POWDER FREE NITRILE PATIENT EXAMINATION GLOVE, BLUE AND WHITE COLORED, NON-STERILE
K123469 · Wear Safe (Malaysia) Sdn Bhd - Facility II · LZA · Feb 19, 2013 · General Hospital
Device Facts
Record ID
K123469
Device Name
POWDER FREE NITRILE PATIENT EXAMINATION GLOVE, BLUE AND WHITE COLORED, NON-STERILE
Applicant
Wear Safe (Malaysia) Sdn Bhd - Facility II
Product Code
LZA · General Hospital
Decision Date
Feb 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable, ambidextrous, non-sterile nitrile examination gloves; available in blue or white; five sizes (XS-XL). Manufactured via polymer coating or chlorination. Used by healthcare professionals in clinical settings to provide a barrier against infectious materials and contaminants. Device is single-use; worn on hands during patient examinations. Benefits include reduction of cross-contamination risk between patient and examiner. No electronic or software components.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against ASTM D6319-10 (dimensions, physical properties, powder residue), ASTM D5151-06 (freedom from pinholes), and biocompatibility standards (ISO 10993-10, 16 CFR 1500).
Technological Characteristics
Nitrile rubber; ambidextrous; beaded cuff; non-sterile; polymer coated or chlorinated. Meets ASTM D6319-10, ASTM D5151-06, and ASTM D6124-06. Biocompatible per ISO 10993-10 and 16 CFR 1500.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner. No specific patient population or contraindications described.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination Glove, White (Non-Colored) and Blue (Colored), Non-Sterile (K093500)
Powder Free Nitrile Patient Examination Glove, Blue Colored and White (Non-Colored), Non-Sterile (K111772)
Submission Summary (Full Text)
{0}------------------------------------------------
### WSM WEAR SAFE (MALAYSIA) SDN BHD
(Company No. 204396-X)
# FDA 510(k), Premarket Notification: 510(k) Summary of Safety and Effectiveness Information
#### 1.0 Submitter:
# K423469
Wear Safe (Malaysia) Sdn Bhd - Facility II LOT 63616, PT 54924, NO 3, Jalan Korporat 7C/KU9, Taman Perindustrian Meru, Mukim Kapar, 42200 Klang, Selangor, Malaysia
| Telephone No.: | +603-3393 6088 |
|----------------|----------------|
| Fax No.: | +603-3393 3688 |
#### 2.0 Contact Person:
| Contact: | Mr Alex Yoong |
|----------------|----------------|
| Telephone No.: | +603-3393 6088 |
| Fax No .: | +603-3393 3688 |
#### 3.0 Preparation Date: 15 February 2013
#### 4.0 -Name of Device:
Trade Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non-Colored). Non-Sterile, Polymer Coated Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non-Colored). Non-Sterile, Without Polymer (Chlorinated) Common Name: Powder-Free Nitrile Patient Examination Glove Classification Name: Patient Examination Glove (21 CFR Part 880.6250, Product Code LZA)
#### 5.0 Identification of the Legally Marketed Device:
Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored), Non-Sterile; Class I Patient Examination Gloves, Nitrile-80LZA, meets all of the requirements of ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application.
Predicate Device 1: K093500, Powder Free Nitrile Examination Glove, White (Non-Colored) and Blue (Colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim).
Predicate Device 2: K111772, Powder Free Nitrile Patient Examination Glove, Blue Colored and White (Non-Colored), Non-Sterile.
{1}------------------------------------------------
# WSM WEAR SAFE (MALAYSIA) SDN BHD
(Company No. 204396-X)
There are no different technological characteristics compared to the Predicate Devices. They are all Powder Free Non-Sterile Nitrile Examination Gloves, one predicate polymer coating (K111772), and the other on-line chlorination (K093500).
#### 6.0 Description of Device:
Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored), Non-Sterile meets all of the requirements of ASTM D 6319-10. The gloves are ambidextrous single-use disposable devices that come in five sizes (XS, S, M, L, XL) in blue or white color.
