Monteris Medical NeuroBlate System

K171255 · Monteris Medical, Inc. · GEX · Jul 25, 2017 · General, Plastic Surgery

Device Facts

Record IDK171255
Device NameMonteris Medical NeuroBlate System
ApplicantMonteris Medical, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 25, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through internal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images. When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis.

Device Story

NeuroBlate System provides MRI-guided laser interstitial thermal therapy (LITT) for neurosurgical procedures. System inputs include MRI images used for trajectory planning and real-time thermographic monitoring. Components include gas-cooled laser delivery probes (SideFire/FullFire), probe drivers (advanced/robotic) for precise positioning, interface platform, electronics rack, and control workstation with M-Vision/M-Vision Pro software. Used in clinical settings by neurosurgeons to ablate intracranial soft tissue (e.g., tumors, epileptic foci). System provides visual feedback via MRI-based thermography; physician interprets output to guide thermal therapy. Benefits include minimally invasive access to deep brain structures. Patient management decisions require physician oversight and should not rely solely on system analysis.

Clinical Evidence

No clinical data. Substantial equivalence supported by labeling clarification and reference to previously provided in-vitro (bench) and in-vivo (animal) data from predicate submissions.

Technological Characteristics

MRI-compatible laser surgical system. Uses 1064 nm laser energy for thermal ablation. Components include gas-cooled laser delivery probes, robotic/advanced probe drivers, and MRI-compatible stereotaxic fixation hardware. Software-based planning and real-time thermographic analysis. System is MRI-conditional.

