K171245 · Christie Medical Holdings, Inc. · MUD · Aug 9, 2018 · Cardiovascular
Device Facts
Record ID
K171245
Device Name
Presygen/si-1
Applicant
Christie Medical Holdings, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Aug 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).
Device Story
Presygen™/si-1 is a non-contact, multispectral optical imaging oximeter; uses four NIR wavelengths (740, 780, 850, 945 nm) to measure StO2 in superficial tissue based on diffuse light reflection. Device includes imaging head, base with wheels, touchscreen monitor, and keyboard. Operated by healthcare professionals in hospital settings. Software performs spectral analysis, compensates for skin melanin levels, and applies Small Motion Correction (SMC) to track regions of interest. Produces 2D color-coded StO2 maps and numeric data on a touchscreen. Used to monitor relative temporal changes in tissue oxygenation; aids clinical decision-making by providing visual and quantitative feedback on tissue perfusion. Includes a sterile single-use drape accessory for contamination control.
Clinical Evidence
Two prospective, unblinded clinical studies (n=55 and n=20) compared Presygen™/si-1 against Kent Camera and T.Ox™ using Vascular Occlusion Tests (VOT). Primary endpoints included Bland-Altman analysis and Deming regression of StO2 measurements. Results showed consistent directional temporal changes and equivalence in StO2 measurement across Fitzpatrick Skin Types I-V. No SAEs or AEs reported.
Technological Characteristics
Multispectral optical imaging; NIR light source (LEDs: 740, 780, 850, 945 nm); 515 nm visible laser for focusing. Mains powered (100-240 Vac). Non-contact imaging head. Software includes melanin calibration and Small Motion Correction (SMC). Class 1 laser product (IEC 60825-07). Photobiological safety Low Risk Group (IEC 62471-06).
Indications for Use
Indicated for adults (22+) in hospital facilities for non-invasive, non-diagnostic measurement of relative temporal variation in superficial tissue oxygen saturation (StO2). Used by healthcare professionals alongside clinical assessment.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
August 9, 2018
CHRISTIE MEDICAL HOLDINGS, Inc. % Diane Horwitz Regulatory Consultant Mandell Horwitz Consultants LLC 2995 Steven Martin Drive Fairfax, Virginia 20031
## Re: K171245
Trade/Device Name: Presygen™/si-1 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: July 11, 2018 Received: July 12, 2018
## Dear Diane Horwitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M & Tillelrem
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K171245
Device Name PresygenTM/si-1
### Indications for Use (Describe)
Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays two-dimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of small squares in shades of green. The plus sign is slightly tilted.
## 510(k) SUMMARY
### 1. GENERAL INFORMATION
### Submitter and 510(k) Owner 1.1
CHRISTIE MEDICAL HOLDINGS, INC. 3175 Lenox Park Boulevard, Suite 200 Memphis, TN 38115
### 1.2 Official Correspondent
Salwa Rezk, Director, Quality & Regulatory
Telephone: 519-741-3794 Fax: 901-721-0350 Email: salwa.rezk(@christiedigital.com
### 1.3 Date of Preparation
August 9, 2018
### NAME OF THE DEVICE 2.
### 2.1.1 Trade/Proprietary Name, 510(k) Number
Presygen™/si-1, K171245
### 2.1.2 Common/Usual Name
Tissue Saturation Oximeter
### 2.1.3 Classification Information
| Classification Name: | Oximeter |
|----------------------------|-----------------|
| Classification Regulation: | 21 CFR 870.2700 |
| Class: | 2 |
| Product Code: | MUD |
| Panel: | Cardiovascular |
### 3. PREDICATE DEVICE AND REFERENCE DEVICE
The predicate device is Kent Camera (Kent Imaging, Inc., K113507). The reference device is T.OxTM (Vioptix, Inc., K042657).
### DESCRIPTION OF THE DEVICE 4.
Presygen™/si-1 is a standalone, non-contact tissue oxygen saturation oximeter based on multispectral optical imaging techniques. The mode of action is to transmit lights at four (4) known
1
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word "CHRISTIE" is a plus sign made up of small green squares. The squares are arranged in a grid pattern, with the color of the squares gradually changing from light green to dark green.
wavelengths through blood in superficial tissue and measure the tissue oxygen saturation (StO2), based on the amount of diffusely reflected light.
