The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCAiO, E-sCAiOV, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients. When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy. These modules are intended for use by qualified medical personnel only.
Device Story
CARESCAPE Respiratory Modules are single-width plug-in parameter modules for modular hospital monitoring systems. Devices use diverting-type gas sampling to continuously draw patient breath for analysis. Modules measure CO2, O2, N2O, anesthetic agents, respiration rate, airway pressure, flow, volume, and gas exchange (VO2, VCO2). Modules acquire sensor signals, calculate parameters, and communicate data to a host monitor. Host monitors manage all physiological and technical alarms based on module data. Used by qualified medical personnel in clinical settings. Benefits include real-time respiratory and gas exchange monitoring to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, design reviews, unit-level verification, system integration testing, and final validation. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 62304, and ISO 21647 standards confirmed.
Technological Characteristics
Diverting-type gas sampling module; plug-in form factor for modular monitoring systems. Measures gas concentrations (CO2, O2, N2O, anesthetic agents) and ventilatory parameters. Software version 2.0. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 62304, IEC 60601-1-6, IEC 62366, ISO 21647, and EN1041.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring monitoring of respiratory parameters (CO2, O2, N2O, anesthetic agents, respiratory rate), ventilatory parameters (airway pressure, flow, volume), and gas exchange (VCO2, VO2). Neonatal/high-rate/low-tidal-volume patients must be monitored within specified measurement accuracy limits.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCAiO and E-sCAIOV and accessories (K123195)
Datex-Ohmeda S/5 Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CaiOV, E-CaiO, E-CaiO, E-COVX, E-COV, E-CO and accessories (K051092)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2015
GE Healthcare Finland Oy Mari Salmenkaita Regulatory Affairs Leader Kuortaneenkatu 2 FI-00510 Helsinki Finland
Re: K150245
Trade/Device Name: Carescape Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, EsCAiO, E-sCAiOV, E-sCAiOVX Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Dated: June 16, 2015 Received: June 17, 2015
Dear Ms. Salmenkaita:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150245
#### Device Name
CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX
Indications for Use (Describe)
The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCAiO, E-sCAiOV, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients.
When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy. These modules are intended for use by qualified medical personnel only.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo, which is a circular emblem with the stylized letters 'g' and 'e' intertwined in the center. The letters are in a cursive font and are surrounded by a dark background. The outer edge of the circle is adorned with decorative, teardrop-shaped elements, adding a touch of elegance to the design.
GE Healthcare Finland Oy Kuortaneenkatu 2, P.O. Box 900 FI-00031 GE
Finland
T: +358 10 39411 F: +358 9 1463310
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | July 17, 2015 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Finland Oy,<br>Kuortaneenkatu 2, Helsinki,<br>FI-00510 FINLAND |
| Primary Contact Person: | Mari Salmenkaita<br>Regulatory Affairs Leader<br>GE Healthcare Finland Oy<br>phone (+358) 10-394-2122<br>fax (+358) 9-272-6532<br>email mari.salmenkaita@ge.com |
| Secondary Contact Person: | Rauno Ruoho<br>Regulatory Affairs Manager<br>GE Healthcare Finland Oy<br>phone (+358) 10-394 3624<br>fax (+358) 9-272-6532<br>email rauno.ruoho@ge.com |
| Device names: | |
| Trade Name: CARESCAPE 101 Respiratory Modules, E-sCO, E-sCOV, E- |
|------------------------------------------------------------------|
| sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX |
Common/Usual Name: Respiratory gas module
{4}------------------------------------------------
| Classification Names: | 21 CFR 868.1400 Analyzer, Gas, Carbon-Dioxide, Gaseous-phase |
|-----------------------|-----------------------------------------------------------------------------------------------|
| | 21 CFR 868.1720 Analyzer, Gas, Oxygen, Gaseous-phase |
| | 21 CFR 868.1850 Spirometer, Monitoring (W/WO alarm) |
| | 21 CFR 868.2600 Monitor, Airway Pressure (Includes gauge and/or alarm) |
| | 21 CFR 868.1700 Analyzer, Gas, Nitrous-Oxide, Gaseous-phase (Anesthetic conc) |
| | 21 CFR 868.1730 Computer, Oxygen-uptake |
| | 21 CFR 868.1500 Analyzer, Gas, Enflurane, Gaseous-phase (Anesthetic conc.) |
| | 21 CFR 868.1500 Analyzer, Gas, Halothane Gaseous-phase (Anesthetic conc.) |
| | 21 CFR 868.1500 Analyzer, Gas, Desflurane, Gaseous-phase (Anesthetic conc.) |
| | 21 CFR 868.1500 Analyzer, Gas, Isoflurane Gaseous-phase (Anesthetic conc.) |
| | 21 CFR 868.1500 Analyzer, Gas, Sevoflurane, Gaseous-phase (Anesthetic conc) |
| Product Code: | CCK, CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP |
| Predicate Device(s): | K123195: CARESCAPET™ Respiratory Modules, E-sCO, E-sCOV, E-sCAiO and E-sCAIOV and accessories |
K051092: Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CaiOV, E-CaiO, E-CaiO, E-COVX, E-COV, E-CO and accessories.