Indications for Use

Indicated for ablation, necrosis, or coagulation of intracranial soft tissue, including brain structures, in neurosurgery using 1064 nm lasers. Intended for use by trained physicians for thermal therapy planning and monitoring under MRI guidance.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a way that they resemble a bird in flight. The profiles are black, and they are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 25, 2017 Monteris Medical, Inc. David Mueller Principal Regulatory Affairs Specialist 14755 27th Avenue North: Suite C Plymouth, Minnesota 55447 Re: K171255 Trade/Device Name: Monteris Medical NeuroBlate System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, HAW Dated: April 27, 2017 Received: April 28, 2017 Dear David Mueller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Michael J. Hoffmann -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171255 Device Name Monteris Medical NeuroBlate System #### Indications for Use (Describe) The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through internal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images. When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------| | <span> <span style="font-size:100%;"> <b> </b> </span> <span style="font-size:100%;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> | | <span> <span style="font-size:100%;"> <b> </b> </span> <span style="font-size:100%;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K171255 Section 5: 510(k) Summary ## a. Device Information: | Category | Comments | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | Monteris Medical<br>14755 27th Avenue North<br>Suite C<br>Plymouth, MN 55446<br>763-253-4710<br>Fax: 763-746-0084<br>www.monteris.com | | Correspondent Contact Information: | David H. Mueller<br>Principal Regulatory Affairs Specialist<br>Monteris Medical<br>TEL: 763-253-2732<br>FAX: 763-746-0084<br>Email: DMueller@Monteris.com | | Device Common Name: | Magnetic Resonance Image Guided Laser<br>Thermal Therapy System | | Device Classification Number: | 21 CFR 878.4810<br>Laser surgical instrument for use in general and<br>plastic surgery and in dermatology<br><br>21 CFR 882.4560<br>Stereotaxic instrument | | Device Classification &<br>Product Code: | Class II, GEX<br>Class II, HAW | | Device Proprietary Name: | Monteris Medical NeuroBlate™ System | ### Predicate Device Information: | Manufacturer | Monteris Medical | |-----------------------------|---------------------------------------------------------------------------------------------------------------------| | Commercial<br>Name | NeuroBlate System | | Common<br>Name: | Laser surgical instrument for<br>use in general and plastic<br>surgery and in dermatology<br>Stereotaxic instrument | | Premarket<br>Notification # | K162762, K170724 | | Regulation | 21 CFR 878.4810<br>21 CFR 882.4560 | | Class/ Product<br>Code | Class II<br>GEX, HAW | # b. Date Summary Prepared July 21, 2017 {4}------------------------------------------------ #### c. Description of Device The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy that is currently indicated for the practice of neurosurgery. As previously described in K162762, K170724, the NeuroBlate System is typically used for the minimally invasive ablation of neurosurgeon identified target tissue (tumors, epileptic foci) in the brain. The NeuroBlate System components consist of: - Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver . controlled energy to a target zone. - . Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone. - . An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes; - . A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation. - . A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems. The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as: - . The AXiiiS stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and - The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools. #### Change Description Monteris seeks clearance for minor modifications to the indications for use' statement to improve clarity and ease of understanding in alignment with the current intended use13, the proposed changed text will read: The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through interstitial irradiation <sup>13</sup> United States Food and Drug Administration, the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], Guidance for Industry and Food and Drug Administration Staff. Washington, D.C .: FDA, 2014. http://www.fda.gov/downloads/MedicalDevices/.../UCM284443.pdf {5}------------------------------------------------ or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The clarified indications were determined to not create a new intended use or raise different questions of safety and effectiveness, and were also determined to not introduce any new or different technical characteristics which could raise different questions of safety and effectiveness, i.e., the clarified indications are more clearly consistent with the existing intended use as well as the exiting indications for use. This 510(k) submission meets the Traditional 510(k) criteria as the proposed changes involve a modification to the language of the indications for use to clarify the current indications. Additionally, the proposed clarified indications were evaluated within the FDA General-Specific Guidance's 44 criteria which determined that the proposed labeling modifications are to provide clarification for the terminology utilized within the existing indications as the device is currently utilized and intended. The proposed changes are not intended to provide specific indications for use which are different than the current language. The proposed clarified indications were also evaluated utilizing several other FDA Guidance Documents 15,16,17,18, which concluded that the modified Indications for Use clarify the terminology utilized within the existing indications as the device is currently used via its Intended use. The proposed changes are not intended to provide specific indications for use which are different than the current language. The modified NeuroBlate has the same intended use and similar indications, and identical technological characteristics and principles of operation as the cleared NeuroBlate System. As explained in the attached 510(k) notice, the minor differences between the subject NeuroBlate and the predicate device do not raise any new questions of safety or effectiveness. Thus, we believe that the modified NeuroBlate System is substantially equivalent. #### d. Indications for Use The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. 15 United States Food and Drug Administration, Deciding When to Submit a 510(k) for a Change to an Existing Device; Draft Guidance for Industry and Food and Drug Administration Staff. Washington, D.C.: FDA, 2016. 