The Presygen™/si-1 software provides user control of the device and access to stored data, through a Graphical User Interface (GUI). The software controls hardware elements of the device such as illuminators, sensors, display and Input/Output, and manages communication between these elements and between processors and memory. The software also performs calculations to derive Tissue Oxygen Saturation (StO2) from optical inputs and display this to the device user in a two-dimension (2D) spatial map.
The Presygen™/si-1 device has been designed with the additional capability of compensating for melanin levels in the skin, since absorption of light by melanin affects the measurement of the oxygenated hemoglobin and deoxygenated hemoglobin through the skin.
The device consists of the following:
- Imaging head: Contains optics, lasers, and electronics to display the %StO2 .
- . Base compartment: Includes the wheels, brakes mechanism, SMB (Smart Multi-Media Board), power switch, certified power supply, fans, filters and cabling
- . Touchscreen monitor display: User interface, display of image, and patient data
- Keyboard: Used to enter user or patient information into the device .
- Accessory: The Presygen™/si-1 Sterilized Drape is a prepackaged single-use third party . drape that is used with Presygen™/si-1 to avoid contamination during the imaging process.
### INDICATIONS FOR USE 5.
The indication for use statement for Presygen™/si-1 is shown below.
Presygen™/si-1 is indicated for use in hospital facilities, along with clinical assessment of signs and symptoms, by healthcare professionals for non-invasive and non-diagnostic measure of relative temporal variation in oxygen saturation (StO2) in superficial tissue. Presygen™/si-1 displays twodimensional color-coded images of tissue oxygenation of the scanned surface and reports tissue oxygenation measurements in selected tissue regions. Presygen™/si-1 is intended for adults only (22 and over).
### INDICATIONS FOR USE COMPARED TO THE PREDICATE 6.
The indications for use statement for Presygen™/si-1 system is similar to the predicate device. The devices share the same target patient population, the same users and conditions of use. Both devices are Rx-only.
| Subject Device | Predicate: K113507 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PresygenTM/si-1<br>(Christie Medical Holdings, Inc.) | Kent Camera<br>(Kent Imaging Inc.) |
| PresygenTM/si-1 is indicated for use in hospital<br>facilities, along with clinical assessment of signs<br>and symptoms, by healthcare professionals for non-<br>invasive and non-diagnostic measure of relative<br>temporal variation in oxygen saturation (StO2) in | The Kent Camera is intended for use by healthcare<br>professionals as a non-invasive tissue oxygenation<br>measurement system that reports an approximate<br>value of:<br>• oxygen saturation (StO2)j, |
### Table 1. Indications for Use Comparison for Presygen™/si-1 Versus the Predicate
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Christie. The logo is blue and green. The word "CHRISTIE" is in blue, and there is a green plus sign to the right of the word.
| Subject Device | Predicate: K113507 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Presygen™/si-1 | Kent Camera |
| (Christie Medical Holdings, Inc.) | (Kent Imaging Inc.) |
| superficial tissue. Presygen™/si-1 displays two-<br>dimensional color-coded images of tissue<br>oxygenation of the scanned surface and reports<br>tissue oxygenation measurements in selected tissue<br>regions. Presygen™/si-1 is intended for adults only<br>(22 and over). | • oxyhemoglobin level (HbO2), and<br>• deoxyhemoglobin (Hb) level<br>in superficial tissue. The Kent Camera displays two-<br>dimensional color-coded images of tissue oxygenation<br>of the scanned surface and reports multispectral tissue<br>oxygenation measurements for selected tissue regions.<br>The Kent Camera is indicated for use to determine<br>oxygenation levels in superficial tissues for patients |
# 7. TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE
A comparison of the technological features between Presygen™/si-1 device and the predicate is shown in Table 1 below.
