Glove (21 CFR 878.4460) |
{4}------------------------------------------------
### Device Description:
The MICRO-TOUCH® DENTA-GLOVE® Nitrile HydraSoft™ are non-sterile, single use only, disposable, powder free examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy. Hydrasoft coating (containing glycerine) is applied on top of the polymer coating on the glove inner surface..
Characteristic:
- Ambidextrous with beaded cuff and straight fingers
- Finger-textured,
- White colored
- Featuring inner coating of polyacrylic polymer coating and HydraSoft™ coating.
- . Five (5) sizes – extra-small, small, medium, large, and extra-large.
High levels of ozone will degrade rubber material of the glove, therefore the glove should be protected from ozone in particular.
The gloves are designed to meets the specifications of ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
### Indications for Use Statement:
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### Technological Characteristics:
Micro Touch Denta Glove with Hydrasoft Coating Nitrile Patient Examination Gloves have the following technological characteristics as compared to ASTM or equivalent standards:
| Characteristics | Standard/Test/ | Result Summary |
|------------------------------|----------------|----------------------------------------------|
| | FDA Guidance | |
| Physical Characteristics : | | |
| Dimensions: | ASTM D6319-10 | Meets ASTM D6319-10 requirements for length, |
| | | width and thickness |
| Length | Minimum 230mm | Minimum 240mm |
| Palm width (mm) | | |
| Size — XS | 70 ± 10 | 75 ± 5 |
| Size — S | 80 ± 10 | 85 + 5 |
| Size — M | 95 ± 10 | 95 + 5 |
| Size — L | 110± 10 | 105 + 5 |
| Size - XL | 120 ± 10 | 115 ± 5 |
| Thickness (mm) - single-wall | | |
| Finger | minimum 0.05 | Finger - 0.11 ± 0.03 |
| Palm | minimum 0.05 | Palm – 0.07 ± 0.02 |
| Cuff | | Cuff -- 0.06 ± 0.02 |
{5}------------------------------------------------
# Pre-market Notification: Traditional 510(k) Submission Micro-Touch Denta-Glove with HydraSoft Coating Nitrile Examination Glove
| Physical Properties: | ASTM D6319-10 | Meets ASTM D6319-10 requirements for tensile<br>strength and ultimate elongation before and after<br>accelerated aging: |
|-----------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Tensile Strength | | |
| Before Aging | minimum 14 MPa | minimum 17 MPa |
| After Aging | minimum 14 MPa | minimum 17 MPa |
| Ultimate Elongation | | |
| Before Aging | minimum 500% | minimum 500% |
| After Aging | minimum 400% | minimum 400% |
| Freedom from holes | ASTM D6319-10<br>ASTM D5151-06 | Meets ASTM D6319-10 and ASTM D5151-06<br>requirements of AQL 2.5 |
| Powder Residual | ASTM D6319-10<br>ASTM D6124-06 | Meets applicable requirement for powder free; ≤<br>2 mg per glove |
| Biocompatibility: | | |
| ISO In Vitro<br>Cytotoxicity | ISO 10993-5:2009 | Under the conditions of the study, undiluted and<br>1:2 dilution was cytotoxic. 1:4, 1:8, 1:16, 1:32 and<br>1:64 are not cytotoxic |
| ISO Skin Irritation<br>Study | ISO10993-10:2010 | Under the conditions of the study, not an irritant |
| ISO Maximization<br>Sensitization Study | ISO 10993-10:2010 | Under the conditions of the study, not a<br>sensitizer |
| ISO acute systemic<br>toxicity | ISO 10993-11: 2006 | Under the conditions of the study, no evidence of<br>systemic toxicity |
# Substantial Equivalence:
| Trade name | Predicate Device | Reference Device | Proposed Subject<br>Device | Substantial<br>Equivalence to<br>Predicate |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| | Powder Free Nitrile<br>Patient Examination<br>Glove, Blue Colored,<br>and White (Non-<br>colored). Non-sterile,<br>Polymer Coated | Encore Sterile Powder-<br>Free Polymer Coated<br>Latex Surgical Glove<br>with Hydrasoft Coating<br>(containing glycerine)<br>and Protein Labeling<br>Claim (50mg or less) | MICRO-TOUCH®<br>DENTA-GLOVE® Nitrile<br>HydraSoft™ Non-<br>Sterile Powder-Free<br>Examination Glove | Not applicable |
| 510k Number | K123469 | K051793 | Pending | Not Applicable |
| Product Owner | Wear Safe Malaysia | Ansell Healthcare | Ansell Healthcare | Ansell Healthcare |
| Product Code | LZA | KGO | LZA | Same |