16 United States Food and Drug Administration, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], Guidance for Industry and Food and Drug Administration Staff. Washington, D.C.: FDA, 2014. http://www.fda.gov/downloads/MedicalDevices/.../UCM284443.pdf <sup>14</sup> United States Food and Drug Administration, Guidance for Industry and Food and Drug Administration Staff, General/Specific Intended Use. Washington, D.C .: FDA, 1998 <sup>17</sup> United States Food and Drug Administration. Determination of Intended use for 5100k) Devices: Guidance for CDRH Staff (Update to K98-1). Washington, D.C.: FDA, 2002. http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm082166.p df <sup>18</sup> United States Food and Drug Administration, Device Labeling Guidance #G91-1. Washington, D.C .: FDA, 1991 {6}------------------------------------------------ The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real-time thermographic analysis of selected MRI images. When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis. ### e. Comparison to Predicate Device The application for the Monteris Medical NeuroBlate™ System with the modified labeling is substantially equivalent (and/ or identical) to the predicate Monteris NeuroBlate™ System (K162762, K170724) in intended use, indications for use, technology, design and physician use. This submission proposes no physical changes to the NeuroBlate System hardware, software, accessories, manufacturing, design or the devices' specific step-by-step directions for use. This submission's intent is not to change the current indication or intended use. but to improve the indication's clarity, ease of understanding, and consistency with the intended use. The indications for use are being clarified to ensure alignment and conformance with the currently cleared NeuroBlate System's intended use. This proposed change clarifies that the existing NeuroBlate (K162762) indications for use are in reference to the ablation of soft "brain" tissue, i.e., - Current K162762. K170724 Indication (excerpt) states. . . to ablate "soft tissue" in "in the . discipline of neurosurgery" while utilizing "stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes". . . - Based on the K162762, K170724 provided Directions for Use, both textual statements ● and graphics demonstrate that the existing NeuroBlate System's intended use is to ablate brain tissue, i.e., brain tissue ablation refers to the ablation of "intracranial soft tissue," including "brain structures". This clarification is based on the considerations that: - . Previous FDA cleared NeuroBlate System 510(k) submissions (e.g., K162762. K170724) demonstrate the device's intended use, based upon the submission contents and the associated proposed labeling and directions for use 19 is the ablation of brain tissue. - . The K162762. K170724 submitted NeuroBlate System content. labeling and directions for use included both textual and graphic brain tissue and skull specific {7}------------------------------------------------ directions for use10 demonstrating the device's intended use1920 is brain tissue ablation. - As "brain soft tissue" is really the "matter" or "tissue" inside the patient's skull (or . cranium), the term "intracranial" is the appropriate medical terminology to describe the "soft tissue" to be ablated, i.e., "intracranial soft tissue". - . When physicians desire to identify specific "intracranial soft tissue" anatomical regions or areas within the brain, medical naming conventions divide the entire human brain into various areas or regions. While all these identified regions are "intracranial soft tissue", the generalized naming terminology for these intracranial soft tissue regions is "brain structures". Thus, as the NeuroBlate System is intended and indicated to ablate soft brain tissue, utilizing both "intracranial soft tissue" and "brain structure" naming conventions is appropriate in order to provide clarity and consistency between the Indications for Use and the Intended use. The proposed labeling modifications: - Do not change the intended use, i.e., they only clarify the indications for use within the ● existing intended use, - . Do not raise new or different questions of safety and effectiveness, - . Do not introduce any new or different technical characteristics that raise different questions of safety and effectiveness, - Present no change to the NeuroBlate System's operating principles, design, . manufacturing, function or specific directions for use. It can be concluded this NeuroBlate™ System application with the modified labeling would meet FDA's "non-significant change" criteria for which "changes to the device name or description that are consistent with the cleared indications for use and changes to improve readability or clarity that do not affect the substance of the indications for use"(21), i.e., a new 510(k) would not be necessary. However, having previously observed inconsistency in interpreting this guidance in coordination with general to specific use questions, a conservative approach was utilized and a new 510k submitted to demonstrate that the clarified indications are substantially equivalent to the described predicate devices. The proposed clarified indications were also assessed utilizing the FDA General-to-Specific Guidance22 with the resulting conclusion being that the proposed Indications for use <sup>19</sup> United States Food and Drug Administration, the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], Guidance for Industry and Food and Drug Administration Staff. Washington, D.C.: FDA, 2014. http://www.fda.gov/downloads/MedicalDevices/.../UCM284443.pdf <sup>20</sup> United States Food and Drug Administration, Determination of Intended use for 510(k) Devices; Guidance for CDRH Staff (Update to K98-1). Washington, D.C.: FDA, 2002. http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/vcm082166.p df <sup>21</sup> United States Food and Drug Administration. K97-1 Deciding When to Submit a 510(k) for a Change to an Existing Device. Washington, D.C .: FDA, 1997. <sup>22</sup> United States Food and Drug Administration, Guidance for Industry and Food and Drug Administration Staff, General/Specific Intended use. Washington, D.C .: FDA, 1998 {8}------------------------------------------------ modifications are to provide clarification for the existing terminology utilized within the existing indications as the device is currently utilized. The proposed changes are not intended to provide specific indications for use which are different than the current language. The application for the NeuroBlate™ System with the modified labeling is substantially equivalent, and/ or identical, to the predicate Monteris NeuroBlate™ System (K162762, K170724) in intended use, technology, design and physician use Thus, the predicate terminology describing brain tissue and brain structure ablation are all synonymous (identical) or equivalent, in intended use, technology, design and physician use. Additionally, this submission proposes no manufacturing changes, process changes, materials changes, or technology proposed changes to the NeuroBlate System. The technical modes of action and technical principles remain the same as the predicate devices. The fundamental functionality and technical characteristics of the proposed modified device is identical and substantially equivalent to the existing NeuroBlate System (K162762, K170724). The conclusion is that the difference in technological characteristics between proposed modified device and the current NeuroBlate System (K162762, K170724) do not raise new or different questions of safety and effectiveness, i.e., the proposed modified device is as safe as the described predicate devices. #### f. Summary of Supporting Data Considering the proposed changes are labeling clarifications, the previously provided in-vitro (bench), and in-vivo (animal) data remains applicable, and are incorporated by reference. Given that the modifications to the indications for use do not create a new intended use or raise new or different questions of safety or efficacy, clinical data are not necessary to demonstrate substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...