| Technological<br>Characteristic | Subject Device<br>Presygen™/si-1<br>(Christie Medical Holdings, Inc.) | Predicate: K113507<br>Kent Camera<br>(Kent Imaging Inc.) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Measured Parameters | • StO2 | • StO2<br>• HbO2<br>• Hb |
| Operating principle | Spectral Analysis at specific<br>wavelengths of light returned from target<br>tissue. | Same |
| Patient Contact | Sensors in imaging head do not touch<br>patient skin. | Same |
| Output Display | Two-dimensional color-coded map<br>(image) of relative % oxygen saturation<br>(%StO2).<br>Numeric data output available. | Same |
| Light Energy<br>Delivered | Multiple wavelengths of Near-infrared<br>light (NIR)<br>Source: Light emitting diodes (LEDs) | Same |
| Energy delivered | Wavelengths: 740, 780, 850, 945 nm<br>Visible lasers for properly focusing<br>device (wavelength 515 nm) | Wavelengths: 670, 735, 890, 940 nm<br>Visible lasers for properly focusing<br>device (wavelength details not<br>available) |
| Power source | Mains powered<br>Rating: 100-240 Vac, 50-60Hz, 198 VA | Mains powered<br>Rating: 90-240 Vac, 50-60Hz, 150<br>Watts |
| Output to user | Color-coded %StO2 map. User capture<br>images for future reference. | Same |
Table 2. Technology Comparison for Presygen™/si-1 Versus the Predicate
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "CHRISTIE" in blue, with a stylized "K" that looks like two triangles. To the right of the word is a plus sign made up of small squares in shades of green. The logo is simple and modern, with a focus on the company name and a visual element that suggests growth or addition.
| Technological<br>Characteristic | Subject Device<br>PresygenTM/si-1<br>(Christie Medical Holdings, Inc.) | Predicate: K113507<br>Kent Camera<br>(Kent Imaging Inc.) |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Monitor view | Immediate live view of the image during<br>a surgical procedure using the touch-<br>screen monitor | No immediate view of the image<br>during an operation in the touch-<br>screen monitor |
| SMC (Small Motion<br>Correction) module in<br>the software | Presygen™/si-1 system uses SMC<br>(Small Motion Correction) to track ROI<br>(regions of interest) using spatial<br>registration to compensate for up to 23 x<br>23 mm translation (X and Y axis) for<br>small motions of the imaging head or<br>subject (e.g., during breathing) | Feature not available. Kent camera<br>acts as a still-image camera. |
| Melanin Calibration | Prior to each session, a melanin<br>calibration is performed for planned<br>imaging areas in the subject to<br>compensate for skin melanin levels in<br>tissue | No calibration for melanin is<br>performed.<br>Note that the ViOptix ODISsey Tissue<br>Oximeter (K042657) is a 510(k)-<br>cleared device with measurements that<br>are independent of melanin<br>concentration. This device is<br>considered a Reference Device. |
| ROI Plot | "Region of Interest" Plot: The user has<br>the option of both saving and displaying<br>StO2 levels from user-selected regions of<br>interest over a maximum of 5 minutes in<br>the form of a line graph | Feature not available |
| Presygen™/si-1<br>Sterilized Drape<br>Accessory | Yes. A sterile single-use drape is<br>obtained sterilized and pre-packaged<br>from a third-party vendor and is used<br>during imaging. | Kent Camera does not have a sterile<br>drape |
### Similarities and Differences in Technology Comparison 7.1
The fundamental technology of the Presygen™/si-1 device is the same as the predicate device: The measured parameters, operating principle, patient contact, output display, materials, light energy delivered, power source and output to user are all the same. The energy delivered (NIR and Visible wavelengths) are similar.
It is acknowledged that there are also some differences: the subject device has features such as live view of images on the monitor, a calibration method to aid in the imaging of different skin tones, a correction for small motion changes of the test location. The subject device includes a sterile single-use drape accessory to cover the optical head and arm. Despite the differences between Presygen™/si-1 and the predicate, extensive bench testing, software verification and validation, usability testing and two comparative clinical performance studies were conducted under design controls to demonstrate that the Presygen™/si-1 device meets its design specifications and that the differences from the predicate do not adversely impact safety or effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word, there is a plus sign made up of small squares in varying shades of green. The logo is simple and modern.
### 8. PERFORMANCE TESTING
The 510(k) submission provided performance data to establish the substantial equivalence of the Presygen™/si-1 to the predicate device. A summary of these performance tests is provided below.