#### Intended Use of the Device: 7.0
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
# 8.0 Summary of the Technological Characteristics of the Device:
Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored). Non-Sterile possesses the following technological characteristic (as compared to ASTM or equivalent standards):
| Characteristic | Standards | Device Performance |
|----------------------------|-----------------------------------------------------------------|-------------------------------|
| Dimensions | ASTM D 6319-10 | Meets |
| Physical Properties | ASTM D 6319-10 | Meets |
| Freedom from pin-<br>holes | ASTM D 5151-06 | Meets |
| | ASTM D 6319-10 | Meets |
| Powder Free Residue | ASTM D 6124-06 | Meets |
| | ASTM D 6319-10 | Meets |
| Biocompatibility | Dermal Sensitization<br>(as per ISO 10993-<br>10:2010) | Not a contact skin sensitizer |
| | Primary Skin<br>Irritation Test<br>(as per 16 CFR Part<br>1500) | Not a primary skin irritant |
### 9.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
The performance test data that support a determination of substantial equivalence are described above.
{2}------------------------------------------------
### WSM WEAR SAFE (MALAYSIA) SDN BHD
(Company No. 204396-X)
#### 10.0 Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
#### 11.0 Conclusion
Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored), Non-Sterile performs according to the glove performance standards referenced in Section 8.0 above. This device is substantially equivalent to current marketed devices per Section 12.0.
#### 12.0 Substantial Equivalence Comparison and Analysis
The gloves are identical to Predicate Device cleared under 510(k) K093500 with Proposed Device not tested for use with chemotherapy drugs. The difference with Chemotherapy Protection Labeling Claim in Predicate Device, does not affect the safety and effectiveness of the Proposed Device that is not tested for Chemotherapy Drugs used.
The gloves are also identical to Predicate Device cleared under 510 (k) K111772, both devices are not tested for use with chemotherapy drugs. Predicate Device is with Polymer Coated Labeling Claim, while Proposed Device is with Polymer Coated optional labeling claim.
There is difference in Blue colorant used in Proposed Device, compared with Predicate Devices. The difference Does Not affect the safety and effectiveness of the Proposed Device, as the Proposed Device Pass biocompatibility test, similar with Predicate Devices.
There is no difference between the Proposed Device and the Predicates with respect to indications for use and technological characteristics.
As such, this device is substantially equivalent to currently marketed devices. The Substantial Equivalence Comparison is summarized per Table below:-
{3}------------------------------------------------
WSM WEAR SAFE (MALAYSIA) SDN BHD
Substantial Equivalence Comparison Table
| Characteristics | Predicate Device 1 | Proposed Device | Predicate Device 2 | Proposed Device |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | K093500, Powder Free<br>Nitrile Examination Glove,<br>White (Non-Colored) and<br>Blue (Colored), Non-Sterile<br>(Chemotherapy Drug<br>Protection Labeling Claim)<br>of Wear Safe (Malaysia)<br>Sdn Bhd | Powder Free Nitrile<br>Patient Examination<br>Glove, Blue Colored,<br>and White (Non<br>Colored), Non-Sterile | K111772, Powder Free<br>Nitrile Patient<br>Examination Glove, Blue<br>Colored and White (Non-<br>Colored), Non-Sterile of<br>Kossan Latex Industries<br>(M) Sdn Bhd. | Powder Free Nitrile Patient<br>Examination Glove, Blue<br>Colored, and White (Non<br>Colored), Non-Sterile |
| Product Code | 80 LZC | 80 LZA | 80 LZA | Identical |
| Intended Use | Intended for medical purposes<br>that is worn on the examiner's<br>hand to prevent contamination<br>between patient and examiner. | Identical | Intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. | Identical |
| Design | Powder Free, Non Sterile,<br>Ambidextrous, Beaded Cuff | Identical | Powder Free, Non Sterile,<br>Ambidextrous, Beaded Cuff | Identical |
| Indications for Use | A powder free nitrile<br>examination glove is a<br>disposable device made of<br>synthetic material intended to<br>be worn on the hand for<br>medical purposes to provide<br>barrier against potentially<br>infectious materials and other<br>contaminants.<br>In addition, this product is<br>tested for use with<br>chemotherapy drugs | Identical | A patient examination glove<br>is a disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. | Identical |