{5}------------------------------------------------
- The CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-Device Description: sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX, measure respiratory parameters (concentrations of Carbon Dioxide, Oxygen, Nitrous Oxide and anesthetic agents in the patient's breath, as well as the patient's respiration rate), ventilatory parameters (airway pressure, flow and breathing volumes) and gas exchange parameters (oxygen consumption and carbon dioxide production) of hospital patients.
Parameters measured by the CARESCAPE Respiratory Modules are CO2, N2O, O2, Anesthetic agents, Agent ID, Spirometry, oxygen consumption (VO2) and carbon dioxide production (VCO2) depending on the model used. The CARESCAPE Respiratory Modules is a family of single-width plug-in parameter modules for modular monitoring systems. The CARESCAPE Respiratory Modules are of the diverting type, which means that a small continuous flow of gas is sampled from the patient's breath to the module for measuring the gas concentrations. The CARESCAPE Respiratory Modules acquire the detected signals from the sensors of the modules, calculate the parameter values and communicate them to the host device. The CARESCAPE Respiratory Modules measure the patient's respiration rate and activate a status signal if no breaths are detected in 20 second time and the modules activate relevant status signals upon detecting failures or anomalies in the operation of the module hardware, software or gas sampling system.
The CARESCAPE Respiratory Modules do not trigger or issue any physiological or technical alarms by themselves. All management of alarms is entirely performed by the host monitors based on parameter and status data received from the modules, as well as on the alarm condition data stored in the host device.
- The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-Intended Use: sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2, VO2) of adult and pediatric patients.
When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used
{6}------------------------------------------------
within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy.
These modules are intended for use by qualified medical personnel only.
- Technology: The modified CARESCAPE Respiratory Modules add models EsCOVX, E-sCAiOVX with the gas exchange calculation to existing predicate device CARESCAPE Respiratory Modules with current software version 1. The gas exchange algorithm is currently available on the predicate device Datex-Ohmeda S/5 Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-COVX (K051092), this 510(k) implements this feature into CARESCAPE Respiratory Modules, E-sCO, EsCOV, E-sCAiO and E-sCAiOV family (K123195). The implemented parameters are oxygen consumption (VO2) and carbon dioxide production (VCO2). The developed software is version 2.
The fundamental technology of the CARESCAPE Respiratory Modules E-sCO. E-sCOV. E-sCOVX, E-sCAiO. E-sCAiO. E-sCAiOV, EsCAiOVX is the same as in the predicate devices.
The CARESCAPE Respiratory Modules E-sCO, E-sCOV, EsCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX is as safe and effective as the predicate devices.
| Determination of | Sub |
|--------------------------|-----|
| Substantial Equivalence: | |
## mmary of Non-Clinical Tests
The CARESCAPE Respiratory Modules and its applications comply with voluntary standards as detailed below. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Final acceptance testing (Validation) ●
- Performance testing (Verification)
- Standard compliance testing (Verification) ●
Extensive non-clinical testing was performed to establish equivalence of the CARESCAPE Respiratory substantial Modules and its applications. Verification and validation testing was performed according to predetermined acceptance criteria,
{7}------------------------------------------------
which concluded that the module and its applications are substantially equivalent to the predicate devices.
The CARESCAPE Respiratory Modules were designed and tested for compliance to the following standards:
- 1. IEC 60601-1:1988, A1:1991, A2:1995, Corr1:1995, electrical Medical Part 1: General Requirements for Safety - Second Edition
- 2. IEC 60601-1-2:2001, A1:2004, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests -Edition 2.0
- 3. IEC 60601-1-2:2007, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- 4. IEC 60601-1-4:1996 + A1:1999, Medical Electrical Equipment - Part 1-4: General Requirements for Safety -Collateral Standard: Programmable Electrical Medical Systems
- 5. IEC 62304:2006, Medical device software Software lifecycle processes
- 6. IEC 60601-1-6:2006, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability - Edition 2.0
- 7. IEC 60601-1-6:2010, Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- 8. IEC 62366:2007, Medical Devices Application of usability engineering to medical devices
- 9. ISO 21647:2004 + C1:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of respiratory gas monitors
- 10. EN1041:2008, Information supplied by the manufacturer of medical devices
Summary of Clinical Tests:
The subject of this premarket submission, CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCOVX, E-sCAiO, EsCAiOV, E-sCAiOVX did not require clinical studies to support substantial equivalence.
GE Healthcare considers the CARESCAPE Respiratory Modules Conclusion:
{8}------------------------------------------------
E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.