Sterilization, Storage, and Packaging: Presygen™/si-1 is a non-sterile imaging device and it is used with the Presygen™/si-1 Sterilized Drape as a pre-packaged single-use accessory from a Third-Party. Verification and validation testing for usage with Presygen™/si-1 together with the Presygen™/si-1 sterile drape met all acceptance criteria. In addition, the sterile drape was used successfully in the usability studies and in the clinical study that support this application.
Presygen™/si-1 device storage has been validated successfully under the proposed storage conditions of -9° to +140° F (-23° to 60°C), and operating conditions of 60° to 86°F (16° to 30°C), which is standard for many electronic medical device products.
Transportation, handling, and packaging strength were tested according to the applicable sections of ASTM D4169-16 by a third party and met the acceptance criteria.
Software Verification and Validation Testing: Software verification and validation testing was conducted according applicable sections of IEC 62304-06 and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. Unit- and system-level software verification and validation testing demonstrated that the system performs as intended. The software successfully passed all requirements.
Electrical and Electromagnetic Compatibility: The Presygen™/si-1 was tested to international electrical and electromagnetic safety testing in accordance with AAMI ANSI ES 60601-1-12 and IEC 60601-1-2-07., and the device met the standards.
Bench (Performance) Testing: Christie Medical Holdings, Inc. performed a series of bench tests to demonstrate that Presygen™/si-1 meets its performance specifications; the device passed all test requirements and met the acceptance criteria for performance. The Bench Performance Tests were as follows:
- . Mechanical component tests
- . Vibration, temperature, and humidity specific conditions tests
- Presygen™/si-1 sterile drape usage functionality device test •
- . Cleaning agents and cleaning process in health care setting tests
- Electrical sensors functionality tests ●
- . Electrical circuit boards functionality tests
- Optical functionality tests .
- . Phantom study
Optical Safety: The system belongs in the Low Risk Group according to IEC 62471-06, Photobiological safety of lamps and lamp systems.
Laser Safety: The system is a laser product of Class 1 according to 21 CFR \$1040.10 and complies with IEC 60825-07.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of nine squares in varying shades of green. The plus sign is smaller than the word "CHRISTIE" and is positioned so that the top of the plus sign is aligned with the top of the word.
Usability Testing: Presygen™/si-1 performed as intended according to AAMI ANSI IEC 62366-1-15. Application of usability engineering to medical devices.
## Phantom Testing:
## Study Objectives:
An independent bench validation phantom study was conducted to evaluate measurement accuracy and spatial uniformity of StO2 values as a secondary validation for (StO2% measurement range of 50%-90%) in which 3 phantom sets were generated with varying %StO2 at high, medium and low levels (target levels of 90%, 70% and 50% StO2, respectively). These phantoms were fabricated using a mixture of human hemoglobin tuned to the target StO2 value and polystyrene microspheres, which act as optical scatterers similar to human tissue. The spectrophotometer measurements for the controlled samples, vs. the measurements of Presygen™/si-1 %StO2 were statistically evaluated to demonstrate the Presygen™/si-1 measurement accuracy in %StO2 and the spatial variations across the imaging field.
## Results:
The results of this testing demonstrated a significant linear trend of %StO2 in a plot of actual vs. theoretical values using a target confidence of 95%. Phantom results indicate that the results are acceptably uniform over the device field of view.
Clinical Studies: A series of two studies were conducted comparing tissue oxygen saturation (StO2) measurements taken with Presygen™/si-1 system versus the Kent Camera [K113507] and with Presygen™/si-1 system and the T.Ox™ Tissue Oximeter [K042657]. The study objectives were:
- To provide clinical evidence to support the claim that Presygen™/si-1 is substantially equivalent to Kent Camera, with a particular focus on subjects in Fitzpatrick Skin Type Scale I and II.
- . To compare the measurement of StO2 between Presygen™/si-1 and T.Ox™ Tissue Oximeter in subjects across Fitzpatrick Skin Type Scales I through V since the device output is not significantly affected by melanin.
## Clinical Study 1
## Clinical Study 1 Objectives:
Study number 1 was a prospective, unblinded, convenience sample population volunteers from the general community (no health information collected) for 55 subjects (199 data points) from ages 22 years and older, male and female. Fifty-two subjects (10 male and 42 female) were included in data analysis and three subjects were excluded due to incomplete data associated with device issues. To evaluate the effect of patient skin type, subjects were stratified across five (5) Fitzpatrick Skin Type Scales. Subjects had static images taken of target areas [on the thenar eminence and ventral forearm] with all three devices.
