Nova StatSensor Creatinine Hospital Meter System

K171059 · Nova Biomedical Corporation · CGL · May 26, 2017 · Clinical Chemistry

Device Facts

Record IDK171059
Device NameNova StatSensor Creatinine Hospital Meter System
ApplicantNova Biomedical Corporation
Product CodeCGL · Clinical Chemistry
Decision DateMay 26, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1225
Device ClassClass 2
AttributesPediatric

Indications for Use

The Nova StatSensor Creatinine Hospital Meter System is intended for the quantitative determination of creatinine in whole blood. The Nova StatSensor Creatinine Hospital Meter System is intended for use by healthcare professionals in a clinical setting. It is an in vitro diagnostic device intended to aid in the assessment of renal function.

Device Story

Nova StatSensor Creatinine Hospital Meter System; handheld point-of-care device for whole blood creatinine measurement. Input: whole blood sample applied to test strip. Operation: electrochemical biosensor technology; meter measures current proportional to creatinine concentration. Output: quantitative creatinine result displayed on screen. Used by healthcare professionals in clinical settings. Modifications include ergonomic housing redesign, larger display font, flat surface design, addition of mechanical test strip eject button, relocation of test strip port and barcode scanner, and keypad simplification. Device provides rapid renal function assessment to support clinical decision-making.

Clinical Evidence

Bench testing only. Verification and validation activities performed based on risk analysis to confirm that ergonomic and physical modifications did not impact device performance or fundamental scientific technology.

Technological Characteristics

Electrochemical biosensor; whole blood creatinine measurement. Physical modifications: plastic housing, flat top surface, relocated test strip port and barcode scanner, mechanical test strip eject button, simplified keypad. Connectivity: 2D barcode scanner added. System remains unchanged in fundamental scientific technology (electrochemical sensing).

Indications for Use

Indicated for healthcare professionals in clinical settings for quantitative determination of creatinine in whole blood to aid in assessment of renal function.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: DOCUMENT NUMBER K171059 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. Nova StatSensor Creatinine Hospital Meter System (Nova StatSensor Creatinine Hospital Meter and Nova StatSensor Creatinine Test Strips), k070068. 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: a. Updates in the ergonomic design of the creatinine meter: - Lighter, plastic outer housing - Larger font on the display - New flat, top surface design that eliminates ridges, recessed corners and raised edges - The addition of an optional use mechanical test strip eject button - Relocation of the test strip port from the top of the meter to the bottom of the meter - The addition of a 2-dimensional bar code scanner which can interpret additional identification formats - Relocation of the bar code scanner from the bottom of the meter to the top of the meter - Elimination of the extra duplicate hard key design of the keypad 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance studies of the Nova StatSensor Creatinine Hospital Meter. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis {1} b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...