| Regulation<br>Number | 21 CFR 880.6250 | 21 CFR 878.4460 | 21 CFR 880.6250 | Same |
| Regulatory Class | 1 | 1 | 1 | Same |
| Regulation Name | Patient Examination<br>Glove | Surgeon's Glove | Patient Examination<br>Glove | Same |
| | | | | Iselin, NJ 088- |
| Intended Uses | The patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's hand<br>to prevent<br>contamination between<br>patient and examiner | The surgical glove is<br>intended to be worn by<br>operating room<br>personnel to protect a<br>surgical wound from<br>contamination. The<br>latex glove contain 50<br>micrograms or less of<br>water extractable per<br>gram | The patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner | Same |
| Material<br>Composition | Synthetic nitrile rubber | Natural rubber latex | Synthetic nitrile rubber | Same |
| Coating | Polyacrylic polymer<br>inner coating to aid<br>donning | Polyurethane polymer<br>inner coating to aid<br>donning | Polyacrylic polymer<br>inner coating to aid<br>donning | Same |
| HydraSoft<br>Coating | N/A | HydraSoft™ Coating<br>coated on the donning<br>surface | HydraSoft™ Coating<br>coated on the donning<br>surface | Different from<br>predicate |
| Design | Non-sterile | Sterile | Non-sterile | Same |
| | Single use | Single use | Single use | Same |
| | Powder-free | Powder-free | Powder-free | Same |
| | Ambidextrous | Hand specific | Ambidextrous | Same |
| | Beaded cuff | Beaded cuff | Beaded cuff | Same |
| Color | White | White | White | Same |
| Shelf Life | 3 years | 3 years | 3 years | Same |
| Performance | | | | Same |
| a. Dimensions | Meets ASTM D6319-10<br>requirements | Meets ASTM 3577<br>requirements | Meets ASTM D6319-10<br>requirements | |
| b. Physical<br>Properties | Meets ASTM D6319-10<br>requirements | Meets ASTM 3577<br>requirements | Meets ASTM D6319-10<br>requirements | Same |
| c. Freedom from<br>holes | Meets ASTM D6319-10<br>requirements of GI, AQL<br>2.5 | Meets ASTM 3577<br>requirements<br>of GI, AQL 1.5 | Meets ASTM D6319-10<br>requirements of GI,<br>AQL 2.5 | Same |
| d. Powder<br>Residual | Meets ASTM D6319-10<br>requirements; Not<br>more than 2.0mg/glove | Meets ASTM 3577<br>requirements; Not more<br>than 2.0mg/glove | Meets ASTM D6319-10<br>requirements; Not<br>more than<br>2.0mg/glove | Same |
| e. Sterility | Non-sterile | Sterile | Non-sterile | Same |
| Biocompatibility | Passes Primary Skin<br>Irritation Test and<br>Dermal Sensitization<br>Test | Passes Primary Skin<br>Irritation Test and<br>Dermal Sensitization<br>Test | "Under the conditions<br>of the study, not an<br>irritant" and "Under<br>the conditions of the<br>study, not a sensitizer" | Same |
{6}------------------------------------------------
## Pre-market Notification: Traditional 510(k) Submission Micro-Touch Denta-Glove with HydraSoft Coating Nitrile Examination Glove
The MICRO-TOUCH® DENTA-GLOVE® Nitrile HydraSoft™ are non-sterile, single use only, disposable, powder free patient examination gloves. The glove is made of nitrile butadiene rubber. A polyacrylic polymer is applied to the inner surface of the glove to make donning easy. Hydrasoft coating glycerine) is applied on top of the polymer coating on the glove inner surface.
{7}------------------------------------------------
The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6319, ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization and irritation, in accord with the ISO 10993 series of standards.
### Performance Data:
A clinical study was not conducted on the subject or predicate devices.
# Substantial Equivalence Statement
The Micro-Touch Denta-Glove with HydraSoft Coating Nitrile Patient Examination Glove is as safe and effective as the predicate device with respect to design, technological characteristics, intended use and conformance to standard requirements.
# Conclusion
The conclusions drawn from the non clinical tests demonstrate that theMicro-Touch Denta-Glove with HydraSoft Coating Nitrile Patient Examination Glove is as safe and effective as the legally marketed predicate device previously cleared under K123469, Class I (21 CFR 880.6250, Product Code: LZA).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.