## Clinical Study 1 Results:
A Vascular Occlusion Test (VOT) was conducted with all devices to evaluate performance during ischemia and reperfusion over a clinically meaningful dynamic range of StO2. The %StO2 results measurement range was 40% to 90%.
Presvgen™/si-1 compared to the Kent camera in Fitzpatrick Scale I with >95% of the observations within the Limits of Agreement on the Bland-Altman with repeated measures and Deming Regression with bootstrap resampling results showing a slope with 95% confidence limits
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in blue, block letters. To the right of the word is a plus sign made up of small green squares. The logo is simple and modern.
including 1. The intercept for the Deming regression equation did not contain 0. however, demonstrating a small positive bias in the measurement of StO2. The assessment of precision found the coefficient of variation was within 2 standard deviations of the mean for >95% of all observations in the measurement of StO2 in Fitzpatrick I and II. Presygen™/si-1, Kent Camera, and T.Ox™ all demonstrate similar reactions to ischemia and reperfusion in the measurement of StO2.
## Clinical Study 1 VOT Results:
VOT data for Presygen™/si-1. Kent Camera, and T.Ox™ Tissue Oximeter all demonstrate similar reactions to ischemia and reperfusion in the measurement of %StO2. The VOT graphs for all three devices show similar patterns of baseline, response to ischemia and reperfusion. Data from the vascular occlusion test shows %StO2 stable at baseline, decreasing through ischemia and rebounding during reperfusion for both Presygen™/si-1 and Kent Camera, and between Presygen™/si-1 and T.Ox™ Tissue Oximeter.
## Clinical Study 1 VOT Re-analysis:
Reanalysis of Clinical Study 1 VOT data was conducted to evaluate the measured temporal StO2 change between Presygen™/si-1 and the reference device (T.Ox™). The measured temporal StO2 change (ΔStO2) was calculated as the measured StO2 value subtracted by the mean StO2 value of the entire time series for a specific subject. The Bland-Altman plot showed most of the data is centered close to a Mean Bias that is close to 0; 91% of the observations are within the Limits of Agreement. It should be noted that the paired VOT measurements were collected on two different arms and at separate time intervals which generates an additional source of variation. Bland-Altman analysis showed a mean difference of 0.110 %StO2 with Limits of Agreement (-9.052, 9.273). The analysis of the correlation between the AStO2 by Deming regression shows equivalence to T.Ox™ with intercepts that are close to 0 and slopes that are close to 1. Deming Regression Analysis showed an intercept of 0.00 %StO2 with a 95% CI (-0.282, 0.278) and a slope of 0.773 with a 95% CI (0.708, 0.845). The histogram plot demonstrates that the differences in the △StO2 are distributed evenly around 0.
Throughout the study there was no SAE (Serious Adverse Events) or AE (Adverse Events) observed.
## Clinical Study 2
Clinical Study 2 Objectives:
Study number 2 was also a prospective, unblinded, convenience sample population who selfreported as being healthy for 20 subjects (6 male, 14 female) from ages 22 years and older, male and female in a laboratory setting. The aim of this study was to provide data to compare StQz values in the left and right forearm using Presvgen™/si-1 (subject device). Kent Camera (predicate device; Fitzpatrick Scales I and II) and T.Ox™ (reference device; Fitzpatrick Skin Type Scales I through V). Test conditions included static measurements and VOT.
Primary effectiveness included evaluation with the predicate device (Kent Camera) and reference device (T.OxTM Tissue Oximeter) using Bland-Altman with repeated measures and Deming Regression with bootstrap resampling. Secondary effectiveness included evaluation across all Fitzpatrick Scales with all three devices (Presygen™/si-1, Kent Camera and T.Ox™ Tissue Oximeter). Presygen™/si-1 and Kent Camera results from this study should be interpreted with caution due to the small sample size and the influence of unusual values from the Kent camera with several observations.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "CHRISTIE" in large, bold, blue letters. To the right of the word is a plus sign made up of small squares in shades of green. The logo appears to be for a company or organization named Christie.