| Characteristics | Predicate Device 1<br>K093500, Powder Free Nitrile Examination Glove, White (Non-Colored) and Blue (Colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim) of Wear Safe (Malaysia) Sdn Bhd | Predicate Device 2<br>K111772, Powder Free Nitrile Patient Examination Glove, Blue Colored and White (Non-Colored), Non-Sterile of Kossan Latex Industries (M) Sdn Bhd. | Proposed Device<br>Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored), Non-Sterile | |
| Construction | Ambidextrous, Chlorinated, Powder Free Nitrile. | Ambidextrous, Polymer Coated, Powder Free Nitrile. | Ambidextrous, Polymer Coated or Chlorinated, Powder Free. | |
| Color Description | Blue and White | Blue and White | Identical | |
| Materials | Nitrile | Nitrile | Identical | |
| Performance<br>I. Sterility | Non-Sterile | Non-Sterile | Identical | |
| II. Freedom from holes | Meets ASTM D6319 | Meets ASTM D6319 | Identical | |
| III. Dimension | Meets ASTM D6319 | Meets ASTM D6319 | Identical | |
| IV. Physical Properties | Meets ASTM D6319 | Meets ASTM D6319 | Identical | |
| V. Powder Free Residue | Meets ASTM D6319 | Meets ASTM D6319 | Identical | |
| Single Use | Yes | Yes | Identical | |
| Biocompatibility Test | Passes<br>i. Primary Skin Irritation Test<br>ii. Dermal Sensitization Test | Passes<br>i. Primary Skin Irritation Test<br>ii. Dermal Sensitization Test | Identical | |
| Characteristics | Predicate Device 1<br>K093500, Powder Free Nitrile<br>Examination Glove, White (Non-Colored) and<br>Blue (Colored), Non-Sterile (Chemotherapy<br>Drug Protection Labeling Claim)<br>of Wear Safe (Malaysia) Sdn Bhd | Proposed Device<br>Powder Free Nitrile<br>Patient Examination<br>Glove, Blue Colored, and<br>White (Non Colored), Non-Sterile | Predicate Device 2<br>K111772, Powder Free Nitrile<br>Patient Examination Glove, Blue<br>Colored and White (Non-Colored), Non-Sterile of<br>Kossan Latex Industries<br>(M) Sdn Bhd. | Proposed Device<br>Powder Free Nitrile Patient<br>Examination Glove, Blue<br>Colored, and White (Non<br>Colored), Non-Sterile |
| Packaging | Packed in Dispenser Boxes | Identical | Packed in Dispenser Boxes | Identical |
| Labeling Claim | Tested For Use with<br>Chemotherapy Drugs<br>Labeling Claim | Not Tested For Use with<br>Chemotherapy Drugs | i. NO Chemotherapy<br>Drugs Labeling Claim<br>ii. Polymer Coated<br>Labeling Claim | Optional Polymer Coated<br>Labeling Claim |
{4}------------------------------------------------
WSM WEAR SAFE (MALAYSIA) SDN BHD
Comment
Page 5 of 6 (Rev 4)
{5}------------------------------------------------
WSM WEAR SAFE (MALAYSIA) SDN BHD
(00000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
·
Page 6 of 6 (Rev 4)
·
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of stylized lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 19, 2013
Mr. Alex Yoong Quality Assurance Manager . Wear Safe (Malaysia) Sdn Bhd - Facility II Lot 63616. PT 54924, NO 3, Jalan Korporat 7C/KU9 Taman Perindustrian Meru Mukim Kapar, Klang Selangor, Malaysia 42200
Re: K123469
Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Colored and White (Non-Colored), Non-Sterile, Polymer Coated Powder Free Nitrile Patient Examination Gloves, Blue Colored and White (Non-Colored), Non-Sterile, Without Polymer (Chlorinated)
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 2, 2013 Received: January 10, 2013
Dear Mr. Yoong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{7}------------------------------------------------
Page 2 - Mr. Yoong
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you aco.il/www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
hori
Theti
Fawshukio.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
K123469
### 510(k) Number (if known):
Device Name:
Powder Free Nitrile Patient Examination Gloves, Blue Colored and White (Non-Colored), Non-Sterile, Polymer Coated Powder Free Nitrile Patient Examination Gloves. Blue Colored and White (Non-Colored), Non-Sterile, Without Polymer (Chlorinated)
## Indications for Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
| Elizabeth-F.Claverie |
|-----------------------------|
| 2013.02.15 22:09:47 -05'00' |
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: *K123469*
Page 1 of _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.