The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients. When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy. These modules are intended for use by qualified medical personnel only.
Device Story
CARESCAPE Respiratory Modules are plug-in parameter modules for modular hospital monitoring systems. They function as diverting-type gas analyzers, sampling a continuous flow of gas from a patient's breath. The modules measure concentrations of CO2, O2, N2O, and anesthetic agents, identify anesthetic agents, and calculate respiration rate, airway pressure, flow, volume, and gas exchange parameters (VO2, VCO2). The modules process sensor signals and communicate parameter values to a host device; they do not trigger alarms independently. The host device manages all physiological and technical alarms. The system includes a sample gas return feature for use with compatible anesthesia machines. Used by qualified medical personnel in clinical settings. Benefits include real-time respiratory monitoring to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, design reviews, module verification, integration testing, validation, and performance/safety testing. Testing specifically evaluated the impact of the gas return connection and filter on measurement accuracy and component lifetime, as well as VOC and particulate matter testing for new materials.
Technological Characteristics
Diverting-type gas analyzer module. Materials: tested for VOC and particulate matter. Connectivity: plug-in module for host monitoring systems. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 62304, IEC 60601-1-6, IEC 62366, ISO 21647, ISO 14971. Software: compliant with IEC 62304.
Indications for Use
Indicated for monitoring respiratory, ventilatory, and gas exchange parameters in adult, pediatric, and neonatal patients. Respiratory/ventilatory parameters apply to all; gas exchange parameters apply to adult and pediatric patients. Contraindicated for use outside specified respiration rate and tidal volume limits for neonates or patients with high respiration rates/low tidal volumes.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a triple-line design representing the bodies.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 17, 2017
GE Healthcare Finland Oy Anna Pehrsson Regulatory Affairs Leader Kuortaneenkatu 2, Helsinki FI-00510 FINLAND
Re: K171028
Trade/Device Name: CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, EsCAiO, E-sCAiOV, E-sCAiOVX and accessories Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Dated: July 5, 2017 Received: July 17, 2017
Dear Anna Pehrsson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Tara A. Ryan -S 2017.08.17 06:34:11 -04'00' for Lori Wiggins Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. | 0910-0120 |
|------------------------|----------------------|
| Expiration Date: | 06/30/2020 |
| See | PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K171028 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | CARESCAPE Respiratory Modules, E-SCO, E-SCOV, E-SCOVX, ESCAIO, E-sCAIOV, E-SCAIOVX and accessories |
| Indications for Use (Describe) | The CARESCAPE Respiratory Modules (E-SCO, E-SCOV, E-SCOVX, E-sCAIO, ESCAIOV, E-SCAIOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients. |
| | When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy. |
| | These modules are intended for use by qualified medical personnel only. |
| Type of Use (Select one or both, as applicable) | <div> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 413-674) EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white teardrop shapes around the outside of the circle. The logo is simple and recognizable, and it is associated with a well-known and respected company.
GE Healthcare Finland Oy
Kuortaneenkatu 2, P.O. Box 900 FI-00031 GE Finland T: +358 10 39411
F: +358 9 1463310
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date:<br>Submitter: | July 5, 2017<br>GE Healthcare Finland Oy,<br>Kuortaneenkatu 2. Helsinki,<br>FI-00510 FINLAND |
|---------------------------|----------------------------------------------------------------------------------------------|
| Primary Contact Person: | Anna Pehrsson |
| | Regulatory Affairs Leader |
| | GE Healthcare Finland Oy |
| | phone (+358) 40-569-5018 |
| | email anna.pehrsson@ge.com |
| Secondary Contact Person: | Rauno Ruoho |
| | Regulatory Affairs Director |
| | GE Healthcare Finland Oy |
| | phone (+358) 10-394-3624 |
| | email rauno.ruoho@ge.com |
| Device names: | |
| Trade Name: | CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-<br>sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessories |
|--------------------|---------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Respiratory gas module and accessories |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are several water droplet shapes surrounding the circle. The logo is simple and recognizable.
Classification Name: 21 CFR 868.1400 Analyzer, Gas, Carbon-Dioxide, Gaseous Phase CCK Product code: Additional Classification 21 CFR 868.1720 Analyzer, Gas, Oxygen, Gaseous-phase Names: 21 CFR 868.1850 Spirometer, Monitoring (W/WO Alarm) 21 CFR 868.2600 Monitor, Airway Pressure (Includes gauge and/or alarm) 21 CFR 868.1700 Analyzer, Gas, Nitrous-Oxide, Gaseous-phase (Anesthetic conc) 21 CFR 868.1730 Computer, Oxygen-uptake 21 CFR 868.1500 Analyzer, Gas, Enflurane, Gaseous-phase (Anesthetic conc.) 21 CFR 868.1620 Analyzer, Gas, Halothane Gaseous-phase (Anesthetic conc.) 21 CFR 868.1500 Analyzer, Gas, Desflurane, Gaseous-phase (Anesthetic conc.) 21 CFR 868.1500 Analyzer, Gas, Isoflurane Gaseous-phase (Anesthetic conc.) 21 CFR 868.1500 Analyzer, Gas, Sevoflurane, Gaseous-phase (Anesthetic conc) Additional Product Codes: CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Predicate Device(s): K150245: CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX Device Description: The CARESCAPE Respiratory Modules E-sCO, E-sCOV, EsCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessories measure respiratory parameters (concentrations of Carbon Dioxide. Oxygen, Nitrous Oxide and anesthetic agents in the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are three water droplet shapes around the circle. The logo is simple and recognizable.
patient's breath, as well as the patient's respiration rate), ventilatory parameters (airway pressure, flow and breathing volumes) and gas exchange parameters (oxygen consumption and carbon dioxide production) of hospital patients.