## Clinical Study 2 Results:
Results of this study were as follows: StO2 values with Presygen™/si-1 and T.OxTM Tissue Oximeter shows equivalence across all Fitzpatrick Scales including within the higher melanin scales of II-V with all of the observations within the Limits of Agreement on the Bland-Altman plot (Table 1). Results of Presygen™/si-1 versus Kent Camera were limited to Fitzpatrick I and II. The %StO2 results measurement range was 67% to 90%.
## Clinical Study 2 VOT Results:
For the VOT study, Presygen™/si-1, Kent Camera and T.Ox™ Tissue Oximeter all demonstrate similar fluctuations in response to ischemia and reperfusion in the measurement of %StO2. Data from the vascular occlusion test shows %StO2 stable at baseline, decreasing through ischemia and rebounding during reperfusion for both Presygen™/si-1 and Kent Camera, and between Presygen™/si-1 and T.Ox™ Tissue Oximeter.
## Clinical Study 2 VOT Re-analysis:
Reanalysis of Clinical Study 2 VOT data was conducted as described above for Clinical Study 1 VOT Re-analysis. The Bland-Altman plot shows most of the data is centered close to a Mean Bias that is close to 0; 90% of the observations are within the Limits of Agreement. It should be noted that the paired VOT measurements were collected on two different arms and at separate time intervals which generates an additional source of variation. Bland-Altman analysis showed a mean difference of 0.016 %StO2 with Limits of Agreement (-8.563, 8.596). The analysis of the correlation between the AStO2 by Deming regression shows equivalence with intercepts that are close to 0 and slopes that are close to 1. Deming Regression Analysis showed an intercept of 0.00 %StO2 with a 95% CI (-0.423, 0.401) and a slope of 0.815 with a 95% CI (0.716, 0.937). The histogram plot demonstrates that the differences in the AStO2 are distributed evenly around 0.
Throughout the study there was no SAE or AE observed.
In conclusion, the weight of the evidence from both clinical study 1 and study 2 shows that Presygen™/si-1 and Kent Camera have consistent directional temporal %StO2 changes, particularly during the VOT. Presygen™/si-1 compared to the Kent camera in Fitzpatrick Scale I with >95% of the observations within the Limits of Agreement on the Bland-Altman plot and the Deming regression results showing a slope with 95% confidence limits including 1 (Study 1). The intercept for the Deming regression equation did not contain 0 however, demonstrating a small positive bias in the measurement of %StO2. Presygen™/si-1 showed equivalence to Kent Camera in subjects with minimal tissue melanin. These studies shows equivalence between Presygen™/si-1 and T.Ox™ Tissue Oximeter %StO2 measurements overall, and within each Fitzpatrick Scale, demonstrating that Presygen™/si-1 shows equivalence with the reference device measurement of %StO2 across Fitzpatrick Scales I through V and is not significantly affected by tissue melanin.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the Christie logo. The word "CHRISTIE" is written in large, bold, blue letters. To the right of the word is a green plus sign made up of smaller squares. The logo is simple and modern.
| | Clinical Study 1 | Clinical Study 2 |
|--------------------------------|-----------------------------|-----------------------------|
| Y-Intercept<br>(95% agreement) | 119.803<br>(98.08, 151.617) | -0.590<br>(-21.244, 16.253) |
| Slope<br>(95% agreement) | -0.734<br>(-1.204, -0.399) | 1.007<br>(0.788, 1.278) |
Table 3. Y-Intercept and Slope from Deming Regression for Presygen vs. Reference Device in Clinical Study 1 vs. Study 2
### Table 4. Summary Table of Bland-Altman Data for Presygen vs. Reference Device in Clinical Study 1 vs. Study 2
| | Clinical Study 1 | Clinical Study 2 |
|----------------------------------|------------------------|----------------------|
| Bias (%StO2),<br>(95% agreement) | 6.4595<br>(4.25, 8.67) | 0.1<br>(-1.64, 1.84) |
| LOA (%StO2) | -31.45, 44.37 | -10.65, 10.85 |
| RMSE (%StO2) | 20.35 | 5.38 |
### CONCLUSIONS 9.
The information presented in this 510(k) submission demonstrates that the Presygen™/si-1 is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.