K162762 · Monteris Medical, Inc. · GEX · Oct 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K162762
Device Name
NeuroBlate System
Applicant
Monteris Medical, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 26, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images. When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.
Device Story
NeuroBlate System provides MRI-guided laser interstitial thermal therapy (LITT) for neurosurgical procedures, such as tumor or epileptic foci ablation. System inputs include MRI images; processed via M-Vision and M-Vision Pro software for trajectory planning and real-time thermographic analysis. Output consists of visual guidance for probe placement and thermal monitoring. Operated by neurosurgeons in an MRI suite; system includes gas-cooled laser delivery probes (SideFire/FullFire), probe drivers (robotic/advanced), interface platform, and electronics rack. Physician interprets thermographic data to guide thermal therapy; aids in precision tissue ablation. Benefits include minimally invasive access to intracranial targets.
Clinical Evidence
No clinical data. Bench testing from previous submissions remains applicable as no physical or technological changes were made.
Technological Characteristics
MRI-compatible laser surgical system; 1064 nm laser source. Components: gas-cooled laser delivery probes, robotic/advanced probe drivers, interface platform, and control workstation. Connectivity: interfaces with MRI systems. Software: M-Vision and M-Vision Pro for planning and thermographic analysis. Sterilization: not specified (components are MRI-compatible).
Indications for Use
Indicated for ablation, necrosis, or coagulation of soft tissue in neurosurgery using 1064 nm lasers; intended for use in patients requiring interstitial thermal therapy under MRI guidance.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a ribbon-like design representing the snakes.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 26, 2016
Monteris Medical, Inc. David Mueller Principal Regulatory Affairs Specialist 14755 27th Avenue North; Suite C Plymouth, Minnesota 55447
Re: K162762
Trade/Device Name: Neuroblate System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, HAW Dated: September 29, 2016 Received: September 30, 2016
Dear David Mueller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Monteris Medical. The logo features the word "MONTERIS" in a large, bold, blue font, with the word "MEDICAL" in a smaller, less bold, blue font underneath. Above the word "MONTERIS" are several teal-colored circles arranged in an arc shape.
### Section 4: Indications for Use
510(k) Number (if known): K162762
Device Name: Monteris Medical NeuroBlate™ System
#### Indications for Use:
The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.
The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.
When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.
Prescription Use x AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font below it. Above the word "MONTERIS" are several teal-colored circles arranged in an arc shape.
# Section 5: 510(k) Summary
# a. Device Information:
| Category | Comments |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Monteris Medical Corp.<br>14755 27th Avenue North<br>Suite C<br>Plymouth, MN 55446<br>763-253-4710<br>Fax: 763-746-0084<br>www.monteris.com |
| Correspondent Contact<br>Information: | David H. Mueller<br>Principal Regulatory Affairs Specialist<br>Monteris Medical, Inc<br>TEL: 763-253-2732<br>FAX: 763-746-0084<br>Email: DMueller@Monteris.com |
| Device Common Name: | Magnetic Resonance Image Guided Laser Thermal<br>Therapy System |
| Device Classification Number: | 21 CFR 878.4810<br>Laser surgical instrument for use in general and plastic<br>surgery and in dermatology<br><br>21 CFR 882.4560<br>Stereotaxic instrument |
| Device Classification &<br>Product Code: | Class II, GEX<br>Class II, HAW |
| Device Proprietary Name: | Monteris Medical NeuroBlate™ System |
#### Predicate Device Information:
| Predicate Device: | NeuroBlate™ System |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Predicate Device Manufacturer: | Monteris Medical |
| Predicate Device Common Name: | Monteris NeuroBlate™ System |
| Predicate Device Premarket Notification # | K120561, K141983, K14347 |
| Predicate Device Regulation: | 21 CFR 878.4810<br>Laser surgical instrument for use in general and<br>plastic surgery and in dermatology |
| | 21 CFR 882.4560 |
| | Stereotaxic instrument |
| | Predicate Device Classification &<br>Product Code: |
# b. Date Summary Prepared
29 Sept 2016
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Monteris Medical. The logo features the company name in a bold, sans-serif font, with "MONTERIS" in a larger font size above the word "MEDICAL". Above the company name, there is a semi-circular arrangement of teal-colored circles, resembling a stylized representation of water droplets or molecules.
#### c. Description of Device
The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.
As previously described in K143457, the NeuroBlate System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.
The NeuroBlate System components consist of:
- Families of gas-cooled Laser Delivery Probe (SideFire & FullFire) to deliver ● controlled energy to a target zone.
- Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon . to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
- . An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes;
- A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation.
- . A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.
The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:
- The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation ● components, and
- The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.
#### d. Indications for Use
The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.
The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in a bold, blue sans-serif font, with the word "MEDICAL" in a smaller font size directly below it. Above the word "MONTERIS" are several teal-colored circles of varying sizes, arranged in a semi-circular pattern.
stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real-time thermographic analysis of selected MRI images.
When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.
#### e. Comparison to Predicate Device
This submission proposes no physical changes to the NeuroBlate System hardware, software, or accessories. The indications for use within the NeuroBlate Systems' Instructions for Use (IFU) have not been changed, however, the product labeling has been modified and corrected as outcomes from the Class 1 Recall's corrective actions as well as from Monteris' response to FDA 483's observation. This submission also provides a listing of historical non-significant changes which have occurred.
The labeling changes include removing references to the discontinued (recalled) 2.2mm SideFire Probes (SFS) as well as providing clarified (or duplicate) warnings and precautions to more easily verify compliance to 21 CFR 1000 -1050 Radiological Health requirements as well as FDA Laser Notice #50 requirements.
Additionally, there are no manufacturing changes, process changes, materials changes, or technology proposed changes to the NeuroBlate System. The technical modes of action and technical principles remain the same as the predicate devices. Considering the proposed changes are labeling related, the previously provided in-vitro (bench) data remains applicable and no additional bench testing was required.
The application for the Monteris Medical NeuroBlate™ System with the modified labeling is substantially equivalent to the predicate Monteris NeuroBlate™ System (K120561, K141983, K143457) in intended use, technology, design and physician use.
As the modifications presented in the current device do not change the intended use, operating principles, or raise any unaddressed safety concerns, it can be concluded the application NeuroBlate™ System with the modified labeling is substantially equivalent to the predicate NeuroBlate™ System.
#### f. Summary of Supporting Data
This submission proposes no physical changes, manufacturing changes, process changes, materials changes, technology changes to the NeuroBlate System. Considering the proposed changes are labeling, the previously provided in-vitro (bench) data remains applicable and no additional bench testing was required.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Monteris Medical. The logo consists of the word "MONTERIS" in a bold, blue sans-serif font, with the word "MEDICAL" in a smaller font below it. Above the word "MONTERIS" are two clusters of teal-colored circles, arranged in a symmetrical pattern.
Reviewing the specific individual 21 CFR 1000 - 1050 requirements, demonstrates that the proposed labeling updates comply with these requirements and are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.