Parameters measured by the CARESCAPE Respiratory Modules are CO2. N2O. O2. Anesthetic agents, Agent ID, Spirometry, oxygen consumption (VO2) and carbon dioxide production (VCO2) depending on the model used. The CARESCAPE Respiratory Modules is a family of single-width plug-in parameter modules for modular monitoring systems. The CARESCAPE Respiratory Modules are of the diverting type, which means that a small continuous flow of gas is sampled from the patient's breath to the module for measuring the gas concentrations. The CARESCAPE Respiratory Modules acquire the detected signals from the sensors of the modules, calculate the parameter values and communicate them to the host device. The CARESCAPE Respiratory Modules measure the patient's respiration rate and activate a status signal if no breaths are detected in 20 second time and the modules activate relevant status signals upon detecting failures or anomalies in the operation of the module hardware, software or gas sampling system.
The CARESCAPE Respiratory Modules do not trigger or issue any physiological or technical alarms by themselves. All management of alarms is entirely performed by the host devices based on parameter and status data received from the modules, as well as on the alarm condition data stored in the host device.
- Intended Use: The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, EsCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients.
When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative, water-like flourishes around the perimeter of the circle, adding a touch of elegance to the design. The logo is simple, recognizable, and represents a well-established global brand.
These modules are intended for use by qualified medical personnel only.
The modified CARESCAPE Respiratory Modules E-sCO. E-Technology: sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessories do not add any new technology or software to existing predicate device CARESCAPE Respiratory Modules.
> The sample gas return feature is added to CARESCAPE Respiratory Modules. The feature may only be used with host device anesthesia machines which have 510(k) clearance for the gas return feature.
> The technology of the CARESCAPE Respiratory Modules and accessories is the same as in the predicate device.
> The CARESCAPE Respiratory Modules and accessories is as safe and effective as the predicate device.
Determination of Substantial Summary of Non-Clinical Tests Equivalence:
The CARESCAPE Respiratory Modules and its applications comply with voluntary standards as detailed below. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews ●
- . Testing on unit level (Module verification)
- Integration testing (System verification)
- Final acceptance testing (Validation) .
- Performance testing (Verification)
- Safety testing (Verification) ●
There are no changes or additions to module software, hardware, mechanics or accessories related to the addition of sample gas return feature. Based on this, testing of the modules is only done to cover any possible effects introduced by the gas return circuit in the host device. An evaluation was performed to account for gas return connection and gas return filter impact on the gas measurement accuracy, sample flow measurement accuracy as well as pump and filter lifetime. System level
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, water-like shapes, also in blue. The overall design is clean and recognizable, representing the brand identity of GE.
Volatile Organic Component (VOC) and particulate matter testing was performed to evaluate new components and materials in the dry gas path.
Changes to the labeling, hardware and components of the CARESCAPE Respiratory Modules since the last clearance of the CARESCAPE Respiratory Modules (K150245) reflect ongoing product maintenance activities and do not affect product safety and performance. It is concluded that the CARESCAPE Respiratory Modules and accessories are substantially equivalent to the predicate device.
The CARESCAPE Respiratory Modules were designed and tested for compliance to the following standards:
- 1. IEC 60601-1:1988, A1:1991, A2:1995, Corr1:1995, Medical electrical equipment Part 1: General Requirements for Safety - Second Edition
- 2. IEC 60601-1-2:2001, A1:2004, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - Edition 2.0
- 3. IEC 60601-1-2:2007, Medical electrical equipment --Part 1-2: General requirements for basic safety and essential standard: Electromagnetic compatibility - Requirements and tests
- 4. IEC 60601-1-4:1996 + A1:1999, Medical Electrical Equipment - Part 1-4: General Requirements for Safety - Collateral Standard: Programmable Electrical Medical Systems
- 5. IEC 62304:2006, Medical device software Software life-cycle processes
- 6. IEC 60601-1-6:2006, Medical electrical equipment-Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability - Edition 2.0
- 7. IEC 60601-1-6:2010. Medical electrical equipment -Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62366:2007. Medical Devices Application of 8. usability engineering to medical devices
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are four water droplet shapes surrounding the circle. The logo is simple and recognizable.
- 9. ISO 21647:2004 + C1:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of respiratory gas monitors
- 10. EN 1041:2008, Information supplied by the manufacturer of medical devices
- 11. ISO 14971: 2007 Medical devices Application of risk management to medical devices
The subject of this premarket submission, CARESCAPE Summary of Clinical Tests: Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessories did not require clinical studies to support substantial equivalence.
> GE Healthcare considers the CARESCAPE Respiratory Conclusion: Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessories to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.